Thermo Fisher Scientific KingFisher Apex Purification System User guide

Type
User guide
For Research Use Only. Not for use in diagnostic procedures.
KingFisher Apex Purification System
USER GUIDE
and BindIx Software version 1.0
Catalog Numbers5400910R, 5400920R, 5400930R, 5400940R
Publication Number MAN0026542
Revision A.0
Life Technologies Holdings Pte Ltd | Block 33 | Marsiling Industrial Estate Road 3 | #07-06, Singapore 739256
For descriptions of symbols on product labels or product documents, go to thermofisher.com/symbols-definition.
The information in this guide is subject to change without notice.
DISCLAIMER: TO THE EXTENT ALLOWED BY LAW, THERMO FISHER SCIENTIFIC INC. AND/OR ITS AFFILIATE(S) WILL NOT BE
LIABLE FOR SPECIAL, INCIDENTAL, INDIRECT, PUNITIVE, MULTIPLE, OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH OR
ARISING FROM THIS DOCUMENT, INCLUDING YOUR USE OF IT.
Rev history: Pub. No.MAN0026542
Revision Date Description
A.0 30 March 2022 New RUO document for the KingFisher Apex Purification System, based on Pub.No.MAN0018970.
Important Licensing Information: These products may be covered by one or more Limited Use Label Licenses. By use of these
products, you accept the terms and conditions of all applicable Limited Use Label Licenses.
TRADEMARKS: All trademarks are the property of Thermo Fisher Scientific and its subsidiaries unless otherwise specified.
©2022 Thermo Fisher Scientific Inc. All rights reserved.
Contents
CHAPTER1Productinformation .................................................. 6
Productdescription ............................................................. 6
Producttype ................................................................... 6
Productcomponents ............................................................ 7
Product overview ............................................................... 8
Instrument overview ......................................................... 8
User interface overview ..................................................... 10
Technology overview ........................................................... 11
Principle of magnetic particle processing ..................................... 11
Upon receiving thedevice ....................................................... 13
Guidelines for unpacking the instrument ..................................... 14
First time instrumentsetup .................................................. 14
Systemcomponents ........................................................... 18
Heating block and corresponding plate ...................................... 18
Magnetichead ............................................................ 20
Tip comb and platecompatibility ............................................ 22
About the Thermo Fisher ConnectPlatform ...................................... 22
Create a Connect Platformaccount .......................................... 22
Connect the instrument to the Internet ....................................... 23
Create a PINnumber ....................................................... 23
Generate a link code fromthe instrument ..................................... 23
Add an instrument to your Connect Platformaccount .......................... 23
Set up a new administrator .................................................. 25
CHAPTER2Methods ............................................................. 26
Protocol access overview ....................................................... 26
Download protocols throughthe instrument ................................... 26
Download protocols from Connect ........................................... 26
Run purificationkit ............................................................. 27
Pause or abortrun ............................................................. 28
Change heat block/magnetichead ............................................... 28
Unload heatingblock ....................................................... 28
Unload magnetichead ..................................................... 29
KingFisher Apex Purification System User Guide 3
CHAPTER3Softwareapplications .............................................. 31
About the user interface ........................................................ 31
Install BindIx Software .................................................... 31
Overview of user interface controls ............................................... 32
Home screen .............................................................. 32
Menu screen .............................................................. 33
Protocol editor screen ...................................................... 34
Settings screen ............................................................ 35
Instrument screen ......................................................... 40
Protocol overview .............................................................. 41
Create a protocol .............................................................. 41
Enter protocolinformation .................................................. 42
Select magnetichead ...................................................... 43
Add protocol steps ........................................................ 43
Define protocol steps ...................................................... 45
Save a protocol ................................................................ 56
Import/Export protocols ........................................................ 56
Import protocol from a USB memory device ................................... 56
Export protocol to a USB memory device ..................................... 57
Run history overview ........................................................... 57
Export run report to a USB memory device ................................... 57
Delete run report .......................................................... 57
Delete multiple run reports .................................................. 57
Delete run report by date ................................................... 58
Delete run report byname .................................................. 58
Telemetrydata ................................................................. 58
Update software ............................................................... 60
APPENDIXAMaintenance ....................................................... 61
Regular and preventive maintenance ............................................. 61
Store the Transport Lock ........................................................ 61
Magnetic head maintenance ..................................................... 62
Decontamination procedure ..................................................... 62
Decontaminateinstrument .................................................. 62
UV decontamination procedure .................................................. 63
UV treat processchamber .................................................. 63
Prepare instrument for transport ................................................. 63
Refit the transportlock ..................................................... 63
Contents
4KingFisher Apex Purification System User Guide
APPENDIXBSpecifications ...................................................... 65
Physicaldimensions ............................................................ 65
Instrumentspecifications ....................................................... 65
Performancespecifications ...................................................... 66
APPENDIXCOrderinginformation .............................................. 68
Accessory products ............................................................ 68
APPENDIXDSafety ............................................................... 70
Safetycompliance ............................................................. 70
Symbols on thisinstrument ..................................................... 71
Standard safetysymbols ................................................... 71
Electromagnetic compatibility (EMC) standards .................................... 72
Class Bnotice ............................................................. 72
EMCcompliance .......................................................... 72
Electricalsafety ................................................................ 73
Chemicalsafety ................................................................ 74
Biological hazardsafety ......................................................... 75
APPENDIXECertificate ofdecontamination ................................... 76
APPENDIXFDocumentation and support ...................................... 77
Customer and technical support ................................................. 77
Limited product warranty ........................................................ 77
Contents
KingFisher Apex Purification System User Guide 5
Product information
Product description
The Thermo Scientific KingFisher Apex Purification System is intended for automated transfer and
processing of magnetic particles in a microplate format. The system utilizes magnetic rods covered with
a disposable, specially designed tip comb, and plates. The instrument functions without any dispensing
or aspiration parts or devices. Before the run, samples and reagents, including magnetic particles,
are dispensed into the plates according to default protocols that are downloaded from the Thermo
Fisher Connect Platform or thermofisher.com/kingfisher. Custom protocols can be created using the
instrument interface or with BindIx Software.
Product type
The KingFisher Apex Purification System consists of the KingFisher Apex instrument, and one of four
dierent magnetic heads and associated heating blocks.
Instrument Catalog No.
KingFisher Apex Purification System with 96 PCR Head 5400910R
KingFisher Apex Purification System with 96 Combi Head 5400920R
KingFisher Apex Purification System with 96 Deep Well Head 5400930R
KingFisher Apex Purification System with 24 Combi Head 5400940R
1
6KingFisher Apex Purification System User Guide
Product components
The contents of the KingFisher Apex Purification System are listed in the following table.
Component Cat. No.
5400910R
Cat. No.
5400920R
Cat. No.
5400930R
Cat. No.
5400940R
Instrument
KingFisher Apex Purification System 1 1 1 1
Accessory tray (See the KingFisher Apex Purification System Installation Guide)
KingFisher Apex Magnetic Head
96 PCR Head
96 Combi Head
96 Deep Well Head
24 Combi Head
1
1
1
1
KingFisher Apex PCR Heating Block 1
KingFisher Apex 96 Heating Block 1 1 1
KingFisher Apex 96 Deep Well Heating Block 1 1
KingFisher Apex 96 Storage Tube Heating Block 1
KingFisher Apex 24 Deep Well Heating Block 1
KingFisher Apex demo package (Contains plastic
consumables) 1111
USB Drive containing KingFisher Apex user
manuals, BindIx Software, and Certificate of
conformance
1111
UV Lamp 2 2 2 2
3 mm flat-tip screwdriver 1 1 1 1
Packing list 1 1 1 1
Chapter1Product information
Product components 1
KingFisher Apex Purification System User Guide 7
Product overview
Instrument overview
The KingFisher Apex Purification System consists of an instrument with touch screen control and a
process chamber with lid to protect samples against environmental contamination.
Figure1KingFisher Apex instrument front view
1Front lid
2Touch screen
3On/o button
4USB port × 3 (Device)
1
346
5
2
Figure2KingFisher Apex instrument rear view
1Power switch
2Power inlet
3Ethernet/LAN port
4USB port (Host) × 2
5RS-232 serial port
6USB port (Device)
Chapter1Product information
Product overview
1
8KingFisher Apex Purification System User Guide
Internal layout
The process chamber of the KingFisher Apex instrument consists of a turntable with eight plate
stations. The loading position has an illuminated display, while the processing position is situated over
the heating block (there are two heating block stations that can be selected for the processing position),
and underneath the processing head.
The processing head consists of two vertically moving platforms—the magnetic head holder, and the
tip comb holder. The magnetic head holder is used to attach up to two magnets composed of 24 or 96
magnetic rods, while the tip comb holder is used to hold plastic tip combs with 24 or 96 tips.
