DRAFT 2/4/02
Service Pendant Hand-Held Terminal
10
[SHIFT] PRSRS
SHOW
VLV ST
Displays the Test Port and System Pressures in the format:
?test port 2.3 (value is in mTorr)
system TC 1 ok (value is in mTorr)
SET
OFFSET
SHOW
Displays the value of the electronic offset of the Preamplifier in the
format:
?OFFSET XX ok
[SHIFT] SET
OFFSET
SHOW
Sets a temporary value of the electronic offset in the
Preamplifier. Enter XX (the desired value), followed by the SHIFT
key, then the function key.
INIT
RPLR
PUT
Sets a temporary new value for Repeller Voltage. Enter XXX
(the desired value), then the function key.
[SHIFT] INIT
RPLR
PUT
Sets a permanent new value for Repeller Voltage. Enter XXX
(the desired value), followed by the SHIFT key, then the
function key.
INIT
ION V
PUT
Sets a temporary new value for Ion Voltage. Enter XXX (the desired
value), then the function key.
[SHIFT] INIT
ION V
PUT
Sets a permanent new value for Ion Voltage. Enter XXX (the
desired value), followed by the SHIFT key, then the function key.
INIT
V FOC
PUT
Sets a temporary new value for Variable Focus Voltage. Enter XXX
(the desired value), then the function key.
[SHIFT] INIT
V FOC
PUT
Sets a permanent new value for Variable Focus Voltage. Enter XXX
(the desired value), followed by the SHIFT key, then the function
key.
INIT
GAIN
PUT
Sets a temporary new value for System Gain. Enter X.X (the
desired value), then the function key.
[SHIFT] INIT
GAIN
PUT
Sets a permanent new value for System Gain. Enter X.X (the
desired value), followed by the SHIFT key, then the function key.
INIT
EMIS
PUT
Sets a temporary new value for Emission Current. Enter XXXX
(the desired value), then the function key.
Table 2 Glossary of Input and Data Keys (continued)
NOTE
Values that are entered using INIT are
permanently stored in non-volatile RAM
unless they are replaced using another
INIT command.
Values that are entered using PUT are
temporary. The parameter reverts to the
previously stored (INIT) value when the
system is shut down.