HP 1GSa/s User manual

Category
Multimeters
Type
User manual
SERVICE MANUAL
HP 16532A
1 GSa/s Digitizing
Oscilloscope
 COPYRIGHT HEWLETT-PACKARD COMPANY,
COLORADO SPRINGS DIVISION 1991-1993
1900 GARDEN OF THE GODS ROAD, COLORADO SPRINGS, COLORADO U.S.A.
ALL RIGHTS RESERVED
Manual Part No. 16532-97002 PRINTED: JULY 1993
GENERAL - This is a Safety Class I instrument (provided with
terminal for protective earthing).
OPERATION - BEFORE APPLYING POWER verify that the
power transformer primary is matched to the available line
voltage, the correct fuse is installed, and Safety Precautions
are taken (see the following warnings). In addition, note the in-
strument’s external markings which are described under
"Safety Symbols."
• Servicing instructions are for use by service-trained per-
sonnel. To avoid dangerous electric shock, do not perform
any servicing unless qualified to do so.
• BEFORE SWITCHING ON THE INSTRUMENT, the protec-
tive earth terminal of the instrument must be connected to
the protective conductor of the (mains) powercord. The
mains plug shall only be inserted in a socket outlet pro-
vided with a protective earth contact. The protective action
must not be negated by the use of an extension cord
(power cable) without a protective conductor (grounding).
Grounding one conductor of a two-conductor outlet is not
sufficient protection.
• If this instrument is to be energized via an auto-trans-
former (for voltage reduction) make sure the common ter-
minal is connected to the earth terminal of the power
source.
• Any interruption of the protective (grounding) conductor
(inside or outside the instrument) or disconnecting the pro-
tective earth terminal will cause a potential shock hazard
that could result in personal injury.
• Whenever it is likely that the protection has been im-
paired, the instrument must be made inoperative and be
secured against any unintended operation.
• Only fuses with the required rated current, voltage, and
specified type (normal blow, time delay, etc.) should be
used. Do not use repaired fuses or short circuited fuse-
holders. To do so could cause a shock or fire hazard.
• Do not operate the instrument in the presence of flamma-
ble gasses or fumes. Operation of any electrical instru-
ment in such an environment constitutes a definite safety
hazard.
• Do not install substitute parts or perform any unauthorized
modification to the instrument.
• Adjustments described in the manual are performed with
power supplied to the instrument while protective covers
are removed. Energy available at many points may, if con-
tacted, result in personal injury.
• Any adjustment, maintenance, and repair of the opened in-
strument under voltage should be avoided as much as
possible, and when inevitable, should be carried out only
by a skilled person who is aware of the hazard involved.
• Capacitors inside the instrument may still be charged
even if the instrument has been disconnected from its
source of supply.
SAFETY SYMBOLS
Instruction manual symbol. The product will be
marked with this symbol when it is necessary for
the user to refer to the instruction manual in order
to protect against damage to the product.
Earth terminal (sometimes used in manual to in-
dicate circuit common connected to grounded
chassis).
The WARNING sign denotes a hazard. It
calls attention to a procedure, practice,
or the like, which, if not correctly per-
formed or adhered to, could result in personal injury. Do not
proceed beyond a WARNING sign until the indicated condi-
tions are fully understood and met.
The CAUTION sign denotes a hazard. It
calls attention to an operating procedure,
practice, or the like, which, if not cor-
rectly performed or adhered to, could result in damage to or
destruction of part or all of the product. Do not proceed be-
yond a CAUTION sign until the indicated conditions are fully
understood or met.
