Brooks GF40, GF80 User manual

Type
User manual

Brooks GF40, GF80: The Brooks® GF40/GF80 Series with EtherCAT® Communications are thermal mass flow controllers and meters designed for precise and reliable flow measurement and control of gases. These devices offer advanced features and capabilities that make them ideal for a wide range of industrial applications, including semiconductor manufacturing, pharmaceutical production, and laboratory research.

With EtherCAT® connectivity, the GF40/GF80 Series can be easily integrated into complex automation systems, enabling real-time data acquisition and control. They feature high accuracy and repeatability, ensuring precise flow measurement and control even in challenging conditions. The devices also provide comprehensive diagnostic and alarm functions, allowing for quick identification and resolution of any issues that may arise.

Brooks GF40, GF80: The Brooks® GF40/GF80 Series with EtherCAT® Communications are thermal mass flow controllers and meters designed for precise and reliable flow measurement and control of gases. These devices offer advanced features and capabilities that make them ideal for a wide range of industrial applications, including semiconductor manufacturing, pharmaceutical production, and laboratory research.

With EtherCAT® connectivity, the GF40/GF80 Series can be easily integrated into complex automation systems, enabling real-time data acquisition and control. They feature high accuracy and repeatability, ensuring precise flow measurement and control even in challenging conditions. The devices also provide comprehensive diagnostic and alarm functions, allowing for quick identification and resolution of any issues that may arise.

Installation and Operation Manual
X-DPT-EtherCAT-GF40-GF80-MFC-eng
Part Number: 541B170AAG
August, 2012
Brooks
®
GF40/GF80 EtherCAT
®
EtherCAT
®
Supplemental Manual for Brooks
®
GF40/GF80 Series Mass Flow Controllers
and Meters
Brooks
®
GF40/GF80 Series
with EtherCAT
®
Communications
Installation and Operation Manual
X-DPT-EtherCAT-GF40-GF80-MFC-eng
Part Number: 541B170AAG
August, 2012
Brooks
®
GF40/GF80 EtherCAT
®
Installation and Operation Manual
X-DPT-EtherCAT-GF40-GF80-MFC-eng
Part Number: 541B170AAG
August, 2012
Brooks
®
GF40/GF80 EtherCAT
®
Dear Customer,
We recommend that you read this manual in its entirety as this will enable efficient and proper use of the
EtherCAT
®
thermal mass flow controllers and meters. Should you require any additional information concerning the
EtherCAT thermal mass flow controllers and meters, please feel free to contact your local Brooks Sales and Service
Office; see back cover for contact information, or visit us on the web at www.BrooksInstrument.com. We appreciate
this opportunity to service your fluid measurement and control requirements, and trust that we will be able to provide
you with further assistance in future.
Yours sincerely,
Brooks Instrument
Installation and Operation Manual
X-DPT-EtherCAT-GF40-GF80-MFC-eng
Part Number: 541B170AAG
August, 2012
Brooks
®
GF40/GF80 EtherCAT
®
THIS PAGE WAS
INTENTIONALLY
LEFT BLANK
i
ContentsInstallation and Operation Manual
X-DPT-EtherCAT-GF40-GF80-MFC-eng
Part Number: 541B170AAG
August, 2012
Brooks
®
GF40/GF80 EtherCAT
®
Paragraph Page
Number Number
Section 1 General Information
1.1 Introduction ...................................................................................................................................1-1
Section 2 Definition of Terms
2.1 Definition of Terms ........................................................................................................................ 2-1
Section 3 Before Starting
3.1 Background & Assumptions...........................................................................................................3-1
3.2 Numbers ........................................................................................................................................3-1
Section 4 Quick Start
4.0 Quick Start.....................................................................................................................................4-1
4.1 Master Hardware ........................................................................................................................... 4-1
4.2 Physical Interfaces........................................................................................................................ 4-1
4.2.1 Power Supply and Analog I/O........................................................................................................4-1
4.2.2 RUN and MOD LEDs .....................................................................................................................4-3
4.2.3 EtherCAT MFC Slave Hardware.....................................................................................................4-3
Section 5 Slave Configuration
5.1 Introduction ...................................................................................................................................5-1
5.2 Outputs (Master Side)....................................................................................................................5-1
5.3 Inputs (Master Side) ......................................................................................................................5-2
5.4 COE Online Attributes ....................................................................................................................5-3
5.4.1 Device Attributes ............................................................................................................................5-3
5.4.2 Flow Sensor...................................................................................................................................5-3
5.4.2.1 Flow Sensor Zero Adjust................................................................................................................5-4
5.4.2.2 Flow Sensor Status ........................................................................................................................5-4
5.4.3 Temperature Sensor ......................................................................................................................5-5
5.4.3.1 Temperature Sensor Status............................................................................................................5-6
5.4.4 Setpoint Controller .........................................................................................................................5-6
5.4.4.1 Setpoint Controller Status ..............................................................................................................5-7
5.4.5 Valve Actuator Attributes ................................................................................................................5-8
5.4.6 Service Transfer Attributes ............................................................................................................5-9
5.4.7 Calibration Object Attributes...........................................................................................................5-9
5.4.8 Exceptions .................................................................................................................................... 5-10
5.4.8.1 Alarm- and Warning Details..............................................................................................
............5-10
5.4.8.2 Exception Status ...........................................................................................................................5-10
Warranty, Local Sales/Service Contact Information....................................................................... Back Cover
ii
Contents
Installation and Operation Manual
X-DPT-EtherCAT-GF40-GF80-MFC-eng
Part Number: 541B170AAG
August, 2012
Brooks
®
GF40/GF80 EtherCAT
®
Figure Figure
Number Number
4-1 EtherCAT Label on Cover .............................................................................................................. 4-1
4-2 M8 Male Device Connector Pin Layout, Pin Side View ................................................................. 1-4
4-3 M8 Female Mating Cable ............................................................................................................... 4-2
4-4 M8 Female Mating Cable Connector Pin Layout ........................................................................... 4-2
5-1 Output PDOs ................................................................................................................................. 5-1
5-2 Input PDOs.................................................................................................................................... 5-2
5-3 Device Attributes............................................................................................................................5-3
5-4 Flow Sensor Attributes ...................................................................................................................5-3
5-5 Flow Sensor Zero Adjust Attribute .................................................................................................. 5-4
5-6 Flow Sensor Status Attributes ........................................................................................................ 5-4
5-7 Temperature Sensor Attributes....................................................................................................... 5-5
5-8 Temperature Sensor Status Attributes ............................................................................................ 5-6
5-9 Setpoint Controller Attributes ......................................................................................................... 5-6
5-10 Setpoint Controller Status Attribute ................................................................................................ 5-7
5-11 Valve Actuator Attributes ................................................................................................................5-8
5-12 Valve Actuator Status Attributes ..................................................................................................... 5-8
5-13 Service Transfer Attributes ............................................................................................................ 5-9
5-14 Calibration Object Attributes........................................................................................................... 5-9
Table Table
Number Number
4-1 Pin Labeling of M8 Male Device and Female Mating Cable Connector.......................................... 4-2
4-2 Wire Labeling of M8 Female Mating Connector ............................................................................ 4-3
4-3 M8 Female Mating Cable Part Numbers ........................................................................................ 4-3
4-4 RUN LED Specification ................................................................................................................. 4-3
4-5 MOD LED Specification................................................................................................................. 4-3
5-1 Valve Override Values (vdOverride) .............................................................................................. 5-11
5-2 Flow Data Units (fmUnits) ............................................................................................................. 5-11
5-3 Temperature Data Units (tmUnits) ................................................................................................. 5-11
5-4 Setpoint Control Mode (fcControlMode)........................................................................................5-11
5-5 Valve Drive Data Units (vdUnits) ................................................................................................... 5-11
1-1
Section 1 Introduction
Installation and Operation Manual
X-DPT-EtherCAT-GF40-GF80-MFC-eng
Part Number: 541B170AAG
August, 2012
Brooks
®
GF40/GF80 EtherCAT
®
1.1 Introduction
Many applications of Flow Controllers/Meters are moving to increasing use
of automation. Automation comes in many forms: PLC’s (Programmable
Logic Controllers such as the Siemens S7 300/4000), DCS’s (Distributed
Control Systems, such as Emerson’s Digital V), PC based solutions
(National Instrument’s Labview
TM
) and Ethernet based field buses. Digital
communications from these varied systems and the devices they measure
and control, are a very effective means of not only accomplishing more
effective and rapid system integration, but also providing greatly improved
system diagnostics and maintainability. EtherCAT is an Ethernet based
communication system and is known for its high cycle time and cost
efficient cabling and master application solutions. Brooks Instrument now
introduces the EtherCAT interface on its GF Series platform.
1-2
Section 1 Introduction
Installation and Operation Manual
X-DPT-EtherCAT-GF40-GF80-MFC-eng
Part Number: 541B170AAG
August, 2012
Brooks
®
GF40/GF80 EtherCAT
®
THIS PAGE WAS
INTENTIONALLY
LEFT BLANK
2-1
Section 2 Definition of Terms
Installation and Operation Manual
X-DPT-EtherCAT-GF40-GF80-MFC-eng
Part Number: 541B170AAG
August, 2012
Brooks
®
GF40/GF80 EtherCAT
Abbreviation Description
MFC/MFM Mass Flow Controller/Meter device
MSB Most Significant Bit
LSB Least Significant Bit
PDO Process Data Object
SDO Service Data Object
CoE CanOpen on EtherCAT
ESI EtherCAT Slave Information (device description in XML format)
ESC EtherCAT Slave Controller
2.1 Definition of Terms
2-2
Section 2 Definition of Terms
Installation and Operation Manual
X-DPT-EtherCAT-GF40-GF80-MFC-eng
Part Number: 541B170AAG
August, 2012
Brooks
®
GF40/GF80 EtherCAT
THIS PAGE WAS
INTENTIONALLY
LEFT BLANK
3-1
Section 3 Before Starting
Installation and Operation Manual
X-DPT-EtherCAT-GF40-GF80-MFC-eng
Part Number: 541B170AAG
August, 2012
Brooks
®
GF40/GF80 EtherCAT
®
3 Before Starting
3.1 Background & Assumptions
This manual is a supplement to the Brooks GF40/GF80 Series installation
and operation manual. It is assumed that the owner of this EtherCAT MFC/
MFM is thoroughly familiar with the theory and operation of this device. If
not, it is recommended that the owner read the installation and operation
manual first before continuing with this supplement.
This manual assumes basic knowledge and understanding of EtherCAT (its
topology and its method of logically accessing the data or parameters
contained within the device). This manual is not intended to be a
replacement to the EtherCAT specifications. It is recommended but not
required for the purposes of this manual, that the user obtains a copy of the
EtherCAT specifications (www.ethercat.org).
This manual does not make any assumptions about any particular
manufacturer of equipment or custom software used by the user to
communicate with the Brooks device, but assumes the user has thorough
understanding of such equipment and any configuration software.
Application Notes and FAQ’s are available at the Brooks Instrument web site
(www.BrooksInstrument.com).
3.2 Numbers
Numeric values used throughout this manual will be clearly denoted as to
the base numeric system it represents. All hexadecimal numbers (base 16)
will be prefixed with a 0x, like 0xA4. All binary numbers (base 2) will be
suffixed with a b, like 1001b. All other numbers not annotated this way will
be assumed decimal (base 10).
3-2
Section 3 Before Starting
Installation and Operation Manual
X-DPT-EtherCAT-GF40-GF80-MFC-eng
Part Number: 541B170AAG
August, 2012
Brooks
®
GF40/GF80 EtherCAT
®
THIS PAGE WAS
INTENTIONALLY
LEFT BLANK
4-1
Section 4 Quick Start
Installation and Operation Manual
X-DPT-EtherCAT-GF40-GF80-MFC-eng
Part Number: 541B170AAG
August, 2012
Brooks
®
GF40/GF80 EtherCAT
®
4 Quick Start
This section assumes the owner of the Digital Series device has a fully
operational and trouble-free communications network with appropriate
power supplies. This section also assumes that an EtherCAT master
application is connected to the network capable of PDO and mailbox data
communication. Both types of data communication modes are supported by
the Brooks GF40/GF80 EtherCAT device
4.1 Master Hardware
Various companies provide EtherCAT master applications, e.g. TwinCAT
from Beckhoff, or offer EtherCAT master stacks to develop a master
application, e.g. Acontis. A PC can be used to run most EtherCAT master
applications but needs dedicated Ethernet hardware to support the high
cycle times and kernel mode operation of the master application, see
wwww.beckhoff.com. Screendumps of master applications used in this
manual are taken from the EtherCAT configurator tool from Beckhoff.
4.2 Physical Interfaces
The available physical interfaces on the EtherCAT device are listed below:
• 5 pin M8 threaded male connector for power and analog I/O, indicated by
PWR
• IN and OUT ports with RJ45 connectors
• ZERO push button, refer to the GF40/80 Series installation and operation
manual for more details
• 2.5mm female jack for RS485 diagnostics port indicated by DIAG, refer to
the GF40/GF80 Series installation and operation manual for more details
Figure 4-1 EtherCAT Label on Cover
4.2.1 Power Supply and Analog I/O
Power needs to be supplied via the M8 connector. This connector also
provides access to analog I/O signals, see Table 4-1.
4-2
Section 4 Quick Start
Installation and Operation Manual
X-DPT-EtherCAT-GF40-GF80-MFC-eng
Part Number: 541B170AAG
August, 2012
Brooks
®
GF40/GF80 EtherCAT
®
M8 mating cables can be purchased as a second line item, details given below.
Figure 4-3 M8 Female Mating Cable
Pin Label Function at Remote Connector
P_COM Power Supply Common
+VPWR Positive Power Supply Voltage
V_AUX_OUT Flow Output 0-5V
AGND Analog I/O Common
V_AUX_IN Auxiliary Input 0-5/10V for Future Use
Table 4-1 Pin Labeling of M8 Male Device and Female Mating Cable Connector
Figure 4-2 M8 Male Device Connector Pin Layout, Pin Side View
Figure 4-4 M8 Female Mating Cable Connector Pin Layout
Mating Side Female Cable Connector
4-3
Section 4 Quick Start
Installation and Operation Manual
X-DPT-EtherCAT-GF40-GF80-MFC-eng
Part Number: 541B170AAG
August, 2012
Brooks
®
GF40/GF80 EtherCAT
®
Wire Color Wire Label Function at Remote Connector
Blue P_COM Power Supply Common
Brown +VPWR Positive Power Supply Voltage
Black V_AUX_OUT Flow Output 0-5V
White AGND Analog I/O Common
Grey V_AUX_IN Auxiliary Input 0-5/10V for Future Use
4.2.2 RUN and MOD LEDs
The device supports a RUN and MOD LED to indicate the status of network
communication and the device. The RUN LED will indicate the following:
Table 4-2 Wire Labeling of M8 Female Mating Cable Connector
Supplier Part Number Description
Brooks 124X049AAA M8 Mating Cable 2m
Instrument 124X050AAA M8 Mating Cable 5m
Table 4-3 M8 Female Mating Cable Part Numbers
The MOD LED will indicate the following:
Flash Code Description
Off The device is in state INIT
Blinking The device is in state PRE-OPERATIONAL
Single Flash The device is in state SAFE-OPERATIONAL
On The device is in state OPERATIONAL
Flickering The device is booting and has not yet entered the INIT state
Triple Flash User can set this state from the master to locate the
specific slave
Table 4-4 RUN LED Specification
4.2.3 EtherCAT MFC Slave Hardware
The main parts of the EtherCAT MFC are:
• Standard Ethernet Physical Layer Components
• EtherCAT Slave Controller (ESC) and EEPROM (ESC configuration data
and application specific data)
• For intelligent slaves with an application controller: Host controller
Flash Code Description
Flashing The device is in the Self-Test mode
Red/Green
Solid Green All self-tests have passed. No faults have been detected
Flashing Red A recovering alarm has been detected
Flashing Green A recoverable warning has been detected
Solid Red An unrecoverable fault has occured
Table 4-5 MOD LED Specification
4-4
Section 4 Quick Start
Installation and Operation Manual
X-DPT-EtherCAT-GF40-GF80-MFC-eng
Part Number: 541B170AAG
August, 2012
Brooks
®
GF40/GF80 EtherCAT
®
THIS PAGE WAS
INTENTIONALLY
LEFT BLANK
5-1
Section 5 Slave Configuration
Installation and Operation Manual
X-DPT-EtherCAT-GF40-GF80-MFC-eng
Part Number: 541B170AAG
August, 2012
Brooks
®
GF40/GF80 EtherCAT
®
5.1 Introduction
Based on the information provided by the EtherCAT Slave Information file
(ESI, device description in XML format) and/ or the EEPROM, master
applications are able to configure the EtherCAT network.
For the EtherCAT network configuration of the GF40/80 Series devices the
following ESI file is provided on the Brooks website
(www.BrooksInstrument.com):
‘Brooks GF04x.xml’ – GF40/80 Series Mass Flow Controller/Meter
5.2 Outputs (Master Side)
The request message, sent from master to slave, consists of the fields
indicated in Figure 5-1, these fields will be described in the sections below.
Figure 5-1 Output PDOs
5-2
Section 5 Slave Configuration
Installation and Operation Manual
X-DPT-EtherCAT-GF40-GF80-MFC-eng
Part Number: 541B170AAG
August, 2012
Brooks
®
GF40/GF80 EtherCAT
®
5.3 Inputs (Master Side)
Figure 5-2 Input PDOs
5-3
Section 5 Slave Configuration
Installation and Operation Manual
X-DPT-EtherCAT-GF40-GF80-MFC-eng
Part Number: 541B170AAG
August, 2012
Brooks
®
GF40/GF80 EtherCAT
®
5.3 COE Online Attributes
5.3.1 Device Attributes
Figure 5-3 Device Attributes
5.3.2 Flow Sensor
Figure 5-4 Flow Sensor Attributes
5-4
Section 5 Slave Configuration
Installation and Operation Manual
X-DPT-EtherCAT-GF40-GF80-MFC-eng
Part Number: 541B170AAG
August, 2012
Brooks
®
GF40/GF80 EtherCAT
®
5.3.2.1 Flow Sensor Zero Adjust
5.3.2.2 Flow Sensor Status
Figure 5-5 Flow Sensor Zero Adjust Attribute
Figure 5-6 Flow Sensor Status Attributes
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30

Brooks GF40, GF80 User manual

Type
User manual

Brooks GF40, GF80: The Brooks® GF40/GF80 Series with EtherCAT® Communications are thermal mass flow controllers and meters designed for precise and reliable flow measurement and control of gases. These devices offer advanced features and capabilities that make them ideal for a wide range of industrial applications, including semiconductor manufacturing, pharmaceutical production, and laboratory research.

With EtherCAT® connectivity, the GF40/GF80 Series can be easily integrated into complex automation systems, enabling real-time data acquisition and control. They feature high accuracy and repeatability, ensuring precise flow measurement and control even in challenging conditions. The devices also provide comprehensive diagnostic and alarm functions, allowing for quick identification and resolution of any issues that may arise.

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI