5
also has measurement records that can be printed out and lled in
manually.
– Enter the room temperature and relative humidity in the
measurement record.
– Normalize the moisture reading (this is necessary if the room
temperature is not 20 °C and the relative humidity is not 50%).
Normalization is performed automatically if the Excel measurement
record is used.
For manual normalization, the table on page 2 of these instructions
should be used. This table is also on the CD-ROM accompanying the
StarterPac. Enter the normalized moisture reading (%MC
N
) in
the measurement record.
5 Evaluation
Compare the normalized moisture reading to the control limits (page 2):
– If the value is within the tolerance range, the instrument has passed
the functional test;
– If the value is outside the tolerance range, there may be a problem with
the instrument or the test conditions may not have been met.
For information on evaluating the result, see the operating instructions
accompanying the StarterPac or visit www.mt.com/smartcal.
Product information
– Material: Molecular sieve (zeolite). This material is not classied as
hazardous. Avoid inhalation or ingestion. For more information, see
the material safety data sheet (MSDS) available at www.mt.com/msds.
– Storage: At room temperature. Do not open the blister pack before use.
– Disposal: Can be disposed of as normal waste. Observe local and
national environmental regulations.
– Shelf life: The expiry date is printed on the blister pack and on the
stick (e.g. Exp08.2013).
– Certicate: Certicate of Analysis (
cSmartCal) and Production Certi-
cate (SmartCal) are available at www.mt.com/smartcal-certicate.