SAFETY CONSIDERATIONS
Indicates hazardous voltages
SC1D984
CONTENTS
SECTION I
General Information
1-1. Introduction .............................................................................................................................. 1-1
1-2. Module Covered by This Manual ............................................................................................ 1-1
1-3. Safety Requirements ............................................................................................................... 1-1
1-4. Product Description ................................................................................................................. 1-2
1-5. Accessories Supplied .............................................................................................................. 1-2
1-6. Accessories Available............................................................................................................... 1-2
1-7. Specifications ........................................................................................................................... 1-2
1-8. Operating Characteristics ....................................................................................................... 1-2
1-9. Recommended Test Equipment ............................................................................................. 1-2
SECTION II
Installation
2-1. Introduction ............................................................................................................................. 2-1
2-2. Initial Inspection ...................................................................................................................... 2-1
2-3. Preparation for Use ................................................................................................................ 2-1
2-4. Power Requirements .............................................................................................................. 2-1
2-5. Safety Requirements .............................................................................................................. 2-1
2-6. Module Installation ................................................................................................................. 2-1
2-7. Operating Environment .......................................................................................................... 2-4
2-8. Storage .................................................................................................................................... 2-4
2-9. Packaging ............................................................................................................................... 2-4
2-10. Tagging for Service ................................................................................................................ 2-4
SECTION III
Performance Tests
3-1. Introduction .............................................................................................................................. 3-1
3-2. Recommended Test Equipment ............................................................................................. 3-1
3-3. Test Record .............................................................................................................................. 3-1
3-4. Performance Test Interval ....................................................................................................... 3-1
3-5. Performance Test Procedures ................................................................................................ 3-1
3-6. DC CAL OUTPUT Test ............................................................................................................. 3-2
3-7. AC CAL OUTPUT Test ............................................................................................................. 3-3
3-8. Input Resistance ...................................................................................................................... 3-4
3-9. Voltage Measurement Accuracy ............................................................................................. 3-6
3-10. Offset Accuracy ....................................................................................................................... 3-9
3-11. Bandwidth ................................................................................................................................ 3-11
3-12. Time Measurement Accuracy .................................................................................................. 3-13
3-13. Trigger Sensitivity .................................................................................................................... 3-15
HP 16532A - Table of Contents
v
SECTION IV
Calibration
4-1. Introduction............................................................................................................................... 4-1
4-2. Safety Requirements................................................................................................................ 4-1
4-3. Calibration Interval.................................................................................................................... 4-1
4-4. Maintaining System Integrity.................................................................................................... 4-1
4-5. Calibration Protection Switch................................................................................................... 4-1
4-6. Recommended Test Equipment..............................................................................................4-2
4-7. Instrument Warmup.................................................................................................................. 4-2
4-8. Calibration................................................................................................................................. 4-2
4-9. Setting the PROTECT/UNPROTECT Switch to UNPROTECT................................................ 4-3
4-10. Equipment Required for Calibration ....................................................................................... 4-4
4-11. Loading the Default Calibration Factors................................................................................. 4-4
4-12. Self Cal Menu Calibrations...................................................................................................... 4-5
4-13. Setting the PROTECT/UNPROTECT Switch to PROTECT.................................................... 4-8
4-14. Sealing the Mainframe............................................................................................................. 4-8
Section V
Replaceable Parts
5-1. Introduction .............................................................................................................................. 5-1
5-2. Abbreviations ........................................................................................................................... 5-1
5-3. Replaceable Parts List ............................................................................................................. 5-1
5-4. Ordering Information ............................................................................................................... 5-1
5-5. Exchange Assemblies ............................................................................................................. 5-1
5-6. Direct Mail Order System ........................................................................................................ 5-2
SECTION VI
Service
6-1. Introduction .............................................................................................................................. 6-1
6-2. Safety Requirements ............................................................................................................... 6-1
6-3. Recommended Test Equipment ............................................................................................. 6-1
6-4. Module Block Diagram and Theory of Operation................................................................... 6-1
6-5. Attenuator Theory of Operation............................................................................................... 6-1
6-6. Main Assembly Theory of Operation....................................................................................... 6-1
