3 Setting Up ToxID Processing Configuration Files
Processing Configuration File Template
14 ToxID User Guide Thermo Scientific
Intensity Threshold
(all experiments)
For MS/MS, and for MS/MS and MS3 experiments, a search
hit is only reported if the signal intensity for the most intense
or most specific fragment in the MS/MS spectra is equal to
or above this value.
For QED MS/MS experiments, a search hit is only reported
if the intensity of the fragment ion collected in an SRM scan
event is equal to or above this value.
For other types of experiments, a hit is reported if the TIC
signal intensity is equal to or above this value.
SI (six experiments) Specifies the minimum search index value for the reported
hits. This parameter is required for the MS/MS spectra,
MS/MS and MS3spectra, MS3spectra, and MS/MS QED
experiments.
RSI (six experiments) Specifies the minimum reverse search index value for the
reported hits. This parameter is required for the MS/MS
spectra, MS/MS and MS3 spectra, MS3 spectra, and MS/MS
QED experiments.
SRM Fragment Ions
(six experiments)
Specifies the fragment and, in parentheses, the relative
intensity of each fragment ion. Separate each ion entry with
a semicolon (;), for example:
191.3(25);245.2(100);345.9(67). This parameter is required
for the SRM and QED. The SRM fragment ions input must
match with the SRM scans, or else it does not return a hit.
Internal Standard
(six experiments)
A reference compound name present in the data file.
Intensity Concentration
Coefficient (for ion trap
experiment only)
A user-defined parameter to convert the intensity ratio to
concentration of analyte. See page 43 for further details.
Area Concentration
Coefficient (for ion trap
experiment only)
A user-defined parameter to convert the area ratio to
concentration of analyte. Use intensity of most MS/MS
scans.
Adduct 1, 2, 3 (Exactive) Ions attached to parent ion during ionization.
Fragment 1, 2, 3 (Exactive) Ions in the HCD scan, which the user can choose for
additional information about the compound match.
Table 3. Compound-specific parameters (Sheet 2 of 2)
Parameter Description