SICK ZIRKOR200 Ex-D Operating instructions

Category
Measuring, testing & control
Type
Operating instructions
MMM
I
MMMOPERATING INSTRUCTIONS
OPERATING INSTRUCTIONS
ZIRKOR200 Ex-D
Oxygen Analyzer
Title page
2
8024252/V1-0/2020-07 | SICKOPERATING INSTRUCTIONS | ZIRKOR200 Ex-D
Subject to change without notice
Described product
Product name: ZIRKOR200 Ex-D
Manufacturer
SICK AG
Erwin-Sick-Str. 1 · D-79183 Waldkirch · Germany
Phone: +49 7641 469-0
Fax: +49 7641 469-1149
Legal information
This work is protected by copyright. Any rights derived from the copyright shall be
reserved for SICK AG. Reproduction of this document or parts of this document is only
permissible within the limits of the legal determination of Copyright Law.
Any modification, shortening or translation of this document is prohibited without the
express written permission of SICK AG.
The trademarks stated in this document are the property of their respective owner.
© SICK AG. All rights reserved.
Original document
This document is an original document of SICK AG.
CONTENTS
3
8024252/V1-0/2020-07 | SICK OPERATING INSTRUCTIONS | ZIRKOR200 Ex-D
Subject to change without notice
1 About this document................................................................................. 7
1.1 Function of this document.............................................................................. 7
1.2 Scope of application ....................................................................................... 7
1.3 User qualification ............................................................................................7
1.4 Further information.........................................................................................7
1.5 Symbols and document conventions .............................................................8
1.5.1 Warning symbols............................................................................. 8
1.5.2 Warning levels and signal words.................................................... 8
1.5.3 Information symbols ....................................................................... 9
1.6 Data integrity...................................................................................................9
2 For your safety..........................................................................................10
2.1 Basic safety information ...............................................................................10
2.2 Warning information on device ....................................................................11
2.3 Intended use .................................................................................................11
2.3.1 Purpose of the device...................................................................11
2.3.2 Operation in potentially explosive atmospheres .........................11
2.4 Responsibility of user....................................................................................13
2.4.1 Designated users..........................................................................13
2.4.2 Correct project planning ...............................................................13
2.4.3 Correct use....................................................................................13
2.4.4 Special local conditions................................................................13
2.4.5 Read the Operating Instructions ..................................................13
3 Product description .................................................................................14
3.1 Product identification ...................................................................................14
3.2 Type plates ....................................................................................................14
3.3 Device versions .............................................................................................16
3.4 Device variants..............................................................................................18
3.5 Options ..........................................................................................................18
3.6 Expansion module.........................................................................................18
4 Layout and function.................................................................................19
4.1 System overview............................................................................................19
4.2 Measuring principle ......................................................................................20
4.3 Scope of application .....................................................................................20
4.4 Potential hazards .........................................................................................20
4.5 Process interruption......................................................................................20
4.6 ATEX / IECEx certification .............................................................................21
4.6.1 Ignition source monitoring of the probe by means of
temperature monitoring ...............................................................21
Contents
Contents
CONTENTS
4
8024252/V1-0/2020-07 | SICKOPERATING INSTRUCTIONS | ZIRKOR200 Ex-D
Subject to change without notice
5 Installation................................................................................................ 22
5.1 Determination of the requirements for explosion protection ..................... 22
5.2 Information on installation in potentially explosive atmospheres.............. 22
5.3 Tools required ............................................................................................... 23
5.4 Material required .......................................................................................... 23
5.5 Preparing the measuring point ................................................................... 23
5.6 Transport....................................................................................................... 23
5.7 Storage information...................................................................................... 24
5.8 Scope of delivery .......................................................................................... 24
5.9 Installation .................................................................................................... 24
5.9.1 Fitting the counterflange on the duct ........................................ 25
5.9.2 Installing the control unit ............................................................. 25
5.9.3 Installing the probe ...................................................................... 26
5.9.4 Installing the measuring probe.................................................... 27
5.9.5 Fitting the counterflange.............................................................. 28
5.9.6 Aligning the V-shield..................................................................... 29
5.9.7 Laying the probe cable and hoses............................................... 29
5.9.8 Installing the probe cable ............................................................ 30
5.10 Electrical installation .................................................................................... 31
5.10.1 Accessing the terminals .............................................................. 33
5.10.2 Instructions for the function of the spring-loaded terminal ....... 33
5.10.3 Potential equalization (PE, ground) ............................................. 34
5.10.4 Fitting the ferrite sleeves (EMC) .................................................. 34
5.10.5 Control unit electrical connections.............................................. 35
5.10.6 Outputs/functions and relay assignment.................................... 36
5.10.7 Digital inputs ................................................................................ 37
5.10.8 Electrical connections on the probe junction unit ...................... 37
5.10.9 Electrical connection plan – ZIRKOR Ex-D .................................. 38
5.11 Pneumatic connections and settings .......................................................... 39
5.11.1 Pneumatic hose requirements ................................................... 39
5.11.2 Preparing the pneumatic hose .................................................... 39
5.11.3 Pneumatic connections, probe .................................................... 39
5.11.4 Pneumatic connections of the electronics.................................. 40
6 Commissioning ........................................................................................ 41
6.1 Safety information on commissioning ......................................................... 41
6.2 Initial commissioning.................................................................................... 42
6.3 Display - heating process ............................................................................. 42
6.4 Display - Measuring mode............................................................................ 42
6.5 Operating elements and display .................................................................. 43
6.6 Status LEDs................................................................................................... 43
6.7 Softkey symbols............................................................................................ 43
6.8 System code ................................................................................................. 43
CONTENTS
5
8024252/V1-0/2020-07 | SICK OPERATING INSTRUCTIONS | ZIRKOR200 Ex-D
Subject to change without notice
7 Menu overview and explanations..........................................................44
7.1 SYS-MENU ....................................................................................................44
7.2 Menu explanations........................................................................................47
7.2.1 O
2
measuring ranges (scaling).....................................................47
7.2.2 Measured value average over [s].................................................47
7.2.3 mA output for system error ..........................................................47
7.2.4 O
2
limit value alarms ....................................................................48
7.2.5 O
2
sensor adjustment values.......................................................48
7.2.6 Time per test gas feed..................................................................48
7.2.7 Stopping time until process (O
2
)..................................................49
7.2.8 Auto. Adjustment (ACAL) ..............................................................49
7.2.9 Settings for Auto. adjustment ......................................................49
7.2.10 REMOTE ........................................................................................50
7.2.11 Units of measure ..........................................................................50
7.2.12 Language.......................................................................................50
7.2.13 Changing the system code ...........................................................51
7.2.14 Loading the factory settings.........................................................51
7.2.15 Service (factory service settings) .................................................51
7.2.16 Adjustment menu ........................................................................51
7.2.17 Display - Adjustment.....................................................................51
7.2.18 1-point adjustment (manual) .......................................................52
7.2.19 2-point adjustment (manual) .......................................................53
7.3 System check ...............................................................................................54
8 Maintenance.............................................................................................55
8.1 Technical knowledge necessary for maintenance work..............................55
8.2 Safety instructions for maintenance work .................................................55
8.3 Information on use in Ex-areas .................................................................56
8.4 Maintenance plan .........................................................................................56
8.5 Tools required ...............................................................................................57
8.6 Maintenance work ........................................................................................57
8.6.1 Replace fuses ...............................................................................57
8.6.2 Test air and reference air volumes ..............................................58
8.6.3 Setting the flow volume (Safe Area control unit).........................58
8.6.4 Setting valves position .................................................................60
8.6.5 Setting the flow volume (19" rack)...............................................60
8.6.6 Replacing the filter head .............................................................61
8.6.7 Replacing the measuring probe ..................................................62
8.6.8 Replacing the O
2
measuring cell ................................................63
8.6.9 Replacing the inner part of the measuring probe .......................64
8.6.10 Stability criteria during adjustment..............................................65
9 Status messages .....................................................................................66
9.1 Error messages .............................................................................................66
9.2 Alarm messages ...........................................................................................68
9.3 Maintenance messages ...............................................................................68
CONTENTS
6
8024252/V1-0/2020-07 | SICKOPERATING INSTRUCTIONS | ZIRKOR200 Ex-D
Subject to change without notice
10 Heating shutdown in case of malfunction ........................................... 69
10.1 General information ..................................................................................... 69
10.2 Functional description.................................................................................. 70
10.3 Maintenance ................................................................................................. 70
11 Troubleshooting ....................................................................................... 71
11.1 Strongly fluctuating O
2
measured value ..................................................... 71
11.2 O
2
display remains at end of measuring range or is higher than
expected ....................................................................................................... 71
11.3 Indications on the display are OK, mA output signal is not correct .......... 71
11.4 O
2
display on 0%, although the operating mode indicates a higher
O
2
value ....................................................................................................... 72
12 Decommissioning....................................................................................73
12.1 Safety information on decommissioning ..................................................... 73
12.2 Shutdown procedure .................................................................................... 74
12.3 Removing the device .................................................................................... 74
12.4 Environmentally compatible disposal.......................................................... 74
13 Technical data..........................................................................................75
13.1 Dimension drawings ..................................................................................... 75
13.1.1 Dimension sheets of the control unit .......................................... 75
13.1.2 Dimension sheets of the probes.................................................. 77
13.1.3 Counterflange dimensions........................................................... 78
13.1.4 Probe components ....................................................................... 79
13.1.5 Gas plans...................................................................................... 81
13.1.6 Installation of the electronics ...................................................... 82
13.1.7 Display board................................................................................ 83
13.2 Technical data .............................................................................................. 84
13.2.1 Technical data of the control unit ............................................... 84
13.2.2 Technical data of the measuring probe ..................................... 85
13.2.3 Technical specification for the gas supply .................................. 85
13.3 Tightening torques for Ex-relevant screw connections ............................... 86
7
8024252/AE00/V1-0/2020-07| SICK OPERATING INSTRUCTIONS | ZIRKOR200 Ex-D
Subject to change without notice
ABOUT THIS DOCUMENT 1
1About this document
1.1 Function of this document
These Operating Instructions describe:
Device components
Installation
Operation
Maintenance work required for reliable operation
Decommissioning
1.2 Scope of application
These Operating Instructions are only applicable for the measuring device described in the
product identification.
They are not applicable for other SICK measuring devices.
The standards referred to in these Operating Instructions are to be observed in the
respective valid version.
1.3 User qualification
These Operating Instructions are intended for persons installing, operating and maintaining
the device.
Table 1: Qualification requirements
1.4 Further information
Observe the supplied documents.
Final inspection record ZIRKOR200 Ex-D
USB stick with technical information and certificates
Work User groups Qualification
Installation Operator / system integrator
E.g. plant operator, untrained in measurement
technology
Qualification for explosion protection
Electrical installation Qualified personnel
Authorized electrician (authorized skilled
electrician or person with similar training)
Qualification for explosion protection
Initial commissioning Authorized operator Ü
General knowledge of measurement
technology, device expertise (customer training
at SICK if necessary)
Qualification for explosion protection
Recommissioning
Decommissioning
Operator / system integrator
Authorized operator Ü
E.g. plant operator, untrained in measurement
technology
Authorized electrician (authorized skilled
electrician or person with similar training)
Qualification for explosion protection
Operation
Maintenance
Troubleshooting
8
8024252/AE00/V1-0/2020-07| SICKOPERATING INSTRUCTIONS | ZIRKOR200 Ex-D
Subject to change without notice
1 ABOUT THIS DOCUMENT
1.5 Symbols and document conventions
1.5.1 Warning symbols
Table 2: Warning symbols
1.5.2 Warning levels and signal words
DANGER:
Risk or hazardous situation which will result in severe personal injury or death.
WARNING:
Risk or hazardous situation which could result in severe personal injury or death.
CAUTION:
Hazard or unsafe practice which could result in less severe or minor injuries.
NOTICE:
Hazard which could result in property damage.
Note:
Hints
Symbol Significance
Hazard (general)
Hazard by electrical voltage
Hazard in potentially explosive atmospheres
Hazard by explosive substances/mixtures
Hazard by oxidizing substances
Hazard by toxic substances
Hazard by unhealthy substances
Hazard by high temperature or hot surfaces
Hazard for environment/nature/organisms
9
8024252/AE00/V1-0/2020-07| SICK OPERATING INSTRUCTIONS | ZIRKOR200 Ex-D
Subject to change without notice
ABOUT THIS DOCUMENT 1
1.5.3 Information symbols
1.6 Data integrity
SICK AG uses standardized data interfaces, such as standard IP technology, in its products.
The focus here is on product availability and features.
SICK AG always assumes that the customer is responsible for the integrity and
confidentiality of data and rights involved in connection with using the products.
In all cases, the customer is responsible for the implementation of safety measures
suitable for the respective situation, e.g., network separation, firewalls, virus protection and
patch management.
Symbol Significance
Information on the nature of the product in relation to Directive 2014/34/EU for devices
and protective systems intended for use in potentially explosive atmospheres
Important technical information for this product
Important information for electrical or electronic functions
10
8024252/AE00/V1-0/2020-07| SICKOPERATING INSTRUCTIONS | ZIRKOR200 Ex-D
Subject to change without notice
2 FOR YOUR SAFETY
2For your safety
2.1 Basic safety information
Work on the device
Escaping gases
DANGER: Danger of explosion
Danger of explosion when working on the device.
Ensure no explosive atmosphere is present when working on the device.
DANGER: Danger of explosion due to ignitable mixture in the components
If the filter and probe junction unit cover are not completely closed after working on the
device, sparks may escape to the outside and cause an explosion.
Always close the filter, probe junction unit cover completely after working on the
device.
DANGER:
Risk for system safety through work on the device not described in these
Operating Instructions
Work on the device not described in these Operating Instructions or associated
documents can lead to unsafe operation of the measuring system and therefore
endanger plant safety.
Only carry out the work on the device described in these Operating Instructions and
associated documents.
DANGER:
Danger of explosion through incorrect performance of work described in these
Operating Instructions
Incorrect performance of work in potentially explosive atmospheres can cause serious
injuries to people and damage during operation.
Maintenance and commissioning tasks as well as checks should only be carried out
by experienced/trained personnel with knowledge of the rules and regulations for
potentially explosive atmospheres, especially:
Ignition protection types
Installation rules
Zone classification
DANGER:
Risk of burns and poisoning from escaping hot and toxic gas in systems with
overpressure conditions
In systems with overpressure, hot and toxic gas can escape from the process
connection. This can lead to burns or damage to health.
Always keep the process connection tightly closed.
Pay attention to hot surfaces.
Wear appropriate protective equipment.
11
8024252/AE00/V1-0/2020-07| SICK OPERATING INSTRUCTIONS | ZIRKOR200 Ex-D
Subject to change without notice
FOR YOUR SAFETY 2
Potential equalization
2.2 Warning information on device
On the probe enclosure:
Fig. 1: Warning information on probe enclosure
Do not open under voltage and when the incinerator is in operation. Observe the waiting
time for all probe components to cool down to 133 °C (min. 1 hour).
Follow the instructions in the Manual.
2.3 Intended use
2.3.1 Purpose of the device
The device is a stationary oxygen measuring device and serves continuous measurement
of oxygen as emission and process monitoring in the industrial sector. The device
measures continuously directly in the gas duct (in-situ).
2.3.2 Operation in potentially explosive atmospheres
The ZIRKOR200 Ex-D measuring probe corresponds to the ATEX category (according to
ATEX 2014/34/EU):
II 2D Ex tb IIIC T133 °C/T141 °C Db
The ZIRKOR200 Ex-D measuring probe fulfills the following IECEx qualification:
Ex tb IIIC T133 °C/T141 °C Db
CAUTION:
Danger of explosion through incorrect or missing grounding
Incorrectly connected potential equalization can generate charges that can lead to
explosions in an Ex-atmosphere.
Connect potential equalization to all points provided on the device components.
Ensure the potential equalization is connected during all work on the device
described in these Operating Instructions.
Ensure there is a ground connection via the power supply.
Regularly check the grounding connections for correctness.
Warning
12
8024252/AE00/V1-0/2020-07| SICKOPERATING INSTRUCTIONS | ZIRKOR200 Ex-D
Subject to change without notice
2 FOR YOUR SAFETY
Location of Ex-relevant assemblies, see “System overview”, page 19.
Do not remove, add or modify any components to or on the device unless described and
specified in the official manufacturer information. Otherwise the approval for the device
for use in potentially explosive atmospheres becomes void.
Adhere to the prescribed maintenance intervals, see “Maintenance plan”, page 56.
Special conditions:
The specification of the surface temperature of T133 °C/T141 °C applies to an ambient
temperature range of -20 °C to +40 °C/70 °C.
The heating voltage of the probes must be switched off by a monitoring device
independent of the controller and certified for this purpose when a limit temperature of
810 °C is reached. This task is performed by the heating monitoring system (see “Heat-
ing shutdown in case of malfunction”, page 69).
The oxygen measuring probe with the associated protective tube may only be used in
flue gases whose composition is not critical with regard to their corrosive effect on the
materials used. If this cannot be ensured, regular recurring checks must be carried out
at sufficiently short intervals (see “Maintenance plan”, page 56).
The flue gas temperature on the probe must not exceed 600 °C.
13
8024252/AE00/V1-0/2020-07| SICK OPERATING INSTRUCTIONS | ZIRKOR200 Ex-D
Subject to change without notice
FOR YOUR SAFETY 2
2.4 Responsibility of user
2.4.1 Designated users
see “User qualification”, page 7
The ZIRKOR200 Ex-D should only be installed and operated by skilled technicians who,
based on their technical training and knowledge as well as knowledge of the relevant
regulations, can assess the tasks given and recognize the hazards involved.
2.4.2 Correct project planning
This Manual presumes that the device has been delivered as specified during project
planning (i.e., based on the SICK application questionnaire) and the relevant delivery state
of the device (delivered System Documentation).
If you are not sure whether the device complies with the planned configuration or the
delivered system documentation:
Contact SICK Customer Service.
2.4.3 Correct use
Only use the ZIRKOR200 Ex-D as described in these Operating Instructions (see
“Intended use”, page 11). The manufacturer bears no responsibility for any other use.
Carry out the prescribed maintenance work (see “Maintenance plan”, page 56).
!
Do not carry out any work or repairs on the ZIRKOR200 Ex-D not described in these
Operating Instructions.
Do not remove, add or modify any components to or on the ZIRKOR200 Ex-D unless
described and specified in the official manufacturer information.
Otherwise:
Any warranty by the manufacturer becomes void.
The ZIRKOR200 Ex-D can become dangerous.
The approval for use in potentially explosive atmospheres is no longer valid.
2.4.4 Special local conditions
Follow all local laws, regulations and company-internal operating directives applicable at
the installation location.
2.4.5 Read the Operating Instructions
Only put the ZIRKOR200 Ex-D into operation after reading and understanding the
Operating Instructions.
Observe all safety information.
If anything is not clear: Contact SICK Customer Service.
These Operating Instructions must be:
Available for reference.
Passed on to new owners.
14
8024252/AE00/V1-0/2020-07| SICKOPERATING INSTRUCTIONS | ZIRKOR200 Ex-D
Subject to change without notice
3 PRODUCT DESCRIPTION
3 Product description
3.1 Product identification
3.2 Type plates
The type plate is located on the unit components as shown in the Figures below.
Fig. 2: Type plates (1)
Product name ZIRKOR200 Ex-D DustEx Oxygen analyzer with probe and electonics
Manufacturer SICK AG
Erwin-Sick-Str. 1 · D-79183 Waldkirch · Germany
Type designation Z200EXD-70********************* for control unit and probe
Z200EXD-71********************* for control unit
Z200EXD-72********************* for probe
Type plate of the control unit - safe area inside
Type plate of the control unit - safe area outside
Type plate of the 19" rack mount electronics
15
8024252/AE00/V1-0/2020-07| SICK OPERATING INSTRUCTIONS | ZIRKOR200 Ex-D
Subject to change without notice
PRODUCT DESCRIPTION 3
Fig. 3: Type plates (2)
Type plate of the Ex-D probe
16
8024252/AE00/V1-0/2020-07| SICKOPERATING INSTRUCTIONS | ZIRKOR200 Ex-D
Subject to change without notice
3 PRODUCT DESCRIPTION
3.3 Device versions
Fig. 4: Type code for Ex device versions
Z200EXD - 7 0 0 0 0 1 1 0 0 0 000A11
Designation of device type
Plausi=
Z200EXD DustEx Oxygen-Analyzer with probe
and electronics
Ex-Approval (EX)
7 ATEX/IECEx Zone 21 [2D]
Configuration (KF)
0 Probe and Electronics
1 Electronics only
2 Probe only
Probe length (required) (LZ)
0 Electronics only
Z Probe immersion depth 540 mm (incl.
integrated solenoid valve, LL-Option
and protection tube with special design
flange)
Y Probe immersion depth 960 mm (incl.
integrated solenoid valve, LL-Option
and protection tube with special design
flange)
Sensor technology (optional)
(ST)
0 No option / electronics only
2 With LongLife (LL) sensor protection for
LongLife^2 (LL^2) protective circuit
Probe version (optional) (SA)
0 Electronics only
1 Standard device
2 SIL2 probe
Reserve (RE)
1 Standard device
Reserve (RY)
1 Standard device
Housing (required) (GE)
0 Probe only
Y Safe area:
ST37 Housing with window, IP 66,IP
66, with failsafe heater control unit for
2D probes
X Safe area:
1.4571 Housing with window, IP 66,
with failsafe heater control unit for 2D
probes
W Safe area:
Glas fiber housing with window, IP 66,
with failsafe heater control unit for 2D
probes
V Safe area:
19" rack electronic for control cabinet,
4HE, IP 20, with failsafe heater control
unit for 2D probes
Power Supply (required) (SV)
0 Probe only
1 115 V ± 10 % 50 Hz
2 230 V ± 10 % 50 Hz
3 115 V ± 10 % 60 Hz
4 230 V ± 10 % 60 Hz
17
8024252/AE00/V1-0/2020-07| SICK OPERATING INSTRUCTIONS | ZIRKOR200 Ex-D
Subject to change without notice
PRODUCT DESCRIPTION 3
Z200EXD - 7 0 0 0 0 1 1 0 0 0 0 0 0 A 1 1
Interfaces (optional) (SS)
0 No Option / probe only
1 RS232 with interface protocol
2 RS485 with interface protocol
3 MODBUS RTU over RS232
4 MODBUS RTU over RS485
5 Hart
6 Fieldbus
Reference Air (optional) (DR)
0 No Option / probe only
1 *I-Airversion*
Integrated flow meter control for
reference air and test gas, enabled
fully- and semi-automatic calibration of
the probe.
Required:
-Instrument air supply (customer)
2 *Pump version*
Integrated flow meter control for
reference air and test gas, enabled
fully- and semi-automatic calibration of
the probe.
Expanded System Option
(optional) (EO)
0 No Option / probe only
1 LongLife^2 (LL^2) sensor protection:
Current mode for reducing gas
atmospheres. Required: Probe with
LongLife-sensor for LongLife^2
Application extension (optional)
(AE)
0 No Option / probe only
1 Additional module for pressure
compensation
Electronics version (optional)
(EA)
A Probe only
B Standard device
C SIL2 electronic
Reserve (RV)
1 Standard device
Reserve (RR)
1 Standard device
Device type (GY)
Standard device
S Special device
M Sample device
Documentation for the special
and sample model (SG)
01
99
18
8024252/AE00/V1-0/2020-07| SICKOPERATING INSTRUCTIONS | ZIRKOR200 Ex-D
Subject to change without notice
3 PRODUCT DESCRIPTION
3.4 Device variants
Probe, standard
3.5 Options
Probe: Solenoid valve for test gas feeding
Counterflange:
Various interfaces (RS232, RS485, HART, Fieldbus)
3.6 Expansion module
The electronics are also available optionally with different interfaces (RS232, RS485,
HART, Fieldbus). If you have ordered one of these options, a separate Manual/Specification
for this interface is included with the delivery.
19
8024252/AE00/V1-0/2020-07| SICK OPERATING INSTRUCTIONS | ZIRKOR200 Ex-D
Subject to change without notice
LAYOUT AND FUNCTION 4
4 Layout and function
4.1 System overview
Fig. 5: System overview
1 Control unit / IP66 A Safe zone
Max. ambient temperature: -20 °C to +55 °C (+50 °C for pump version)
(-4 °F to +131 °F (+122 °F for pump version))
2 In-situ measuring probe / IP6X B Hazardous area, zone IIIC T133 °C/T141 °C
Max. ambient temperature:-20 °C to +70 °C (-4 °F to +158 °F)
3 Pneumatic cable C Hazardous area (DustEx), flue gas duct / combustion chamber
4O
2
probe signal cable D Flue gas direction - max. flue gas temperature: 600 °C (1112 °F)
5 Counterflange (optional)
6 Insulation: Customer*
7Duct wall
8 Probe junction unit a Power supply, power voltage 230 V/115 V ±10% 50 to 60 Hz, max. 400 VA
b Test gas, inlet
c Output signals (analog and digital)
d Instrument air, inlet
* Note: Risk of corrosion due to lack of insulation
-1° ... -
a b c d
A
B
C
D
1
3
4
5
2
6
7
8
20
8024252/AE00/V1-0/2020-07| SICKOPERATING INSTRUCTIONS | ZIRKOR200 Ex-D
Subject to change without notice
4 LAYOUT AND FUNCTION
4.2 Measuring principle
The ZIRKOR200 Ex-D oxygen measuring system consists of an in-situ probe, which is
installed on the duct with the gas to be analyzed, and a control unit for voltage and gas
supply as well as for signal processing. An oxygen sensor regulated to 800 °C which
functions according to the zirconium oxide principle is fitted on the probe tip. An mV signal
is measured between the reference gas side of the sensor (inside, instrument air 20.95%
O
2
) and the measuring gas side, which depends logarithmically on the relation of the
oxygen partial pressures of both sides. The mV signal is converted to the oxygen partial
pressure in the sample gas using the Nernst equation, which determines the O
2
concentration in the sample gas.
4.3 Scope of application
The ZIRKOR200 Ex-D DustEx oxygen measuring system is suitable for measuring oxygen
(O
2
) in flue gases.
4.4 Potential hazards
4.5 Process interruption
The oxygen measuring system must remain in operation even when the process is
interrupted or the system is temporarily switched off (e.g. at night or on weekends).
Frequent cooling and heating of the measuring probe leads to a thermal load on hot
measuring probe components (heater, thermoelement and O
2
sensor) and shortens their
service life. SICK shall not be liable for any resulting damage.
NOTICE:
The system must not be used to determine the oxygen content of combustible gases
because the measuring accuracy cannot be guaranteed when combustible gases occur.
NOTICE:
Device damage with regular O
2
concentrations below 0.5%. Under normal process
conditions, the minimum concentration of O
2
in the flue gas should not be less than
0.5%.
The cell protection circuit option is recommended when the O
2
concentration
regularly falls below 0.5%. Called LL
2
(LongLife
2
), so that the O
2
sensor is
permanently protected.
WARNING: Danger of burns on hot components which are in the process gas
The temperature of the probe filter head and all parts in the process gas is 150 °C to
800 °C (302 °F to 1472 °F) during operation. Directly touching hot parts for
disassembly or maintenance causes serious burns!
Use heat protection gloves when removing the probe.
Switch the electronics supply voltage off before removing the probe.
Place the probe in a safe, protected area after removal and wait until the probe
temperature has cooled down to the ambient temperature.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88

SICK ZIRKOR200 Ex-D Operating instructions

Category
Measuring, testing & control
Type
Operating instructions

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI