Endres+Hauser Liqui-box a 2 / liqui 2 Portable water sampler Operating instructions

Type
Operating instructions
liqui-box a 2 / liqui-compact a 2
Portable water sampler
Installation and operating instructions
BA 019R/09/e/10.95
Part No. 500 47779
Nothing beats know-how
Endress Hauser
LIQUI - BOX A 2 / LIQUI - COMPACT A 2
Portable water sampler
Installation and operating instructions
Please fill in and always give these details when reordering or on any queries:
Unit number ................................................................................
Date ................................................................................
Operator ................................................................................
Wetzer Meßtechnik Printed in W. Germany
Modifications reserved
liqui-box a 2 / liqui-compact a 2
Complete delivery Check: That the delivery note and delivery contents correspond!
Delivery should include:
1 x liqui-box a 2 or liqui-compact a 2
1 x Mains cable or plug with 12 VDC version
1 x Signal plug
1 x Sample hose connection, elbow including
O-ring
2 x Wall fixing brackets (accessories) *
1 x Angle bracket (accessories) *
6 x Philips head screws DIN 7985 M4 x 8 V2A
(accessories) *
1 x Jubilee clip
2 x Short conductivity electrodes (accessories)
1 x Setting up instructions
Further accessories dependent on delivery note.
* = Not included in delivery of "liqui-compact a 2"
Transport insurance Check the package and contents for external damage.
Should there be any visible damage you should immediately inform both the
transport agency as well as your supplier. If this is not done any later claims
cannot be handled under the guarantee.
Please take note of the following characters:
Hint: Hints for better installation.
Attention: Ignoring this note can lead to damage of the
device or faulty operation.
Danger: Ignoring this warning can lead to personal injury.
Should the "liqui-box a 2" be in storage for more than 6 months please take
note of the storage details in chapter
"Maintenance, general".
liqui-box a 2/liqui-compact a 2 Please note
Construction and operating elements
Operating elements and display
2
25
24
10 9 8
16
18
17
76543
23
22 21 20 19 11
12
26
13
1
15
14
liqui-box a 2/liqui-compact a 2 Construction and operating elements
"liqui-box a 2" construction A Carrying handle
B Display and operating components
C Power supply socket
D Elbow connector (for suction hose)
E Signal in and output socket
F Hose clamp
GDosing system
K Housing
L Lead sealing opening
MLegend plate (unit number, supply voltage, power consumption
of functional module)
Operating elements and
display 1 MAN-push button: Immediate sample start
2 AUT-push button: Automatic sample sequence start
3 ON-push button: Switches unit on
4 OFF-push button: Switches unit off
5 push button: Operation mode selection
6 - push button: Reduce target value
7 + push button: Increase target value
8 INFO- display field: Information (pump running times etc.)
9 TEMP- display field: Cabinet temperature display
(only on heater option)
10 IMP- display field: Quantity proportional sampling
11 TIME- display field: Time proportional sampling
12 PRESET Sample and temperature target values
13 SAMPLE Sample selection (heading)
14 Sampling display fields (total)
15 STATUS Sample and temperature actual values
16 push button Operation mode selection
17 + push button Increase target value
18 - push button Reduce target value
19 INFO- display field: Information (samples not taken)
20 END- display field: Sample sequence end or continuous
operation
21 SUM- display field: Number of samples per container
22 TIME-display field: Fill time per container
23 STATUS Actual value for sample sequence end
24 PRESET Target value for sample sequence end
25 DISTRIBUTION Sample distribution (heading)
26 Sample sequence end and distribution display fields (total)
liqui-box a 2/liqui-compact a 2 Construction and operating elements
Page
Please note:
- If the delivery is complete
- Transport insurance
liqui-box a 2
- Construction, - dimensional drawing
General notes 2
- Safety 2
Mechanical installation 3
- General 3
- Wall mounting 3
- Drilling template 3
- Stand mounting (option) 4
- Suspension system (option) 4
- Hose connection/installation 5
- Fixing sample outlet hose 5
Electrical installation 6
- Power supply 6
- Mains AC version 6
- DC Version 6
- Option “Battery pack/charger” 6
- In / outputs 7
- Signal plug connections 7
- Outputs 7
- Inputs 7
- Connection examples 8
- Signal input 8
- Alarm output 8
- What happens on power up ? 8
Operating components and
setting up principle 9
Page
What can the “liqui-box a 2do ? 10
Sampling principle 10
Setting up sample volume 11
Methods of sampling 15
- Manual sampling 15
- Time proportional 15
- Setting up sample sequence end 16
- Example 17
- Quantity proportional 18
- Example 19
- Event controlled sampling 19
- Countdown for sample
sequence start 20
- Changes during operation 20
- Heater 20
Additional information 21
- Information that can be called up 21
- Faults 21
- Number of not taken samples 21
Maintenance 22
- General 22
- Repairs 22
Problems and solutions 23
- Fault messages 23
- Spare parts 25
Technical data 26
Appendix: LIQUI-COMPACT A 2
liqui-box a 2/liqui-compact a 2 Contents
1
Safety This unit is constructed and tested according to EN 61010-1 /
VDE 0411 Part 1 and left our works in perfect and safe condition. In order to
maintain this condition and operate safely the user must take note of the
following safety information and warnings contained in these instructions.
First check that the power supply to be used corresponds with that on the
unit legend plate.
This unit is supplied with a loose power cable including plug and socket
arrangement and is therefore classified to protection class I.
The power supply plug must only be connected to a socket with an earth
protection connection. This protection must be continued when using
extension leads. Any breakage of the earth conductor within or outside the
unit or loosening the earth connections can make the unit potentially
dangerous. Intentional disconnection or an open circuit of this earth
connection is not permissible.
There are no components in the unit that can be repaired by the user. All
repairs must be made by trained service personnel.
Removing covers or components, except where this can be done by hand,
must only be carried out by skilled personnel.
If it is assumed that the unit cannot be safely operated it must be
immediately taken out of operation and secured against unintentional use.
It can be assumed that the unit cannot be safely operated,
- if the unit is visibly damaged
- if the unit no longer operates
- if the unit has been in storage under adverse conditions for a longer
period of time
- after long transport under adverse conditions.
The manufacturer does not accept liability for any damage that has been
made due to the unit not having been used in accordance with these safety
instructions.
liqui-box a 2/liqui-compact a 2 General notes
2
GeneralThe ”liqui-box a 2” should be installed in an area where there is no
additional heating from external sources (eg. radiators etc.). Protect the unit
from direct sun rays. Do not install the "liqui-box a 2" close to producers of
large magnetic fields (eg. motors, transformers etc.). Do not install the
"liqui-box a 2" in areas where it can be subject to high mechanical vibration.
When transporting the "liqui-box a 2" please avoid heavy shocks.
The suction hose must be laid in such a way that it always rises from the
sampling point to the "liqui-box a 2"! Syphons must be avoided between
the sampling point and sampler! The "liqui-box a 2" must not be
connected to a pressurised system! For sample less than 2 m we
recommend using a 15mm hose. Both 13mm and 15mm connections are
delivered. The liquid conductivity must not be less than 30 microsiemens!
Maximum height difference: 6m between sampling point and
"liqui-box a 2".
Maximum sampling hose length: 30m
Always mount the sampler upright, do not tilt !
In order to connect the suction hose to the sampler use the jubilee clip
supplied.
The suction hose must be a spiral reinforced type
Wall mounting
(not for liqui-compact a 2)
Fix the wall mounting brackets and angle bracket to the rear wall of the
unit using the screws supplied.
Drilling template
(hole dia. 7.5mm)
liqui-box a 2/liqui-compact a 2 Mechanical installation
3
Stand mounting (Option)
All weather roof (accessories)
"liqui-box a 2"
Battery pack (Option with "liqui-box a 2" 12 VDC version)
Mounting bracket
Sample outlet hose
Composite container (accessories)
Legs (3)
Construct stand as shown in diagram. Use wing nuts supplied to secure
components.
If required connect battery pack with "liqui-box a 2".
Suspension unit with
composite container (Option)
Suspension system Mounting bracket Battery pack (12 VDC option)
Liqui-box a 2 14 l container.
liqui-box a 2/liqui-compact a 22 Mechanical installation
4
Fix suction hose
= Nut
= Suction hose connector elbow
= Suction hose
= Jubilee clip
= O-Ring
= Nipple
What to do:Loosen
Pull off
Plug into
Push approx 3 cm overand secure with
Check that is correctly positioned inside
Tighten on
Note
The spiral strengthened suction hose is normally supplied by the user
(maximum length 30 metres).
It is not within the normal delivery conditions.
liqui-box a 2/liqui-compact a 2 Mechanical installation
5
Power supply Plug in the mains plug into socket C and tighten using nut.
Mains version (AC)
Plug connections:Numbers are engraved on the plug
1= Black(L), 2 = free, 3 = Blue(N) = Green/yellow (PE)
Direct current version (12 VDC) After having connected a 12 VDC power supply (see connections). Connect
the 4 pole plug to the socket C and tighten using nut.
Connections:
1=free
2 = Plus pin
3=free
= Minus pin
Always maintain the batteries used (i.e. regularly charged, recharge
straightaway after long term use). Please take note of safety notes in
separate instructions ! Disconnect "liqui-box a 2" from battery pack when
storing. Store battery pack in a cool area and recharge at least every 6
months.
Option "Battery pack / charger"
"liqui-box a 2" 12 VDC: Follow connection diagram on the battery pack housing. Make connections !
liqui-box a 2/liqui-compact a 2 Electrical installation
6
In/outputs
Signal socket connections
(Signal socket E)
White = 1 = Auxiliary voltage (-)
0V (common)
Brown = 2 = Auxiliary voltage (+)
8 to 19 V output
Green = 3 = Flow impulse input
Yellow = 4 = External stop input
Grey = 5 = Do not connect
Pink = 6 = Alarm output
Blue = 7 = Sample sequence
end output
Red = 8 = do not connect
Cable type LiYY8 core
(approx.1,5 m long)
Single cores x 0,23 / 0,25
Outputs2 outputs for alarm and sequence end are available from the controller.
These are in the form of an open-collector output and are not switched
when active (in alarm) or during a power failure.
Sequence endSequence end is signalled:
When the preset sampling sequence has finished.
On power failure.
Sequence end is de-activated once a new sampling
sequence programme is started.
AlarmThe output signals alarm, fleeting contact 30sec. :
If a fault occurs during sampling.
The respective error message appears on the display.
See chapter “Problems and solutions”.
Inputs
Flow1 Impulse input (opto-coupler) max. 25Hz. For connection to an external
quantity measurement).
External stop1 Stop input (opto-coupler). A voltage of between +7 to +27 Volts on the
input stops all "liqui-box a 2" functions.
0 to +3 Volts initiates restart of the unit (The status display restarts at
zero and the next sample is taken once the target value is reached). For
further description of this function see chapter "Interrupt / restart
sample cycle".
150 cm.
liqui-box a 2/liqui-compact a 2 Electrical installation
7
Connection examples Alternatives: In example "Flow input".
Input
Output Alternatives: In example "Alarm output"
What happens on power up ? a) The display shows "INIT", unit does a self-check (start up). Display
shows "OFF", start “liqui-box a 2 by operating push button "ON".
b) Sort power failures (up to approx. 500 hours). The "liqui-box a 2"
continues to operate normally. Longer loss of power leads to resetting the
unit as described under "a)".
c) Only when using "liqui-box a 2" in 12 VDC version ! When connection
operational voltage (12 VDC power source, accumulator or battery) to the
unit the display shows "EOA Akku". This must be acknowleged by
operating the "OFF" followed by the "ON" push button.
+
+
Imax=50mA
Imax=50mA b
a
a: Using an external voltage source b: Using the internal auxiliary voltage
b
a
a: Using an external voltage source b: Using the internal auxiliary voltage
liqui-box a 2/liqui-compact a 2 Electrical installation
8
Operating elementsSee drawings at the front of these instructions
("Operating components and display")
Operating principle
"ON" push button
Push button ”ON” (3) activates the
"liqui-box a 2"
controller
- The display shows the last operation
sequence values.
"OFF" push buttonPush button "OFF" (4) switches the
"liqui-box a 2"
off.
- The display (14) shows "OFF"
” push buttonPush buttons (5 and 16) change the mode of operation
- Display arrowhead changes to the next field.
- Display arrowhead flashes for a few seconds,
then the new settings are accepted (always wait
for the arrowhead to stop flashing).
"+" and "-" push buttonsPush buttons ”+” (7 and 17) and (6 and 8) change the target
value (12 and 24)
- Simultaneous operation of "+" and "" changes
the display to auf 0001 (Reset).
- If the target value is changed the arrowhead
flashes.
- Once the arrowhead is stable the new values are
accepted and stored.
"AUT" push buttonPush button ”AUT” (2) starts an automatic sampling sequence.
Count down"Count down" to sample sequence starts:
If the "AUT" (2) push button is operated for more than approx. 5 seconds a
preset counter appears in the display. Once this counter has been set the
arrowhead flashes.
Once the arrowhead is stable the counter counts down in a minute cycle.
The sampler is now blocked. Once 0000 is reached the sampler starts
automatic sample operation using the preset sampling data.
"MAN" push buttonPush button "MAN" (1) initiates an immediate sample cycle.
liqui-box a 2/liqui-compact a 2 Operating elements and principle
9
Sampling principle
Vacuum principle
4 The hose clamp is released
and the sample flows into
the composite container or
bottles if operating using
sample distribution.
3 The sample is now dosed to
the preset volume (VP). This
is dependent on the dosing
tube position (D). Excess
liquid flows back to the
sampling point due to a
syphonic effect.
1 The dosing system is
pneumatically sealed at the
beginning of each sample
cycle. The diaphragm pump
then blows the suction hose
free of obstructions via the
dosing chamber
2 A fresh sample is then
sucked into the dosing
chamber until the
conductivity level switch is
activated (sensors in the
dosing chamber flange)
liqui-box a 2/liqui-compact a 2 Sampling principle
10
Varying sample volumeFollow the next steps.
1. Open controller doorLift the catch on the left side of the
unit forwards
2. Switch unit offOperate the - OFF- (4) push button
at the "liqui-box a 2".
Dosing system:
= Elbow = Dosing tube
= Dosing chamber = Piping clamp
3. Remove air hose
1
2
3
4
liqui-box a 2/liqui-compact a 2 What can the liqui-box a 2 do ?
11
4. Remove dosing system Lift clamp from pipe elbow.
Pull dosing system forwards and
remove from controller.
5. Remove flange Release the bayonet fitting by turning
the lower part of the flange as shown.
Remove the flange from the dosing
chamber.
6. Set sample volume Set the dosing tube to the required
sample volume by pushing it in or out.
Take note of the engraved quantities
on the tube. (The further the tube is
pulled out of its retainer the smaller
the sample volume)
(Only move the dosing tube. NEVER
loosen the nut and NEVER move the
upper elbow.)
Use the following calculation in order to avoid overfilling the sample
containers:
Preset sample volume x number of samples = Container volume
300
350
400
450
500
liqui-box a 2/liqui-compact a 2 What can the liqui-box a 2 do ?
12
7. Dosing system
= Nipple
= Suction pipe elbow
= Hose clamp
= Contacts
= Spring contacts
= Fixing clamp
= Silicon hose
Push silicon hose into the hose clamp .
Push suction pipe elbow onto the nipple . (Make sure that the
spring contacts and flange contacts are made).
Push fixing bracket downwards.
"The contacts and contact springs must be made
(otherwise faults can occur).
We cannot be held responsible for damage caused by not complying with
these instructions !
8. Replace air hose
liqui-box a 2/liqui-compact a 2 What can the liqui-box a 2 do ?
13
Setting up example
Time proportional sampling. Requirement:
Composite sample with sample cycle intervals of 10 min.
Switch unit on by operating the ”ON” push button.
Using the push button select TIME”.
Set 0010 using the - and/or + push buttons.
Sample sequence end ? Without sample sequence end (continuous operation)
Using push button select ”END”.
SetOFF” using the - and/or + push buttons.
Start the sequence by operating the AUT” push button.
Or: Sample seuquence end after
24hours
Using push button select ”END”.
SetON” using the - and/or + push buttons.
Using push button select ”TIME”. Set ”1440min” using the - and/or +
push buttons. Start the sequence by operating the ”AUT” push button.
Or: Sample sequence end after
150 samples
Using push button select ”END”.
SetON” using the - and/or + push buttons.
Using push button select ”SUM”. Set ”0150min” using the - and/or +
push buttons. Start the sequence by operating the ”AUT” push button.
0010
0000
ON
0150
0000
OFF
1440min
0000min
liqui-box a 2/liqui-compact a 2 What can the liqui-box a 2 do ?
14
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34