  • Hello! I am an AI chatbot trained to assist you with the Extech Instruments RH35 User manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!

USERGUIDE
HygroThermometerAlert
ModelRH35
EXTECHINSTRUMENTS
2RH35enEU_v1.17/14
Introduction
Thankyouforselecting the ExtechInstrumentsModel RH35. Thisdeviceisshippedfullytested
and calibrated and, with proper use, will provide years of reliable service. Please visit our
website(www.extech.com)tocheckforthelatestversionofthisUserGuide,ProductUpdates,
andCustomerSupport.
Safety
InternationalSafetySymbols
Thissymbol,adjacenttoanothersymbolorterminal,indicatestheusermustrefertothemanualfor
furtherinformation.
Doubleinsulation
SafetyNotes
o Removethebatteryifthedeviceistobestoredforlonge rthan60days.
o Neverdisposeofbatteriesinafire.Batteriesmayexplodeorleak.
o Nevermixbatterytypes.Alwaysinstallnewbatteriesofthesametype.
Cautions
o If the equipment is used in a manner not specified by the manufacturer, the protection
providedbytheequipmentmaybeimpaired.
o Batteriescancontainharmfulacids.Lowbatteriesshouldbechangedassoonaspossibleto
preventdamagecausedbyleaking.
o Neveruseacombinationofold
andnewbatteriestogether,orbatteriesofdifferenttypes.
o Wearchemicalresistantprotectiveglovesandglasseswhenhandlingleakingbatteries.
o Donotplaceyourproductnearextremetemperatures,vibrations,orshocks.
o Protectdevicefrommoisture.
EXTECHINSTRUMENTS
3RH35enEU_v1.17/14
1
3
5
4
6
2
1
2
3
7
4
5
6
MeterDescription
MeterFront
1. 24hourmoldradaranimation
2. Indoortemperature(°C/°F)andMIN
MAX/Dewpointtemperature
3. FacialIconsforcomfortlevelstatus
4. MINMAXButton
5. DewPointButton
6. RelativeHumidity%display
7. WarningLEDwhenriskofmold
MeterBack
1. Wallmount
2. Batterycompartment
3. Batterycompartmentlatch
4. AlarmSoundON/OFFswitch(SpeakerSymbol)
5. RadarAnimationON/OFFswitch(Arrow
Symbol)
6. Supportstand(removestandtoaccessAlarm
soundandRadaranimationswitches)
EXTECHINSTRUMENTS
4RH35enEU_v1.17/14
Operation
GettingStarted
Removetheprotectivefoilfromthedisplay.
Openthebatterycompartmentandremovethebatteryinsulationstrip.Thedevicewill
soundthealarmandallLCDsegmentswillbedisplayedfortwosecondsasthedevice
switchesON.
Theindoortemperature,relativehumidity,andcorrespondingfacialsymbolwill
appear
onthedisplay.Thisindicatesthatthemeterisnowreadyforuse.
Note:DonotattachthesupportstanduntiltheswitchesfortheRadarAnimationandAlarm
Soundaresetasdesired.Thesetwoswitchesareconcealedbythesupportstandwhenitis
attached.Referto
theDescriptionsectionofthisguideforthelocationoftheswitchesand
refertothededicatedsectionsbelowformoredetail.
24HourMoldRadarDisplay
The device measures the relative humidity % every 30 seconds and calculates the average
valueevery10minutes.Theinnovativemoldradardisplaygraphicallydepictstheaverageon
a24hourtimeline.
Thetime “0 h” representsthe valuefor the current hour. 1” represents the last hour, 3”
forthepast23hours,6”forthepast46hours,12”forthepast712hours,and24”for
thepast1324hours.
Therelativehumidityvaluesareshownongridlinesfrom15to85%.
Areadinghigherthan65%
RHindicatesapotentialformold.
RadarAnimationON/OFFSwitch
Removethesupportstandtoaccessthe RadarAnimationON/OFF switch on the back of the
unit(theswitch is on the rightidentifiedbythearrowicons).Slidetheswitch UP to activate
theRadarAnimation.SlidetheswitchDOWNtodeactivatetheRadarAnimation.
