• This Affirma Lubricant Analysis System can be used to measure aqueous and organic liquids
and light oils such as transmission and hydraulic engine oils. (Not intended for viscous
materials such as gels, greases and heavy gear oils.)
• Use one of the sample workflows in the Affirma FTRN In-Service Lubricant group to run a
sample analysis.
Note Sample workflows based on a differential (or subtraction) require a reference
spectrum. Use the Collect Reference workflow in the FTRN Setup category to acquire a
spectrum from your reference material.
• To inject a sample for measurement, follow these steps:
CAUTION
Splash hazard. Avoid applying excessive pressure to the flow-through cell
(via the syringe). Sudden leaks can cause a skin and eyes splash hazard.
Draw 1 t o 2 ml of sample into a clean 10 ml Luer-Lok syringe. Screw the end of the syringe
into the adapter on the flow cell inlet tubing and slowly inject enough sample to fill the cell (0.3
to 0.4 ml). Continue injecting sample until it starts to flow into the outlet tubing. If bubbles
appear in the outlet tubing, continue injecting sample until no more bubbles are expelled.
Start the measurement. Do not disconnect the syringe until the measurement has completed.
• Always clean the cell immediately after you finish the analysis. See Clean the flow-through
cell for details.
Clean the flow-through cell
Make sure the flow-through transmission cell is clean before and after taking sample
measurements with the Affirma Lubricant Analysis system.
Materials required:
• Small volume of heptane or isooctane
• Clean 10 ml Luer-Lok syringe (latex and silicone-oil free)
16
Nicolet Summit X FTRN Quick Start Guide
Clean the flow-through cell