Front Panel Menus - At a Glance
ADDRESS 7 Sets the GPIB Address
INTF GPIB Selects an interface (GPIB or RS232)
BAUDRATE 300 Selects baud rate (300, 600, 1200, 2400, 4800, 9600)
PARITY NONE Selects message parity (NONE, EVEN, ODD, MARK, SPACE)
FLOW NONE Selects flow control (XON-XOFF, RTS-CTS, DTR-DSR, NONE)
LANG SCPI Selects language (SCPI or COMP)
*RCL 0 Recalls instrument state
*SAV 0 Saves present instrument state
ERROR 0 Displays errors in SCPI error queue
12.000V 0.204A Measures output voltage and current
12.500V MAX Measures peak output voltage1
1.000V MIN Measures minimum output voltage1
12.330V HIGH Measures the high level of a voltage pulse waveform1
0.080V LOW Measures the low level of a voltage pulse waveform1
12.000V RMS Measures rms voltage1
0.350A MAX Measures peak output current1
0.050A MIN Measures minimum output current1
0.400A HIGH Measures the high level of a current pulse waveform1
0.012A LOW Measures the low level of a current pulse waveform1
0.210A RMS Measures rms current1
VOLT 20.000 Sets the output voltage
CURR 2.000 Sets the output current
OC -- -- -- -- Protection status (example shows overcurrent tripped)
*RST Places the dc source in the factory-default state
PON:STATE RST Select the power-on state command (RST or RCL0)
PROT:DLY 0.08 Sets the output protection delay in seconds
RI LATCHING Sets the remote inhibit mode (LATCHING, LIVE, or OFF)
DFI OFF Sets the discrete fault indicator state (ON or OFF)
DFI:SOUR OFF Selects the DFI source (QUES, OPER, ESB, RQS, or OFF)
PORT RIDFI Sets the output port functions (RIDFI or DIGIO)
DIGIO 7 Sets and reads the I/O port value (0 through 7)
RELAY ON Sets the output relay state (ON or OFF) 2
RELAY NORM Sets the output relay polarity (NORM or REV) 2
VOLT:PROT 22 Sets overvoltage protection level
CURR:RANG HIGH Sets current range (HIGH, LOW, or AUTO)
CURR:DET ACDC Sets current measurement detector (ACDC or DC)1
CAL ON Accesses calibration menu (See User’s Guide).
Use and to select menu parameters. Use to exit any menu and return to metering mode.
1Not available on Agilent 6631B - 6634B or in Compatibility mode. 2Not available on Agilent 6631B.