AlarmSoundON/OFFSwitch
RemovethesupportstandtoaccesstheAlarmSoundON/OFFswitchonthebackoftheunit
(the switch is on the left identified by the audio icons). Slide the switch UP to activate the
AlarmSoundbeeper.SlidetheswitchDOWNtodeactivatetheAlarmSoundbeeper.
If the relative humidity measurement is higher than 65%, the alarm beeper sounds for one
minute and the red warning light (front panel, right) flashes. Press any button to stop the
alarm sound. The warning light continues flashing until the measured humidity falls below
65%.
ComfortLevel
The device displays one of two facial indicators representing the comfort levelof the indoor
climate. A smiling face indicates an ideal relative humidity range (4060%). A frowning face
indicatesanoverlywetordrycondition(<40%or>60%).
EXTECHINSTRUMENTS
5RH35enEU_v1.17/14
DewPointTemperatureDisplay
PresstheDewPointbuttontodisplaythe DewPointTemperaturefor5seconds.Thedisplay
willreverttotheAirTemperature/RelativeHumiditydisplaymodeafter5seconds,
automatically.
TemperatureUnitsSelection
PressandholdtheDewPointbuttonfor5secondstotogglethetemperatureunits(°Cand°F).
Note: In MIN/MAX mode pressing and holding the Dew Point button clears the MIN or MAX
memory (depending on which one is selected) and does not change the temperature units.
EnsurethatMIN/MAXmodeisnotactivatedwhenchangingthetemperatureunits.
MinimumandMaximumValues
PresstheMIN/MAXbuttonandtheMAXiconwillappearalongwiththehighestindoor
temperatureandhumidityreadingscapturedsincethelastreset.
TodisplaytheMINvalues,presstheMIN/MAXbuttonagain.TheMINiconwillappear
alongwiththelowestindoortemperatureandhumidityreadingscapturedsincethelast
reset.
TheinstrumentwillautomaticallyexittheMIN/MAXmodeifnobuttonispressed.
Toexittothenormaldisplaymodemanually,presstheMIN/MAXbuttonagain(theMIN
andMAXindicatorswillswitchOFF).
TocleartheMINorMAXmemory,pressand
holdtheDewPointbuttonforthree
secondswhiletheMINorMAXiconisshowing. TheMINortheMAXindicatorwillflash
asitsmemoryiscleared.
TableTopandWallmountingConsiderations
The unit can be wall mounted using the wall mount access hole located on the rear of the
device.Usethesupportstandfortabletopuse.Notethatthehumiditylevelcanvarygreatly
accordingtolocation;mountthe unitasclosetotheareaofconcernaspossible.
EXTECHINSTRUMENTS
6RH35enEU_v1.17/14
Maintenance
Cleaning
Cleantheinstrumentwithasoftdampcloth.Donotusesolventsorabrasives.
Batteryreplacement
Removethebatteriesiftheproductisnottobeusedforlongperiods.Changethebatteries
whenthebatterysymbolappearsemptyonthedisplay.Opentherearbatterycompartmentand
inserttwonewbatteries1.5VAAAobservingcorrectpolarity.Closethebatterycompartment
beforeuse.
BatterySafetyReminders
o Pleasedisposeofbatteriesresponsibly;observelocal,state,andnationalregulations.
o Neverdisposeofbatteriesinafire;batteriesmayexplodeorleak.
o Nevermixbatterytypes;installnewbatteriesofthesametype.
o
Neverdisposeofusedbatteriesorrechargeablebatteriesin
householdwaste.
As consumers, users are legally required to take used batteries to appropriate
collection sites, the retail store where the batteries were purchased, or wherever
batteriesaresold.
Disposal: Do not dispose of this instrument in household waste. The user is obligated to take
endoflife devices to a
designated collection point for the disposal of electrical and electronic
equipment.
Specifications
TemperatureRange‐10°Cto60°C(14°Fto140°F)
TemperatureAccuracy ±1°C(±1.8°F)from0to50°C(32to122°F),otherwise±1.5°C(2.7°F)
RelativeHumidityRange 10to99%
RHAccuracy ±4%from30%to80RH%,otherwise±5%
Batteries Batteries2x1.5VAAA
Dimensions 98x23[50]x98mm
(3.86x0.91[2.04]x3.86in.)
Weight 87g(3.1oz.)
Copyright©2014FLIRSystems,Inc.