10
• Power Supply: Connect a 3-wire power cable
to the terminal strip, while paying attention to
the correct line (L), earth (PE) and neutral (N)
terminal connections.
Power: 115VAC - 100 mA / 230VAC - 50 mA.
Line Contact: fused inside 400 mA.
PE must be connected to ground; leakage current 1 mA.
• Electrode: Connect the pH or ORP electrode to the BNC
socket (#10 at page 6).
To benefit from the differential input, connect the proper
electrode wire (if available)
or a cable with a potential
matching pin (grounding bar) to the relevant terminal
(#5 at page 7).
Note When it is not possible to immerse the Potential Matching Pin
together with the pH electrode in the solution, disable the
differential input by connecting the Connection for Potential
Matching Pin (#5 page 7)
with the Connection for
Electrode Reference (#6 on
page 7) with a jumper wire.
• Pt 100 Terminals: these contacts (#3 at page 7) connect
the Pt 100 temperature sensor for automatic temperature
compensation of pH measurement. In the case of shielded
wire, connect the shield to pin 4.
In the case of a 2-wire sensor
connect the Pt 100 to pins 1 and
3, and short pins 2 and 3 with a
jumper wire.
If the Pt 100 has more than 2 wires,
connect the two wires of one end to
pins 2 and 3 (pin 2 is an auxiliary
input to compensate for the cable resistance) and one wire
from the other end to pin 1. Leave the
fourth wire unconnected, if present.
• Power Supply Output: these terminals
provide +5V and -5V DC (max current: 10 mA) signals
to supply power to amplified electrodes.
Note All cables connected to the panel should end with cable
lugs.