BD Alaris™ MRI Capsule Quick start guide

Type
Quick start guide
BD Alaris™ MRI Capsule
Quick Reference Guide – en
BDQR00070_Issue2.indd 1 20/11/2018 08:22:26
BDQR00070 Iss. 2 2/14
Features of the MRI Capsule
This BD Alaris™ MRI Capsule (hereinafter referred to as MRI
Capsule) guide supports the following Pumps:
Alaris™ CC Syringe Pump
Alaris™ GH Syringe Pump
Alaris™ PK Syringe Pump
Alaris™ VP Volumetric Pump
This guide is not intended to be comprehensive instructions for the
set-up and operation of the BD Alaris™ MRI Capsule.
For complete BD Alaris™ MRI Capsule information, refer to the
Directions For Use.
All persons operating the BD Alaris™ MRI Capsule should consult
the Directions For Use.
Height adjustable
Pole with bag
hanger hooks
Pump Alarm
Location Beacon
Keypad with Indicators
Door handle
with a lock
Grooves for
Infusion Sets from
Infusion Pumps
Grooves for
Infusion Sets to
Volumetric Pumps
Manual Brakes
Automatic Brakes
Shielded window
BDQR00070_Issue2.indd 2 20/11/2018 08:22:28
BDQR00070 Iss. 2 3/14
Magnetic Indicator
Handle used to position
MRI Capsule with lever
for automatic brakes
AC Power cable
storage hooks
AC Power Inlet
Features of the MRI Capsule continued
BDQR00070_Issue2.indd 3 20/11/2018 08:22:30
BDQR00070 Iss. 2 4/14
Controls and Indicators
Power
on / o
AC Power
Indicator
System Fault
Indicator
Battery
Indicator
On Status
Indicator
A Status Indicator
(Software indicator)
B Status Indicator
(Communication
indicator)
The MRI Capsule is in position for
safe operation.
Move the MRI Capsule until
the Magnetic Indicator green
LED flashes.
Move the MRI Capsule until the
Magnetic Indicator green LED
flashes. If the Pump continues
to infuse without alarming,
it should be checked at the
earliest opportunity.
BDQR00070_Issue2.indd 4 20/11/2018 08:22:30
BDQR00070 Iss. 2 5/14
Positioning the MRI Capsule
Fringe field of
MR Scanner
Typical values for distances in Z-direction to
opening of magnet bore
MR Scanner: 3.0 Tesla MR Scanner: 1.5 Tesla
200mT / 2000G 0.6m 0.5m
70mT / 700G 0.9m 0.8m
40mT / 400G 1.1m 1.0m
30mT / 300G 1.2m 1.1m
20mT / 200G 1.4m 1.2m
10mT / 100G 1.7m 1.5m
5mt / 50G 2.1m 1.8m
1. Release the foot operated brakes.
2. Activate the hand operated brake lever.
3. Manoeuvre the MRI Capsule into position, ensuring the Magnetic Indicator
remains green.
4. Lock the trolley in position with the MRI Capsule window facing the control room
window. Ensure the automatic brakes and two manual brakes are engaged.
BDQR00070_Issue2.indd 5 20/11/2018 08:22:31
BDQR00070 Iss. 2 6/14
Switching On
Note: Magnetic Indicator is permanently on and is not aected by switching
on/o the MRI Capsule.
1. Connect the AC Power cord from the AC Power Supply to the AC Power inlet socket
at the back of the MRI Capsule.
2. Verify the AC Power indicator is illuminated.
3. Press the key once to switch the MRI Capsule on.
Note: When powered on the MRI Capsule will illuminate the alarm location
beacon, red then amber, and generate an audible tone.
4. After initially switching on the MRI Capsule, it may take up to 90 seconds to
become fully operational.
Do not switch off the MRI Capsule during this initial 90 second period.
If any of the verification checks fail when the MRI Capsule is switched
on, suspect a fault. Remove the MRI Capsule from service and contact
Qualified Service Personnel.
Operating the MRI Capsule
BDQR00070_Issue2.indd 6 20/11/2018 08:22:31
BDQR00070 Iss. 2 7/14
Operating the MRI Capsule continued
Switching O
Press the key and hold for two seconds to switch the MRI Capsule o.
Resetting the MRI Capsule
In the unlikely event that the MRI Capsule needs to be reset, press and hold the
key for at least four seconds until the ON Status Indicator is extinguished, release the
key then press again to switch the MRI Capsule back on.
If the MRI Capsule still fails to operate correctly after resetting, remove
the MRI Capsule from service and contact Qualified Service Personnel.
Resetting the MRI Capsule is also required to clear any System Fault
Indicator.
BDQR00070_Issue2.indd 7 20/11/2018 08:22:31
BDQR00070 Iss. 2 8/14
To open the door
1. Lift the door handle
2. Turn the door handle in a anti-
clockwise direction
3. Open the MRI Capsule door
To close the door
1. Close the MRI Capsule door
2. Turn the door handle in a clockwise
direction
3. Lower the door handle
Operating the door
BDQR00070_Issue2.indd 8 20/11/2018 08:22:33
BDQR00070 Iss. 2 9/14
Only dock the compatible Pumps into the MRI Capsule.
1. Align the rotating cam on the rear of the Pump with the rectangular bar on the
MRI Capsule tile.
2. Hold the Pump horizontally, push the Pump firmly onto the rectangular bar.
3. The Pump should click into position when fitted to the bar.
4. Verify Pump is secure by gently pulling the Pump away from the MRI Capsule tile
without using the release lever. When the Pump is securely attached, the AC Power
Indicator LED on the pump will be lit.
Docking the Pumps
BDQR00070_Issue2.indd 9 20/11/2018 08:22:34
BDQR00070 Iss. 2 10/14
Hand positioning for docking Pump on top left MRI Tile
Hand positioning for undocking Pump on top left MRI Tile
Best practice for docking and undocking Alaris™ VP Plus Guardrails Volumetric Pumps
BDQR00070_Issue2.indd 10 20/11/2018 08:22:37
BDQR00070 Iss. 2 11/14
w
Warning: Positioning of hand, as shown below, when handling the
Pump in the MRI Capsule could result in user harm.
Best practice for docking and undocking Alaris™ VP Plus Guardrails Volumetric Pumps continued
BDQR00070_Issue2.indd 11 20/11/2018 08:22:37
BDQR00070 Iss. 2 12/14
3 Volumetric and 3 Syringe Pumps
4 Volumetric Pumps
4 Syringe Pumps
Available Configurations
BDQR00070_Issue2.indd 12 20/11/2018 08:22:38
BDQR00070 Iss. 2 13/14
Pump Alarm Priority Colour Frequency
HIGH Priority Red Flashing faster than Medium Priority
MEDIUM Priority Yellow Flashing or solid
LOW Priority Yellow Solid
Pump Alarm Location Beacon
BDQR00070_Issue2.indd 13 20/11/2018 08:22:38
bd.com
BD, the BD Logo and Alaris are trademarks of Becton, Dickinson and Company.
© 2018 BD and its subsidiaries. All rights reserved.
BDQR00070_Issue2.indd 14 20/11/2018 08:22:38
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14

BD Alaris™ MRI Capsule Quick start guide

Type
Quick start guide

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI