Troubleshooting
2-7
played to indicate that the tests are being run in
the manual single looping mode. If a failure oc-
curs, a star (*) appears at the right hand end of
the display and remains on for the remainder
of the tests in the series. This testing process au-
tomatically stops after the last test in the series
is completed. This test process can also be
stopped by pressing EXIT. When EXIT is
pressed, any test in process will be allowed to
Þnish before aborting the testing process.
B. CONTINUOUS looping continuously repeats
all the tests in the speciÞed series until the test-
ing process is manually stopped. During test-
ing, the ÒMCÓ message is displayed to indicate
that tests are being run in the manual continu-
ous looping mode. If a failure occurs, a star (*)
appears at the right hand end of the display
and remains on for the remainder of the tests in
the series. This test process can be stopped by
pressing EXIT. When EXIT is pressed, any test
in process will be allowed to Þnish before
aborting the testing process.
C. STEP looping is used to perform one test at a
time. Each press of the ENTER key performs
the displayed test. The ÒMSÓ message is dis-
played to indicate that tests are being run in the
manual step looping mode. If a failure occurs,
a star (*) appears at the right hand end of the
display and remains on for the remainder of
the tests in the series. The instrument automat-
ically aborts the testing process after the last
test in the series is run. If you do not wish to
run all the tests in the series, simply press EXIT
after the desired test is run.
8. After the testing process is stopped, the following
message is displayed:
All tests complete *
Press ENTER to review or EXIT
The star (*) is only displayed if a failure occurs.
9. In the event of no test failures, press any key to re-
turn to the BUILT-IN TEST menu. If you wish to
run more tests, repeat steps 4 through 8.
In the event of a failure, press ENTER to display
the Þrst test that failed. Other test failures can be
displayed by using the and keys. The INFO
key can be used to provide a brief summary of each
displayed test failure. Paragraph 2.10.3 provides
detailed documentation for troubleshooting the
defective circuit. When Þnished, press EXIT to re-
turn to the BUILT-IN TEST menu. If you wish to
run more tests, repeat steps 4 through 8.
10. When Þnished with BUILT-IN TEST, use the EXIT
key to back out of the menu structure.
2.6 Diagnostics
The Model 2001 has diagnostic test modes which allow
you to ÒfreezeÓ instrument operation to allow you to
check logic levels on the DC_STB control registers
(U303, U300, U800 and U801). The known bit pattern at
these registers can then be used for signal tracing
through the unit. Table 2-10 provides a brief descrip-
tion of each register bit.
Perform the following steps to use DIAGNOSTICS:
1. Select the desired function and range to be
checked. Note that there are no range selections for
FREQ and TEMP.
2. Display the MAIN MENU by pressing the MENU
key.
3. Using the or key to place the cursor on TEST
and press ENTER to display the SELF-TEST
MENU.
4. Place the cursor on DIAGNOSTICS and press EN-
TER. The Þrst diagnostic test mode (Signal Phase
or Ohms Sense High) is selected (displayed).
5. Perform the following steps to determine the bit
pattern at the control registers:
A. Refer to one of the following DIAGNOSTIC
Test Modes tables to determine the bit pattern
designator (A through X) for the selected func-
tion/range:
Table 2-2 Ñ All functions except
Ω
4
Table 2-3 Ñ
Ω
4 function; 20
Ω
and 200
Ω
ranges
Table 2-4 Ñ
Ω
4 function, 2k
Ω
, 20k
Ω
and 200k
Ω
ranges
B. Once the bit pattern designator is determined,
use Table 2-5 to determine the logic state of
each register bit.
Example: Assume the 20VDC range is selected
and the instrument is in the ÒSignal PhaseÓ of
DIAGNOSTICS. From Table 2-2, the bit pattern
designator is C. Table 2-5 provides the logic
states for bit pattern C.
6. Use the cursor keys to select the other diagnostic
test modes. The key scrolls forward through the