Figure3Process chamber layout
1Turntable with plate stations (1–8)
2Processing position
3Loading position
4Tip comb holder
5Magnetic head holder
6Shield plate
7Barcode reader
8UV light clip × 2
9UV light socket × 2
Chapter1Product information
Product overview 1
KingFisher Apex Purification System User Guide 9
User interface overview
Symbol Function
Open Notifications screen (See page 32 for more details)
Swipe down from the top of the touch screen
Displays the following information:
Notifications
Internet status
Storage devices
Open Menu screen (See page 33 for more details)
Select Menu
Gives access to the following items:
Home
Protocol editor
Protocol library
Run history
Import/Export
Settings
Instrument
Open Settings screen (See page 35 for more details)
Select Menu4Settings
Gives access to the following items:
Localization
Protocols
Sound
Loading position display
Connect
Network share
About
Reset factory settings
Troubleshoot
View EULA
Open Instrument screen (See page 40 for more details)
Select Menu4Settings
Gives access to the following items:
Tools
Magnets
Heating blocks
Chapter1Product information
Product overview
1
10 KingFisher Apex Purification System User Guide
Technology overview
Principle of magnetic particle processing
The KingFisher Apex instrument operates on inverse magnetic particle processing (MPP) technology.
Instead of transferring reagents in and out of plates, as might occur with an external magnet method,
the magnetic particles are moved through a series of plates containing specific reagents with the aid of
magnetic rods covered with a disposable, specially designed plastic tip comb.
4 7
3
5
6
2
1
Figure4Inverse magnetic particle processing
1Magnet
2Tip
3Slow up and down collecting movement
4Well 1: Reagent 1 and magnetic particles
5Transfer
6Fast up and down releasing movement
7Well 2: Reagent 2 and magnetic particles
Working with magnetic particles can be divided into five separate processes:
Collecting magnetic particles
Releasing magnetic particles
Washing magnetic particles
• Incubation
• Concentration
Chapter1Product information
Technology overview 1
KingFisher Apex Purification System User Guide 11
Collecting magnetic particles
During the collection of the magnetic particles, the magnetic rod is fully inside the tip. The magnetic
rods together with the tip comb move slowly up and down in the plate and the magnetic particles are
collected onto the edge of the tips. The magnetic rods together with the tip comb, having collected the
magnetic particles, can be lifted out of the plate and transferred into the next plate.
Releasing magnetic particles
After collection of the magnetic particles, the magnetic rods together with the tip comb are lifted from
the plate, the magnetic rods are lifted o and the tip comb is lowered into the next plate containing a
reagent.
Magnetic particles are released by moving the tip comb up and down several times at considerably high
speed until all the particles have been mixed with the substance in the next reaction.
Washing magnetic particles
Washing the magnetic particles is a frequent and an important processing phase. Washing is a
combination of the release and collection processes in a plate filled with washing solution.
To maximize washing eciency, the magnetic rods together with the tip comb are designed to have
minimized liquid-carrying properties.
Incubation
The temperature of the reaction mixture is elevated to obtain a sucient level of specific binding. To
keep the magnetic particle suspension evenly mixed in long-running reactions, the tip comb can be
moved up and down in the solution.
Chapter1Product information
Technology overview
1
12 KingFisher Apex Purification System User Guide
Concentration
Magnetic particle processing allows changes in volume to be performed during the procedure.
Transferring the magnetic particles from a plate with a larger volume to a plate with a smaller volume
concentrates the sample.
47
3
5
6
2
1
Figure5Concentration step during magnetic particle processing
1Magnet
2Tip
3Slow up and down collecting movement
4Well 1: Collection of magnetic particles from a large-volume well
5Transfer
6Fast up and down releasing movement
7Well 2: Release of magnetic particles into a low-volume well
Upon receiving the device
Check the enclosed packing list against order.
Visually inspect the transport package, the instrument and the accessories for any damage
incurred during transit.
If the carton has been damaged in transit, it is particularly important that you retain it for inspection
by the carrier in case there has also been damage to the instrument.
Chapter1Product information
Upon receiving the device 1
KingFisher Apex Purification System User Guide 13
Any damage claims must be filed with the carrier. Neither the manufacturer nor its agents can be
held responsible for any damage incurred in transit, but the manufacturer will make every eort to
help obtain restitution from the carrier. Upon receipt of the carrier's inspection report, arrangements
will be made for repair or replacement.
Guidelines for unpacking the instrument
To prevent condensation, the instrument should be left in its protective, antistatic plastic wrapping
until the ambient temperature has been reached.
The instrument weighs 56 kg (123 lbs) and requires at least two people to lift. Use the proper
precautions when lifting the instrument to avoid injury.
Retain the original packaging and packing material for future transportation. The packaging is
designed to assure safe transport and minimize transit damage. Use of alternative packaging
materials may invalidate the warranty. Also retain all instrument-related documentation provided by
the manufacturer for future use.
Refer to the installation guide for detailed instructions on how to unpack the instrument.
First time instrument setup
1. Connect the power supply cable to the power inlet.
The instrument operates at voltages of 100–240 VAC and the frequency range of 50/60 Hz. Ensure
that the local supply voltage in the laboratory conforms to that specified on the type label on the
back of the instrument
2. Remove the transport lock screw from the magnet holder.
Chapter1Product information
Upon receiving the device
1
14 KingFisher Apex Purification System User Guide
3. Remove the transport lock screws from the transport lock plate over the tip comb holder.
3
2
4. Remove the upper transport lock plate, then lift the tip comb holder and remove the lower
transport lock plate.
Do not discard the parts of the transport lock. The lock is required in case of future transportation
of the instrument (see “Store the Transport Lock” on page61).
Install the UV light bulbs
1. Insert the UV light bulb into the light clip.
2. Slide the light bulb back and into the light socket.
1
2
Chapter1Product information
Upon receiving the device 1
KingFisher Apex Purification System User Guide 15
Turn on the instrument
1. Turn on the power switch at the rear of the instrument.
2. Press the on/o button to turn on the instrument.
Install heating block
1. Select Menu4Instrument.
2. Select Insert for the position (1 or 2) to install the heating block.
3. Scan the heating block to be installed with the barcode reader or perform manual selection with
the menu.
4. Place the heating block on the heating platform. Fit the pins on the block into the slots in the
platform. Press the block until it clicks properly in position.
Note: Select Information for details on heating block installation.
5. Select Next
Chapter1Product information
Upon receiving the device
1
16 KingFisher Apex Purification System User Guide
Install magnetic head
1. Select Menu4Instrument.
2. Select Insert for the position (1 or 2) to install the magnet.
3. Scan the magnetic head to be installed with the barcode reader or perform manual selection with
the menu.
4. Place the magnetic head onto the holder. Fit the pins on the head into the slots in the holder.
5. Turn the handle to lock the magnet in place.
Note: Select Information for details on magnetic head installation.
5
4
6. Select Next
Chapter1Product information
Upon receiving the device 1
KingFisher Apex Purification System User Guide 17
System components
Heating block and corresponding plate
The instrument uses specially designed barcoded plates for optimal processing. Use of other types of
plates may damage the instrument and compromise the warranty, as well as result in below optimal
performance.
The heating blocks are specifically designed for the plates listed in the following table to ensure even
heating during sample processing.
See “Change heat block/magnetic head” on page28 for details on changing a heating block.
Heating block Plate
KingFisher Apex 24 Deep-Well Heating Block,
barcoded
KingFisher 24 deep well plate, barcoded (30 µL to
5mL [1])
KingFisher Apex 96 Deep-Well Heating Block,
barcoded
KingFisher 96 Deep-Well Plate, barcoded (15–
1000µL [1])
KingFisher Apex 96 Heating Block, barcoded KingFisher 96-Well Microplate, barcoded (15–
200µL [1])
Chapter1Product information
System components
1
18 KingFisher Apex Purification System User Guide
(continued)
Heating block Plate
KingFisher Apex PCR Heating Block, barcoded PCR plate (10–80µL [1]), skirted (Cat. No.
AB2396 [2]), or semi-skirted (Cat. No. AB2496 ) [2]
KingFisher Apex 96 Storage Tube Heating Block,
barcoded
96 Storage Tube (Nunc 374086), 30–200µL
KingFisher Apex 24 Storage Tube Heating Block,
barcoded
24 Storage Tube (Nunc 364323), 200–1000µL
Chapter1Product information
System components 1
KingFisher Apex Purification System User Guide 19
(continued)
Heating block Plate
Adapter
KingFisher Apex 24 Storage Tube Adapter KingFisher Apex 96 PCR Semiskirted Plate Adapter
[1] The fill volume depends on the type of tip comb being used.
[2] Recommended PCR plate
Magnetic head
WARNING! This product contains very strong permanent magnets. People wearing a pacemaker
or metallic prostheses should not use this product. A pacemaker or prostheses may be aected or
damaged if it comes in close contact with a strong magnetic field.
There are four kinds of interchangeable magnetic heads with corresponding disposable plastic tip
combs available for the instrument. The magnetic head holder has two positions that can be used to
hold dierent types of magnetic heads.
IMPORTANT! Keep magnetic heads away from each other and other magnets at all times. Clashing of
the magnets can cause serious damage to the magnets.
Do not place the magnetic head on top of the instrument or on any metal surfaces.
Do not use metal tools when handling magnetic heads.
Do not place the magnetic head in close proximity to magnetic tapes, computer discs or other
magnetic storage systems, such as credit cards, as damage can occur due to the strong magnetic
field of the magnetic heads.
Do not place the magnetic heads near a PC display, as this may cause damage to the display.
When not in use, keep the magnetic heads in their respective storage boxes.
See “Change heat block/magnetic head” on page28 for details on changing a magnetic head.
Chapter1Product information
System components
1
20 KingFisher Apex Purification System User Guide
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78

Thermo Fisher Scientific KingFisher Apex Purification System User guide

Type
User guide

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI