Zeiss Correlative Cryo Workflow User manual

Category
Microscopes
Type
User manual

Zeiss Correlative Cryo Workflow is a specialized system for preparing and examining frozen-hydrated samples in a cryo-electron microscope, enabling high-resolution imaging of biological structures in their near-native state. With its advanced cryo-preparation techniques and integrated imaging capabilities, it streamlines the cryo-workflow, enhancing the quality and efficiency of cryo-EM studies.

Zeiss Correlative Cryo Workflow is a specialized system for preparing and examining frozen-hydrated samples in a cryo-electron microscope, enabling high-resolution imaging of biological structures in their near-native state. With its advanced cryo-preparation techniques and integrated imaging capabilities, it streamlines the cryo-workflow, enhancing the quality and efficiency of cryo-EM studies.

Instruction Manual
ZEISS Cryo Trap
ZEISS Cryo Trap
Original Manual
Carl Zeiss Microscopy GmbH
Carl-Zeiss-Promenade 10
07745 Jena
Germany
www.zeiss.com/microscopy
Carl Zeiss Microscopy Ltd
1030 Cambourne Business Park
CB23 6DW Cambridge
United Kingdom
Document Name: ZEISS Cryo Trap Instruction Manual
Order Number: 346000-8166-000
Revision: 2
Language: en-US
Effective from: 11/2019
© 2019 Without the prior written consent of ZEISS, this document or any part of it must neither be translated nor reproduced or
transmitted in any form or by any means - including electronic or mechanic methods, by photocopying, recording or by any in-
formation or filing system. The right to make backup-copies for archiving purposes shall remain unaffected thereby. Any viola-
tions may be prosecuted as copyright infringements.
The use of general descriptive names, registered names, trademarks, etc. in this document does not imply that such names are
exempt from the relevant intellectual property laws and regulations and therefore free for general use. This shall also apply if
this is not specifically referred to. Software programs shall entirely remain the property of ZEISS. No program or subsequent up-
grade thereof may be disclosed to any third party, copied or reproduced in any other form without the prior written consent of
ZEISS, even if these copies or reproductions are destined for internal use at the customer's only, the only exception being one
single back-up copy for archiving purposes.
ZEISS reserves the right to make modifications to this document without notice.
ZEISS  Table of Contents
Table of Contents
1 About this Document............................................................................................ 4
2 Text Conventions and Link Types ......................................................................... 5
3 Definition of Terms ............................................................................................... 6
4 Safety .................................................................................................................... 7
4.1 Intended Use ........................................................................................................... 7
4.2 Safety Information ................................................................................................... 7
4.3 Warning Labels and Lights....................................................................................... 8
4.3.1 Warning Label on Cryo Trap Dewar.............................................................. 8
5 Description............................................................................................................ 9
5.1 Cryo Trap................................................................................................................. 9
5.2 Principle of Operation .............................................................................................. 10
5.3 Technical Data ......................................................................................................... 11
5.4 System Requirements............................................................................................... 11
5.5 Detector Compatibility ............................................................................................. 12
5.6 Stage Movement Limitiation .................................................................................... 12
6 Installation............................................................................................................ 13
7 Operation .............................................................................................................. 14
7.1 Filling the Dewar with Liquid Nitrogen ..................................................................... 15
7.2 Venting the Specimen Chamber............................................................................... 15
7.3 Dismounting the Cooling Plate................................................................................. 16
7.4 Installing the Cooling Plate ...................................................................................... 16
7.5 Defrosting the Dewar .............................................................................................. 16
8 De-installation and Disposal ................................................................................ 17
9 Troubleshooting.................................................................................................... 18
Instruction Manual ZEISS Cryo Trap | en-US | Rev. 2 | 346000-8166-000 3
1 About this Document ZEISS
1 About this Document
This instruction manual is considered to be part of the Cryo Trap. It offers topics which are specific
to the Cryo Trap. For information regarding the ZEISS field emission scanning electron microscope
(FESEM), please refer to the relevant manual before starting to work with the Cryo Trap.
Read the instructions carefully. Keep the instruction manual nearby the Cryo Trap and hand it over
to future owners of the instrument.
This instruction manual is designed for users who have been trained by an authorized Carl Zeiss
Microscopy expert to operate the Cryo Trap, which is built into the FESEM. Operators must not
deviate from the instructions provided in this document.
4 Instruction Manual ZEISS Cryo Trap | en-US | Rev. 2 | 346000-8166-000
ZEISS 2 Text Conventions and Link Types
2 Text Conventions and Link Types
The following text conventions and link types are used:
Text convention Meaning
Click Start.
Press the STANDBY button.
Press [Enter] on the keyboard.
The names of controls and important infor-
mation are shown in bold letters.
Press <Ctrl+Alt+Del> Press several keys on the keyboard simultane-
ously.
Select Tools > Goto Control Panel > Air-
lock.
Follow a path in the software.
Text input
Text output
§Text to be entered by the user
§Text displayed by the system
Programming and Macros Anything typed in literally during program-
ming, including, for example, macro codes,
keywords, data types, method names, vari-
ables, class names, and interface names.
Tab.1: Text convention
Link type Meaning
See: Text Conventions and Link Types [
}
5].See page 5 for further information on text
conventions and link types.
https://www.zeiss.com/corporate/int/
home.html
Link to a website on the internet.
Tab.2: Link types
Instruction Manual ZEISS Cryo Trap | en-US | Rev. 2 | 346000-8166-000 5
3 Definition of Terms ZEISS
3 Definition of Terms
The following terms are used in this manual:
Cooling Plate Aluminum plate close to specimen and objective lens
Cooling Rod Copper rod connecting the cooling plate to the dewar
Contamination Unwanted absorbation of volatile hydrocarbon compounds at speci-
men and instrument surfaces
Cryo Trap Reduces the contamination in the specimen chamber
Dewar Isolated can, containing the liquid nitrogen
FESEM Field emission scanning electron microscope
6 Instruction Manual ZEISS Cryo Trap | en-US | Rev. 2 | 346000-8166-000
ZEISS 4 Safety | 4.1 Intended Use
4 Safety
4.1 Intended Use
The Cryo Trap is used to cool down the area below the objective lens while the object chamber is
at vacuum. This decreases contamination by binding the volatile hydrocarbons.
The Cryo Trap is an option for the following ZEISS field emission scanning electron microscopes
(FESEM) with airlock:
§GeminiSEM 300
§GeminiSEM 450
§GeminiSEM 500
Using the Cryo Trap for any other purpose is not allowed and could be hazardous.
4.2 Safety Information
This manual only shows the hazards which are specific to the Cryo Trap. However the Cryo Trap
must be used together with a ZEISS electron microscope. For the risks regarding the operation of
the ZEISS electron microscope please refer to the relevant instruction manual before starting to
work with the Cryo Trap.
WARNING
Burning hazard
Danger of burns or frostbites when liquid nitrogen gets in contact with the skin.
4Wear proper protection equipment when handling the liquid nitrogen and Cryo Trap.
4Do not touch the cooling rod and the cooling plate inside the chamber as long as the de-
war is not empty.
4The dewar has to be in upright position. Never unscrew the dewar as long as liquid nitro-
gen is in it.
4Use suitable tools when filling up the dewar.
4The dewar must be empty and must have room temperature before exchanging the cool-
ing plate of the Cryo Trap.
WARNING
Explosion hazard
Explosion risk if the pressure inside the dewar rises due to evaporating liquid nitrogen.
4Only use the cover which is provided by ZEISS for the dewar filling hole.
4Regularly check that ice does not build up at or around the cover for the dewar filling hole.
WARNING
Suffocation hazard due to lack of oxygen
Danger of nitrogen accumulation in low-lying areas or subterranean rooms.
4Ensure the room is ventilated properly.
4Ensure the room is equipped with a ventilation system with bottom suction and air ex-
changer.
Instruction Manual ZEISS Cryo Trap | en-US | Rev. 2 | 346000-8166-000 7
4 Safety | 4.3 Warning Labels and Lights ZEISS
WARNING
Fire hazard
Fire risk due to concentration of oxygen in liquid nitrogen.
4Cover the dewar so that the evaporating nitrogen can escape, but the penetration of air is
inhibited and no flammable material can fall into the dewar.
4Store and transport liquid nitrogen only in suitable dewars.
4Do not smoke or have open fire in the same room the dewar is located.
NOTICE
Damage of components inside the chamber
In case the chamber is vented before the Cryo Trap is warmed up, the water from ambient air
will condense at cold surfaces inside the chamber.
4Do not vent the specimen chamber as long as the dewar is filled with liquid nitrogen.
4Before you vent the specimen chamber, wait until the dewar is empty and the cooling rod
and cooling plate are warmed up to room temperature.
4Always exchange the specimen via airlock.
4.3 Warning Labels and Lights
All locations that may pose specific hazards are marked with additional warning labels (pic-
tograms) on the Microscope System. These warning labels indicate potential hazards and form
part of this Instruction Manual. They are to be kept in clean and easily legible condition. Damaged
or illegible warning labels must be replaced immediately. Always consider all warning labels on
the Microscope System.
4.3.1 Warning Label on Cryo Trap Dewar
WARNING Liquid Nitrogen
Risk of freeze burns, suffocation or explosions.
8 Instruction Manual ZEISS Cryo Trap | en-US | Rev. 2 | 346000-8166-000
ZEISS 5 Description | 5.1 Cryo Trap
5 Description
5.1 Cryo Trap
Purpose For SEM imaging, especially for low kV imaging, hydrocarbon contamination is very critical as it
massively influences image quality. The Cryo Trap is used to minimize contamination in the speci-
men chamber caused by volatile hydrocarbon compounds.
Position The Cryo Trap is a purely mechanical device that is attached to the EDS3 port of the GeminiSEM
chamber. The Cryo Trap assembly includes a dewar vessel mounted outside, an adapter flange
connecting it to the specimen chamber, a thermoconducting rod plus a metal plate reaching near
the electron beam.
6
4
2
3
1
5
Fig.1: Cryo Trap assembly
1Blind Flange (currently not used) 2Lid (with borehole)
3Dewar 4Adapter Flange
5Cooling Plate 6Cooling Rod
Function The chamber vacuum provides the thermal isolation of a dewar vessel, which is filled with liquid
nitrogen. The extreme cold will be transferred into the specimen chamber by a copper rod that is
attached to the dewar bottom. A metal plate is attached to the tip of this copper rod. This metal
plate is cooled down to -170°C. It acts as a trap for hydrocarbon contamination. Hydrocarbon
compounds stick to the cold surface and are thus prevented from depositing on the specimen sur-
face whilst the system remains cold.
LN2 continuously evaporates and leaves the dewar through the bore hole of the lid as gaseous ni-
trogen.
There is no electronics and software integrated to the Cryo Trap. There is no measurement of the
final temperature, and there is no monitoring of the filling level of LN2 in the dewar.
Instruction Manual ZEISS Cryo Trap | en-US | Rev. 2 | 346000-8166-000 9
5 Description | 5.2 Principle of Operation ZEISS
5.2 Principle of Operation
There are always volatile hydrocarbon compounds
1
in the specimen chamber. These com-
pounds deposit on the specimen surface
2
. The specimen becomes contaminated. This may
influence the image quality.
1
2
The temperature of the cooling plate
3
decreases and the hydrocarbons change their physical
state, from volatile to solid. The compounds are deposited at the cooling plate. The cold surface
of the cooling plate acts as a trap for hydrocarbons and hence decreases the risk of contamina-
tion.
3
10 Instruction Manual ZEISS Cryo Trap | en-US | Rev. 2 | 346000-8166-000
ZEISS 5 Description | 5.3 Technical Data
5.3 Technical Data
Specification
Capacity 0.6 l
Size of dewar assembly overall height 334 mm, maximum diameter
152 mm
Cooling agent liquid nitrogen (LN2)
Cooling-down time 30 minutes to reach -130°C
45 minutes to reach -170°C
Temperature of cooling plate down to -180°C
Standing time After refilling the dewar the temperature of
the cooling plate stays below -130°C for
10…12 hours.
Warming-up time At regular system vacuum:
With a refilled dewar it takes about 2 days to
reach room temperature again (36 hours to
reach 20°C).
CryoWarmUp macro:
A macro allows to reduce the warming-up
time. With a refilled dewar it only takes
about 6 hours to reach room temperature
again (4.5 hours to reach 20°C).
System vacuum Improvement of an approximate factor of 2
5.4 System Requirements
§Airlock
§Liquid nitrogen, has to be provided by the customer
Instruction Manual ZEISS Cryo Trap | en-US | Rev. 2 | 346000-8166-000 11
5 Description | 5.5 Detector Compatibility ZEISS
5.5 Detector Compatibility
All the detectors, including the moveable ones, are compatible. BsD, NanoVP and CC should be
mounted on port 66 mm MP1 or MP2.
IMPORTANT: The EDS detector must be mounted on port EDS1.
Fig.2: Ports of GeminiSEM 300/500
5.6 Stage Movement Limitiation
With cooling plate in position, untilted stage movement is uncompromised.
A stage tilt of 45° is possible, but the possible range of WD strongly depends on sample/sample
holder.
Fig.3: Tilted stage with cooling plate and detectors
12 Instruction Manual ZEISS Cryo Trap | en-US | Rev. 2 | 346000-8166-000
ZEISS 6 Installation
6 Installation
The Cryo Trap is either factory-installed or retrofitted by authorized ZEISS service personnel.
Instruction Manual ZEISS Cryo Trap | en-US | Rev. 2 | 346000-8166-000 13
7 Operation |  ZEISS
7 Operation
WARNING
Burning hazard
Danger of burns or frostbites when liquid nitrogen gets in contact with the skin.
4Wear proper protection equipment when handling the liquid nitrogen and Cryo Trap.
4Do not touch the cooling rod and the cooling plate inside the chamber as long as the de-
war is not empty.
4The dewar has to be in upright position. Never unscrew the dewar as long as liquid nitro-
gen is in it.
4Use suitable tools when filling up the dewar.
4The dewar must be empty and must have room temperature before exchanging the cool-
ing plate of the Cryo Trap.
WARNING
Explosion hazard
Explosion risk if the pressure inside the dewar rises due to evaporating liquid nitrogen.
4Only use the cover which is provided by ZEISS for the dewar filling hole.
4Regularly check that ice does not build up at or around the cover for the dewar filling hole.
WARNING
Suffocation hazard due to lack of oxygen
Danger of nitrogen accumulation in low-lying areas or subterranean rooms.
4Ensure the room is ventilated properly.
4Ensure the room is equipped with a ventilation system with bottom suction and air ex-
changer.
WARNING
Fire hazard
Fire risk due to concentration of oxygen in liquid nitrogen.
4Cover the dewar so that the evaporating nitrogen can escape, but the penetration of air is
inhibited and no flammable material can fall into the dewar.
4Store and transport liquid nitrogen only in suitable dewars.
4Do not smoke or have open fire in the same room the dewar is located.
NOTICE
The detectors can be damaged
The cooling plate may collide with the detectors during mounting because there is very little
space between cooling plate and detectors.
4The EDS detector should be retracted during installation.
4Take care of the nearby detectors during the whole mounting procedure.
Info
14 Instruction Manual ZEISS Cryo Trap | en-US | Rev. 2 | 346000-8166-000
ZEISS 7 Operation | 7.1 Filling the Dewar with Liquid Nitrogen
Always exchange the specimen via the airlock, when the cryo trap is cooled down.
Info
Ensure, that the borehole in the lid does not face the specimen chamber.
7.1 Filling the Dewar with Liquid Nitrogen
Prerequisite üThe chamber must be evacuated.
1. Open the lid.
2. Fill in liquid nitrogen as necessary with a suitable container (dewar) and funnel.
3. Wait 10 to 15 minutes.
4. Refill.
5. Close the lid (white Teflon cover).
6. Make sure that the lid is correctly seated and
the borehole is not facing towards the speci-
men chamber.
7. For extended operating hours (>10 hours), refill liquid nitrogen as necessary.
7.2 Venting the Specimen Chamber
1. Start the CryoWarmUp macro with the button
CryoWarmUp in the SmartSEM Toolbar.
àThis will stop the turbo pump, execute a partial vent and keep the system chamber at a
pressure of about 30 mbar, followed by a regular system pump-down.
àThe cooling plate will be warmed up to 20°C within 6 hours.
9After that you can open the specimen chamber.
Instruction Manual ZEISS Cryo Trap | en-US | Rev. 2 | 346000-8166-000 15
7 Operation | 7.3 Dismounting the Cooling Plate ZEISS
7.3 Dismounting the Cooling Plate
1. NOTICE!Take care of collisions with the cooling plate.
Retract the EDS detector from working position.
2. Open the knurled thumb screw that fixes the
cooling plate
2
.
2
3. Carefully pull the cooling plate down.
7.4 Installing the Cooling Plate
Prerequisite üThe pin is mounted onto the cooling plate.
1. Insert the pin
1
of the cooling plate into the
borehole on the tip of the cooling rod.
1
2. Set the bottom surface of the cooling plate to the same level as the bottom surface of the
end cap of the objective lens.
3. Fix the cooling plate with the knurled thumb
screw
.
2
4. Measure the resistance between SE2 collector grid and the cooling plate with a DVM.
These two parts should be isolated from each other.
INFO: The cooling plate is not allowed to touch any detector or the end cap of the objec-
tive lens.
5. Also measure the resistance between VPSE collector grid and the cooling plate with a DVM.
These two parts should be isolated from each other.
6. NOTICE!Take care of collisions with the cooling plate.
Insert the EDS detector back to working position.
7.5 Defrosting the Dewar
Defrost the dewar to avoid icing in the vacuum chamber.
The procedure is the same as the procedure for venting the specimen chamber [
}
15].
16 Instruction Manual ZEISS Cryo Trap | en-US | Rev. 2 | 346000-8166-000
ZEISS 8 De-installation and Disposal
8 De-installation and Disposal
The Cryo Trap is de-installed and disposed of by authorized ZEISS service personnel only.
Instruction Manual ZEISS Cryo Trap | en-US | Rev. 2 | 346000-8166-000 17
9 Troubleshooting ZEISS
9 Troubleshooting
Symptom Cause Recommended action/s
Corrosion Condensation of water Defrost the dewar.
Icing The specimen chamber has been
vented while the temperature of
the Cryo Trap was far below room
temperature.
Do not attempt to pump the cham-
ber again until all components are
warm and fully dry.
Defrost the dewar.
Image drift Fluctuating temperature in the
specimen chamber
Check position of borehole of the
lid.
Long pumping
time
The specimen chamber has been
vented while the temperature of
the Cryo Trap was far below room
temperature.
Defrost the dewar.
18 Instruction Manual ZEISS Cryo Trap | en-US | Rev. 2 | 346000-8166-000
Carl Zeiss Microscopy GmbH
Carl-Zeiss-Promenade 10
07745 Jena
Germany
phone: +49 3641 64 3161
fax: +49 3641 64 3439
www.zeiss.com/microscopy
Instruction Manual ZEISS Cryo Trap | en-US | Rev. 2 | Printed in Germany
Modifications reserved.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20

Zeiss Correlative Cryo Workflow User manual

Category
Microscopes
Type
User manual

Zeiss Correlative Cryo Workflow is a specialized system for preparing and examining frozen-hydrated samples in a cryo-electron microscope, enabling high-resolution imaging of biological structures in their near-native state. With its advanced cryo-preparation techniques and integrated imaging capabilities, it streamlines the cryo-workflow, enhancing the quality and efficiency of cryo-EM studies.