Operation
Pinless Moisture Measurements
1. Press the power button to turn the meter on.
2. As described below, you must calibrate the meter before first use.
3. To calibrate, press the MEAS button immediately after powering on the meter. The display will
audibly beep and visually flash the CAL display icon. Keep the measurement sensor at least 3
to 4” (8 to 10 cm) from any object during calibration.
4. When the calibration is complete the beeping and flashing stop. The calibration reading must
not exceed 0.5. If so, re-calibrate ensuring that hands or other objects are kept clear of the
measurement sensor.
5. The meter should be calibrated whenever it is moved from one measurement area to another.
6. To take a measurement, hold the meter and position the sensor so that it is very near or
directly touching the surface under test. The meter can sense moisture to a depth of 0.79 to
1.57” (20 to 40mm).
7. IMPORTANT NOTE: The position of the hand can greatly affect a reading. Do not move or
reposition the hand that holds the meter while taking measurements or when calibrating the
unit.
8. The LCD indicates a relative moisture reading in numeric digits from 0.0 to 100.0 and with
programmable alarm alerts (DRY, RISK, and WET).
9. Press the MEAS button to HOLD (freeze) a reading on the LCD (the HOLD icon will appear on
the LCD). The meter will hold the reading for 30 seconds before automatically shutting the
meter down.
10. The MAX and MIN display areas indicate the highest and lowest readings encountered since
the measurement session was started.
11. If the battery symbol appears or if the meter does not switch ON, replace the batteries.
Alarm Settings
Two high alarm points can be set for Moisture measurements: A RISK Alarm (0 to 50 setting range
with a default setting of 30 units) and a WET Alarm (50 to 100 setting range with a default setting of
60 units).
Moisture Alarm Setup Procedure:
1. Press the MEAS button to place the meter in HOLD mode to change the Alarm settings.
2. Press the SET button. The RISK display icon will appear flashing on the LCD indicating that a
RISK alarm can now be programmed.
3. Use the up/down buttons to select the desired RISK Alarm value.
4. Press the SET button when the desired setting is displayed.
5. The WET icon will appear flashing on the LCD indicating that the WET Alarm is ready to be
programmed.
6. Use the up/down buttons to select the desired Wet Alarm setting.
7. Press SET to complete the programming.
8. Now, if the moisture measurement is higher than the RISK setting, the meter will beep once
every two seconds and if the moisture measurement is higher than the WET setting the meter
will beep four times per second.
9. Press the MEAS button to exit HOLD mode.