Number Description
35070 Pierce Fast Western Blot Kit, SuperSignal West Dura Substrate, Mouse, contains sufficient
reagents to perform 10 blots (8 × 10cm) probed with a mouse primary antibody
Kit Contents:
Fast Western Antibody Diluent, 200mL
Fast Western 10X Wash Buffer, 100mL
Fast Western Mouse Optimized HRP Reagent, Dura, 10mL
SuperSignal West Dura Luminol/Enhancer Solution, 50mL
SuperSignal West Stable Peroxide Solution, 50mL
35071 Pierce Fast Western Blot Kit, SuperSignal West Dura Substrate, Rabbit, contains sufficient
reagents to perform 10 Western blots (8 × 10cm) probed with a rabbit primary antibody
Kit Contents:
Fast Western Antibody Diluent, 200mL
Fast Western 10X Wash Buffer, 100mL
Fast Western Rabbit Optimized HRP Reagent, Dura, 10mL
SuperSignal West Dura Luminol/Enhancer Solution, 50mL
SuperSignal West Stable Peroxide Solution, 50mL
35075 Pierce Fast Western Blot Kit, SuperSignal West Dura Substrate, Mouse Trial Size, contains
sufficient reagents to perform two blots (8 × 10cm) probed with a mouse primary antibody
Kit Contents:
Fast Western Antibody Diluent, 40mL
Fast Western 10X Wash Buffer, 20mL
Fast Western Mouse Optimized HRP Reagent, Dura, 2mL
SuperSignal West Dura Luminol/Enhancer Solution, 10mL
SuperSignal West Stable Peroxide Solution, 10mL
35076 Pierce Fast Western Blot Kit, SuperSignal West Dura Substrate, Rabbit Trial Size, contains
sufficient reagents to perform two blots (8 × 10cm) probed with a rabbit primary antibody
Kit Contents:
Fast Western Antibody Diluent, 40mL
Fast Western 10X Wash Buffer, 20mL
Fast Western Rabbit Optimized HRP Reagent, Dura, 2mL
SuperSignal West Dura Luminol/Enhancer Solution, 10mL
SuperSignal West Stable Peroxide Solution, 10mL
Storage: Upon receipt store at 4°C. Product shipped at ambient temperature.
2173.1
35070 35071 35075 35076
Pierce® Fast Western Blot Kit,
SuperSignal® West Dura Substrate
Pierce Biotechnology
PO Box 117
(815) 968-0747
www.thermoscientific.com/pierce
3747 N. Meridian Road
Rockford, lL 61105 USA
(815) 968-7316 fax
Introduction
The Thermo Scientific Pierce Fast Western Blot Kit, SuperSignal West Dura Substrate contains optimized reagents that
shorten the time to perform a typical Western blot from 4 hours to ~60 minutes. The kit provides all the reagents necessary to
complete a Western blot that was probed with a mouse or rabbit primary antibody. The protocol requires minimal hands-on
time and yields results comparable to classic Western blotting. This kit includes SuperSignal West Dura Chemiluminescent
Substrate, a highly sensitive HRP substrate that is detected using photographic or other imaging methods. This substrate
produces a signal that is stable for several hours, enabling repeated film exposure to obtain optimal results.
Important Product Information
The Pierce Fast Western Blot Kit reagents are optimized to function together. Use the primary antibody at the
concentration typically used in Western blotting procedures with SuperSignal West Dura Substrate (i.e., 0.02-1.0µg/mL).
The Pierce Fast Western Blot Kit is optimized for blots that are not pre-blocked. Pre-blocking the membrane can cause a
decrease in assay sensitivity.
Shake the Fast Western Antibody Diluent well before use. The antibody diluent is a saturated solution and settling may occur.
Use a clean incubation tray for each step of the blotting procedure. Trays do not need to be changed between washes, but
it is critical to use a clean or new tray when beginning the wash steps.
For optimal results, use a shaking platform during incubation steps.
Do not handle membrane with ungloved hands. Always wear gloves or use clean forceps to handle the blot.
The stability of primary antibodies diluted in the Fast Western Antibody Diluent varies. For best results, prepare the
antibody working dilution immediately before use.
All equipment must be clean and free of foreign material. Metallic devices (e.g., scissors) must have no visible signs of
rust. Rust causes speckling and high background.
The Substrate Working Solution is stable for up to 8 hours at room temperature.
We offer a variety of protein transfer membranes, primary antibodies and X-ray film. Please consult our web site or
catalog for product and ordering information.
Additional Materials Required
Membrane with transferred protein
Primary Antibody: Mouse or rabbit antibody that is specific to the target protein(s)
X-ray film, film cassette, developing and fixing reagents for film processing, or an imaging instrument such as a CCD
camera
Rotary platform shaker
Pierce Biotechnology
PO Box 117
(815) 968-0747
www.thermoscientific.com/pierce
3747 N. Meridian Road
Rockford, lL 61105 USA
(815) 968-7316 fax
Material Preparation
1X Wash Buffer Mix 1 part of the Fast Western 10X Wash Buffer with 9 parts of water.
Example: Mix 10mL of 10X Fast Western Wash Buffer with 90mL of water. Prepare at least 60mL for
each 8 × 10cm blot.
Primary Antibody
Solution Shake the Fast Western Antibody Diluent well before use. Dilute the primary antibody
(0.02-1.0µg/mL) with Antibody Diluent. Use ~0.125mL of antibody per cm2 of membrane (e.g., 10mL
per 8 × 10cm blot). The stability of diluted primary antibodies varies depending on the antibody. For
best results, prepare working dilution just before use.
Example: To prepare 1µg/mL from a stock concentration of 1mg/mL, mix 10µL of the primary
antibody with 10mL of the Fast Western Antibody Diluent.
Optimized HRP
Reagent Working
Dilution
Mix 1 part of Optimized HRP Reagent with 9 parts of Antibody Diluent. Use 0.125mL per cm2 of
membrane (e.g., 10mL per 8 × 10cm blot). For best results, use this solution within 1 hour.
Example: Mix 1mL of the Optimized HRP Reagent with 9mL of the Antibody Diluent.
Note: If using the SNAP i.d.™ System, use 0.2-1mL of Optimized HRP Reagent and adjust to the
appropriate volume with the Antibody Diluent. Reagent volume and incubation time might require
optimization.
SuperSignal West
Dura Working
Solution
Mix SuperSignal West Dura Luminol/Enhancer Solution and SuperSignal West Stable Peroxide
Solution at 1:1. Use 0.125mL of Working Solution per cm2 of membrane (e.g., 10mL per 8 × 10cm
blot). For best results, prepare Working Solution just before use (Step 8). The Working Solution is
stable for up to 8 hours at room temperature.
Example: Mix 5mL of Luminol/Enhancer Solution with 5mL of Stable Peroxide Solution.
Fast Western Blotting Procedure
1. Remove blot from the transfer apparatus and place in a clean incubation tray.
2. Briefly wash blot in 1X Wash Buffer to remove transfer buffer.
3. Add the Primary Antibody Solution to the blot and incubate for 30 minutes at room temperature (RT) with shaking.
Note: Primary antibody incubation time may be reduced to 10 minutes at RT or increased to an overnight incubation at
C. Evaluate each specific antibody/antigen to determine compatibility with incubation time.
4. Discard the primary antibody solution from the tray, or place blot in a new incubation tray.
5. Add the Optimized HRP Reagent Working Dilution and incubate for 10 minutes at RT with shaking. Higher sensitivity
can be obtained by increasing the incubation time to 15 minutes; however, one or more additional washes will be
required to reduce background.
6. Remove blot from the HRP solution and place it in a clean incubation tray.
Note: To reuse the tray, remove the HRP solution and completely fill the tray containing the blot with ultrapure water
and decant. Repeat this wash twice.
7. Wash membrane by suspending it in approximately 20mL of Fast Western 1X Wash Buffer and agitating for 5 minutes.
Repeat this wash twice.
8. Remove blot and place it in a clean incubation tray. Add the SuperSignal West Dura Working Solution and incubate for
5 minutes at RT.
9. Remove blot from SuperSignal West Pico Working Solution and place it in a plastic sheet protector or clear plastic wrap.
Use an absorbent tissue to remove excess liquid and to carefully press out any bubbles from between the blot and the
membrane protector.
10. Expose the blot to film or use your preferred imaging method.
Pierce Biotechnology
PO Box 117
(815) 968-0747
www.thermoscientific.com/pierce
3747 N. Meridian Road
Rockford, lL 61105 USA
(815) 968-7316 fax
Troubleshooting
Problem
Possible Cause
Solution
High background Incubation tray is
contaminated with HRP Use a clean incubation tray after every step of the procedure
If reusing the tray, rinse tray and blot with ultrapure water three
times before washing with Fast Western 1X Wash Buffer
Used too much primary
antibody Reduce primary antibody concentration to 0.02-1µg/mL
Insufficient washing Use a minimum of 20mL of 1X Wash Buffer for each wash
Use a clean incubation tray to begin the wash steps
Add an additional wash cycle for a total of four 5 minute washes
Overexposed film Decrease exposure time or use Thermo Scientific Pierce
Background Eliminator (Product No. 21065)
Omitted the brief pre-wash
Wash membrane in 1X Wash Buffer briefly before beginning the
protocol
Weak signal Antigen or primary antibody
amounts were not optimal Strip and re-probe blot using different primary antibody
concentration use the primary antibody at 0.02-1µg/mL
Optimize the amount of sample applied to the gel
Inefficient protein transfer Optimize transfer conditions
Spots within the
protein bands Inefficient protein transfer Optimize transfer conditions
Unevenly hydrated
membrane Hydrate membrane according to manufacturer’s instructions
Bubble between x-ray film
and membrane Remove all bubbles before exposing blot to film
Speckling Over-heating during
electrophoresis or transfer Control temperature during electrophoresis and transfer
Additional Information
Visit our web site for additional information relating to this product including the following:
Tech Tip #67: Chemiluminescent Western blotting technical guide and protocols
Tech Tip #23: Strip and reprobe Western blots
Tech Tip #24: Optimize antigen and antibody concentrations for Western blots
Tech Tip #32: Guide to enzyme substrates for Western blotting
Tech Tip #43: Protein stability and storage
Western Blotting Handbook and Troubleshooting Guide
Related Thermo Scientific Products
37071 SuperSignal West Dura Chemiluminescent Substrate, Trial Kit
34075 SuperSignal West Dura Chemiluminescent Substrate, 100mL
34076 SuperSignal West Dura Chemiluminescent Substrate, 200mL
34089 CL-XPosure™ Film, 7 × 9.5in (18 × 24cm) sheets, 100/pkg
34090 CL-XPosure Film, 5 × 7in (13 × 18cm) sheets, 100/pkg
34091 CL-XPosure Film, 8 × 10in (20 × 25cm) sheets, 100/pkg
34097 CL-XPosure Film, 9.5 × 11.8in (24 × 30cm) sheets, 100/pkg
34099 CL-XPosure Film, 14 × 17in (35 × 40cm) sheets, 100/pkg
Pierce Biotechnology
PO Box 117
(815) 968-0747
www.thermoscientific.com/pierce
3747 N. Meridian Road
Rockford, lL 61105 USA
(815) 968-7316 fax
21059 Restore Western Blot Stripping Buffer, 500mL
46430 Restore PLUS Western Blot Stripping Buffer, 500mL
21065 Pierce Background Eliminator Kit, for eliminating background from X-ray film
88018 Nitrocellulose Membrane, 0.45µm, 33cm × 3m, 1 roll
77010 Nitrocellulose Membrane, 0.45µm, 8 × 12cm, 25/pkg
88025 Nitrocellulose Membrane, 0.45µm, 8 × 8cm, 15/pkg
88585 PVDF Transfer Membrane, 0.45µm, 10 × 10cm, 10/pkg
88518 PVDF Transfer Membrane, 0.45µm, 26.5cm × 3.75m, 1 roll
88605 Western Blotting Filter Paper, Extra Thick, 7cm × 8.4cm, 50 sheets
88610 Western Blotting Filter Paper, Extra Thick, 8.5cm × 9cm, 50 sheets
88615 Western Blotting Filter Paper, Extra Thick, 8cm × 13.5cm, 50 sheets
88620 Western Blotting Filter Paper, Extra Thick, 20cm × 20cm, 50 sheets
88600 Western Blotting Filter Paper, 10 × 10.5 cm, 100 sheets
35035 Pierce Fast Semi-Dry Transfer Buffer, 10X, methanol-free, 500mL
88217 Pierce Fast Semi-Dry Blotter
SNAP i.d. is a trademark of Millipore Corporation.
This product (“Product”) is warranted to operate or perform substantially in conformance with published Product specifications in effect at the time of sale,
as set forth in the Product documentation, specifications and/or accompanying package inserts (“Documentation”) and to be free from defects in material and
workmanship. Unless otherwise expressly authorized in writing, Products are supplied for research use only. No claim of suitability for use in applications
regulated by FDA is made. The warranty provided herein is valid only when used by properly trained individuals. Unless otherwise stated in the
Documentation, this warranty is limited to one year from date of shipment when the Product is subjected to normal, proper and intended usage. This
warranty does not extend to anyone other than the original purchaser of the Product (Buyer”).
No other warranties, express or implied, are granted, including without limitation, implied warranties of merchantability, fitness for any particular
purpose, or non infringement. Buyer’s exclusive remedy for non-conforming Products during the warranty period is limited to replacement of or
refund for the non-conforming Product(s).
There is no obligation to replace Products as the result of (i) accident, disaster or event of force majeure, (ii) misuse, fault or negligence of or by Buyer, (iii)
use of the Products in a manner for which they were not designed, or (iv) improper storage and handling of the Products.
Current product instructions are available at www.thermoscientific.com/pierce. For a faxed copy, call 800-874-3723 or contact your local distributor.
© 2011 Thermo Fisher Scientific Inc. All rights reserved. Unless otherwise indicated, all trademarks are property of Thermo Fisher Scientific Inc. and its
subsidiaries. Printed in the USA.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5

Thermo Fisher Scientific Pierce Fast Western Blot Kit SuperSignal West Dura Substrate User guide

Type
User guide

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI