Gotop RigiScan Plus User manual

Type
User manual
User Guide
CAUTION: Federal (USA) law restricts this device to sale by or on the order of a physician.
-2-
RIGISCAN
®
(User Guide)
©2016 GOTOP Medical
This manual pertains only to the GOTOP Medical RigiScan® Plus System. All rights reserved. Printed in the
U.S.A. U.S. Patent Nos. 4,515,166 and 4,766,909 apply.
Le présent manuel ne s’applique qu’au système RigiScan® Plus d’’GOTOP Medical . Tous droits réservés. Im-
primé aux Etats- Unis d’Amérique. Les brevets U.S. n° 4,515,166 et 4,766,909 sont applicables.
Dieses Handbuch behandelt ausschließlich das GOTOP Medical RigiScan® Plus Rigidity Assessment System.
Alle Rechte vorbehalten. Gedruckt in den U.S.A. U.S. Patent Nr. 4,515,166 und 4,766,909 kommen zur
Anwendung.
Questo manuale riguarda soltanto il sistema di valutazione della rigidità RigiScan® Plus della GOTOP Medical.
Tutti i diritti sono riservati. Stampato negli USA. Sono validi i brevetti statunitensi No. 4, 515, 166 e 4,766,909.
Este manual concierne sólo al Sistema de Evaluación de Rigidez RigiScan® Plus de GOTOP Medical . Se
reservan todos los derechos. Impreso en los Estados Unidos. Se aplican las patentes estadounidenses núm.
4,515,166 y 4,766,909.
Deze handleiding heeft alleen betrekking op het GOTOP Medical RigiScan® Plus-systeem voor evaluatie van
stijfheid. Alle rechten voorbehouden. Gedrukt in de V.S. Octrooinummers (V.S.) 4,515,166 en 4,766,909 zijn
van toepassing.
Este manual refere-se apenas ao Sistema de Avaliação de Rigidez RigiScan® Plus da GOTOP Medical . Todos
os direitos reservados. Coberto pelas Patentes Nos. 4.515.166 e 4.766.909 dos E.U.A.
-3-
Table of Contents
Chapter 1: Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
1.1 Contraindications, Warnings, Cautions, and Side Effects . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
1.2 Principles of Operation .........................................................7
1.3 The RigiScan Plus Monitoring System ...............................................8
Chapter 2: Setting Up the RigiScan Plus System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
2.1 System Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
2.2 Setting Up the Hardware ........................................................9
2.3 Connecting the Event Marker (Optional) ...........................................10
2.4 Installing the RigiScan Plus Software ..............................................11
Chapter 3: Software Operation...........................................................13
3.1 Starting RigiScan Plus..........................................................13
3.2 The RigiScan Plus Home Screen..................................................14
3.3 RigiScan Control Window ......................................................15
Initializing the Monitor.........................................................16
3.4 Patient Management Window ...................................................18
3.5 Report Manager Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25
Filtering Available Patients ......................................................26
Report Selector ..............................................................27
3.6 Data Management Window ....................................................28
3.7 RigiScan Settings Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .33
General Tab ................................................................33
Rx Codes Tab . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .36
Drug Codes Tab . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .37
Prosthesis Codes Tab ..........................................................38
About Tab . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .39
Chapter 4: Preparing the RigiScan Plus System for Patient Testing .............................. 40
4.1 Installing the Tension Guides ................................................... 40
4.2 Removing the Tension Guides .................................................. 41
4.3 Installing the Loop Covers ..................................................... 42
4.4 Removing the Loop Covers .................................................... 43
4.5 Installing the Batteries ........................................................ 43
4.6 Attaching the Leg Strap ....................................................... 44
Chapter 5: Patient Testing ...............................................................45
5.1 Nocturnal and Provocative Collection Modes .......................................45
Nocturnal Testing.............................................................45
Provocative Testing ...........................................................45
5.2 Instructing the Patient .........................................................46
Removing the Loops Quickly ....................................................46
5.3 Testing Patients Using the Real-time and Ambulatory Modes ............................47
Real-time Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .47
Ambulatory Testing ...........................................................48
-4-
RIGISCAN
®
(User Guide)
Table of Contents (continued)
Beep Cues .... ...............................................................50
5.4 The Characterization Session . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .50
5.5 Downloading Data from the Monitor..............................................50
Chapter 6: RigiScan Plus Reports..........................................................51
6.1 The Session Scan Summary Report................................................51
6.2 The Summary Analysis Statistic Report .............................................52
6.3 Printing Session Scan Summary and Summary Analysis Statistics Reports ...................53
6.4 Patient Summary Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .53
6.5 Treatment Summary Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .54
6.6 General Session Info Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .54
6.7 Base Data Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .55
6.8 Tip Data Report ..............................................................55
Chapter 7: Maintenance ................................................................56
7.1 Maintenance ................................................................56
Glossary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .57
Appendix A: Product Specications Physical Dimensions .......................................60
Appendix B: Patient Instructions ..........................................................61
Appendix C: References ................................................................65
Appendix D: Electromagnetic Declaration ... . .. . .. . .. . .. . .. . .. . .. . .. . . . . . . . . . . . . . . . . . . . . . . .66
Appendix E: RigiScan Nomogram .........................................................70
Appendix F: Limited Warranty ............................................................71
-5-
Contraindications
Not for use in evaluating erectile capability in children.
Do not use the RigiScan Plus Monitor in men who have communicable skin or venereal dis-
eases or men who have any rash or lesions on the penis or surrounding area.
Do not use in men with impaired mental capacity unless the testing period is directly super-
vised by appropriate medical personnel.
Chapter 1: Introduction
The RigiScan Plus System allows you to measure the duration, frequency, and degree of rigidity and
tumescence of the penis. This device provides a graphical and tabular display of base and tip penile
rigidity and tumescence data for the physician’s interpretation. The design of the RigiScan Plus System
makes it simple for you and your patients to use, yet it offers a sophisticated range of capabilities and
applications. The system is suitable for nocturnal data collection or data collection during various
provocative interventions that are intended to produce an erectile response. Testing may be conduct-
ed in a stand-alone Ambulatory mode or in a Real-time monitoring mode when connected to your
computer.
This chapter contains important information which must be observed to ensure patient safety and
proper operation of the RigiScan Plus system.
Warnings
Adequate patient education is required to assure safe operation of the RigiScan Plus by a patient for
in-home nocturnal testing.
DO NOT use the RigiScan Plus Monitor if the tension guide boots are loose, the cable is
kinked or curled, or the tension guide is crushed or broken. Use of worn or damaged tension
guides may result in patient injury.
DO NOT use the RigiScan Plus Monitor with improperly or incompletely installed tension
guides. The tension guides must be inserted in the monitor with the tension guide pin fully
seated in the slot on the top of the housing. Use of a RigiScan Plus with improperly inserted
tension guides may result in patient injury.
DO NOT operate the RigiScan Plus Monitor near sources of intense pulsed or conducted
electrical or magnetic elds such as Magnetic Resonance Imaging (MRI), CT scan, X-ray,
-6-
RIGISCAN
®
(User Guide)
electrocautery, electrocoagulation, diametry or telemetry equipment, high power transmit-
ters, or household appliances such as electric blankets, electric razors, power tools, micro-
wave ovens, hair dryers or cellular phones. Items such as these may interfere with operation
of the monitor and cause erroneous readings.
Patients MUST NOT bath or shower while wearing the RigiScan Plus Monitor. Getting the
monitor wet may cause electrical shock to the patient or damage to the monitor.
Patients MUST NOT engage in any sexual activity, such as intercourse, oral sex or mastur-
bation, during RigiScan Plus testing. Doing so may result in patient injury or erroneous data.
Patients with impaired penile sensation are at greater risk of exposure to maximum compres-
sive forces should the system fail to operate properly.
DO NOT turn the RigiScan Plus Monitor “on” until the loops have been placed on the penis.
Turning the monitor on before placing the loops will cause continuous rigidity sampling and
may result in patient discomfort, pain, irritation or bruising.
The RigiScan Monitor must be turned “off” prior to urination. The monitor may be turned
off for a continuous period of up to 15 minutes without interrupting the monitoring session.
Urinating with the monitor on may cause pain, urethral irritation or hematuria.
The RigiScan Monitor must be turned “off” prior to readjusting the position of the loops on
the penis. Adjusting the loops with the monitor on may cause continuous rigidity sampling
and result in patient discomfort, pain, irritation or bruising.
Reuse of disposable loop covers may result in patient injury or transmission of communica-
ble disease.
When the RigiScan Plus System is to be used in a Real-time Monitoring Mode (RTM), the host
PC must use an isolation transformer manufactured to EN60601 standards. Failure to do so
may result in a safety hazard.
Cautions
Ingestion of coffee, tea, cola, alcoholic beverages, sedatives, tranquilizers, muscle relaxants
or sleeping pills in the period immediately preceding nocturnal testing with the RigiScan Plus
may affect test results by interfering with normal sleep patterns.
Warnings (continued)
-7-
Care must be taken to remove the tension guides slowly and gently to avoid damage to the
cables or internal mechanisms. Removing the tension guides too quickly may damage the
monitor or result in an “out-of-phase” condition and inability to re-insert the tension guides.
USE ONLY the communications cable supplied with the RigiScan Plus system. Alternate
communications cables are not compatible with the system and may damage the RigiScan
Plus Monitor.
USE ONLY Duracell 9-volt disposable alkaline or GOTOP approved rechargeable 9-volt bat-
teries. Other batteries may not have sufcient power to complete a Monitoring session.
Federal (USA) law restricts this device to the sale or use by or on the order of a physician.
Side Effects and Complications
The side effects and complications listed below have been reported with use of the device:
Penile pain, redness, bruising, abrasion or cuts of the penile skin from action of the loops
Urethral bleeding
Hematuria
1.2 Principles of Operation
The RigiScan Plus Monitor is an ambulatory data logging unit that measures and records penile rigidity
and tumescence. It is linked by cable to an IBM-compatible computer, which downloads and process-
es the data, stores it for review, and prints it on command. The ambulatory monitor uses tip and base
penile loops that adjust by tightening slightly at discrete time intervals to measure and record penile
rigidity and tumescence. Each loop contains a cable that moves freely inside a conduit. Each loop
takes a measurement every 15 seconds. The loop gently tightens with a linear force of 4 ounces (114g)
and then immediately releases and the tissue rebounds to its unloaded state. The RigiScan Plus Monitor
then takes a tumescence measurement. After the tumescence measurement is taken, it is compared to
the previous samples. When a 6 mm increase in tumescence is detected, representing possible erectile
activity, the RigiScan Plus Monitor takes a second measurement every 30 seconds. After tumescence
is measured, the loops tighten a second time around the circumference of the penis with a linear force
of 10 ounces (283.5g). The RigiScan Plus Monitor takes a measurement when this force is applied to
record a cross sectional response to radial compression. This is how rigidity is measured.
-8-
1.3 The RigiScan Plus Monitoring System
The RigiScan Plus System consists of the following components and accessories:
RigiScan Plus Monitor
RigiScan Plus Program (USB)
Carrying case
Box of Leg straps
Communications cable
Event Marker cable (optional)
RigiScan Plus User’s Manual
Set of tension guides
Box of loop covers
Box of Duracell 9-volt Batteries
Optional Accessories:
Rechargeable Batteries and Charger are available for purchase from GOTOP Medical Inc.
NOTE: Excluding the customer provided EN60950 compliant PC, all components of the RigiS-
can system are suitable for use in the patient environment.
RIGISCAN
®
(User Guide)
Fig. 1-1
-9-
Chapter 2: Setting Up the RigiScan Plus System
2.1 System Requirements
Before you can install the RigiScan Plus Software, you must assemble the following hardware com-
ponents:
EN 60950 Compliant IBM-compatible personal computer (PC) equipped with:
Microsoft Windows XP Service Pack 3 or newer (Windows 7, 8, 8.1, and 10)
Hard-disk with at least 1 GB of free disk space
At least 1 GB of random-access memory (RAM)
Available USB 2.0 port
VGA compatible graphics card
VGA monitor
Mouse and keyboard
Printer
Supplied USB Flash Drive
Note: Windows DPI setting should be 100% for optimum display. “Unsupported DPI” warning may
appear at launch when using other DPI setting.
2.2 Setting Up the Hardware
The RigiScan Plus Monitor connects to your computer with the supplied communication cable. The
15-pin male connector attaches to the RigiScan Plus Monitor (Fig. 2-1) and the USB plug attaches to
an available USB 2.0 port on the computer (Fig. 2-2).
Fig. 2-1 Fig. 2-2
-10-
RIGISCAN
®
(User Guide)
2.3 Connecting the Event Marker (Optional)
You can use the optional Event Marker to mark pertinent points in the patient data.
To connect the Event Marker cable:
Connect the small, round 4-pin connector on the communications cable to the round connector on the
Event Marker cable. To disconnect the Event Marker cable, pull back on the metal connector sheath.
-11-
2.4 Installing the RigiScan Plus Software
To install RigiScan Plus Software:
1. Insert the USB Flash Drive into an available USB port on your computer.
2. Allow Windows time to recognize the USB Flash Drive and install driver if necessary.
3. Use Windows Explorer to navigate to the Flash Drive’s root directory.
4. Double click the RigiScan-Inst (v5.03).exe excecutable le.
5. When prompted, click Next in the installer window to begin installation.
The RigiScan Plus software will begin installing as indicated by a progress bar.
Note: You may need Administrator Priveleges to install software in Windows.
-12-
RIGISCAN
®
(User Guide)
When the progress bar reaches 100%, installation of the USB to Serial Driver will automatically
begin as indicated by the pop-up window below.
When installation of the USB to Serial driver is complete, click Exit to nish installation.
-13-
Chapter 3: Software Operation
3.1 Starting RigiScan Plus
To start the RigiScan Plus application:
1. Double-click the RigiScan icon on the desktop
or
1. Click the Start button on the task bar
2. Navigate to All Programs -> GOTOP Medical RigiScan PLUS
3. Click RigiScan
When running RigiScan for the rst time you may be prompted to enter licensing information.
Enter the name of the business or individual to whom the software is licensed, then click Save.
-14-
RIGISCAN
®
(User Guide)
3.2 The RigiScan Plus Home Screen
When you start the RigiScan Plus software, the RigiScan Plus Home Screen will appear in a new
window.
A column of buttons, along with functional descriptions of each are provided.
RigiScan - opens the RigiScan Control window (see Section 3.3)
Patient Mgmt - opens the Patient Management window (see Section 3.4)
Reports - opens the Report Manager window (see Section 3.5)
Data Mgmt - opens the Data Management window (see Section 3.6)
Settings - opens the RigiScan Settings window (see Section 3.7)
-15-
3.3 RigiScan Control Window
Clicking the RigiScan button on the RigiScan Plus Home Screen opens the RigiScan Control window.
Here the user may download session data from a previous ambulatory or real-time monitoring session,
initialize the device for a new patient, or start a Real-Time Monitoring session.
If clicking the RigiScan button on the RigiScan Plus Home screen opens the Error Communicating
pop-up window, click OK, verify that the monitor is turned on and properly connected to the PC, and
check that the correct communication port is selected in the Setting window (see Section 3.7). It may
be necessary to power cycle the device.
-16-
Initializing the Monitor
To Initialize the Device (New Patient):
1. Click the Initialize Device button in the RigiScan Control window. This opens the Device Ini-
tialization window.
2. Click the Add New button to add a new patient. This opens the Patient Information window.
3. Enter pertinent information into the Patient Information window (see Section 3.4 for details).
4. Click the button.
5. Click the button in the Patient Information window to return to the Device Initialization
window.
6. When prompted with the Conrm Close pop-up window, click OK.
7. The Patient ID you just entered should now be displayed in the Device Initialization window.
8. Select either Nocturnal or Provocative from the Test Method drop-down menu.
9. Click OK to complete initialization and return to the RigiScan Control window.
To Initialize the Device (Existing Patient):
1. Click the Initialize Device button in the RigiScan Control window. This opens the Device Ini-
tialization window.
2. Click the Select Existing button. This opens the Patient Management window.
3. Use the Search and Quick Search features as necessary to locate the desired patient record (see
Section 3.4 for Search and Quick Search details).
4. Double-click the desired patient record to close the Patient Management window and return to
the Device Initialization window. The Patient ID you selected should now be displayed in the
Device Initialization window.
5. Select either Nocturnal or Provocative from the Test Method drop-down menu.
6. Click OK to complete initialization and return to the RigiScan Control window.
RIGISCAN
®
(User Guide)
-17-
To Start Monitoring Session:
See Chapter 5: Patient Testing for details on starting Real-Time and Ambulatory Monitoring Sessions.
To Download Patient Data (following a monitoring session):
1. Following an ambulatory or real-time monitoring session, connect the device to the PC via the
supplied communication cable.
2. Turn the device on, if it isn’t already. It may be necessary to power cycle the device to re-establish
communications.
3. Navigate to the RigiScan Control window.
4. If the device contains monitoring data not yet downloaded to the PC database, the RigiScan Con-
trol window will display “NOT Downloaded” in the Download Status area.
5. Click the Download Now button.
6. A Data Download Options pop-up window will appear with options to Auto Print Graphs
and Auto Print Summary Stats. These options will be set with the default settings. If they
are changed, they will only be true for this particular download. The default settings can be
changed in the Settings screen (see Section 3.7 for RigiScan Settings Window details).
7. Then, click Download to begin downloading session data.
8. A progress bar will appear indicating the downloading status between the PC and the device.
-18-
RIGISCAN
®
(User Guide)
3.4 Patient Management Window
Clicking the Patient Mgmt button on the RigiScan Plus Home Screen opens the Patient Management
window. Here, the user may add, view, edit, delete, import, and export patient records.
To aid in viewing, editing, and deleting patient records, a search feature as well as quick search but-
tons are available along the left side of the window.
To Search for an individual patient record:
1. Type a patient ID number into the Patient ID text box.
2. Click the Search button.
To view all patients who visited on or after a specic day:
1. Click the Recently Visited button.
2. When prompted with the Last Visit Date pop-up
window, select a date.
3. Click OK.
To view all Archived patients:
1. Click the All Archived button.
To view all Active patients:
1. Click the All Active button.
-19-
To Add A New Patient (Manual Entry):
1. Click the to open the Patient Information window.
2. Fill in pertinent patient information on the General tab.
3. Click the Treatment History tab.
4. Click Add New to open the Treatment Record pop-up window.
5. Use text entry elds and drop-down menus to ll in the pertinent treatment history information.
6. Click the button to save treatment record and return to the Patient Information window.
7. Click the button in the Patient Information window to save patient data.
8. Click the button to return to the RigiScan Plus Home Screen.
9. When prompted with the Conrm Close pop-up window, click OK.
-20-
RIGISCAN
®
(User Guide)
To Import a New Patient From File:
1. Click the button to open the Patient Information window.
2. Click the button.
3. Use the dialog box to navigate to and select the le to import data from (usually a le with
extension “.dat”).
4. Click Open in the dialog box.
5. Click the button to save patient record.
6. Click the button to return to the RigiScan Plus Home Screen.
7. When prompted with the Conrm Close pop-up window, click OK.
To Edit an Existing Patient Record:
1. Click the Patient ID of the record you wish to edit, then click the button to open the Pa-
tient Information window.
or
Double-click the Patient ID of the record you wish to edit to open the Patient Information
window.
2. Modify elds in the General tab as necessary.
3. Click the button to save patient record.
4. In the Treatment History tab, use the Add New and Edit Selected buttons to create or modify
treatment records as done in To Add A New Patient (Manual Entry).
5. Use the Delete Selected button to delete existing treatment records.
6. Click the button to save changes made to treatment records.
7. In the RigiScan tab, use the Delete Selected Group button if you wish to remove any RigiScan
data from the patient record.
8. Click the button to save changes made to session records.
9. Click the button to return to the RigiScan Plus Home Screen.
10. When prompted with the Conrm Close pop-up window, click OK.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74