2
CANNON
®
Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup Guide
Version 1.3g— August, 2011; CANNON
®
Instrument Company
2139 High Tech Road • State College, PA 16803 • USA
The compressor must be capable of producing a constant air flow of 3
cfm (5 m
3
/hr) at 40 psi (275 kPa) minimum. Line pressure should be set
at 75-100 psi (517-690 kPa). If the CSU-200 Service Unit is not being
used, vacuum must be supplied by the user to the rear panel VACUUM
connection on the Bath Unit. Vacuum should be approximately -8 in Hg
to -10 in Hg, and should be suitable for the solvent/samples being run.
3.___Sample/waste disposal
The customer must provide for the containment and disposal of all used
solvents and waste oils. Collection, storage, and disposal of all waste
from the CAV is the responsibility of the customer, and must be done in
accordance with all local safety and environmental regulations.
4.___Cooling water
Instruments maintaining bath temperatures lower than 40°C require a
source of cooling water. Cooling water connections are located at the rear
of the CAV unit. The cooling water should be maintained at 15-20°C
below the bath temperature setting.
NOTE
Temperature baths operating at 40°C may require circulating cooling
water, particularly if the room temperature of the laboratory exceeds
24°C (75°F).
5.___Ventilation
All models of the CAV are provided with a 4” (10.16 cm) diameter
ventilation shroud in the rear panel of the unit. A ventilation system
having a continuous airflow adequate to ensure removal of solvent fumes
and vapors should be connected to the CAV. CANNON
®
Instrument
Company does not supply ventilation fans, hoods, or systems.
CAUTION
The CAV must be operated with ventilation precautions appropriate for
your facility and the sample materials and solvents being used.
CANNON
®
recommends that the CAV be connected to an external
ventilation system. The 4” (10.16 cm) diameter opening on the back
panel of the CAV will accommodate a standard 4" dryer hose. If any
solvent vapors are detected around the CAV, verify the operation of the
exhaust ventilation system.
6.___Laboratory environment
It is important that the CAV be kept in a suitable laboratory environment.
The room temperature of the laboratory must be maintained between
15°C and 27°C (60-80°F). The laboratory must be free of dust and dirt,
which could possibly taint the samples. The laboratory must be equipped
with all of the standard safety features normally associated with use of
solvents, corrosives, etc.