Roche Diagnostics
Status Notifier for cobas t 711 coagulation analyzer · Software version 1.0 · Quick Reference Guide · Version 1.0
About the Status Notifier 5
About the Status Notifier
The Status Notifier is a feature that informs the user of
the instrument status from a distance.
It is composed of a status light and an acoustic alarm.
Colors, sound, and statuses The following table describes how color and sound of the
Status Notifier correspond to the analyzer statuses.
Specifications
Sound pulse
Status indicator Statuses Description
"Normal"
Green
o Operating
o Standby
o Pausing
o Maintenance
No user action required
"Alarm"
Red
o Error
o Warning
The alarm status indicates that an immediate user action is
required. The analyzer is not ready for operation.
o The status light is red and the sound is active.
"Warning"
Yello w
o Analyzer offline
o Paused
o Sampling stop
The warning status indicates that a user action is required to
enter normal operation. If no user action is done, the status may
change to alarm.
o No acoustic alarm
"Alternating Status"
three lamp colors
o Initializing
o Shutting down
o No acoustic alarm
"Service"
Flashing yellow
o Diagnostics In order to perform tests and special maintenance actions, the
Roche Service representative sets the instrument to Diagnostics
mode.
o No acoustic alarm
y Colors, Sound, and states
Status light R/Y/G (LED)
Diameter of status light 70 mm
Height of pole 500 mm
Acoustic signal
integrated
Volume (70–80 dB)
(selectable by switch)
Length of USB cable 2.0 m
Installation to table Mounting bracket for tables
(thickness: 15 mm up to 58 mm)
Power USB port
Rhythm Buzzer pt4
Pitch A 2093.0 Hz
Pitch B 1760.0 Hz
Duration of a pulse 1.5 sec
Interval between pulses 1.0 min
First pulse after status
change
The first pulse starts as a double
pulse with an interval of 0.5 s