Dealer: This manual MUST be given to the end
user.
User:
Before using this cylinder, read this
manual and save for future reference.
For more information regarding
Invacare products, parts, and services,
please visit www.invacare.com
Operator’s Manual
HomeFill
Oxygen Conserving
Cylinder
HF2PCL6
HF2PC9
HF2PCL4
WARNING
HomeFill
2 Part No 1112648
WARNING
DO NOT OPERATE THIS PRODUCT OR ANY
AVAILABLE OPTIONAL EQUIPMENT WITHOUT
FIRST COMPLETELY READING AND
UNDERSTANDING THESE INSTRUCTIONS AND
ANY ADDITIONAL INSTRUCTIONAL MATERIAL
SUCH AS OWNER’S MANUALS, SERVICE
MANUALS OR INSTRUCTION SHEETS SUPPLIED
WITH THIS PRODUCT OR OPTIONAL
EQUIPMENT. IF YOU ARE UNABLE TO
UNDERSTAND THE WARNINGS, CAUTIONS, OR
INSTRUCTIONS, CONTACT A HEALTHCARE
PROFESSIONAL, DEALER OR TECHNICAL
PERSONNEL BEFORE ATTEMPTING TO USE
THIS EQUIPMENT - OTHERWISE INJURY OR
DAMAGE MAY OCCUR.
ACCESSORIES WARNING
Invacare products are specifically designed and
manufactured for use in conjunction with Invacare
accessories. Accessories designed by other manufacturers
have not been tested by Invacare and are not
recommended for use with Invacare products.
NOTE:Updatedversionsofthismanualareavailableon
www.invacare.com.
TABLE OF CONTENTS
Part No 1112648 3 HomeFill
TABLE OF CONTENTS
SPECIAL NOTES ..........................................................4
RECEIVING INSPECTION .............................................5
INTENDED USE ...........................................................5
SECTION 1—GENERAL GUIDELINES .............................6
Operating Information.............................................................................................6
Maintenance ................................................................................................................8
SECTION 2—FEATURES ...............................................9
SECTION 3—TYPICAL PRODUCT PARAMETERS ........10
SECTION 4—INSPECTION AND COMPRESSOR
INTERFACE ..............................................................11
Cylinder Prefill Inspection.................................................................................... 11
Connecting/Disconnecting the Cylinder to/from the Compressor....... 14
SECTION 5—OPERATING INSTRUCTIONS .................19
SECTION 6—MAINTENANCE ....................................21
SECTION 7—CYLINDER FILL TIMES ............................22
Troubleshooting...................................................................................................... 22
Replacement Parts - Accessories ...................................................................... 22
LIMITED WARRANTY ...............................................23
SPECIAL NOTES
HomeFill
4 Part No 1112648
SPECIAL NOTES
Signalwordsareusedinthismanualandapplytohazardsor
unsafepracticeswhichcouldresultinpersonalinjuryor
propertydamage.Refertothefollowingtablefordefinitions
ofthesignalwords.
NOTICE
Theinformationcontainedinthisdocumentissubjectto
changewithoutnotice.
DANGER
DONOTSMOKEwhileusingthisdevice.Keepall
matches,litcigarettesorothersourcesofignitionoutofthe
roominwhichthisproductislocatedandawayfromwhere
oxygenisbeingdelivered.
NOSMOKINGsignsshouldbeprominentlydisplayed.
Textilesandothermaterialsthatnormallywouldnotburn
areeasilyignitedandburnwithgreatintensityinoxygen
enrichedair.Failuretoobservethiswarningcanresultin
severefire,propertydamageandcausephysicalinjuryor
death.
SIGNAL WORD MEANING
DANGER Danger indicates an imminently hazardous
situation which, if not avoided, will result in
death or serious injury.
WARNING Warning indicates a potentially hazardous
situation which, if not avoided, could result
in death or serious injury.
CAUTION Caution indicates a potentially hazardous
situation which, if not avoided, may result in
property damage or minor injury or both.
RECEIVING INSPECTION
Part No 1112648 5 HomeFill
CAUTION
“Caution:Federallawrestrictsthisdevicetosalebyorrental
orontheorderofaphysician,oranyotherpractitioner
licensedbythelawofthestateinwhichhe/shepracticesto
useorordertheuseofthisdevice.”
RECEIVING INSPECTION
Removeproductfrompackageandinspectfordamage.If
thereisanydamage,DONOTuse.Contactyourdealer/
HomeCareequipmentprovider.
INTENDED USE
Tostore,regulate,anddeliversupplementaloxygento
patientswhomayhavedifficultyextractingoxygenfromthe
airtheybreathe.Itisforpatientswhowouldnormallyreceive
theoxygenviaanasalcannula.ThedevicedeliversUSP93%
oxygenatanumberofdifferentflow/pulsesettings.Itis
intendedtobeusedasanoxygensavingdevicethatmay
reducethedryingoutoftheairways.
SECTION 1—GENERAL GUIDELINES
HomeFill
6 Part No 1112648
SECTION 1—GENERAL GUIDELINES
Inordertoensurethesafeinstallation,assemblyand
operationoftheHomeFillOxygenConservingCylinderthese
instructionsMUSTbefollowed.
WARNING
This section contains important information for the safe operation
and use of this product.
Operating Information
Onlypersonnelinstructedandtrainedintheuseofthe
HomeFillO2ConservingCylindershouldoperatethis
product.
Productcontainsmagnetic,ferrousmaterialthatmayaffect
theresultsofanMRI.
Besureallconnectionsaretightandleakfree.
DONOTautoclave.
DONOTgassterilizewithethyleneoxide.
DONOTcleanwitharomatichydrocarbons.
DONOTimmerseproductinanykindofliquid.
Storeproductinacleanareawhennotisuse.
OnlyuseInvacareCorporationHF2PCL6carryingbag,
HF2PC9carryingbag,orHF2PCL4carryingbag.
NEVERsmokeinanareawhereoxygenisbeing
administered.
ALWAYSconfirmprescribeddosebeforeadministeringto
patientandmonitoronafrequentbasis.
DONOTuseifdirtorcontaminantsarepresentonoraround
cylinder,valve,productorconnectingdevices.
SECTION 1—GENERAL GUIDELINES
Part No 1112648 7 HomeFill
NEVERusenearanytypeofflameorflammable/explosive
substances,vaporsoratmosphere.
ALWAYSfollowANSIandCGAstandardsforMedicalGas
ProductsandFlowmeters(E7)andOxygenHandling(G4).
NOOXYGENisdeliveredwhenthepointerisalignedwith
“Off”.
DONOTblocktheoutletfittingorkinkthecannulawhenthe
productisinuse,thismaydamagetheunit.
Theuseofthisdeviceislimitedtotheoxygenpatient.
CylindersMUSTbeusedonlybytheoxygenpatientandare
nottobedistributedtoanyotherindividualforanypurpose.
ThisproductisNOTtobeusedbypatientswhobreathe
throughtheirmouths.
DONOTuseoils,greases,organiclubricantsorany
combustiblematerialsonornearthisproduct.Washanddry
handsproperlypriortousage.
DONOTallowoilfromyourhandsorothersourcestocome
intocontactwiththecompressorcouplingorbottlefittings.
Thesesubstancesmaybecomeflammableinthepresenceof
oxygenandcauseinjury.Avoidtouchingthebottlefitting
wheneverpossible.Exampleofitemstoavoidarehand
creamsorlotions,cookingoils,suntanoilsorsunblocksand
similaroilyproducts.Ifyouusethesetypesofproducts,wash
yourhandspriortousingtheHomeFillunit.Ifthebottle
fittingshouldcomeincontactwithsuchsubstancesitshould
becleanedoffwithadampclothpriortoconnectingittothe
compressor.
NEVERoilorlubricatethecompressorcouplerorbottle
fittings.NEVERuseanypenetratingoilorlubricantsuchas
WD40
®
or3in1oil
®
.Ifconnectiondifficultyisexperienced
verifythatthebottlefittingisbeinginsertedstraightdown
intothecompressorcoupler.Ifcontinueddifficultyis
experiencedcontactyoursupplierforassistance.
SECTION 1—GENERAL GUIDELINES
HomeFill
8 Part No 1112648
DONOTuseahumidifierwiththisproduct.
DONOTallowcylinderstotiporfall.Securegascylindersso
theycannotfall.Ensurecylinderisalwaysinuprightposition.
DONOTstorecylindersnearsourcesofheatorflame.
Thisproductisequippedwithareliefvalve.Ifyouhearaloud
hissingorpoppingsoundcomingfromtheproduct,
discontinueuseofproduct,turntheflowselectortothe“Off”
position,andcontactyourdealer/HomeCareequipment
provider.
DONOTusewhilesleepingwithoutconsultingyour
healthcareprovider.
Thisproductisdesignedtooperatewithastandardsingle
lumen,adultoxygennasalcannulawithamaximumlengthof
sevenfeet.
Thecannulaisforsinglepatientuseonly.
CannulaMUSTbeinstalledonoutletconnectionpriortouse.
DONOTplacetheunitunderclothingwhileinuse.When
thisproductisinuseasmallamountofoxygenisvented.
Wearingtheunitunderclothingmaysaturatefabricswith
oxygenandcausethemtoburnrapidlyifexposedtosparks
orflame.Itmaytakeseveralhoursforoxygenlevelsinfabrics
toreturntonormal.
Maintenance
DONOTusealiquidleakdetectortotestforleaks.
DONOTattempttorepairproduct.
SECTION 2—FEATURES
Part No 1112648 9 HomeFill
SECTION 2—FEATURES
DANGER
Thisproductisnotintendedasalifesustainingor
lifesupportingdevice.
FIGURE 2.1 Product Diagram
Outlet
Connection
Cannula
Flow Dial
Cylinder
Gauge
Cylinder Fill
Port
SECTION 3—TYPICAL PRODUCT PARAMETERS
HomeFill
10 Part No 1112648
SECTION 3—TYPICAL PRODUCT
PARAMETERS
CONSERVER S/N RANGE 1-69,999 75,000 +
INLET PRESSURE RANGE: 300 - 2100 psig 300 - 2100 psig
INTERNAL REGULATED
PRESSURE: 50 psig 50 psig
DIMENSIONS:
HF2PCL6: Weight 4.3 lbs (1.95 kg) 4.1 lbs (1.86 kg)
HF2PCL6: Overall Length 11.1 in (28 cm) 10.9 in (27.7 cm)
HF2PCL6: Diameter 4.4 in (11.1 cm) 4.4 in (11.1 cm)
HF2PC9: Weight 5.3 lbs (2.4 kg) 5.1 lbs (2.31 kg)
HF2PC9: Overall Length 14.2 in (36 cm) 14 in (35.5 cm)
HF2PC9: Diameter 4.4 in (11.1 cm) 4.4 in (11.1 cm)
HF2PCL4: Weight N/A 3.6 lbs (1.63 kg)
HF2PCL4:Overall Length N/A 8.9 in (22.2 cm)
HF2PCL4:Diameter. N/A 4.4 in (11.1 cm)
PULSE SETTINGS: 1,2,3,4, & 5
FLOW SETTINGS: 2 LPM Continuous
SAVINGS RATIO: Up to 4.3:1 Up to 5.7:1
CANNULA
REQUIREMENT:
4 ft recommended (MS0204) (7 ft long max)
std. adult single lumen oxygen nasal cannula.
OPERATING CONDI-
TIONS TEMPERATURE:
32°F (0°C) to 105°F
(40.6°C)
35°F (1.7°C) to 105°F
(40.6°C)
ALTITUDE: 0 to 5280 ft (0
to1600 m)
0 to 10,000 ft (0 to
3071 m)
STORAGE CONDITIONS
TEMPERATURE:
-40°F (-40°C) to
140°F (60°C)
-40°F (-40°C) to 140°F
(60°C)
MAXIMUM HUMIDITY: 95%
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE
Part No 1112648 11 HomeFill
SECTION 4—INSPECTION AND
COMPRESSOR INTERFACE
NOTE:RefertotheHomeFillCompressorOwner’sManualfor
compressoroperatinginstructions.
Cylinder Prefill Inspection
WARNING
AllcylindersMUSTbeinspectedbeforeattemptingtofill;
otherwise,injuryordamagemayoccur.
NEVERsmokeinanareawhereoxygenisbeing
administered.
NEVERusenearanytypeofflameorflammable/explosive
substances,vaporsoratmosphere.
NOTE:Forthisprocedure,refertoFIGURE4.1.
The Hydrostatic Testing Date
AluminumcylindersMUSTundergotestingeveryfiveyears.
WARNING
DONOTfillcylindersthathavenotbeentestedinthepast
fiveyears.Otherwise,seriouspersonalinjurymayresult.
Contactyourdealerforreplacement.
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE
HomeFill
12 Part No 1112648
External Examination
1. Examinetheoutsideofthecylinderforthefollowing
conditions,andreplacethecylinderiftheyexist:
Dentsordings
ArcBurns
OilorGrease
Anyothersignsofdamagethatmightcauseacylinder
tobeunacceptableorunsafeforuse.
2. Examinethecylinderforevidenceoffireorthermal
damage.Evidenceincludescharringorblisteringofthe
paint,orotherprotectivecoatingorheatsensitive
indicator.Iffireorthermaldamageisfound,replacethe
cylinder.
3. Inspectthecylinderfillportforthefollowing,andreplace
iffound:
Debris,oilorgrease
Noticeablesignsofdamage
Signsofcorrosioninsidethevalve
Signsofexcessiveheatorfiredamage
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE
Part No 1112648 13 HomeFill
FIGURE 4.1 External Examination
Improper filling or abuse of this cylinder, or failure to follow these safety
instructions may cause serious bodily injury or death. High pressure
oxidizing gas vigorously accelerates combustion.
DO NOT alter or modify cylinder or related components.
DO NOT use oil or grease. Keep cylinder, cylinder valve, all equipment
and connections free of oil and grease to avoid violent ignition resulting
in serious bodily injury or death.
For medical applications. Use only as directed by a licensed
practitioner. Uninterrupted use of high concentrations of oxygen over
a long duration, without monitoring its effect on oxygen content of
arterial blood, may be harmful.
Oxygen Cylinders filled by the Invacare HomeFill oxygen
system shall be used for personal use only. "Not to be
filled for resale or use by professional users."
This cylinder is to be used with the Invacare HomeFill
oxygen system. This cylinder contains 93 ± 3% oxygen
produced by an oxygen concentrator using the pressure swing
adsorption process.
DO NOT handle cylinder or use contents until you are professionally
trained, including emergency procedures.
Use in accordance with Invacare HomeFill oxygen system Operator’s
Manual. Open valve slowly. Close valve after each use and when
empty. Secure cylinder during storage and use. No smoking in cylinder
area. Keep away from heat, flame and spark. Keep out of reach of
children. DO NOT drop.
Filling of this gas is performed by Invacare HomeFill oxygen system
Only.
Cylinders that have been refinished and/or exposed to elevated
temperatures must be hydrostatically tested before filling. Cylinder
temperature should not exceed 125°F (52°C). Aluminum cylinders
subject to the action of fire or heated to temperatures in excess of
350°F must be withdrawn from service and condemned by trained
personnel. DO NOT use caustic paint strippers. Valve and safety
relief devices must be removed and replaced by trained and
authorized personnel. DO NOT alter or change valve assembly.
DO NOT REMOVE THIS PRODUCT LABEL
Not for Commercial Resale
Part Number 1125090 Rev D
Invacare Corporation
2101 E. Lake Mary Blvd.
Sanford, FL 32773
1-800-832-4707
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE
HomeFill
14 Part No 1112648
Connecting/Disconnecting the Cylinder
to/from the Compressor
Connecting the Cylinder to the Compressor
WARNING
NEVERusetoolsofanykindtoconnect/disconnectthe
cylinderandthecompressor.Otherwise,severeinjuryand/or
damagemayoccur.
DONOTdropoxygencylinders.Usetwohandswhen
handling/transportingoxygencylinders.Otherwise,injuryor
damagemayoccur.
DONOTusealiquidleakdetectortotestforleaks.
NOTE:Forthisprocedure,refertoFIGURE4.2andFIGURE4.3.
1. Examinethecylindergauge.Ifthecylinderpressureisless
than1500psig.,proceedtoSTEP2(FIGURE4.3).
NOTE:Ifthecylinderpressureisgreaterthan1500psig.,DO
NOTattempttotopoffthiscylinder.Itmaynotfill.
2. SettheregulatorflowdialonthecylindertoOFF(FIGURE
4.3).
3. Removethecylinderandconnectorfillportcovers(if
present).
CAUTION
DONOTconnecttheoutlet(cannula)connectiontothe
HomeFillunit,otherwisetheconservingcylinderwillnot
workproperly.
4. MomentarilypushDOWNontheouterring(sleeve)ofthe
connectorfillporttoresettheconnectorasshownin
FIGURE4.2.
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE
Part No 1112648 15 HomeFill
NOTE:Iftheouterring(sleeve)isintheUPposition
(GREENdotsnotvisible),theconnectorfillportwillnot
beabletoacceptthecylinderfillport.PushingDOWN
momentarilywillresettheconnectorfillport(GREEN
dotsvisible)toacceptthecylinderfillport.
FIGURE 4.2 Resetting Connector Fillport
5. Graspthecylinder/regulatorassemblyintheareabehind
thecylindergauge(FIGURE4.3).
6. Positionthecylinderinthecompressorcradle(FIGURE
4.3).
7. Alignthecylinderfillportwiththeconnectorfillport
(FIGURE4.3).
8. PullUPontheouterring(sleeve)oftheconnectorfillport
whilepushingDOWNonthecylinder/regulatorassembly
tocouplethecylinderfillportintotheconnectorfillport
(FIGURE4.3).
NOTE:Thecylinderisproperlyconnectedwhenanaudible“click”
isheard.
BEFORE
COUPLING
CYLINDER,
PUSH DOWN
ON SLEEVE
UNTIL
GREEN DOTS
ARE VISIBLE.
Connector Fillport
Outer Ring (Sleeve)
PUSH
DOWN
GREEN Dots
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE
HomeFill
16 Part No 1112648
FIGURE 4.3 Connecting the Cylinder to the Compressor
Cylinder/Regulator
Assembly
Compressor
Cylinder
Gauge
Cylinder
Fillport
Connector
Fillport
PULL UP
Outer Ring
(Sleeve)
Cylinder
Compressor
Cradle
Cylinder
Gauge
Regulator
Flow Dial
Outer Ring
(Sleeve)
Connector
Fillport
Outlet (Cannula)
Connection
Power
Switch
Regulator
Cylinder
Fillport
NOTE:Fillportcoversnotshownforclarity.
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE
Part No 1112648 17 HomeFill
Disconnecting the Cylinder from the Compressor
WARNING
NEVERusetoolsofanykindtoconnect/disconnectthe
cylinderandthecompressor.Otherwise,severeinjuryand/or
damagemayoccur.
NOTE:Forthisprocedure,refertoFIGURE4.4.
1. PressthecompressorpowerswitchtotheOff(O)position.
2. Graspthecylinder/regulatorassemblyintheareabehind
thecylindergauge.
3. Withtheotherhand,grasptheouterring(sleeve)ofthe
connectorfillportandpushDOWN.
4. Liftuponthecylinder/regulatorassemblytoremovefrom
theconnectorfillport.
FIGURE 4.4 Disconnecting the Cylinder from the Compressor
Cylinder/Regulator
Assembly
Compressor
Cylinder
Gauge
Cylinder
Fillport
Connector
Fillport
Outer Ring
(Sleeve)
PUSH DOWN
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE
HomeFill
18 Part No 1112648
5. Whenthecylinderfillportisdisconnectedfromthe
connectorfillport,releasetheouterring(sleeve)ofthe
connectorfillportandusetwohandstoremovethe
cylinderfromthecompressorcradle.
6. Placethefillportcoversontotheconnectorandcylinder
fillports.
NOTE:Thefillportcoversshouldbeusedwhenevercylindersare
notbeingfilled.
SECTION 5—OPERATING INSTRUCTIONS
Part No 1112648 19 HomeFill
SECTION 5—OPERATING
INSTRUCTIONS
WARNING
NEVERsmokeinanareawhereoxygenisbeing
administered.
NEVERusenearanytypeofflameorflammable/explosive
substances,vaporsoratmosphere.
NOTE:Forthisprocedure,refertoFIGURE5.1.
1. Priortoeachuse,inspectproductforvisibledamage.DO
NOTuseifanydamageisfound.
NOTE:Ifforanyreason,anylabelbecomesillegibleorlost,contact
yourdealer/HomeCareequipmentprovider.
2. Besureregulatorisinthe“Off”position.Align“Off”with
theindicatingpointer.
3. Positionthecylindersotheoxygencylindervalveoutletis
pointingawayfromtheuserandanyotherperson(s).
WARNING
DONOTdirectflowofoxygenatanyperson,orflammable
materialwhenadjustingtheflowdial.
4. Attachastandardadultsinglelumenoxygennasal
cannula,nolongerthansevenfeettotheproduct’soutlet
accordingtothecannulamanufacturersinstructions.DO
NOTusepediatric,lowflownasalcannulasoroxygen
maskswiththisproduct.
NOTE:Aunitwithoutcannulaattachedmaypulseifsetatposition
15.
SECTION 5—OPERATING INSTRUCTIONS
HomeFill
20 Part No 1112648
5. Placethecannulaover
yourearsandposition
theprongsinyournose
asinstructedbyyour
HealthCareprovideror
cannulamanufacturer.
RefertoFIGURE5.1.
6. Turndialuntilindicating
pointerisalignedwith
theprescribedsetting.
FIGURE 5.1 Cannula
Placement
WARNING
Whenpointeroncontentsgaugeentersredzone,itis
recommendedtorefillthecylinderontheInvacareHomeFill
Compressor.
NOOXYGENisdeliveredinbetweensettings.Toobtain
desiredflow,theindicatingpointermustalignwithaspecific
numberonthedial.
Toavoidinjurytopatient,ALWAYSconfirmprescribed
settingbeforeadministeringtopatientandmonitorflowona
frequentbasis.
OnlyuseInvacareCorporationHF2PCL6carryingbag,the
HF2PC9carryingbag,ortheHF2PCL4carryingbag.
7. Breathethroughthenoseandfeelapulseofoxygenasyou
inhale.
NOTE:Whenonapulsesetting,thereisfloworapulseonlyatthe
beginningofeachbreath.Ifyoudonotfeelthepulseatthebeginning
ofeachbreath,checkthesetting.
NOTE:Becauseeachpatient’sbreathingpatternisdifferentandthe
environmentvaries,itmaybedifficulttofeelsomelowsetting
pulses.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24

Invacare HomeFill HF2PCL6 User manual

Category
Oxygen Equipment
Type
User manual
This manual is also suitable for

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI