Midmark Digital Spirometer Reference guide

  • Hello! I am an AI chatbot trained to assist you with the Midmark Digital Spirometer Reference guide. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
01 Click on the Settings button from the IQmanager
®
main screen and navigate to the Devices tab.
02 Select the Spirometry Settings button.
03 Under the Conguration section, select PCP Mode
from the Conguration Prole dropdown.
04 Click the Save button on the bottom left hand side of
the screen. This will save the setting change and return
you to the IQmanager Congurations Settings screen.
05 Click the Back Arrow in the Conguration Settings screen. This
will take you back to the IQmanager Opening Screen.
06 Search for a patient or select New Patient to start a
spirometry test.
Quick Reference Guide
Instructions for putting the Midmark Digital Spirometer
in PCP mode
© 2019 Please refer to the Midmark Digital Spiromter Operation Manual #56-78-0001 for further details and troubleshooting tips.
Interpretation Decision Tree
And low vital capacity, cannot rule out superimposed restriction.
FVC Below 80% of predicted?
START
Acceptable FVC? Repeat
Best of 3?
Ratio
Abnormal
FVC Normal
Restriction
Obstruction
Normal Spirometry
70
%
<
FVC
<
LLN
of predicted?
FEV
1
>
100
%
of predicted?
34
%
<
FVC
<
50
%
of predicted?
50
%
<
FEV
1
<
60
%
of predicted?
60
%
<
FVC
<
70
%
of predicted?
70
%
<
FEV
1
<
100
%
of predicted?
34
%
<
FEV
1
<
50
%
of predicted?
50
%
<
FVC
<
60
%
of predicted?
60
%
<
FEV
1
<
70
%
of predicted?
FVC
<
34
%
of predicted?
FEV
1
<
34
%
of predicted?
no
no no
no
no
no
no
no
no
no
no
no
no
May be a
Physiological
Variant
Mild
Restriction
Severe
Restriction
Moderate
Restriction
Very
Severe
Restriction
Moderately
Severe
Restriction
Moderately
Severe
Obstruction
Mild
Obstruction
Severe
Obstruction
Moderate
Obstruction
Very
Severe
Obstruction
yes
yes
yes
yes yes
yes yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
Midmark Corporation | 690 Knox Street, Suite 100, Torrance, CA 90502 | Support Services 1-844-856-1230 #2 | midmark.com 003-10243-00 Rev. A
/