6-7. Self-Tests .................................................................................................................................. 6-4
6-8. Troubleshooting ....................................................................................................................... 6-10
6-9. Oscilloscope Module Removal and Replacement ................................................................. 6-18
6-10. Trigger Cable Removal and Replacement ............................................................................. 6-20
6-11. Attenuator Removal and Replacement .................................................................................. 6-21
HP 16532A - Table of Contents
vi
LIST OF TABLES
TABLE TITLE PAGE
1-1. HP 16532A Specifications........................................................................................................ 1-3
1-2. HP 16532A Operating Characteristics..................................................................................... 1-4
1-3. Recommended Test Equipment..............................................................................................1-9
3-1. Performance Test Record ........................................................................................................ 3-17
4-1. Calibration Sequence............................................................................................................... 4-2
5-1. Reference Designator and Abbreviations................................................................................ 5-3
5-2. Replaceable Parts List.............................................................................................................. 5-5
LIST OF FIGURES
FIGURE TITLE PAGE
2-1. Endplate Overlap - Top Sequence........................................................................................... 2-2
2-2. Endplate Overlap - Bottom Sequence..................................................................................... 2-3
3-1. Voltage Measurement Accuracy Test........................................................................................ 3-7
4-1. Endplate Overlap - Top Sequence........................................................................................... 4-3
4-2. Location of the PROTECT/UNPROTECT Switch..................................................................... 4-3
4-3. Default Calibration Display....................................................................................................... 4-4
4-4. Ext Trig Null Calibration Connections...................................................................................... 4-7
4-5. End of Calibration Display........................................................................................................ 4-8
4-6. Endplate Overlap - Top Sequence .......................................................................................... 4-8
5-1. Parts Identification.................................................................................................................... 5-4
6-1. HP 16532A Oscilloscope Block Diagram................................................................................ 6-2
6-2. Startup Screen.......................................................................................................................... 6-4
6-3. Load Test System Screen........................................................................................................ 6-5
6-4. Test System Screen.................................................................................................................. 6-5
6-5. Functional Tests Screen........................................................................................................... 6-6
6-6. Main Test Menu ........................................................................................................................ 6-6
6-7. A/D Test Run Screen................................................................................................................ 6-7
6-8. Stop Field.................................................................................................................................. 6-7
6-9. Exit Test System Screen.......................................................................................................... 6-8
6-10. Troubleshooting Flowchart...................................................................................................... 6-13
6-11. Endplate Overlap - Top Sequence........................................................................................... 6-18
6-12. Endplate Overlap - Bottom Sequence..................................................................................... 6-19
6-13. Oscilloscope Module Exploded View...................................................................................... 6-20
HP 16532A - Table of Contents
vii
TABLE OF CONTENTS
General Information
1-1. Introduction...........................................................................................................................1-1
1-2. Module Covered by Manual.................................................................................................1-1
1-3. Safety Requirements.............................................................................................................1-1
1-4. Product Description..............................................................................................................1-2
1-5. Accessories Supplied...........................................................................................................1-2
1-6. Accessories Available...........................................................................................................1-2
1-7. Specifications........................................................................................................................1-2
1-8. Operating Characteristics.....................................................................................................1-2
1-9. Recommended Test Equipment ..........................................................................................1-2
SECTION I
General Information
1-1. Introduction
The service policy for this instrument is replace-
ment of defective assemblies. This service guidel
contains information for finding a defective
assembly by testing and servicing the HP 16532A
1 GSa/s Oscilloscope Module. Also included in
the guide are installation procedures and a list of
recommended test equipment. This guide is
divided into six sections as follows:
I - General Information
II - Installation
III - Performance Tests
IV - Calibration
V- Replaceable Parts
VI - Service
The General Information Section includes safety
requirements, a product description, and a list of
accessories supplied and of accessories available.
Also included are tables listing specifications and
operating characteristics, and a list of
recommended test equipment.
To complete the service documentation for your
HP 16500 Logic Analysis System, place this
service manual in the 3-ring binder with your other
logic analysis system Service Guides.
1-2. Module Covered by This Manual
The information covered in this manual is for the
HP 16532A 1 GSa/s Oscilloscope Module.
1-3. Safety Requirements
Specific warnings, cautions, and instructions are
placed wherever applicable throughout the
manual. These must be observed during all
phases of operation, service, and repair of the
module. Failure to comply with them violates
safety standards of design, manufacture, and
intended use of this module.
Hewlett-Packard assumes no liability for the failure
of the customer to comply with these safety
requirements.
HP 16532A - General Information
1-1
1-4. Product Description
The HP 16532A Oscilloscope Module is a 1 GSa/s
digitizing oscilloscope. Some of the main features
are:
• 1 GSa/s digitizing for 250 MHz bandwidth sin-
gle shot analysis.
• 8000 samples per channel.
• Automatic pulse parameters, displays time be-
tween markers, acquires until specified time
between markers is captured, performs statis-
tical analysis on time between markers.
• Lightweight miniprobes.
1-5. Accessories Supplied
The following accessories are supplied with the
HP 16532A Oscilloscope module. Quantity one
unless shown otherwise.
• User’s Reference and Programming Refer-
ence set
• Service Guide
• HP 10441A 10:1 probes, Qty 2
• Right angle BNC adapters, Qty 2
• BNC miniprobe adapter
• Master-slave trigger cable
• Operating System Software
1-6. Accessories Available
• 10:1, 100:1, 50 Ω, 10 pf resistive divider probe
set, 1.2 m (HP 10020A)
• BNC-to-BNC cable, 1.2 m (HP 10503A)
• 24-pin IC test clip (HP 10211A)
• BNC-to-BNC ac coupling capacitor
(HP 10240B)
10:1 Probe:
• 1-MΩ, 7.5-pf miniprobe, 1 m (HP 10435A)
1:1 Probes:
• 40-pf miniprobe, 1 m (HP 10438A)
• 64-pf miniprobe, 2 m (HP 10439A)
• 50-Ω miniprobe, 2 m (HP 10437A)
100:1 Probe:
• 10-MΩ, 2.5-pf miniprobe, 2 m (HP 10440A)
1-7. Specifications
Module specifications are listed in table 1-1. These
specifications are the performance standards
against which the module is tested.
1-8. Operating Characteristics
Table 1-2 is a listing of the module operating
characteristics. The operating characteristics are
not specifications, but are the typical operating
characteristics included as additional information
for the user.
1-9. Recommended Test
Equipment
Equipment required to test and maintain the
HP 16532A Oscilloscope Module is listed in table
1-3. Other equipment may be substituted if it
meets or exceeds the critical specifications listed
in the table.
HP 16532A - General Information
1-2
Table 1-1. HP 16532A Specifications
HP 16532A SPECIFICATIONS
Type: 2-channel simultaneous acquisition
Bandwidth
(*,1)
: dc to 250 MHz (real time, dc-coupled)
Maximum Sample Rate: 1 Gigasample per second
Number of Channels: 2, 4, 6, 8 simultaneous channels using the same time base setting(s)
OR
Up to 10 channels with independent time bases for each pair of channels.
Up to 18 channels with the HP 16501A Expansion Frame.
Rise Time
(2)
: 1.4 ns
ADC: 8-bit real time
Vertical Resolution: 8 bits over 4 vertical divisions (± 0.4%)
Waveform Record Length: 8000 points
Time Interval Measurement Accuracy
(*,3,6)
: ± [(0.005% X ∆t)+ (2 x 10
−
6
x delay setting)+ 150 ps]
Vertical (dc) Gain Accuracy
(4)
: ± 1.5% of full scale
DC Offset Accuracy
(*)
: ± (1.0% of channel offset + 2.0% of full scale)
Voltage Measurement Accuracy
(*,5)
: ± [(1.5% of full scale + offset accuracy) + (0.008 x V/div)]
Trigger Sensitivity
(*)
: dc to 50 MHz: 0.063 x full scale, 50 to 250 MHz: 0.125 x full scale
Input Coupling: 1 MΩ: ac and dc, 50 Ω: dc only
Input R (selectable)
(*)
: 1 MΩ: ± 1%, 50 Ω: ± 1%
Input C: Approximately 7 pF
___________________________________________________________________________________________
NOTES:
* Specifications (valid within ± 10° C of auto-calibration temperature, excluding bandwidth− see
note 1 for bandwidth specification.)
1. Upper bandwidth reduces by 2.5 MHz for every degree C above 35° C.
2. Rise time is calculated from the formula:
t
r
=
0.35
bandwidth
3. Specification applies to the maximum sampling rate. At lower rates, the specification is:
± [(0.005% x ∆t)+ (2 x 10
-6
x delay setting) + (0.15 x sample interval)] for bandwidth limited signals
(t
r
= 1.4 x sample interval). Sample interval is defined as
1
sample
rate
4. Vertical gain accuracy decreases 0.08% per degree C from software calibration temperature.
5. Digitizing level = (#vertical divisions)(
1
2
)(
1
LSB
), where LSB= 2
# bits in ADC
6. The Time Interval Measurement Accuracy deteriorates across multiple modules connected as one
unit with each added module.
HP 16532A - General Information
1-3
Table 1-2. HP 16532A Operating Characteristics
HP 16532A OPERATING CHARACTERISTICS
VERTICAL (at BNC)
Vertical Sensitivity Range (1:1 Probe):
4 mV/div to 10 V/div in 1-2-4 increments
DC Offset Range (1:1 Probe):
Vertical Sensitivity Available Offset
4 mV/div to 100 mV/div ± 2 V
> 100 mV/div to 500 mV/div ± 10 V
> 500 mV/div to 2.5 V/div ± 50 V
> 2.5 V/div to 10 V/div ± 250 V
Probe Factors:
Any integer ratio from 1:1 to 1000:1
Maximum Safe Input Voltage:
1 MΩ: ± 250 V [dc + peak ac (< 10 KHz)]
50 Ω: ± 5 VRMS
Channel-to-Channel Isolation:
dc to 50 MHz: 40 dB, 50 MHz to 250 MHz: 30 dB
TIMEBASE
Range:
1 ns/div to 5 s/div
Resolution:
20 ps
Delay Pre-trigger Range:
Time/div Setting Available Delay
1 µs to 5 s/div -8 x (s/div)
1 ns to 500 ns/div -4 µs
Delay Post-trigger Range:
Time/div Setting Available Delay
100 ms to 5 s/div 2.5 ks
1 µs to 50 ms/div 33,500 x (s/div)
1ns to 500 ns/div 16.7 ms
HP 16532A - General Information
1-4
Table 1-2. HP 16532A Operating Characteristics (cont.)
TRIGGERING
Trigger Level Range:
Within display window (full scale and offset)
Trigger Modes:
Immediate:
Triggers immediately after arming condition is met.
Edge:
Triggers on rising or falling edge of any internal channel or external trigger,
count adjustable from 1 to 32,000.
Pattern:
Triggers on entering, exiting, duration greater than, duration less than, and
duration in range for a specified pattern of both internal channels and external
trigger, count adjustable from 1 to 32,000. Duration time range is adjustable
from 20 ns to 160 ms in 10 ns steps. Duration accuracy is approximately ± 3%
± 2 ns.
Auto-Trigger:
If enabled, the module will self-trigger if no trigger condition is found within
approximately 50 ms after arming.
Events Delay:
The trigger can be set to occur on the nth edge or pattern, as specified by the
user. The number of events (n) can be set from 1 to 32,000 events. Maximum
count frequency is 70 MHz.
Intermodule:
Arms another measurement module or triggers the rear panel BNC.
External:
If enabled, the oscilloscope will trigger on an ECL level signal (-1.3 V).
DIGITIZER
Resolution: 8 bits (1 part in 256)
Digitizing Rate: Up to 1 Gigasample per second
Digitizing Technique: Real-time digitizing; each 8000 samples are acquired on a single acquisition
Acquisition Memory Size: 8000 samples per channel
HP 16532A - General Information
1-5
Table 1-2. HP 16532A Operating Characteristics (cont.)
WAVEFORM DISPLAY
Displayed Waveforms:
Eight waveform windows maximum, with scrolling across 96 waveforms.
Display Formats:
Waveforms can be displayed in an overlapping and/or non-overlapping format.
Display capability of A− B and A+ B is also provided.
Display Resolution:
500 points horizontal, 240 points vertical
DISPLAY MODES
Normal (Single):
New acquisition replaces old acquisition on screen.
Accumulate:
New acquisition is displayed in addition to previous acquisitions until screen is
erased.
Average:
New acquisitions are averaged with old acquisitions, with the updated waveform
displayed until erased. Maximum number of averages is 256.
Overlay:
Up to 8 acquired waveforms from separate channels can be overlayed in the
same display area.
Connect-the-dots:
Provides a display of the sample points connected by straight lines.
Waveform Reconstruction:
When there is insufficient data to fill every horizontal location, a post acquisition
reconstruction filter fills in missing data points for time base < 50 ns/div.
Waveform Math:
Display capability of A− B and A+ B functions is provided.
HP 16532A - General Information
1-6
Table 1-2. HP 16532A Operating Characteristics (cont.)
MEASUREMENT AIDS
Time Markers:
Two vertical markers, X and O, are provided for measurements of time and
voltage. Capabilities are: measure voltage at point where X and O markers
cross each analog waveform; measure time from trigger to X, trigger to O, and
X to O; automatic marker placement by specifying percentage of edge, edge
number, and rising or falling edge type; run until X to O > than, < than, in
range, and not in range provides selective event search; X to O statistics (mean,
max, and min) provide analysis of time interval deviation.
Automatic Search:
Searches for a percentage or an absolute voltage level at a positive or negative
edge, count adjustable from 1 to 100.
Auto Search Statistics:
Mean, maximum, and minimum values for elapsed time from X to O markers for
multiple runs. Number of valid runs and total number of runs available.
Automatic Measurements:
The following pulse parameter measurements can be performed automatically:
Frequency Rise time + pulse width
Period Fall time − pulse width
V p-p Preshoot V
amplitude
*
Overshoot
(*gives difference between top and base
voltages; only available over bus)
Grid:
Graticules can be displayed in backround of waveform.
Voltage Markers:
Two horizontal markers, A and B, are provided for voltage measurements.
These markers can each be placed on any acquisition channel trace. A delta
voltage display provides delta V between markers on the same channel or
between two channels.
HP 16532A - General Information
1-7
Table 1-2. HP 16532A Operating Characteristics (cont.)
SETUP AIDS
Autoscale:
Autoscales the vertical and horizontal ranges, offset, and trigger level to display
the input signals. Requires a frequency between 50 Hz and 250 MHz.
Presets:
Scales the vertical range, offset, and trigger level to predetermined values for
displaying ECL or TTL waveforms.
Calibration:
Vertical, trigger, delay, and all defaults. Calibration factors stored in NV-RAM on
the circuit board.
Probe Compensation Source:
External BNC supplies a square wave approximately 0.0 mV to -800 mV into the
open circuit at approximately 1000 Hz.
OPERATING ENVIRONMENT
Temperature:
Instrument, 0° to 55° C (+ 32° to 131° F). Probes and cables, 0° to 65° C (+ 32°
to 149° F).
Humidity:
Instrument, up to 95% relative humidity at + 40° C (+ 104° F). Recommended
disk media, 8% to 80% relative humidity at + 40° C (+ 104° F).
Altitude:
To 4600 m (15,000 ft).
Vibration
Operation:
Random vibration 5-500 Hz, 10 minutes per axis, ~ 0.3 g (rms).
Non-operating:
Random vibration 5-500 Hz, 10 minutes per axis, ~ 2.41 g (rms); and swept
sine resonant search, 5-500 Hz, 0.75 g (0- peak), 5 minute resonant dwell @ 4
resonances per axis.
HP 16532A - General Information
1-8
Table 1-3. Recommended Test Equipment
EQUIPMENT CRITICAL SPECIFICATIONS RECOMMENDED MODEL USE*
Signal Generator Frequency: 1 - 250 MHz at approx .
170 mV RMS resolution
Output Accuracy: ± 1 dB
1 MHz time base accuracy 0.25 ppm
HP 8656B Option 001 P
Dc Power Supply Range: − 35.000 to + 35.000 Vdc, ± 1 mV HP 6114A P
Digital Multimeter 0.1 mV resolution
Accuracy: better than 0.005%
Resistance measurement: better than
0.25% accuracy
HP 3458A P
Power Meter/Sensor 1 - 250 MHz ± 3% accuracy HP 436/8482A P
Power Splitter Outputs differ by < 0.15 dB HP 11667B P
Blocking Capacitor 50 Ω BNC (m-to-f), 0.18 µF, ± 200 V HP 10240B P
Adapter 50 Ω BNC (m)(m) HP 1250-1236 C
Adapter 50 Ω Type N(m)-to-BNC(m) HP 1250-0082 P
Adapter 50 Ω BNC(f)-to-Dual Banana Plug HP 1251-2277 P
Adapter 50 Ω Type N(m)-to-BNC(f) HP 1250-0780 P
Tee Adapter 50 Ω BNC (m)(f)(f) HP 1250-0781 P,C
Cable Type N(m-to-m) 24-inch HP 11500B P
Cable 50 Ω BNC (m-to-m) 48-inch HP 10503A P
Cable 50 Ω BNC (m-to-m) 9 inch HP 10502A C
Cable 50 Ω BNC-SMB HP 16532-61601 C
*P= Performance Tests C= Calibrations
HP 16532A - General Information
1-9
1-10
TABLE OF CONTENTS
Installation
2-1. Introduction ......................................................................................................................... 2-1
2-2. Initial Inspection .................................................................................................................. 2-1
2-3. Preparation for Use ............................................................................................................. 2-1
2-4. Power Requirements .......................................................................................................... 2-1
2-5. Safety Requirements ........................................................................................................... 2-1
2-6. Module Installation .............................................................................................................. 2-1
2-7. Operating Environment ...................................................................................................... 2-4
2-8. Storage ................................................................................................................................ 2-4
2-9. Packaging ........................................................................................................................... 2-4
2-10. Tagging for Service ............................................................................................................. 2-4
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90

HP 1GSa/s User manual

Category
Multimeters
Type
User manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI