Varian Star 800 Driver Manual

  • Hello! I am an AI chatbot trained to assist you with the Varian Star 800 Driver Manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
Varian, Inc.
2700 Mitchell Drive
Walnut Creek, CA 94598-1675/USA
Varian, Inc. 2005 03-914947-02:2
Star 800 MIB TTL I/O Control
Galaxie Driver Manual
Star 800 MIB TTLIO Control Manual - 1 - 03-914947-02:2
Table of Contents
Introduction.............................................................................................. 2
General................................................................................................................................2
Installation and Configuration................................................................ 3
Installing the Driver Software...............................................................................................3
Configuring the Driver..........................................................................................................3
Configuring the 800 MIB TTLIO....................................................................................3
Configuring the Synopsis..............................................................................................7
Building a Method ................................................................................... 9
Creating a New Method.......................................................................................................9
Control Method Section for the 800 MIB TTLIO................................................................12
Channel Value Control................................................................................................14
Prerun Control.............................................................................................................15
Injection Control..........................................................................................................17
Run Control.................................................................................................................19
Postrun Control...........................................................................................................21
Starting and Viewing an Acquisition.................................................... 23
Quick Start and Sequence.................................................................................................23
Quick Start (Single Acquisition) ..................................................................................23
Sequence....................................................................................................................25
Running Acquisition Status................................................................................................26
Viewing a Running Acquisition....................................................................................26
Status of the 800 MIB TTLIO ......................................................................................26
Star 800 MIB TTLIO Control Manual - 2 - 03-914947-02:2
Introduction
General
This document describes the configuration of the Varian 800
MIB TTLIO driver. It also explains how to build a control method
for this driver, how to start and view a running acquisition, and
what actions are available during an analysis.
Star 800 MIB TTLIO Control Manual - 3 - 03-914947-02:2
Installation and
Configuration
Installing the Driver Software
By default the driver software is installed in Galaxie 1.8. No
further installation is required for that driver.
Configuring the Driver
The driver software configuration is done in the Galaxie
Configuration Manager when a new chromatographic system is
created.
Configuring the 800 MIB TTLIO
Click on the button corresponding to the 800 MIB TTLIO
driver and the configuration screen appears.
Up to nine sub-sections allow you to configure the instrument:
Star 800 MIB, Channel #X, Channel Value and Miscellaneous.
The toolbar allows you to easily scroll through the configuration
windows:
Star 800 MIB TTLIO Control Manual - 4 - 03-914947-02:2
to select the first configuration window.
to select the previous configuration window.
to select the next configuration window.
to select the final configuration window.
Star 800 MIB Window
Click on the
button to select the 800 MIB box with the
TTLIO card (the Star 800 MIB box must be previously
configured on the acquisition server). The name of the selected
800 MIB box will then be displayed in the Hercule Name field
Star 800 MIB TTLIO Control Manual - 5 - 03-914947-02:2
Press the button to automatically
get the configuration from the instrument.
Channel #x
Window
Check or uncheck the Use Channel #X box to activate or
deactivate this channel.
In the Name Field, enter the name of the selected channel.
Physical Name: Displays the physical name of the relays of the
selected channel.
Output Name: Enter in this field the name of the relays of the
selected channel.
Star 800 MIB TTLIO Control Manual - 6 - 03-914947-02:2
Channel Value Window
In this screen users can predefine some channel configurations
saved in the system configuration that can be used in methods.
Use the following buttons to edit the table:
to add a line to the table.
to delete a line in the table.
to clear the entire table.
In the column Configuration Name, enter the name of the
defined channel configuration.
In the Channel Name column, select the channel to configure.
In the Out X columns, select if the relay of the selected channel
is opened
or closed .
Star 800 MIB TTLIO Control Manual - 7 - 03-914947-02:2
Miscellaneous Window
Check or uncheck the Allow remote control in status window to
activate or deactivate this option. If it is checked, the user will
be able to open and close relay from the status.
Check or uncheck the Allow remote control in status window
during run to activate or deactivate this option. If it is checked,
the user will be able to open and close relay from the status
during run.
Configuring the Synopsis
In the Overview part of the system configuration, a synopsis can
be defined for the system. This synopsis depends on the
configured devices.
1. To save the configuration of these devices, click on the OK
button.
Star 800 MIB TTLIO Control Manual - 8 - 03-914947-02:2
2. Right-click on the system name in the bottom part of the
Galaxie Configuration Manager screen. A pop-up menu
appears: click on Stop. Then right-click on the system name
again, and click on Start.
3. Go back to the System Configuration screen (Properties
wizard). A graphic of the devices that comprise the system
is displayed in the synopsis screen. To configure the
synopsis as the real instrument configuration, move each
module, and then connect it appropriately. To do this, select
one anchor for each of the two modules (the selected
anchors become red) and click on the
button; a
connection tube will appear between the two anchors.
Click on the
to delete a selected tube; click on the
button to delete all the tubes.
Click on
to define a background color, or on to
make the background transparent.
Click on the
icon to change the grid color.
4. Once the synopsis configuration is complete, it will be
displayed in the Galaxie Chromatography Data System, in
the Systems tab. In this tab, the Overview part will show
the synopsis of the system with some parameters
displayed (temperature, flow, signal, etc.) depending on the
installed devices.
NOTE: Once the system has been created and correctly configured, it must be
associated with one or more projects in order to start some acquisitions. For
more details about the creation and configuration of a system, refer to the
Galaxie Configuration Manager User’s Guide.
Click on the
button to print the current configuration.
Star 800 MIB TTLIO Control Manual - 9 - 03-914947-02:2
Building a Method
Creating a New Method
The method contains parameters for instrument control, data
acquisition, chromatogram processing, and editing of final
results. To initially set up a method, it is not necessary to define
all of the sub-method sections; however, you must fill in the
instrument control section prior to starting an acquisition.
1. To create a new method, select the FILE / NEW METHOD
option from the Galaxie Chromatography Data System main
menu.
A wizard guide will appear and provide assistance during
the first steps of method creation:
Star 800 MIB TTLIO Control Manual - 10 - 03-914947-02:2
2. Choose the system (chromatograph) associated with the
method. A method is created for a particular system. When
starting an acquisition, the name of the system that
performs the acquisition must be correct, since method
access is limited only to those methods associated with this
system.
3. Once this field is completed, click on the Next button to
move to the second step of the method creation:
Star 800 MIB TTLIO Control Manual - 11 - 03-914947-02:2
4. Enter the New Method Name in the first field. In the
Description field, enter information concerning the method
(optional).
5. Once these two fields have been completed, click on the
OK button. The new method is created and is opened with
default parameters in the Galaxie Chromatography Data
System. The next step is the definition of each of the
method sections.
Star 800 MIB TTLIO Control Manual - 12 - 03-914947-02:2
The instrument control section must be defined in order to
start an acquisition. In the following part of this guide, only
the control section for the 800 MIB TTLIO is detailed. Other
parameters such as Injection Volume, Vial and Rack
number may be necessary and are defined in the
acquisition section.
NOTE: For more details about the programming of each method section, turn to the
Galaxie Chromatography Data System User’s Guide.
Control Method Section for the 800 MIB TTLIO
Once the new method is created, select the control section in
the lower part of the browser.
The following screen appears:
Star 800 MIB TTLIO Control Manual - 13 - 03-914947-02:2
Press the button to display the System Control Method
Advanced Tools window:
to send the current method to the instrument.
to get the active method from the instrument.
to print the current method.
Click on the
button to display the 800 MIB TTLIO control
method section. It contains five sub-sections: Channel value,
Prerun, Injection, Run and Postrun. Click on the required sub-
section to define the corresponding parameters that will be
used during the chromatographic run.
The toolbar allows you to easily scroll through the control
method windows:
to select the first window of the control method.
to move back to the previous window of the control
method.
to move forward to the next window of the control
method.
to select the final window of the control method.
to expand the control method section tree view.
to compact the control method section tree view.
Star 800 MIB TTLIO Control Manual - 14 - 03-914947-02:2
Channel Value Control
In this screen users can define some channel configurations
that can be used in the method.
Use the following buttons to edit the table:
to add a line to the table.
to delete a line in the table.
to clear the entire table.
In the column Configuration Name, enter the name of the
defined channel configuration.
Star 800 MIB TTLIO Control Manual - 15 - 03-914947-02:2
In the Channel Name column, select the channel to configure.
In the Out X columns, select if the relay of the selected channel
is opened
or closed .
Prerun Control
In this window users can time program the state of each relay
during the Prerun of an acquisition.
Star 800 MIB TTLIO Control Manual - 16 - 03-914947-02:2
Use the following buttons to edit the table:
to add a line to the table.
to delete a line in the table.
to clear the entire table.
In the column Selection Type, select the type of action that will
be done one the relays. Three choices are available: Channel,
Config ch.Value and Method ch. Value.
If Channel is selected the user can define the state of a specific
relay.
If Config ch. Value is selected the user can use at the specified
time a channel value configured in the system configuration.
If Method ch. Value is selected the user can use at the specified
time a channel value configured in the method.
In the column Cfg. Name select the name of the method
channel value or configuration channel value to use.
In the column Channel Name, select the name of the channel to
use.
In the column Output Name, select the name of the relay to use.
In the column Action, select the type of action to do on the
relay. Four actions are available: Open, Close, Invert and Pulse.
The open or close actions will open or close the relay. The
invert action will invert the state of the relay and the pulse will
pulse the relay for a defined time.
In the column Pulse, enter the length of the pulse in minutes.
Star 800 MIB TTLIO Control Manual - 17 - 03-914947-02:2
Injection Control
In this window users can program the state of each relay at the
injection of an acquisition.
Use the following buttons to edit the table:
to add a line to the table.
to delete a line in the table.
to clear the entire table.
Star 800 MIB TTLIO Control Manual - 18 - 03-914947-02:2
In the column Selection Type, select the type of action that will
be done one the relays. Three choices are available: Channel,
Config ch.Value and Method ch. Value.
If Channel is selected the user can define the state of a specific
relay.
If Config ch. Value is selected the user can use at the specified
time a channel value configured in the system configuration.
If Method ch. Value is selected the user can use at the specified
time a channel value configured in the method.
In the column Cfg. Name select the name of the method
channel value or configuration channel value to use.
In the column Channel Name, select the name of the channel to
use.
In the column Output Name, select the name of the relay to use.
In the column Action, select the type of action to do on the
relay. Four actions are available: Open, Close, Invert and Pulse.
The open or close actions will open or close the relay. The
invert action will invert the state of the relay and the pulse will
pulse the relay for a defined time.
In the column Pulse, enter the length of the pulse in minutes.
Star 800 MIB TTLIO Control Manual - 19 - 03-914947-02:2
Run Control
In this window users can time program the state of each relay
during the Run of an acquisition.
Use the following buttons to edit the table:
to add a line to the table.
to delete a line in the table.
to clear the entire table.
/