33. Next choose the Date/Time tab, and set the correct Date
and Time.
34. Select the Back icon to return to the Home Screen.
35. Select the System Menu icon in the bottom right corner of
the screen to open the Menu screen.
36. Select Video, and watch the videos labelled “SMART 6
Intro” and “How to Run a Sample”
37. After reviewing the process for running a sample, run the monitor solution provided with the SMART 6, following the
Monitor Testing instructions provided in the accessory kit to ensure the SMART 6 is performing within proper
specification.
38. Once the SMART 6 has been validated, and all users are
familiar with the system, ensure that all of the methods
required for full testing are on the SMART 6. The SMART 6
already contains methods that cover a range of basic
products. Choose the One Touch tab to access these
methods.
a. If your sample results do not match or run
properly under these methods, visit
www.cem.com/smart_6/resource-library.html
and log in (see instructions below) for a full
library of optimized One Touch Methods
b. If your sample does not work with any methods
from this library, please either contact the CEM
Process Applications group, or follow the
instructions in the Operation Manual for
developing and optimizing your own methods.
39. Once the correct methods are located, they can be
downloaded to a USB stick, which will then be plugged into
the USB port on the side of the unit and imported into the
SMART 6. To import, choose either the One Touch or
Classic library, and choose the Import icon on the top of the
screen.
40. After importing and validating all methods and sample types, the SMART 6 is ready for analysis.