Welch Allyn XENON 300 Light Source is a powerful illumination device designed for surgical procedures. Its 300-watt power output ensures optimal brightness, while the adjustable light intensity control allows you to tailor the illumination to your specific needs. The XENON 300 is compatible with various fiber optic bundles for ACMI, Wolf, Olympus, and Storz systems, providing versatility for different applications. Additionally, the cooling fans prevent overheating, ensuring reliable performance during extended use.

Welch Allyn XENON 300 Light Source is a powerful illumination device designed for surgical procedures. Its 300-watt power output ensures optimal brightness, while the adjustable light intensity control allows you to tailor the illumination to your specific needs. The XENON 300 is compatible with various fiber optic bundles for ACMI, Wolf, Olympus, and Storz systems, providing versatility for different applications. Additionally, the cooling fans prevent overheating, ensuring reliable performance during extended use.

XENON 300 LIGHT SOURCE
Thank you for purchasing the Welch Allyn Xenon 300 Light Source. By following the use and care guide-
lines given in this booklet, you will be rewarded with years of dependable, trouble-free service from your
new light source.
WARNING: All users of this product should be thoroughly trained in the medical procedures appropriate
to the equipment. Furthermore, they should read and understand the instructions for all other equipment
used in conjunction with this product.
i
Contents
1 - Conventions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Warnings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Indications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Contraindications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2 - Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
System Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Set-Up and Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
3 - Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Cleaning Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Lamp Care and Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Fuse Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Basic Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
4 - Technical Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
5 - Warranty and Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Limited Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
ii Contents Welch Allyn Xenon 300 Light Source
WARNING:
Extremely high energy light. Some fiber optic bundles will not withstand the light
intensity of this light source. USE ONLY WELCH ALLYN FIBER OPTIC BUNDLE or
SUNOPTICS TECHNOLOGIES (formerly CUDA FiberOptics - 904-737-7611) CLAD
ROD (CR) MODEL FIBER OPTIC BUNDLES.
1
1
Conventions
Symbols
Power “ON”
Protective Earth Ground
Type BF
CAUTION; consult Directions
For Use
Power “STANDBY”
~
Alternating Current
Warning
2 Chapter 1 Conventions Welch Allyn Xenon 300 Light Source
Warnings
A warning statement in this manual identifies a condition or practice which, if not
corrected or discontinued immediately, could lead to patient injury, illness, or death.
WARNING It is the user’s responsibility to check the specifications of any
device attached to this light source to verify that it can transmit the intense light
output without generating high temperatures and heat build-up in the device that
can cause serious burns.
WARNING High intensity light can cause burns to tissue even if the tip of the
fiber optic cable or attached terminal device (lighted retractor, for example) is
cool. Burns are caused by the absorption of light by tissue and the subsequent
conversion of that light energy into thermal energy. There is serious risk of
igniting fires if the energized light cable fibers are placed on flammable materials,
and full-thickness burns can be caused by prolonged exposure to concentrated
light energy. Anesthetized patients are particularly susceptible to burn injuries via
this mechanism. IT IS THE USER’S RESPONSIBILITY TO DETERMINE AND
MAINTAIN THE MINIMUM SAFE DISTANCE BETWEEN THE END OF THE
ENERGIZED FIBER-OPTIC LIGHT PATH AND ANY LIVING TISSUE OR
FLAMMABLE MATERIAL.
WARNING When used to power the Welch Allyn MFI Headlight system, we
recommend a standard 16” working distance separating the headlight and the
patient. Under no circumstances should the energized headlight be placed closer
than 10” to any object.
WARNING There is a risk of injury to tissue from exposure to the intense
illumination. The risk of injury is reduced by using the minimum level of
illumination necessary, by minimizing exposure time, and by avoiding close
stationary viewing.
WARNING The high intensity light at the front of the light source and at the tip
of a fiber optic bundle may give rise to high temperatures. To minimize the risk of
injury, avoid contact with tissues.
WARNING As with any bright light, do not look into the beam. Permanent eye
injury may result.
WARNING Total system risk current - excluding the patient applied instrument -
should not exceed 100 microamps. an isolation transformer is required if total
system risk current exceeds 100 microamps when accessories are
interconnected.
WARNING POSSIBLE EXPLOSION HAZARD. Equipment not suitable for use in
the presence of a flammable anesthetic mixture with air or with oxygen or
nitrous oxide.
WARNING The cover of the light source should never be removed.
Electrical shock hazard due to high internal voltage. there are no user serviceable
parts inside this light source except for the lamp and fuse, which are accessible
through the bottom and side door panels. refer all service to Welch Allyn.
WARNING Replace fuses as marked. See Fuse Replacement Section.
T8.0L 250V
Directions for Use Chapter 1 Conventions 3
Cautions
A caution statement in this manual identifies a condition or practice which, if not
corrected or discontinued immediately, could lead to equipment failure, equipment
damage, or data loss.
Caution Consult users manual for important information.
Caution This system should only be used by qualified medical personnel.
Caution To reduce the chance of direct visual exposure to the light, set the
light intensity knob to the minimum level and plug the fiber optic bundle into the
light source before turning the power ON.
Caution When the unit is on, never disconnect the fiber optic bundle from the
light source or the accessory from the end of the fiber bundle. High intensity light
from the end of the fiber optic bundle and light source can ignite flammable
materials (i.e. drapes).
Caution Grounding reliability is achieved only when connected to hospital-use
or hospital-grade receptacles. Inspect electrical plug and cord routinely. Do not
use if damage is discovered.
Caution PROVIDE VENTILATION TO PREVENT OVERHEATING. Do not
cover or drape the light source.
Caution Do not use the turret as a light attenuator or with the turret
misaligned with the light port, overheating will result.
Caution The ends of the fiber bundle get hot during use. Allow to cool
before removing.
Caution Do not operate light source without lamp module in place.
Caution If this light source is used during invasive surgery, the failure of a
component within the system, causing loss of illumination, may directly cause
risk to the patient. Therefore, it is recommended to have alternative means of
illumination available.
Caution Since the light source uses a custom lamp, always have a spare lamp
available for replacement. Use only Welch Allyn replacement lamp number
04980. Use caution because lamp may be hot. Allow the lamp to cool before
replacing.
Caution IPXØ - Equipment not protected against the ingress of water. Do not
use or store liquids above or on the light source.
4 Chapter 1 Conventions Welch Allyn Xenon 300 Light Source
Indications
For use with the Welch Allyn Headlight System
Contraindications
Neonate transillumination
Ophthalmic procedures
Any procedure where close or direct exposure to high intensity light could cause tissue
injury or damage (see “Warnings” on page 2).
2
5
Components
System Description
The Welch Allyn 49801 light source is intended to be used as a 300 watt light source to
supply illumination for surgical fiber optic headlights.
DO
OPERATION.
TO
THE
MODULE.
LAMP
DO
IN
FOR
WITH PART NUMBER 04980, 14 VDC 300 W
LAMP REPLACEMEN T, REPLACE
PLACE.
NOT OPERATE
THE
LIGHTSOUR CE WIT HOUT LAM P MODULE
OR LAMP MODULE.
READ
INSTRUCT IONS
BEFORE REPLACING
SAME
TYPE AND
RATING OF
LAMP
OR
LAMP
PREVENT OVERHEATING, REPLACE ONLY WITH
ALLOW LAMP
TO
COOL
5
MINUTES.
NOT REMOVE LAMP
IMMEDIATELY AFTER
SERIAL
/
200-240V 50
60Hz, 6.5A
UL2601
NO.
601.1
C22.2
CAN/CSA
IEC601-2-18
C
IEC601-1
/
~
100 - 120
N344
T8.0L250V
74227
NUMBER
REF 49801
R
L
I
S
T
E
D
R
L
I
S
T
E
D
R
R
S
T
O
Z
R
W
O
L
F
A
C
M
I
O
L
Y
M
P
U
S
1
3
2
4
6
5
9
7
8
DO
OPERATION.
TO
THE
MODULE.
LAMP
DO
IN
FOR
WITH PART NUMBER 04980, 14 VDC 300 W
LAMP REPLACEMEN T, REPLACE
PLACE.
NOT
OPERATE
THE
LIGHTSOURCE
WITHOUT LAMP MODULE
OR LAMP MODULE.
READ
INSTRUCT IONS BEFO RE REPLACING
SAME
TYPE
AND RATING OF
LAMP
OR L AMP
PREVENT
OVERHEATING, REPLACE
ONLY WITH
ALLOW
LAMP TO COOL 5
MINUTES.
NOT REMOVE LAMP IMMEDIATELY AFTER
SERIAL
/
200-240V 50
60Hz, 6.5A
UL2601 NO.
601.1
C22.2
CAN/CSA
IEC601-2-18
C
IEC601-1
/
~
100 - 120
N344
T8.0L250V
74227
NUMBER
REF 49801
R
L
I
S
T
E
D
R
L
I
S
T
E
D
R
R
S
T
O
Z
R
W
O
L
F
A
C
M
I
O
L
Y
M
P
U
S
1
3
2
4
6
5
9
7
8
DO
OPERATION.
TO
THE
MODULE.
LAMP
DO
IN
FOR
WITH PART NUMBER 04980, 14 VDC 300 W
LAMP REPLACEMEN T, REPLACE
PLACE.
NOT
OPERATE
THE
LIGHTSOURCE
WITHOUT LAMP MODULE
OR LAMP MODULE.
READ
INSTRUCT IONS BEFO RE REPLACING
SAME
TYPE
AND RATING OF
LAMP
OR L AMP
PREVENT
OVERHEATING, REPLACE
ONLY WITH
ALLOW
LAMP TO COOL 5
MINUTES.
NOT REMOVE LAMP IMMEDIATELY AFTER
SERIAL
/
200-240V 50
60Hz, 6.5A
UL2601 NO.
601.1
C22.2
CAN/CSA
IEC601-2-18
C
IEC601-1
/
~
100 - 120
N344
T8.0L250V
74227
NUMBER
REF 49801
R
L
I
S
T
E
D
R
L
I
S
T
E
D
R
R
S
T
O
Z
R
W
O
L
F
A
C
M
I
O
L
Y
M
P
U
S
1
3
2
4
6
5
9
7
8
Power Switch - “l”
indicates “ON” position
Fiber Optic Bundle Turret -
accepts fiber optic bundles for
ACMI, Wolf, Olympus and Storz.
Cooling Fans - provides
cooling to light source
Lamp Module Access Door -
access for lamp replacement.
Handle - means to carry light source.
Active light port
Fuse Holders - contains
the main fuses.
Power Module - accepts
AC power cord.
Brightness Control - increases or
decreases light output intensity level.
6 Chapter 2 Components Welch Allyn Acuity LT Hardwired
Set-Up and Operation
1. Unpack unit and visually inspect it for damage that may have occurred during
shipment.
2. Plug power cord into light source power module and the remaining end into a hospital
grade receptacle.
3. Select type of fiber optic bundle port (ACMI, Wolf, Olympus, Storz) and rotate turret
to align with light port.
4. Insure that the ends of the fiber optic bundle are clean before each use. Any residue
on the ends of the fiber optic will absorb the high intensity light and will overheat or
damage the fiber optic bundle.
5. Insert fiber optic bundle into light port and connect the headlight luminare. Set the
light intensity to minimum level. This prevents inadvertent shining into the patient’s or
user’s eyes when turned on.
6. Turn the unit on.
7. Set light intensity to minimum level necessary for adequate visualization.
8. Turn unit off when not in use.
9. Allow the light source to cool before removing fiber optic bundle. The ends of the
fiber optic bundle get hot during use.
With the output of high intensity (Xenon) light sources, it is possible to convey a large
amount of light energy and to concentrate this energy in a very small area. This may result
in thermal injury to the tissue.
The risk of injury is increased under the following conditions:
Prolonged close contact or close stationary viewing of the mucosa or tissue.
Using a high intensity light source at high output levels.
Using multiple instruments to illuminate the same area.
The following recommendations will reduce the risk of thermal injury:
Use the minimum level of illumination necessary for adequate visualization.
When possible, avoid close stationary viewing.
Minimize exposure time.
Caution When the unit is on, NEVER disconnect the fiber optic bundle from the
light source or the headlight from the fiber optic bundle. High intensity light from
the end of the fiber optic bundle and light source can ignite flammable materials
(i.e. drapes).
3
7
Maintenance
Cleaning Instructions
1. Turn power off and disconnect the power cord from both the wall outlet and the
Welch Allyn Xenon 300 Light Source.
2. Wipe the external surfaces clean with a cloth dampened with mild soap and water or
use one of the following cleaning/disinfecting solutions. DO NOT IMMERSE.
70% isopropyl alcohol
10% mild bleach solution
3. Wipe the power cord clean using the same cleaning/disinfecting solutions listed
above. Wipe dry with a clean cloth. Take care not to get the plug prongs wet. DO NOT
IMMERSE.
4. Do not plug the unit into a wall outlet until it is thoroughly dry.
Note
Do not sterilize or immerse the Welch Allyn Xenon 300 Light Source.
Caution Clean the ends of the fiber optic bundle before each use. Any residue
(including cleaning and disinfecting solutions) will absorb high intensity light on
the surface of the fiber optic. Absorbed light will overheat the residue causing
damage to the fiber optic bundle.
8 Chapter 3 Maintenance Welch Allyn Xenon 300 Light Source
Lamp Care and Replacement
The lamp in the Welch Allyn Xenon 300 Light Source will typically provide more than 500
hours of service. The symptoms indicating the need for replacement are:
Low light output
The unit requires more than four ignition strikes to start the lamp going into operation.
The lamp life indicator reads more than 500 hours.
The light source is designed for simple and quick changing of the lamp module, which
houses the lamp.
Removing the Lamp
1. Turn unit off and unplug power cord from its power source.
2. Allow the lamp to cool.
3. Unlock and open access door located on the side of the light source.
4. Grasp the two handles and pull the entire lamp module out of the light source.
Inserting a Lamp
:
1. Orient the new lamp module so that the window of the lamp is facing the front of the
light source.
2. Carefully align the connection pins with the brass sockets on the lamp module and
push firmly to insure proper connection.
3. Close and lock the access door.
+
500HR
0 HR
HOURMETER ASSY
TO REMOVE LAMP
PULL OUT.
DO NOT REMOVE
IF LAMP IS HOT.
SERIAL NUMBER
8AA 1104
FRONT
175 Hours
Caution DO not incinerate used lamps to avoid risk of explosion.
S
T
O
Z
R
W
O
L
F
A
C
M
I
O
L
Y
M
P
U
S
Directions for Use Chapter 3 Maintenance 9
Fuse Replacement
1. Turn unit off prior to checking the fuses.
2. Unplug the light source.
3. Use a flat blade screwdriver to remove fuse covers and check for blown fuse.
4. Replace the blown fuses with the same size and rating.
5. Replace fuse covers.
Ordering Information
Part Information
Replacement Parts
DO
OPERATION.
TO
THE
MODULE.
LAMP
DO
IN
FOR
WITH PART NUMBER 04980, 14 VDC 300 W
LAMP
REPLACEMENT, REPLACE
PLACE.
NOT OPERATE
THE
LIGHTSOURCE WITHOUT LAMP MODULE
OR LA MP MOD ULE.
READ INSTRUCTIONS BEFORE REPLACING
SAME TYPE AND RATING OF LAMP OR L AMP
PREVENT
OVERHEATING, REPLACE
ONLY WITH
ALLOW
LAMP TO COOL 5
MINUTES.
NOT REMOVE
LAMP IMMEDIATELY AFTER
49800 Welch Allyn Xenon 300 Light Source
100-120/200-240 VAC, 50-60 Hz, 5.5A
50-60 Hz
USA Plug type
04980-U Replacement Lamp Module
76400 USA Power Cord
10 Chapter 3 Maintenance Welch Allyn Xenon 300 Light Source
Basic Troubleshooting
Symptom Possible Problem Remedy
Power does
not come on.
Power plug is not plugged into the receptacle.
Power switch is off.
Fuse is blown.
Lamp door is not closed tightly.
Lamp module is not installed incorrectly.
Plug in power cord.
Turn on power switch.
Replace fuse.
Close lamp door completely.
Reinstall lamp module correctly.
No light
output.
Power plug is not plugged into receptacle.
Power switch is off.
Fuse is blown.
Lamp door is not closed tightly.
Lamp module is installed incorrectly.
Turret misaligned with light port.
Lamp burned out.
Fiber bundle is not connected.
Light attenuator is set to minimum.
Plug in power cord.
Turn on power switch.
Replace fuse.
Close lamp door completely.
Reinstall lamp module.
Set turret to detent position (light port
aligned).
Replace lamp.
Connect fiber bundle correctly.
Set the light attenuator to minimum level
needed for adequate visualization.
Lamp
flickers or
dims.
Lamp is getting old. Replace lamp.
Light source
turns off
after a few
minutes of
operation.
Obstructed air intake; overheating causes
thermal switch to trip.
Fan not running; overheating causes thermal
switch to trip.
Allow unit to cool (5-10 minutes). Remove
obstruction(s).
Contact Welch Allyn.
Low light
output
Accessory damaged.
Lamp is old.
Light transmission rod is damaged or
contaminated.
Replace accessory.
Replace lamp.
Clean with isopropyl alcohol and allow to dry.
Use care not to chip edges.
Contact Welch Allyn for repair.
Several
ignition
attempts
needed to
start lamp.
Three or fewer ignition attempts are normal. Replace lamp if four or more ignition attempts
are required.
4
11
Technical Information
Specifications
Electrical
Dimensions
Fiber Optic Bundle Interface
Environment
Fuses
Lamp Operating Power:
Lamp Current:
Input Voltage:
Lamp:
Lamp Life:
300 watts DC, nominal
21 Amps DC, nominal
100-120/200-240 VAC, 50/60 Hz, 5.5A
Cermax® Xenon Lamp
500 hours
Size:
Weight:
12.75” (324mm)W x 4.75” (121mm) H x 12.5” (318mm) D
12.5 lbs (5.7kg)
Turret Fiber Bundle Port accepts ACMI, Wolf, Olympus, and Storz
Operating:
Transport/Storage:
10
o
C to 40
o
C, 95% R.H., 500hPa - 1060 hPa, Altitude
-20
o
C to 49
o
C, 95% R.H. Max, 500hPa - 1060hPa, Altitude
5mm fuse, T 8.0L 8 amp 250V.
12 Chapter 4 Technical Information Welch Allyn Acuity LT Hardwired
Approvals
Equipment Classification
Patents
NOTE: This light source complies with current required standards for electromagnetic
interference and should not present problems to other equipment or be affected by other
devices. As a precaution, avoid using this device in close proximity to other equipment.
Equipment Disposal: Do not incinerate this equipment at the end of useful life for risk of
lamp explosion. Equipment contains no hazardous materials and can be discarded without
environmental risks.
UL 2601-1
CAN/CSA C22.2
No601.1-M90
IEC 601-1
IEC 601-1-2
The CE mark on this product indicates it has been tested to and conforms
with the provisions noted within the 93/42/EEC Medical Device Directive.
Declarations of Conformity are held by Welch Allyn’s European
Representative.
European Regulatory Manager
Welch Allyn LTD.
Navan Business Park
Dublin Road
Navan, County Meath, Republic of Ireland
Tel +353 46 67700
FAX +353 46 27128
Class 1 Equipment
Continuous Operation
Patents Pending
5
13
Warranty and Service
Limited Warranty
The Welch Allyn Xenon 300 Light source is guaranteed by Welch Allyn against all
manufacturing defects. Welch Allyn will repair or replace, free of charge, any parts of its
own manufacture that prove to be defective for reasons other than misuse, neglect,
damage in shipment, or normal wear. This warranty is in effect for one year from original
date of purchase.
Service
Service Policy
A Welch Allyn Service Center must perform or approve all repairs on products under
warranty. Unauthorized repairs will void the warranty
. Qualified electronics personnel or a
Welch Allyn Service Center should repair products out of warranty.
Technical Assistance
If you have an equipment problem that you cannot resolve, you may call the Welch Allyn
Service Center nearest you for assistance. Call the service center on normal business
days at the phone numbers listed on page 14.
If you are advised to return a product to Welch Allyn for repair or routine maintenance,
schedule the repair with the service center nearest you.
Before returning a product for repair you must obtain authorization from Welch Allyn. Our
service personnel will give you a Service Notification number. Note this number on the
outside of your shipping box. Service centers do not accept deliveries without an Service
Notification number.
14 Chapter 5 Warranty and Service Welch Allyn Xenon 300 Light Source
Welch Allyn Service Centers
For Service or Repair
USA Customers
Welch Allyn, Inc.
4341 State Street Road
Skaneateles Falls, NY 13153 USA
Tel: 800-535-6663 (USA ONLY) or 315-685-4560
Fax: 315-685-3361
Canada Customers
Welch Allyn Canada Limited
Technical Service Centers
160 Matheson Blvd., East
Mississauga, Ontario L4Z 1V4
Tel: 800-561-8797 (CANADA ONLY) or 905-890-0001
Fax: 905-890-0008
International Customers
Welch Allyn GmbH Co. + KG
Postfach 31
Zollerstrasse 2-4
72417 Jungingen
Germany
Tel: 0-7477-92710
Fax: 0-7477-927190
Welch Allyn Australia Pty. Ltd.
Technical Service Centers
Unit 5, 38-46 South Street
Rydalmere NSW 2116
Tel: 0 2 9638 3000
Fax: 0 2 9638 3500
Welch Allyn Ltd. Singapore
10 Hoe Chiang Road
#19-03/04
Keppel Towers
Singapore 089315
Tel: 6419 8100
Fax: 6291 5780
Welch Allyn UK Ltd.
Navan Business Park
Dublin Road
Navan, County Meath
Ireland
Europe Tel: + 353 46 906 7790
Europe Fax: + 353 46 906 7754
UK Tel: 0 207 365 6780
UK Fax: 0 207 365 9694
Welch Allyn China
Unit 1011 Central Plaza
No. 227 HuangPi North Road
Shanghai, China 200003
Tel: 0 21 6327 9631
Fax: 0 21 6327 9632
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22

Welch Allyn XENON 300 User manual

Type
User manual

Welch Allyn XENON 300 Light Source is a powerful illumination device designed for surgical procedures. Its 300-watt power output ensures optimal brightness, while the adjustable light intensity control allows you to tailor the illumination to your specific needs. The XENON 300 is compatible with various fiber optic bundles for ACMI, Wolf, Olympus, and Storz systems, providing versatility for different applications. Additionally, the cooling fans prevent overheating, ensuring reliable performance during extended use.

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI