  • Hello! I am an AI chatbot trained to assist you with the Mettler Toledo Test Procedure for Moisture Analyzer Routine Testing Operating instructions. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
Routine Testing
SOP for Periodic SmartCal
Test
of Moisture Analyzers
Title: SOP for Periodic SmartCal test with Halogen
Moisture Analyzers
Document No.:
Author: METTLER TOLEDO Laboratory & Weighing Technologies
Number of Pages: 4
Controls
Replacement Document: N/A
Reason for Revision: N/A
Release Date: 21. Aug 2013
Release for Routine Operation
Reviewed by:
Date:
Signature:
Approved by:
Date:
Signature:
2 of 4
SOP SmartCal
METTLER TOLEDO
2 of 4
Routine Testing of Moisture Analyzers
General
The SmartCal test is a rapid test which checks the moisture analyzer’s overall functionality. SmartCal is a
granular substance which releases a defined amount of moisture depending on the selected drying tempera-
ture. Intended use is the verification of METTLER TOLEDO Halogen Moisture Analyzers.
Basic Rules for Handling Moisture Analyzers
•Beforeusingamoistureanalyzer,makesuretheinstrumentwasleftonpowerforasufficientperiodoftime
(mentioned in the user manual).
•Makesurethemoistureanalyzerisleveled.
•Minimizeenvironmentalinfluences,e.g.draftordirectsunlight.
•Distributesamplesgentlyandevenlyonthesamplepan.
Basic Rules for Handling SmartCal
Important to know
•Thegranulartestsubstanceisdouble-packedinastickinsideablisterpack.
•Thestickshouldonlyberemovedfromtheblisterpackimmediatelybeforeuse.
•Donotuseanydamagedorexpiredsticks.
•AftertheSmartCaltest,theusedtestsubstancecanbedisposedofasnormalwaste.
How to Store SmartCal
•SmartCalshouldbestoredinthesameroomasthemoistureanalyzertheyareusedwith,sincetemperature
differences between the test substance and their surroundings can lead to measurement errors.
How to Handle SmartCal
• TheSmartCaltestresultrequiresanormalizationtoambientconditions(%RHandroomtemperature).
• WhenperformingaSmartCaltest,usethesamplepanhandler.
• Usethecompletecontentsofthestick(approx.weight8.5g).
3 of 4
SOP SmartCal
METTLER TOLEDO
3 of 4
1)
– Values within the control limit: no action is necessary.
–Valuesbeyondthecontrollimitshowthatthemoistureanalyzerisnolongerundercontrolandimmediateactionisthereforerequired.
SmartCal Routine Test
Preparation
•Beforethetestisperformed,SmartCaltestsubstanceandthethermohygrometermustbeacclimatizedtothe
ambient temperature of the moisture analyzer.
•Whenthetestisperformed,theheatingunitofthemoistureanalyzershouldbeatroomtemperature.
•Ifasensitivitytestneedstobedoneaswell,pleaseperformthesensitivitytestfirst(toavoidwaitingtimefor
cooling down).
•UseinstrumentpredefinedtestproceduresorsetinstrumenttotheSmartCaltestparameters:
–Switch-offtime:10minutes
– Drying program: standard drying
–Display:%MC
–Dryingtemperature:70,100,130or160°C
(Select the temperature closest to the drying temperature normally used)
Test Procedure
•Ifrequiredbyinstrumentmenuenterambienttemperatureandambienthumidity.
•Placethesamplepanhandlerwiththealuminumsampledishintheinstrumentandtare.
•RemoveaSmartCalstickfromtheblisterpack,openitanddistributetheentirecontentsevenlyoverthe
sample pan (whole pan covered with granules).
•Startthetestmeasurementimmediatelyafterdistributingthesample.
•Readthedisplayedresult(%MC,%MC
N
) from the instrument.
•IftheresultisshownasMC%recordambienttemperatureandambienthumidityaswell.
Evaluation
•Iftheunitofthedisplayedresultis%MC
N
,nonormalizationisneeded.Iftheunitofthedisplayedresultis
MC%normalizethedisplayedresult–usingtheambientconditionsrecordedabove–eithermanuallywith
normalizationtableorwiththeExcelmeasurementreport.
•Evaluatewhetherthenormalizedresultexceedsthedefinedcontrollimits.
1)
Deviation
Control Limit
1)
•Ifthecontrollimitisexceeded,reduceerrorsources(seeSmartCalusermanual)andrepeatthetest
(after cooling down of the moisture analyzer).
•Ifthecontrollimitisexceededagain,reportsupervisorthatthecontrollimitwasnotmet.
•Letthemoistureanalyzercooldown,performweightandthentemperatureadjustment,letthemoisture
analyzer cool down and repeat the SmartCal test.
•Ifthecontrollimitisstillexceeded,reporttheproblemtothesupervisor.Optionally,contactMETTLER
TOLEDO’s service organization for advice.
•Markthemoistureanalyzeras“outofcontrollimits”.
4 of 4
SOP SmartCal
METTLER TOLEDO
Recommended Control Limits for SmartCal Test
Control limits are valid for METTLER TOLEDO Halogen Moisture Analyzers.
Drying Temperature cSmartCal SmartCal
70°C 3.3-4.3%MC
N
3.2-4.4%MC
N
100°C 5.3-6.3%MC
N
5.2-6.4%MC
N
130°C 7.5-8.7%MC
N
7.4-8.8%MC
N
160°C 10.0-11.6%MC
N
9.9-11.7%MC
N
Mettler-Toledo AG