Mettler Toledo SevenCompact™ S210 Operating instructions

  • Hello! I am an AI chatbot trained to assist you with the Mettler Toledo SevenCompact™ S210 Operating instructions. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
Switching the instrument on/off
Press and
release to switch
on, press for 2s
and release to
switch off.
Changing the mode
Mode
Pressing MODE
changes the
mode: pH, mV,
rel. mV.
Performing a pH calibration
Switch to
pH mode.
Insert electrode
into first pH
buffer.
Blinking A: mea
suring
Change pH buffer group
if needed.
q
Value stable
q
Insert electrode
into second pH
buffer.
Value stable
q
Press END to end calibration or repeat the last steps for a
multi-point calibration.
q
Calibration
result.
q
Press SAVE to save calibration.
SevenCompact™ S210
QuickGuide - English
Switch display view
Display with all
information.
q
Display with
most important
information.
(uFocus™)
Performing a measurement
Insert electrode
into sample.
Blinking A: mea
suring
q
Value stable
Endpoint type
can be set to
automatic, man
ual or timed in
the menu set
tings.
Storing measurement data
Store
q
Storage number
is incremented.
Note: automatic storage to memory can be switched on in
the setup menu.
Accessing menu settings
Menu
Press to enter the
menu settings,
e.g. to enter
userID, sam
pleID and Sen
sorID.
*30130406A*
Mettler-Toledo AG, Analytical
CH-8603 Schwerzenbach, Switzerland
Tel. +41 (0)44 806 77 11
Fax +41 (0)44 806 73 50
www.mt.com/pH
Subject to technical changes.
© Mettler-Toledo AG 02/2014
30130406A
/