Mettler Toledo (Software Version 2.2) Transmitter Module PA 700(X) Operating instructions

Category
Measuring, testing & control
Type
Operating instructions
Measuring Module
M 700
®
PA 700(X)
Communication Unit
for PROFIBUS PA
Trademarks
The following registered trademarks are used in this instruction manual
without further marking
SMARTMEDIA
®
is a registered trademark of Toshiba Corp., Japan
FOUNDATION FIELDBUS
is a trademark of Fieldbus Foundation, Austin, USA
Warranty
Defects occurring within 1 year from delivery date shall be remedied free of charge at our plant
(carriage and insurance paid by sender). Sensors, fittings, and accessories: 1 year.
©2007 Subject to change without notice
Return of products under warranty
Please contact our Service Team before returning a defective device. Ship the cleaned device to
the address you have been given. If the device has been in contact with process fluids, it must be
decontaminated/disinfected before shipment. In that case, please attach a corresponding certificate,
for the health and safety of our service personnel.
Disposal
Please observe the applicable local or national regulations concerning the disposal of
“waste electrical and electronic equipment”.
Mettler-Toledo AG,
Process Analytics, Industrie Nord, CH-8902 Urdorf,
Tel. +41 (44) 729 62 11 Fax +41 (44) 729 26 36
Subject to technical changes.
3
4
Contents
M 700 PA 700 Module
Warranty .....................................................................................................2
Disposal ....................................................................................................... 2
Trademarks ..................................................................................................2
Contents .................................................................................................... 4
Intended Use ...............................................................................................6
Conformity with FDA 21 CFR Part 11 ..........................................................6
Safety Information .......................................................................................7
Application in Hazardous Locations: PA 700X Module .................................7
Software Version ........................................................................................8
Modular Concept ........................................................................................ 9
Short Description ......................................................................... 10
Short Description: FRONT Module ..............................................................10
Short Description: Menu Structure .............................................................11
Short Description: BASE Module ................................................................13
PROFIBUS Technology .................................................................. 14
Terminal Plate PA 700(X) Module ............................................... 16
Attaching the Terminal Plates .............................................................. 16
Inserting the Module ................................................................... 17
PROFIBUS PA Installation ............................................................ 18
Communication Model ..............................................................................20
Physical Block (PB) ......................................................................................20
Analog Input blocks ...................................................................................21
For Copy: Individual Settings ......................................................................24
Configuration with PROFIBUS ....................................................................25
Device database file (GSD file) ...................................................................25
Cyclic Data Communication .......................................................................26
Structure of Cyclic Input Data Telegram ................................... 26
DI Function Blocks .....................................................................................27
DI 1: EC 400 status ............................................................................. 27
DI 2: CONTACTS / LOCK Status / ENABLE Request .............................. 27
DI 3: EC 400 Messages ....................................................................... 28
Explanation of EC 400 Messages: Maintenance request ...................... 28
5
Contents
M 700 PA 700 Module
Explanation of EC 400 Messages: Failure ............................................. 29
DI 4: EC 400 Step ............................................................................... 29
DO Function Blocks ....................................................................................30
DO 1: HOLD Control ........................................................................... 30
DO 2: PARSET...................................................................................... 30
DO 3: EC 400 Control ......................................................................... 31
DO 4: LOCK Control ........................................................................... 31
Configuration Data ....................................................................................32
Analog Input blocks (1 ... 8) ...................................................................... 33
Discrete Input Blocks (1 ...4) ......................................................................34
Discrete Output Blocks (1 ... 4) .................................................................. 35
PA Slot Model ............................................................................................ 36
PB Block Parameters ..................................................................................37
TB Analyzer Block Parameters ....................................................................39
AI Function Block Parameters .....................................................................41
Defaults & Writable Ranges ....................................................................... 41
DI Function Block Parameters .....................................................................44
DI Function Block Parameters .....................................................................45
DO Function Block Parameters ...................................................................46
AO Function Block Parameters ...................................................................48
Calibration Record Parameters ................................................................... 49
Menu Selection ............................................................................ 61
Passcode Entry ............................................................................. 62
Changing a passcode .......................................................................... 62
Passcode lost ....................................................................................... 62
Function Control Matrix ............................................................................63
Pressure Compensation via Bus (AO1) ....................................... 64
Diagnostics: Bus Monitor ...........................................................................65
Bus Monitor ..............................................................................................66
Specifications ............................................................................... 67
Index ............................................................................................. 73
6
Intended Use
In their directive “Title 21 Code of Federal Regulations, 21 CFR Part 11,
Electronic Records; Electronic Signatures“ the US American health agency
FDA (Food and Drug Administration) regulates the production and processing
of electronic documents for pharmaceutical development and production.
This results in requirements for measuring devices used for correspond-
ing applications. The following features ensure that the M 700(X) modular
process analysis system meets the demands of FDA 21 CFR Part 11:
Electronic Signature
Access to the device functions is regulated and limited by individually adjust-
able codes – “Passcodes“. This prevents unauthorized modification of device
settings or manipulation of the measurement results. Appropriate use of
these passcodes makes them suitable as electronic signature.
Audit Trail Log
Every change of device settings can be automatically recorded and docu-
mented in the Audit Trail Log on the SmartMedia card. The recording can be
encrypted.
Conformity with FDA 21 CFR Part 11
The module is a communication unit for PROFIBUS-PA and allows digital
communication via current modulation.
The PA 700X module is intended for operation in locations subject to explo-
sion hazards which require equipment of Group II, device category 2(1), gas/
dust.
7
Safety Information
Application in Hazardous Locations
Caution!
Never try to open the module! If a repair should be required, return the
module to our factory.
If the specifications in the instruction manual are not sufficient for assessing
the safety of operation, please contact the manufacturer to make sure that
your intended application is possible and safe.
Application in Hazardous Locations:
PA 700X Module
When using the PA 700 X module, the stipulations for electrical installations
in hazardous areas (EN 60079-14) must be observed. When installing the
device outside the range of applicability of the 94/9/EC directive, the appro-
priate standards and regulations in the country of use must be observed.
The module has been developed and manufactured in compliance with the
applicable European guidelines and standards.
Compliance with the European Harmonized Standards for use in hazardous
locations is confirmed by the EC-Type-Examination Certificate.
Compliance with the European guidelines and standards is confirmed by the
EC Declaration of Conformity.
There is no particular direct hazard caused by the operation of the device in
the specified environment.
Be sure to observe during installation:
Switch off power supply before replacing or inserting a module.
Before commissioning it must be proved that the device may be connected
with other equipment.
8
Software Version
PA 700(X) Module
Device Software M 700(X)
The PA 700(X) module is supported by software version 5.0 or higher.
Module software version 2.x requires device software version 7.x.
Module Software COMPA 3400(X)-081
Software version 1.3 01.06.2004
Software version 2.2 02.04.2007
Menu Display Device description
Provides information about all
modules installed: Module type and
function, serial number, hardware
and software version and device
options.
Select the different modules (FRONT,
BASE, slots 1 - 3) using the arrow
keys.
Device description
Return
22.7 °C
0.003mS/cm
Operating panel
Module FRONT 700-011
Hardware: 1, Software: 7.0
Serial number 0000815
BASE
Module
FRONT
Query actual device/module software
When the analyzer is in measuring mode:
Press menu key, open Diagnostics menu.
Options
9
Modular Concept
Basic Unit, Measuring Module, Additional Functions
Documentation
The basic unit is accompanied by a CD-ROM containing the complete
documentation.
Latest product information as well as instruction manuals for earlier software
releases are available at www.mtpro.com.
The M 700(X) is an expandable modular process analysis system.
The basic unit (FRONT and BASE modules) provides three slots which can be
equipped by the user with any combination of measuring or communication
modules. The software capabilities can be expanded by additional functions
(options). Additional functions must be ordered separately. They are supplied
with a device-specific TAN for function release.
3 module slots
for free combination of
measuring and
communication modules
Measuring modules
pH / ORP / Temp
0
2
/Temp
Noncontacting conductivity/Temp
Contacting conductivity/Temp
Additional
functions
Activation via device-
specific TAN
SmartMedia card
Data recording
M 700(X) Modular Process Analysis System
Communication modules
OUT (additional switching and
current outputs)
PID (analog and digital controller)
Profibus PA
Foundation Fieldbus
EC 400 probe controller
10
Short Description
Short Description: FRONT Module
Transflective LC graphic display
(240 x 160 pixels)
white backlighting, high resolution
and high contrast.
Red LED
signals failure (On) or
maintenance request/function check
(flashing) according to NE 44.
4 captive screws
for opening the analyzer
(Caution! Make sure that the gasket between
FRONT and BASE is properly seated and clean!)
Green LED
Voltage supply okay
5 self-sealing cable glands
M20 x 1.5
for entry of voltage supply and signal lines
2 softkeys
with context-sensitive functions.
User interface
with plaintext menus as
recommended by NAMUR.
Menu texts can be switched to:
German, English, French, Italian,
Swedish, and Spanish.
Intuitively acquirable menu logic,
based on Windows standards.
Control panel
3 function keys
(menu, meas, enter)
and 4 arrow keys for menu selection
and data entries
Measurement display
Secondary displays
%Air
°C
24.0°C
25.8°C
11
Measuring
Short Description: Menu Structure
Basic Functions: Calibration, Maintenance, Parameter Setting, Diagnostics
Calibration
Maintenance
Parameter setting
Diagnostics
Menu groups
1
2
3
1147
2958
1246
Operator level
1989
Administrator level
Passcode:
Selection of
further menu
items:
Module 1
Module 2
Module 3
BASE
Module 1
Module 2
Module 3
SYSTEM
FRONT
BASE
Module 1
Module 2
Module 3
Message list
Point of meas
description
Logbook
Device description
FRONT
BASE
Module 1
Module 2
Module 3
4
5
6
Legend:
Pressing the menu key accesses menu selection.
Pressing the meas key returns to measurement.
Menu groups are selected using the arrow keys.
Press enter to confirm, enter passcode.
Further menu items are displayed.
Selected functions of the Diagnostics menu can be
recalled via softkey even when in measuring mode.
1)
2)
3)
4)
5)
6)
12
Short Description: FRONT Module
View into the open device (FRONT module)
Terminal plates
of “hidden” modules
Each module comes with an adhesive
label containing the contact assignments.
This label should be sticked to the inner
side of the front (as shown). Then, the
terminal assignments remain visible even
if further modules are inserted.
Slot for SmartMedia card
Data recording
The SmartMedia card expands the
measurement recorder capacity to
> 50000 records.
Exchange of parameter sets
5 parameter sets can be stored on
the SmartMedia card. The 2 internal
parameter sets can be switched by
remote control.
Configurations can be transmitted
from one analyzer to the other.
Function expansions
are possible with additional software
modules, which are released using
transaction numbers (TAN)
Software updates
The circumferential sealing
guarantees IP 65 protection and allows
spray cleaning / disinfection.
Caution! Keep clean!
Replacing the front module
Pull off power cord and ground wire.
To separate the FRONT module
from the BASE module, turn the
retaining screws of the pivot hinge by
90°.
13
Short Description: BASE Module
View into the open device (BASE module, 3 function modules installed)
BASE module
2 current outputs (free assignment of
process variable) and 4 relay contacts,
2 digital inputs.
VariPower broad-range power supply,
20 ... 265 V AC/DC, suitable for all public
mains supplies in the world.
Power supply units, IS version:
100 ... 230 V AC or
24 V AC/DC
Important notice concerning SmartMedia card
The SmartMedia card may be inserted or replaced with the power supply
switched on. Before a memory card is removed, it must be “closed” in the
maintenance menu. When closing the device, make sure that the sealing is
properly seated and clean.
Module equipment
Module identification: Plug & Play.
Up to 3 modules can be combined as
desired. Several input and communication
modules are available.
Warning!
Do not touch the terminal compartment, there may be
dangerous contact voltages!
14
PROFIBUS Technology
PROFIBUS is a digital communication system that connects different field
devices over a common cable and integrates them into a control system. In
the long term, PROFIBUS will replace the 4-20mA technology, which only
supplies pure measured values. Advantages of the PROFIBUS technology are:
easy and cost-saving cabling
convenient operation over a central control station
transmission, evaluation, and control of high amounts of data from field
device to control station.
devices installed in hazardous locations are configured and maintained
from the control station
PROFIBUS is the leading open fieldbus system in Europe. Its application range
covers manufacturing, process, and building automation. As open fieldbus
standard to EN 50170 and IEC 61158, PROFIBUS ensures communication of
different devices over one bus. The PROFIBUS User Organization (PNO) pro-
vides for further development and maintenance of the PROFIBUS technology.
It combines the interests of users and manufacturers.
Variants and Basic Characteristics
PROFIBUS determines the technical and functional characteristics of a serial
bus system. There are three PROFIBUS variants:
PROFIBUS-DP (decentralized peripherals)
Tailored for communication of automation systems and distributed periph-
erals.
RS 485 standard with transmission rates up to 12 MBits/sec
PROFIBUS-PA (process automation)
Dedicated to the process industry. It permits connection of sensors and
actuators to a common bus even in hazardous locations. PROFIBUS-PA has
a transmission rate of 31.25 kBits/sec.
15
PROFIBUS Technology
PROFIBUS distinguishes between two types of devices:
Masters
Control the data traffic on the bus. They send messages without external
request.
Slaves
Peripheral devices such as valves, drives, transmitters, and analyzers. They
can react acyclically to servicing, configuration and diagnostic tasks of the
master. The central controller cyclically reads the measurement data with
status.
Definitions for PROFIBUS-PA
The bus protocol defines type and speed of the data exchange between
master and slave devices and determines the transmission protocol of the
respective PROFIBUS system.
PROFIBUS-PA permits cyclic and acyclic services.
Cyclic services are used for transmission of measurement data and actuating
commands with status information.
Acyclic services are used for device configuration, maintenance and diag-
nostics during operation.
The PA 3.0 device profile defines the device class and typical functionalities
with parameters, ranges, and limit values.
The FISCO model developed by the German PTB for hazardous locations per-
mits connection of several devices to one common bus and defines permis-
sible limits for device and cable parameters.
16
Terminal Plate
PA 700(X) Module
Attaching the Terminal Plates
The terminal plates of the lower modules
can be sticked to the inner side of the door.
This facilitates maintenance and service.
Terminal Plate PA 700 Module:
Terminal Plate PA 700X Module:
17
Inserting the Module
Switch off power supply
Open the device (loosen the 4 screws at the front)
Place module in slot (D-SUB connector)
Tighten fastening screws of the module
Connect signal lines
Close device, tighten screws at the front
Switch on power supply
Assigning process variables to AI blocks on the device
Set parameters
Make sure that the cable
glands are tightly closed to
protect against humidity.
Thanks to the staggered arrangement of
connectors and fastening screws the termi-
nal strips of all modules are easy to access.
1.
2.
3.
4.
5.
6.
7.
8.
9.
18
PROFIBUS PA Installation
Basic build-up of a PROFIBUS system:
e.g. M 700(X)
with module
PA 700(X)
Shield
12
13
PA 700(X) Module
PROFIBUS cable
PA- (green)
PA+ (red)
14
Electrical connection between module and PROFIBUS PA is in accordance
with the PROFIBUS Guideline, Order No. 2.092 (www.profibus.com).
PROFIBUS-DP
PROFIBUS-PA
Slave 1
Control room
Slave 2 Slave n
...... ..... .....
AI 1
Analog Input
Function Block
Physical Block
Global Status
AI 2
Analog Input
Function Block
AI 3
Analog Input
Function Block
AI 4
Analog Input
Function Block
Module Selection
Process Variable Selection
Protocols
Calibration
AO
Analog Output
DI 1
DI 2
DI 3
DI 4
DO 1
DO 2
DO 3
DO 4
Meas. Value Correction
EC 400 Status
CONTACT Status
EC 400 Message
EC 400 Step
HOLD Control
PARSET
EC 400 Control
LOCK Control
PROFIBUS PA
Module I
Module II
Calc. Block I
Calc. Block II
ATB 1
Analyzer Transducer
Block
ATB 2
Analyzer Transducer
Block
ATB 3
Analyzer Transducer
Block
ATB 4
Analyzer Transducer
Block
AI 5
Analog Input
Function Block
AI 6
Analog Input
Function Block
AI 7
Analog Input
Function Block
AI 8
Analog Input
Function Block
Module Selection
Process Variable Selection
Protocols
Calibration
Module I
Module II
Calc. Block I
Calc. Block II
ATB 5
Analyzer Transducer
Block
ATB 6
Analyzer Transducer
Block
ATB 7
Analyzer Transducer
Block
ATB 8
Analyzer Transducer
Block
20
Communication Model
See diagram on previous side
The device parameters are sorted in three types of blocks:
Physical Block (PB)
This block contains the general parameters which apply to the whole device.
Transducer Blocks (TB 1 ... TB 8)
8 analog blocks. They contain measurement parameters (process variable,
temperature) according to the PROFIBUS-PA Profile 3.0 specification.
Function Blocks
8 analog input blocks (AI1..4, AI5..8, for scaling measured values),
4 digital output blocks (DO 1 ... DO 4, for control signals)
4 digital input blocks (DI 1 ... DI 4, for status messages).
1 analog output block (AO 1) for analog compensation signals,
e.g. O
2
process pressure.
Physical Block (PB)
This block contains the device-specific parameters (model designation, manu-
facturer ID, serial number...) and controls basic device functions such as:
Write protection
(“WRITE_LOCKING” parameter) Enables or locks acyclic services
(maintenance, configuration).
Blocking operator access to the device
(“LOCAL_OP_ENA” parameter)
Enables or locks access via the user interface on the device.
Notice:
When communication fails for more than 30 seconds,
the device automatically switches to local access.
Reset
(“FACTORY_RESET” parameter)
Caution – data loss!
Resets all configuration values to factory setting.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80