Q-S222

Vernier Q-S222 User manual

  • Hello! I am an AI chatbot trained to assist you with the Vernier Q-S222 User manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
- 1 -
Q-S222 Galvanic Skin Response (GSR) Sensor
Manufactured and Distributed by Qubit Systems Inc.
613-384-1977, [email protected]
www.qubitsystems.com
October 2015
- 2 -
TableofContents
OverviewofGalvanicSkinResponse3
ComponentsforuseofQS222GSRSensor4
QuickStartUpSteps4
ExperimentswiththeGSRSensor6
RapidResponses 6
BasalResponses 7
DataAcquisitionUsingtheLabQuestMiniInterface
LabQuestMiniInterface8
InstallingandRunningLoggerPro39
C404CustomSetupFiles 9
GraphRanges,RunTimeanddataCollection 10
DataAnalysiswithLoggerPro3 11
SpecificationsofoQS222GSRSensor11
QubitSystemsWarrantyInformation 12
ReturnProcedure11
IMPORTANTNOTICE
TheGalvanicSkinResponseSensordescribedinthismanualisforeducationaland
researchuseonly.Itisnottobe
usedasaclinicalordiagnostictool.
- 3 -
OverviewofGalvanicSkinRes p onse
Galvanic SkinResponse (GSR) is used as an indicator of stress in human subjects.When
stressed,subjectsperspiremore,andthis,togetherwithotherphysiologicalevents,alters
skin conductance. This change in conductance is measured by placing electrodes in
contact with the skin.This is done most conveniently by attaching electrodes to the
fingers,thoughotherbodyparts,suchasthebackoftheneck,havebeenused.Unlikein
other electrophysiological measurements,no electrode gel should be used between the
electrodesandtheskin,sincethiswouldalterorbufferskinconductivitychanges.
Shortterm stress events may be elicited in response to psychological shocks, such as
mentioningtheword“spider”toasubjectwhohasanaversiontoarachnids,orshowinga
hypodermic needle to someone with an intense dislike of injections. The responses
appear as peaks in the voltage output that occur within seconds of the stimulus being
presented. This type of response is most often looked for in polygraph (lie detector)
analysis. The response is quite small, appearing as brief peaks in the voltage output.
Typicalpeaksizesarenomorethanafewhundredmillivolts.
The basal response measures slower changes in skin conductance that occur with
changes in mood.Increased stress or excitement is indicated by a gradual increase in
voltageoutput,whereasrelaxationisseenasagradualdeclineinvoltageoutput.
With the Qubit QS222 GSR sensor, two cables are attached by alligator clips to tab
electrodes on two fingers of one hand, and the conductivity of the skin between the
electrodesismeasured.Thesecondandthirdfingersorsecondandfourthfingersofone
handaremostcommonlyusedininvestigations.Theelectrodesshouldbeattachedwith
lightpressure.Thepolarityoftheelectrodesisnotimportant.The0to5Vanalogsignals
from the GSR sensor are recorded by a LabQuest Mini data interface and displayed in
LoggerProsoftwareinmetersandtimegraphs. Skinconductance
ismeasuredinunitsof
microSiemens(µS).
Thesensorhas 2analogoutputs, a high range(40 µS)and a lowrange(5µS).The5 µS
outputcanbeoffsetwithapotentiometertopermithighsensitivitymeasurementsover
theentirerangefrom0to45µS.
- 4 -
ComponentsforuseoftheQS222GSRSensor
QS222GSRSensor
Powersupply
A263TabElectrodes(pkg.of100)
C610LabQuestMinidataacquisitioninterface(notincluded)
C901LoggerProDataAcquisitionSoftware(notincluded)
C404CustomizedSetupFiles(notincluded)
Manual
QuickStartUpSteps
1. LoadLoggerProsoftwareontothecomputer(pg.9)
2. LoadC404customizedfiles(pg.9)
3. ConnectthepowersupplytotheQS222GSR
4. Connectthe40µSdatacabletoanalogchannel1ontheLabQuestMini
5. Connectthe5µSoutputdatacabletoanalogchannel2
6. ConnecttheUSBcablebetweentheLabQuestMiniandthecomputer
7. Turnthe5µSoffsetpotentiometerontheQS222off(i.e.fullyclockwise)
8. Washthesubject'shandswithsoap.Donotwipefingerswithalcoholsincethis
willdrytheskinand
alteritsconductance.Electrodegelshouldnotbeusedasit
willalsochangeskinconductance.
9. Pressthestickytabelectrodesontothefingerpadsasshowninthephotobelow.
10. AttachthealligatorclipsontheinputwiresoftheQS222tothetabsofthe
electrodes.
- 5 -
11. Openthe“QS222GSRSetup”filetostarttheLoggerProsoftware.Notefor
convenience,ashortcuttothisfilecanbeplacedonthedesktop.
12. ASensorconfirmationscreenwillopeninLoggerProasshownbelow."Connect"
thetwoGSRoutputsonchannels1and2oftheLabQuestMini.
Note:TheqS222GSRhasbeenfactorycalibratedinunitsofmicroSiemens(µS)and
doesnotrequirefurthercalibrationbytheuser.
13. Letthesubject'shandrestgentlyonthesurfaceofthebenchwiththepalm
facingdownwards.
Itisimportantthatthefingersarenotmovingduringanexperiment,sincechangesin
electrodecontactwillgreatlyaffectthesignal.
- 6 -
14. Startdatacollectionbyclickingthegreenbuttonandrelaxforafewminutesuntil
thesignalsshownonthegrapharestable.
15. Adjust the offset potentiometer for the 5 µS output so that the signal is in the
middle of the graph. Further relaxation will now be seen
as a reduction in µS
output,andexcitationwillbeseenasanincreaseintheoutput.
Youarenowreadytostartthetreatment.
16. To test the system, start data collection, and when a stable voltage is attained,
exert a pressure with the stomach muscles while keeping
the hand still. A
transient peak in voltage output should be observed.Alternatively, start data
collection with the subject seated in front of the computer with an assistant
standingbehindthesubject.Atatimeunanticipatedbythesubject,theassistant
shouldtouchthesubjectontheshoulder,orrunafingeragainstthebackofthe
subject’sneck.Thisshouldevokeatransientvoltagespikeirrespectiveofwhether
thesubjectfindsthestimuluspleasurableornot.
ExperimentswiththeGSRSensor
TheGSRSensormaybeusedforavarietyofinvestigationsthatinvolvethestudyof
emotionalstatesintestsubjects.Theseinclude,butarenotlimitedtothefollowing.
RapidResponses
Theseresponsesaretransitoryresponsesoccurringoverafewseconds,andwithinafew
secondsoftheonsetof thestimulus.Responsesare typicallysmall (afewhundredmV)
thereforeobservableinthelowrange(i.e.sensitive)5µStrace.Thesubject’sreactionto
variousstimulicanbeassessed.Examplesinclude:
Touching the subject unexpectedly.The magnitude of the response should vary
depending on the type and
location of the t ouch, as well as with the relationship
betweenthesubjectandthepersonapplyingthestimulus.
Showing the subject objects that might provoke a response (e.g. pictures of insects,
syringes etc.).Bigger responses should be seen in subjects that are completely
unaware of the nature of the objects before the experiment starts.As the subject
becomesconditionedtotheobjectwithrepeatedexposuretheresponsemaydecline.
Askingthesubjectquestionsofaprovocativenature.Itisrecommendedthatcaution
and sensitivity be observed with respect to the subject’s rights to privacy when
conductingsuchinvestigations.NotethattheGSRsensormeasuresemotionalstate,
andsimilarresponsescanbeobtainedforcircumstancesthatprovokeeitherpositive
- 7 -
or negative emotional reactions.As with polygraph tests, numerous physiological
parameters need to be measured (not just GSR) for proper psycho physiological
assessments.
Note:Channel3isfreeontheLabQuestMiniforrecordinganotherphysiologicalsignal.
Alsomultiple Minismay beconnectedto LoggerPro to recordmoreanalog (0to 5
V) signals. Contact Qubit Systems for advice on other physiological sensors (eg.
heartrate,skintemperature,bloodpressure,EKG/EMG,respirationrate).
BasalResponses
Basalemotionalstatesdonotchangerapidlywithbriefemotionalchangessuchasthose
elicitedbyashockstimulus.Thebasalresponseismostoftenusedinbiofeedbackstudies
forthecontrolofstress,andalsotodeterminefactorsthatcauseemotionalexcitationor
relaxation over the longer term (several minutes rather than a few seconds).
Investigations may include biofeedback studies in which the subject uses relaxation
techniques to reduce stress. In this case,
the subject sets an initial voltage output to
representthe starting point, and then attempts to reducethe voltage valueconsciously
by relaxation techniques such as deep breathing, imagery or listening to music or a
calming voice. Often, a temporary increase in voltage is observed at the start of the
experiment
asthesubjectrecognizestheexperimentitselfasastress.
The basal GSRoutput can also beused to assess a subject’sresponse to various images
andcolours.Afterarelaxationphasewitheyesclosed,inwhichthebasalsignaldeclines,
the subject is asked to open the eyes and look at an image or a large card of a specific
colour. Usually, opening the eyes causes an increase in GSR signal, but the rapidity and
degreeoftheincrease isgreatlyinfluencedbytheimageorthecolourthatisimmediately
observed.Provocativeimages causearapidincrease,andbrightredtonescauseamore
rapidincreasethanmutedgreenorbluetones.
Anexampleofdataobtainedbyperiodsofrapidbreathingfollowedbyrelaxedbreathing
isshownbelow.
- 8 -
DataAcquisitionUsingtheLabQuestMiniInterface
LabQuestMiniInterface
TheLabQuestMiniinterfacehas3analogand2digitalchannels.Theinterfaceis
connectedtothecomputer(PCorMacintosh)byaUSBcable.Theinterfaceconvertsthe
analogvoltages(12bitAtoD)todigitalsignalswhicharetransmittedtothecomputervia
USBandprocessedbytheLoggerProsoftware.ThegreenLEDindicatesthatthe
LabQuestMiniisrecognizedbythedriversandisreadytocollectdata.
- 9 -
InstallingandRunningLoggerPro3
PCUsers:
(1) To start, a complete copy of Logger Pro 3 must be installed on the computer.Before
starting the installation, make sure all USB cables are disconnected from the computer.
Failuretodosomaycauseanerrorintheinstallationofthedrivers.
(2) Runtheinstallationanddonotchange thedefaultdestination directory.LoggerPro3will
beinstalledinC:/ProgramFiles/VernierSoftware/LoggerPro3.
(3) Thesetupprocesswillautomatically loadtheUSBdriversforconnectingtheLabQuestMini
orotherinterfacestothecomputer.
(4) If QuickTime 6 (or greater) is not installed on the computer, install it when prompted.
QuickTimewillallowuseofthepictureandmoviefeaturesofLoggerPro3.
(5) YouwillbepromptedtoconnecttheLabQuestMiniorotherinterfacestothecomputervia
theUSBconnection.
(6) Click‘Finish’tocompletetheinstallationprocess.
(7) ProceedtotheC404installation(below)beforeopeningtheLoggerPro.
(8) YoucanstartloggerprodirectlyfromtheexperimentalfilethatislocatedontheC404disk
and can be transferred to user specified location and a shortcut may be created on the
desktopforeasyaccess.Doubleclickthe“QS222GSRSetup”file.
(9) IfLoggerProdetectstheLabQuestMiniinterface,theLoggerProscreenwillappearwitha
star(iconforLabQuestMini)inthetopleftcorner.

(9) IfLoggerProcannotdetecttheLabQuestMiniInterface,amessagewillappear“nodevice
connected”.The LEDontheMinisshouldbe green.AredLEDindicatesthat power ison
(supplied by the computer through the USB) but there is no communication between the
interfacesandthecomputer.In
thiscase,exitthe“QS222GSRSetup”fileandunplugthe
USBcablefromthecomputer.ReconnecttheUSBcable.Allowthecompu tertorecognize
theMini(belltone)andthenreopenthe“QS222GSRSetup”file.
C404CustomSetupFiles:
QubitSystems’C404CustomSsetupFilesCDcontainsallofQubitExperimentfiles
(designedbyQubitSystems)including“QS222GSRSetup”.TheExperimentfilescontain
thesetup(i.e.graphs,table,calculationsetc)forthevariousexperiments,aswellasthe
- 10 -
calibrationconstantsfortheQubitsensors.TheC404diskalsocontainsmanualsforthe
differentQubitsensorsandpackages.Thesefilescanbecopiedtouserspecifiedlocation
onthecomputerandtheexperimentalfileofinterestforthissensor“QS222GSR
Setup”shouldbeplacedinanaccessible
locationorhaveashortcutcreatedonthe
desktoptothefile.
GraphRanges,RunTimeandDataCollection
Tomodifytherangeofthexandyaxes, click on the maximumand/orminimum values
andtypeinthenewvalues.
If you wish to alter the maximum time
allotted to your experiment or the rate of data
sampling,select“Experiment”,then“DataCollection”orclickontheclockiconinthetop
menu.
Type in the duration time of your experiment in the box labelled “Duration”. Note that
the maximum value you select for the Timeaxis limits the time over which you can
collect data for a particular run.If your experiment exceeds the allotted time, select
“Experiment”,then“StoreLatestRun”.Youmaythenrestartdatacollection.Datafrom
thefirstpartofyourexperimentwillremainonthescreenasafainttraceandnewdata
willbeplottedin a bolded shade of the same colour.You can collectnumerous runs in
thisway.Eachrunwillbecollectedtoaseparatedatatable.Whenyousaveyourdata,
alloftherunswillbesavedunderthesamefilename.
- 11 -
You may also alter the "Sampling Rate" at which you collect data.The default rate for
GSRis10 samples/s.To startdata collection,clickon the“Collect” greenicon abovethe
graphs.This will change to a “Stop” red icon.Click on the “Stop” icon to stop data
collection.
DataAnalysiswithLoggerPro3
(1) Position the cursor atthe beginning of thedat a setthat you wish to analyze, then
clickandholdwiththemouse.
(2) Drag the cursor across the data set and unclick.A greyed box will appear on the
screentoshowthedatarangetobeanalyzed.
(3) Select “Analyze”, and then choose from a wide variety of analysis options.For
example,use:
“Examine”toscrollthroughthedatapointsatspecifictimeintervals.
“Integral”tointegratetheselecteddata.
“Statistics”toautomaticallyperformavarietyofstatisticalanalyses.
You can also fit curves and tangents to the selected data by choosing the appropriate
menuoption.Notetheanalysisfunctionscanalsobeaccessedfromiconsinthemenu.
SpecificationsofQS222GSRSensor
Power Supply: 9V, 560mA
Units: (uS) uSiemens
Resolution: 40uS Range = 0.0097uS, 5uS Range = 0.0012uS
Accuracy: 0.5% for Full Scale, or better
Range: 40uS and 5uS
Offset: 5uS Range can be offset from 0 to 40 uS
Gain: System Gain is 1980 and 15840 for 40uS and 5uS
Cut-off Frequency fc High= 7.2 Hz
Patient Isolation: 1500Vrms
Size (h/w/l) cm: 2.8/6.2/9.2
Weight (g): 200
Materials: plastic (black)
Warranty: 1 years
- 12 -
QubitSystemsWarrantyInformation
QUBIT warrants all its instruments to be free from defects in materials or workmanship
foraperiodofoneyearfromthedateofinvoice/shipmentfromQUBIT.
Ifatanytimewithinthiswarrantyperiodtheinstrumentdoesnotfunctionaswarranted,
returnitandQUBITwillrepairorreplaceitatnocharge.Thecustomerisresponsiblefor
shipping and insurance charges (for the full product value) to QUBIT.QUBIT is
responsibleforshippingandinsuranceonreturnoftheinstrumenttothecustomer.
Nowarrantywillapplytoanyinstrumentthathasbeen(i)modified,altered,orrepaired
bypersonsunauthorized by QUBIT;(ii)subjected to misuse, negligence, or accident; (iii)
connected,installed,adjusted,orusedotherwisethaninaccordancewiththeinstructions
suppliedbyQUBIT.
Thewarranty is returntobaseonly, and doesnotincludeonsiterepair chargessuch as
labour,travel,orotherexpensesassociatedwiththerepairorinstallation ofreplacement
partsatthecustomer’ssite.QUBITrepairsorreplacesthefaultyinstrumentsasquicklyas
possible; maximum time is one month. QUBIT will keep spare parts or their adequate
substitutesforaperiodofatleastfiveyears.
Returned instruments must be packaged sufficiently so as not to assume any transit
damage.Ifdamageiscausedduetoinsufficientpackaging,theinstrumentwillbetreated
asanoutofwarrantyrepairandchargedassuch.
QUBITalsooffersoutofwarrantyrepairs.Theseareusuallyreturnedtothecustomeron
acash ondeliverybasis.Wear & Tear Items are excluded from this warranty.The term
Wear & Tear denotes the damage that naturally and inevitably occurs as a result of
normal use or aging even when an item is used competently and with care and proper
maintenance.
ReturnProcedure
BeforereturninganyinstrumenttoQUBIT:
ConsulttheoperatingmanualorcontactQubittoensurethatthe instrument(s)isinfact
faultyandhasnotjustbeensetupimproperly.
ContactQUBITbeforesendinganythingback.WewillissueanRMAnumberandprovide
shipping instructions.QUBIT will refuse any goods
that are returned without an RMA
number,orwhicharesentinamanneroutsideofQUBIT’Sstipulations.
Ifyouhaveencounteredaprogramfailure,weneedaprintedcopyofanyfaultsyouhave
seen,including howto reproduce them.Includethese in the returnpackage alongwith
yourmailingaddress.
IncludeacopyoftheInvoiceonwhichtheproductwasshippedtoyou.
Allreturnsmustbeshippedprepaid.Unpaidpackageswillnotbeaccepted.
IncaseofquestionscontactQUBITby
Email:[email protected],
byphone:(01)6133841977,
orbyfax:(01)613384‐9118.
/