HPP27XMOUNTINGSET2

Vaisala HPP27XMOUNTINGSET2, HPP271MOUNTINGSET2, HPP272 User manual

  • Hello! I am an AI chatbot trained to assist you with the Vaisala HPP27XMOUNTINGSET2 User manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
Through-Wall Installation Kit for HPP271 and HPP272
The optional through-wall installation kits for HPP271 (HPP271MOUNTINGSET2) and HPP272
(HPP272MOUNTINGSET2) allow installing the Vaisala HPP270 series H
2
O
2
probes into the monitored
process through the process wall.
The figure below shows an example installation for HPP272. The HPP271 installation kit is otherwise the
same, but does not include the HPP272-specific temperature probe lead-through. The figure includes
the flange drilling dimensions.
3
60 ... 100 mm
4
Ø 5 mm
Ø 27 mm
40 mm
60 mm
10
mm
2
1
Metal parts: AISI 316L
Seals and O-rings: FPM
Figure 1 Through-Wall Installation Example for HPP272
1
Maximum wall thickness: 4 mm (0.16 in)
2 Screws for tightening the flange in place (4 pcs, Ø 5 mm (0.19 in))
3 Nut for mounting the temperature probe gland (HPP272 only)
4 Temperature probe gland diameter: 8 mm (0.31 in) (HPP272 only)
Seal the lead-throughs on the metal body of the probes.
HPP272 only: Install the main probe body and the temperature probe 60 … 100 mm (2.36 … 3.93 in)
apart from each other. The temperature probe must be far enough to prevent heat from the main probe
body from aecting the temperature measurement, but close enough so that the two probes measure
the same environment. When installing the probes horizontally, place the temperature probe
downstream of the warmed main probe body.
M212194EN-B
Quick Guide
Mounting Accessory
HPP271MOUNTINGSET2 / HPP272MOUNTINGSET2
PUBLISHED BY
Vanha Nurmijärventie 21
FI-01670 Vantaa, Finland
© Vaisala 2019
All rights reserved. Any logos and/or product names are trademarks of Vaisala or its individual partners.
Any reproduction, transfer, distribution or storage of information contained in this document is strictly
prohibited. All specifications — technical included — are subject to change without notice.
HPP271MOUNTINGSET2/HPP272MOUNTINGSET2
Dimensions
Figure 2 (page 2) shows the dimensions of the flange through-wall installation accessory
HPP271MOUNTINGSET2/HPP272MOUNTINGSET2 in millimeters and [inches].
Ø 12 [0.47]
Ø 60 [2.36]
40 [1.57]
60 [2.36]
Ø 18.7 [0.74]
M4
10 [0.39]
mm
[in]
24 [0.94]
14 [0.55]
Ø 12
[0.47]
Ø 8 [0.31]
Ø 5 [0.196]
Figure 2 HPP271MOUNTINGSET2/HPP272MOUNTINGSET2 Dimensions
Quick Guide
M212194EN-B
2
/