Flexim FLUXUS F60 Series Operating

Category
Measuring, testing & control
Type
Operating
FLUXUS F60*
UMFLUXUS_F60xV5-0EN
Operating instruction
FLEXIM GmbH
Boxberger Straße 4
12681 Berlin
Germany
Tel.: +49 (30) 936 67 660
Fax: +49 (30) 936 67 680
www.flexim.com
FLUXUS is a registered trademark of FLEXIM GmbH.
Operating instruction for
FLUXUS F60*
UMFLUXUS_F60xV5-0EN, 2017-10-16
Firmware version: V6.28
Copyright (©) FLEXIM GmbH 2017
Subject to change without prior notice.
FLUXUS F60* Table of contents
Table of contents
3
UMFLUXUS_F60xV5-0EN, 2017-10-16
1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2 Safety instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.1 General safety instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.2 Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.3 Not intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.4 Safety instructions for the user . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.5 Safety instructions for the operator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.6 Safety instructions for electrical work . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.7 Safety instructions for transport . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.8 Recommended procedure in hazardous situations . . . . . . . . . . . . . . . . . . . . . 12
3 General principles . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.1 Measurement principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.2 Measurement arrangements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.3 Acoustic penetration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
3.4 Undisturbed flow profile . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
4 Product description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
4.1 Measuring system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
4.2 Handling concept . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
4.3 Navigation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
4.4 Keyboard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
5 Transport and storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
5.1 Transport . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
5.2 Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
6Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
6.1 Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
6.2 Transducers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
6.3 Temperature probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
7 Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
7.1 FLUXUS *601 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
7.2 FLUXUS *608 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
8 Start-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
8.1 Start-up settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
8.2 Switching on/off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
Table of contents FLUXUS F60*
2017-10-16, UMFLUXUSF60xV5-0EN
4
8.3 Program branches . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
8.4 HotCodes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
8.5 Language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
8.6 Initialization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
8.7 Time and date . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
8.8 Instrument information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
9 Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
9.1 Parameter input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
9.2 Measurement settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
9.3 Start of the measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102
9.4 Displaying measured values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
9.5 Execution of special functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 110
9.6 Determination of the flow direction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111
9.7 Stop of measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111
10 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 112
10.1 Problems with the measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113
10.2 Measuring point selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114
10.3 Maximum acoustic contact . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114
10.4 Application-specific problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115
10.5 Significant deviations of the measured values . . . . . . . . . . . . . . . . . . . . . . . 115
10.6 Problems with the totalizers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116
10.7 Problems during the heat flow measurement . . . . . . . . . . . . . . . . . . . . . . . . 116
11 Maintenance and cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117
11.1 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 118
11.2 Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 118
12 Dismounting and disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119
12.1 Dismounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119
12.2 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119
13 Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120
13.1 Configuration of an output using the adapter for the active current input . . . 120
13.2 Installation of a binary output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 122
13.3 Configuration of a frequency output as pulse output . . . . . . . . . . . . . . . . . . . 125
13.4 Activation of a binary output as pulse output . . . . . . . . . . . . . . . . . . . . . . . . . 127
14 Inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 129
14.1 Assignment of the temperature inputs to the measuring channels . . . . . . . . 129
FLUXUS F60* Table of contents
5
UMFLUXUS_F60xV5-0EN, 2017-10-16
14.2 Assignment of other inputs to the measuring channels . . . . . . . . . . . . . . . . 131
14.3 Activation of the inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 132
14.4 Temperature correction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 133
15 Data logger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 136
15.1 Activating/deactivating the data logger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 136
15.2 Setting the storage rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 137
15.3 Configuration of the data logger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 137
15.4 Measurement with activated data logger . . . . . . . . . . . . . . . . . . . . . . . . . . . 140
15.5 Deletion of measured values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 140
15.6 Information relating the data logger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 140
16 Data transmission . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 142
16.1 Online transmission . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 142
16.2 Offline transmission . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 142
16.3 Data format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143
16.4 Transmission parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 144
16.5 Online transmission of data to a terminal program . . . . . . . . . . . . . . . . . . . . 145
16.6 Offline transmission of data to a terminal program . . . . . . . . . . . . . . . . . . . . 146
16.7 Data structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 146
17 Advanced functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 149
17.1 Totalizers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 149
17.2 NoiseTrek parallel beam mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 152
17.3 HybridTrek mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 152
17.4 Upper limit of the flow velocity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 154
17.5 Cut-off flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 155
17.6 Uncorrected flow velocity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 156
17.7 FastFood mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 157
17.8 Calculation channels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 159
17.9 Diagnosis by means of the snap function . . . . . . . . . . . . . . . . . . . . . . . . . . . 163
17.10 Modification of the limit for the inner pipe diameter . . . . . . . . . . . . . . . . . . . 165
17.11 Activation of a binary output as alarm output . . . . . . . . . . . . . . . . . . . . . . . . 165
17.12 Behavior of the alarm outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 170
18 SuperUser mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 174
18.1 Activation/deactivation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 174
18.2 Transducer parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 175
18.3 Definition of the flow parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 175
18.4 Limit of the signal amplification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 178
Table of contents FLUXUS F60*
2017-10-16, UMFLUXUSF60xV5-0EN
6
18.5 Upper limit of the sound speed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 179
18.6 Detection of long measurement failures . . . . . . . . . . . . . . . . . . . . . . . . . . . . 181
18.7 Number of decimal places of the totalizers . . . . . . . . . . . . . . . . . . . . . . . . . . 181
18.8 Temperature-based heat flow cut-off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 182
18.9 Manual reset of the totalizers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 183
18.10 Display of the totalizer sum . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 183
18.11 Display of the last valid measured value . . . . . . . . . . . . . . . . . . . . . . . . . . . . 184
18.12 Displays during the measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 184
19 Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 185
19.1 Dialogs and menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 185
19.2 Measurement settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 189
19.3 Working with parameter records . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 192
19.4 Libraries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 194
19.5 Contrast settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 197
20 Wall thickness measurement (optional) . . . . . . . . . . . . . . . . . . . . . . . . . . 198
20.1 Orientation of the wall thickness probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 199
20.2 Activation of the wall thickness measurement . . . . . . . . . . . . . . . . . . . . . . . . 199
20.3 Parameter input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 200
20.4 Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 201
21 Heat flow measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 206
21.1 Calculation of the heat flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 207
21.2 Normal measuring mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 208
21.3 BTU mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 211
21.4 Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 214
21.5 Two independent heat flow measurements . . . . . . . . . . . . . . . . . . . . . . . . . . 215
21.6 Steam in inlet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 215
Annex
A Menu structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 219
B Units of measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 229
C Reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 235
D Conformity declarations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 241
1 Introduction
FLUXUS F60*
7
UMFLUXUS_F60xV5-0EN, 2017-10-16
1 Introduction
This operating instruction has been written for users operating the ultrasonic flowmeter
FLUXUS. It contains important information about the measuring equipment, how to han-
dle it correctly, and how to avoid damages. Read the safety instructions carefully. Make
sure you have read and understood this operating instruction before using the measuring
equipment.
Any work on the measuring equipment has to be carried out by authorized and qualified
personnel in order to detect and avoid possible risks and dangers.
Presentation of warnings
This operating instruction contains warnings marked as follows:
Danger!
Type and source of danger
danger with high level of risk, which if not avoided, can lead to death or
serious injuries
measures of prevention
Warning!
Type and source of danger
danger with medium level of risk, which if not avoided, can lead to death
or serious injuries
measures of prevention
Caution!
Type and source of danger
danger with low level of risk, which if not avoided, can lead to minor or
moderate injuries
measures of prevention
Notice!
This text contains important information about the use of the transmitter. The infor-
mation describes situations which can lead to material damage.
1 Introduction
FLUXUS F60*
2017-10-16, UMFLUXUSF60xV5-0EN
8
Storage of the operating instruction
The operating instruction must permanently be available at the place where the measur-
ing equipment is used. It must always be available to the user.
User comments
All reasonable effort has been made to ensure the correctness of the content of this oper-
ating instruction. If you however find some erroneous information or miss information,
please inform us.
We will be grateful for any suggestions and comments regarding the concept and your
experience working with the measuring equipment. If you have any suggestions about
improving the documentation and particularly this operating instruction, please let us
know so that we can consider your comments for future reprints.
Copyright
The contents of this operating instruction are subject to changes without prior notice. All
rights reserved. No part of this operating instruction may be reproduced in any form with-
out FLEXIM's written permission.
2 Safety instructions
FLUXUS F60* 2.1 General safety instructions
9
UMFLUXUS_F60xV5-0EN, 2017-10-16
2 Safety instructions
2.1 General safety instructions
Prior to any work, read the operating instruction carefully and in full.
Failure to comply with the instructions, in particular with the safety instructions, poses a
risk to health and can lead to material damages. For further information, contact FLEXIM.
During installation and operation of the measuring equipment, observe the ambient and
installation conditions specified in the documentation.
The measuring equipment has to be checked for proper condition and operational safety
before each use. If disturbances or damages have occurred during installation or opera-
tion of the measuring equipment, please inform FLEXIM.
It is not allowed to make unauthorized modifications or alterations to the measuring
equipment.
If the measuring point is within an explosive atmosphere, the danger zone and present
explosive atmosphere have to be determined. The transmitter, transducers and accesso-
ries have to be appropriate and approved for the conditions in this zone.
The personnel has to be suitably trained and experienced for the work.
Observe the "Safety instructions for the use in explosive atmospheres", see document
SIFLUXUS_608. Observe the instructions for hazardous substances and the respective
safety data sheets. Observe the regulations for the disposal of electrical equipment.
2.2 Intended use
The measuring equipment is intended for the measurement of fluid properties in closed
pipes. By means of connected transducers, the transit times of the ultrasonic signals in
the fluid and the pipe as well as other related properties, such as temperature and pres-
sure, are measured and evaluated.
The transmitter uses these values to calculate the sought physical quantities, e.g., volu-
metric flow rate, mass flow rate, heat quantity, density and concentration. Through com-
parison with the values stored in the transmitter further physical quantities can be deter-
mined. The physical quantities are provided via configurable outputs and the display.
• All instructions of this operating manual have to be observed to ensure intended use.
• Any use beyond or other than the intended use is not covered by warranty and can
present a danger. Any damage arising from not intended use shall be solely the liability
of the operator or user.
• The measurement is carried out without direct contact to the fluid in the pipe. The flow
profile is not influenced.
• The transducers are fixed to the pipe using the supplied transducer mounting fixture.
• Observe the operating conditions, e.g., environment, voltage ranges. For the technical
data of the transmitter, transducers and accessories, see Technical specification.
2 Safety instructions
2.3 Not intended use FLUXUS F60*
2017-10-16, UMFLUXUSF60xV5-0EN
10
2.3 Not intended use
Not intended use in terms of a misuse means:
• any work on the measuring equipment without observing all instructions in this operat-
ing instruction
• use of transmitter, transducer and accessories combinations not intended by FLEXIM
• installation of the transmitter, transducers and accessories in explosive atmospheres
they are not approved for
any work on the measuring equipment (e.g., installation, dismounting, connection, start-
up, operation, service and maintenance) carried out by unauthorized and untrained per-
sonnel
• storage, installation and operation of the measuring equipment outside the specified
ambient conditions, see Technical specification
2.4 Safety instructions for the user
Any work on the transmitter has to be carried out by authorized and qualified personnel.
Observe the safety instructions in the operating instruction. For the technical data of the
transmitter, transducers and accessories, see Technical specification.
• Observe the safety and accident prevention regulations applicable on the site of opera-
tion.
• Only use the supplied mounting fixtures and transducers as well as the intended acces-
sories.
• Always wear the required personal protective equipment.
2.5 Safety instructions for the operator
The operator shall qualify the personnel to perform their assigned tasks. They shall pro-
vide the required personal protective equipment and oblige the personnel to wear it. It is
recommended to risk assess the workplace.
• Besides the safety instructions in this operating instruction, the health, safety and envi-
ronment regulations applicable for the range of application of the transmitter, transduc-
ers and accessories have to be observed.
• With the exceptions stated in chapter 10, the measuring equipment is maintenance-
free. Any components and spare parts may only be replaced by FLEXIM. The operator
shall carry out periodic checks for changes or damages that can present a danger. For
further information, contact FLEXIM.
• Observe the specifications for the installation and connection of the transmitter, trans-
ducers and accessories, see chapter 6 and 7.
2 Safety instructions
FLUXUS F60* 2.6 Safety instructions for electrical work
11
UMFLUXUS_F60xV5-0EN, 2017-10-16
2.6 Safety instructions for electrical work
• Electrical work may only be carried out if there is enough space.
• The degree of protection of the transmitter is only ensured if the cables are firmly con-
nected to the transmitter and all unused connections are covered.
• The condition and tight fit of the electrical connections have to be checked at regular in-
tervals.
• The power supply unit for charging the battery, may only be connected to networks up
to overvoltage category II. Use only the supplied power supply unit. For the power sup-
ply via power cable and power adapter, observe the safety instructions in the sections
7.1.2 and 7.2.2.
• The transmitter and the power supply unit must not be disassembled, see Fig. 2.1. The
transmitter does not contain any components to be maintained by the user. For repair
and service work, please contact FLEXIM.
• Observe the safety and accident prevention regulations for electrical systems and
equipment.
2.7 Safety instructions for transport
• If you detect a transport damage when unpacking the delivery, please contact the sup-
plier or FLEXIM immediately.
• The transmitter is a sensitive electronic measuring instrument. Avoid shocks or im-
pacts.
• Handle the transducer cable with care. Avoid sharp bends or kinks. Observe the ambi-
ent conditions.
• Select a solid surface to put the transmitter, transducers and accessories on.
• The transmitter, transducers and accessories have to be properly packed for transport:
Use, if possible, the original packaging by FLEXIM or an equivalent cardboard box.
Position the transmitter, transducers and accessories in the middle of the cardboard
box.
Fill any voids with appropriate packaging material (e.g., paper, foam, bubble wrap).
Protect the cardboard box against humidity.
Fig. 2.1: Transmitter
2 Safety instructions
2.8 Recommended procedure in hazardous situations FLUXUS F60*
2017-10-16, UMFLUXUSF60xV5-0EN
12
2.8 Recommended procedure in hazardous situations
Fire fighting measures
• If possible, disconnect the transmitter from the power supply.
• Prior to extinguishing, protect any electrical parts that are not affected by the fire (e.g.,
using a cover).
• Select a suitable extinguishing agent. Avoid, if possible, conductive extinguishing
agents.
• Observe the applicable minimum distances. The minimum distances differ depending
on the used extinguishing agent.
3 General principles
FLUXUS F60* 3.1 Measurement principle
13
UMFLUXUS_F60xV5-0EN, 2017-10-16
3 General principles
For the ultrasonic measurement of the flow rate, the flow velocity of the fluid in a pipe is
determined. Further physical quantities are derived from the flow velocity and from addi-
tional physical quantities, if necessary.
3.1 Measurement principle
The flow velocity of the fluid is measured in the TransitTime mode using the ultrasonic
transit time difference correlation principle. When measuring with a high proportion of
gas or solid particles, the transmitter can toggle into the NoiseTrek mode.
3.1.1 Terms
Flow profile
Distribution of flow velocities over the cross-sectional pipe area. For an optimal measure-
ment, the flow profile has to be fully developed and axisymmetrical. The shape of the flow
profile depends on whether the flow is laminar or turbulent and is influenced by the condi-
tions at the inlet of the measuring point.
Reynolds number Re
Coefficient describing the turbulence behavior of a fluid in the pipe. The Reynolds num-
ber Re is calculated from the flow velocity, the kinematic viscosity of the fluid and the in-
ner pipe diameter.
If the Reynolds number exceeds a critical value (usually approx. 2300, if the fluid flows in
a pipe), a transition from a laminar flow to a turbulent flow takes place.
Laminar flow
A flow without any turbulence. There is no mixing between the parallel flowing layers of
the fluid.
Turbulent flow
A flow with turbulences (swirling of the fluid). In technical applications, the flow in the pipe
is mostly turbulent.
Transition range
The flow is partly laminar and partly turbulent.
Sound speed c
Speed of the propagating sound. The sound speed depends on the mechanical proper-
ties of the fluid or the pipe material. In pipe materials and other solid materials, a distinc-
tion is made between the longitudinal and transversal sound speed. For the sound speed
of some fluids and materials, see annex C.
Flow velocity v
Average value of all flow velocities of the fluid over the cross-sectional pipe area.
3 General principles
3.1 Measurement principle FLUXUS F60*
2017-10-16, UMFLUXUSF60xV5-0EN
14
Acoustic calibration factor ka
ka =
The acoustic calibration factor ka is a transducer parameter which results from the sound
speed c within the transducer and the angle of incidence. According to Snell's law of re-
fraction, the angle of propagation in the adjoining fluid or pipe material is:
ka = = =
Fluid mechanics correction factor kRe
With the fluid mechanics correction factor kRe, the measured value of the flow velocity in
the area of the sound beam is converted into the value of the flow velocity across the
whole cross-sectional pipe area. In case of a fully developed flow profile, the fluid me-
chanics correction factor only depends on the Reynolds number and the roughness of
the inner pipe wall. The fluid mechanics correction factor is recalculated by the transmit-
ter for each new measurement.
Mass flow rate
= · ρ
The mass of the fluid that passes through the pipe per unit time. The mass flow rate is
calculated from the product of the volumetric flow rate and the density ρ.
Transit time difference t
Is the transit time difference of the signals. In the TransitTime principle, the transit time
difference of the signals in and against the flow direction is measured, in the NoiseTrek
mode, the time difference of the signal from the transducer to the particle and vice versa.
The flow velocity of the fluid flowing in the pipe is determined from the transit time differ-
ence.
Volumetric flow rate
= v · A
The volume of the fluid that passes through the pipe per unit time. The volumetric flow rate
is calculated from the product of the flow velocity v and the cross-sectional pipe area A.
Heat flow
The heat quantity that is transferred per unit time. For the calculation of the heat flow see
chapter 21.
c
sin
------------
c
sin
------------
c
sin
------------
c
sin
-----------
m
·
m
·
V
·
V
·
V
·
V
·
3 General principles
FLUXUS F60* 3.1 Measurement principle
15
UMFLUXUS_F60xV5-0EN, 2017-10-16
3.1.2 Measurement of the flow velocity in the TransitTime mode
The signals are emitted and received by two transducers alternatively in and against the
flow direction. If the fluid is flowing, the signals propagating in the fluid are displaced with
the flow.
This displacement causes a reduction in distance for the signal in flow direction and an
increase in distance for the signal against the flow direction of the receiving transducer,
see Fig. 3.1 and Fig. 3.2.
This causes a change in the transit times. The transit time of the signal in flow direction is
shorter than the transit time against the flow direction. This transit time difference is pro-
portional to the average flow velocity.
The average flow velocity of the fluid is calculated as follows:
v = kRe · ka ·
where
v average flow velocity of the fluid
kRe fluid mechanics correction factor
ka acoustic calibration factor
t transit time difference
tγ transit time in the fluid
Fig. 3.1: Sound path of the signal in the flow direction
c sound speed
1 transducer (emitter)
2 transducer (receiver)
3 pipe wall
cβ
β
α
γ
cγ
3
21
cαcαdistance reduction
within the transducer
flow direction of
the fluid
sound path with flow
sound path without
flow
3 General principles
3.1 Measurement principle FLUXUS F60*
2017-10-16, UMFLUXUSF60xV5-0EN
16
Fig. 3.2: Sound path of the signal against the flow direction
c sound speed
1 transducer (emitter)
2 transducer (receiver)
3 pipe wall
Fig. 3.3: Transit time difference t
1 signal in the flow direction
2 signal against the flow direction
cβ
flow direction of
the fluid
sound path with flow
sound path without
flow
β
α
γ
cγ
distance increase
within the transducer 3
12
cαcα
1 2
t
3 General principles
FLUXUS F60* 3.1 Measurement principle
17
UMFLUXUS_F60xV5-0EN, 2017-10-16
3.1.3 Measurement of the flow velocity in the NoiseTrek mode
If the proportion of gas or solids in the fluid is high, the damping of the ultrasonic signal
can be so high that the complete propagation of the fluid and therefore a measurement in
TransitTime mode is not possible anymore. In this case the NoiseTrek mode has to be
used.
The NoiseTrek mode uses the presence of gas bubbles and/or solids in the fluids.
Ultrasonic signals are sent from a transducer into the fluid at short intervals, reflected by
the gas bubbles and/or the solids particles and again received by the same transducer.
The measurement setup used in the TransitTime mode does not need to be changed.
The transit time difference t of two consecutive ultrasonic signals is determined. It be-
haves proportionately to the distance the solid particle is covering between two consecu-
tive pulses and thus, to the average flow velocity of the fluid, see Fig. 3.4.
The average flow velocity of the fluid is calculated as follows:
v=k
Re ·k
a·
where
Depending on the attenuation of the ultrasonic signal, the error of measurement in the
NoiseTrek mode can be greater than in the TransitTime mode.
Fig. 3.4: Measurement of the flow velocity in the NoiseTrek mode
v average flow velocity of the fluid
kRe fluid mechanics calibration factor
ka acoustic calibration factor
tp time difference between 2 consecutive pulses
t transit time difference of ultrasonic signals S1 and S2 (t = t2 - t1)
flow direction
transducer 1 transducer 1
flow direction
ultrasonic signal S1
gas bubble or
solids particle
ultrasonic signal S2
pulse at
time t
gas bubble or
solids particle
pulse at
time t + tp
transit time t1transit time t2
t
2tp
----------------
3 General principles
3.1 Measurement principle FLUXUS F60*
2017-10-16, UMFLUXUSF60xV5-0EN
18
3.1.4 Measurement of the flow velocity in the HybridTrek mode
The HybridTrek mode combines the TransitTime and the NoiseTrek mode. During a
measurement in the HybridTrek mode, the transmitter automatically toggles between the
TransitTime and the NoiseTrek mode depending on the gaseous or solid content.
The measurement setup is the same as the one in the TransitTime mode.
3.1.5 Measurement of the flow velocity in the NoiseTrek parallel beam
mode
Pipes with a small pipe diameter or fluids which strongly attenuate the ultrasonic signal
can cause a reduction of the transit time in the fluid with the result that the signal quality
is no longer sufficient. In this case the NoiseTrek parallel beam mode has to be used.
The NoiseTrek parallel beam mode works the same way the NoiseTrek mode does, the
only difference is that the sending and receiving process do not occur in the same trans-
ducer. This leads to a better signal quality. The transducers are mounted in parallel on
the pipe at a small angle, see Fig. 3.5.
This measurement arrangement cannot be used in the TransitTime mode.
Fig. 3.5: Measurement arrangement in the NoiseTrek parallel beam mode
3 General principles
FLUXUS F60* 3.2 Measurement arrangements
19
UMFLUXUS_F60xV5-0EN, 2017-10-16
3.2 Measurement arrangements
3.2.1 Terms
Sound path
The distance covered by the ultrasonic signal after crossing the pipe once. The number
of the sound paths is:
• odd if the measurement is carried out in diagonal arrangement
• even if the measurement is carried out in reflection arrangement
Beam
The path covered by the ultrasonic signal between the transducers, i.e., the transducer
emitting the ultrasonic signal and the transducer receiving it. One beam consists of 1 or
several sound paths.
Diagonal arrangement Reflection arrangement
The transducers are mounted on opposite
sides of the pipe.
The transducers are mounted on the same
side of the pipe.
Fig. 3.6: Diagonal arrangement with
2 beams and 3 sound paths
Fig. 3.7: Reflection arrangement with
1 beam and 4 sound paths
3 General principles
3.2 Measurement arrangements FLUXUS F60*
2017-10-16, UMFLUXUSF60xV5-0EN
20
Transducer distance
The transducer distance is measured between the inner edges of the transducers.
Sound beam plane
Plane containing 1 or several sound paths or beams.
reflection arrangement
diagonal arrangement
(positive transducer distance)
diagonal arrangement
(negative transducer distance)
a – transducer distance
Fig. 3.8: 2 beams in 1 plane Fig. 3.9: 2 sound paths in 1 plane
a
a
a
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133
  • Page 134 134
  • Page 135 135
  • Page 136 136
  • Page 137 137
  • Page 138 138
  • Page 139 139
  • Page 140 140
  • Page 141 141
  • Page 142 142
  • Page 143 143
  • Page 144 144
  • Page 145 145
  • Page 146 146
  • Page 147 147
  • Page 148 148
  • Page 149 149
  • Page 150 150
  • Page 151 151
  • Page 152 152
  • Page 153 153
  • Page 154 154
  • Page 155 155
  • Page 156 156
  • Page 157 157
  • Page 158 158
  • Page 159 159
  • Page 160 160
  • Page 161 161
  • Page 162 162
  • Page 163 163
  • Page 164 164
  • Page 165 165
  • Page 166 166
  • Page 167 167
  • Page 168 168
  • Page 169 169
  • Page 170 170
  • Page 171 171
  • Page 172 172
  • Page 173 173
  • Page 174 174
  • Page 175 175
  • Page 176 176
  • Page 177 177
  • Page 178 178
  • Page 179 179
  • Page 180 180
  • Page 181 181
  • Page 182 182
  • Page 183 183
  • Page 184 184
  • Page 185 185
  • Page 186 186
  • Page 187 187
  • Page 188 188
  • Page 189 189
  • Page 190 190
  • Page 191 191
  • Page 192 192
  • Page 193 193
  • Page 194 194
  • Page 195 195
  • Page 196 196
  • Page 197 197
  • Page 198 198
  • Page 199 199
  • Page 200 200
  • Page 201 201
  • Page 202 202
  • Page 203 203
  • Page 204 204
  • Page 205 205
  • Page 206 206
  • Page 207 207
  • Page 208 208
  • Page 209 209
  • Page 210 210
  • Page 211 211
  • Page 212 212
  • Page 213 213
  • Page 214 214
  • Page 215 215
  • Page 216 216
  • Page 217 217
  • Page 218 218
  • Page 219 219
  • Page 220 220
  • Page 221 221
  • Page 222 222
  • Page 223 223
  • Page 224 224
  • Page 225 225
  • Page 226 226
  • Page 227 227
  • Page 228 228
  • Page 229 229
  • Page 230 230
  • Page 231 231
  • Page 232 232
  • Page 233 233
  • Page 234 234
  • Page 235 235
  • Page 236 236
  • Page 237 237
  • Page 238 238
  • Page 239 239
  • Page 240 240
  • Page 241 241
  • Page 242 242
  • Page 243 243
  • Page 244 244

Flexim FLUXUS F60 Series Operating

Category
Measuring, testing & control
Type
Operating

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI