BDQR00070 Iss. 2 6/14
Switching On
Note: Magnetic Indicator is permanently on and is not aected by switching
on/o the MRI Capsule.
1. Connect the AC Power cord from the AC Power Supply to the AC Power inlet socket
at the back of the MRI Capsule.
2. Verify the AC Power indicator is illuminated.
3. Press the key once to switch the MRI Capsule on.
Note: When powered on the MRI Capsule will illuminate the alarm location
beacon, red then amber, and generate an audible tone.
4. After initially switching on the MRI Capsule, it may take up to 90 seconds to
become fully operational.
Do not switch off the MRI Capsule during this initial 90 second period.
If any of the verification checks fail when the MRI Capsule is switched
on, suspect a fault. Remove the MRI Capsule from service and contact
Qualified Service Personnel.
Operating the MRI Capsule
BDQR00070_Issue2.indd 6 20/11/2018 08:22:31