Thermo Fisher Scientific HID Validation: The 3500 Series Genetic Analyzers Owner's manual

Type
Owner's manual
HID Validation:
The 3500 Series Genetic Analyzers
2
HID Validation:
The 3500 Series Genetic Analyzers
Outline
Scope
Materials and Reagents
Validation Elements
Selection of Validation Data
Additional Sources of Validation Information
User Bulletin
3
HID Validation:
The 3500 Series Genetic Analyzers
Validation Scope
Hardware and Reagents
Instruments
(3) 8 Capillary Instruments (3500)
(3) 24 Capillary Instruments (3500xL)
Size Standard
GeneScan™ 600 LIZ®Size Standard v2.0
GeneScan™ 500 ROXSize Standard
Profiler Plus® / Cofiler® do not use normalization
Polymer and Array
POP-4and 36 cm Capillary Array
4
HID Validation:
The 3500 Series Genetic Analyzers
Validation Scope
Sample Summary
Test Name Samples Input Replicates
Genotype
Concordance and
Reproducibility
40 male and 42 female gDNA samples
4 racial groups
Controls
1-2 ng total DNA 1 replicate per sample
2 injections
Sizing Precision
and Accuracy Allelic ladder 1 µL per well See below
Sensitivity and
Normalization 007 control DNA
3 gDNA samples STR: 0.125ng, 0.5ng,
1ng, 6ng 5 replicates for DNA
samples
3 replicates for NTC
4 injections
Minifiler: 0.125ng,
0.25ng, 0.5ng, .75ng
Mixture Analysis 4 pairs of male and female gDNA
Mixture ratios
1:0, 1:1, 1:3, 1:5, 1:9, and 0:1
Recommended
concentrations per kit 1 replicate per sample
Resolution Allelic Ladder
Research material 1 µL per well 24 replicates
5
HID Validation:
The 3500 Series Genetic Analyzers
Validation Scope
AmpFSTR®Kits
Identifiler®
Minifiler
SGM Plus®
Yfiler®
Sinofiler
SEfiler Plus
Profiler Plus®
Cofiler®
Identifiler®Direct
Identifiler®Plus
NGM
Precision and Accuracy
7
HID Validation:
The 3500 Series Genetic Analyzers
Precision and Accuracy
Example Configurations for Standard Deviation
Within Injection -3500
8 data points per mean
12 injections per plate
12 columns
3 Instruments
12 data points per mean
4 injections per plate
8 rows
3 Instruments
Across Injections 3500xL
8
HID Validation:
The 3500 Series Genetic Analyzers
3500 Sizing Precision Study:
Within an Injection of Identifiler®Allelic Ladder
9
HID Validation:
The 3500 Series Genetic Analyzers
3500 Sizing Precision Study:
Across Multiple Injections of Identifiler®Allelic Ladder Samples
96 Replicates of Identifiler®Allelic Ladder
3500 -12 Injections
3500xL -4 Injections
Introduction to Normalization
11
HID Validation:
The 3500 Series Genetic Analyzers
Sources of CE Variation (3100)
Numerous optical
components,
capillary array
Electrophoresis,
injection
Sample and
sample
preparation
Raman Intensity, 3100
file: INSTRUMENT DATA Mod2B.xls
y = -2.7202x + 6042.8
R
2
= 0.0059
0
1000
2000
3000
4000
5000
6000
7000
8000
9000
020 40 60 80 100 120 140
Instrument
A/D Units
Series1
Linear (Series1)
median= 6026.50
Stdev = 808.54
Average = 6054.11
Stdev/ average = 0.13
Almost 2-fold difference in signal from over 100 instruments from the factory floor
12
HID Validation:
The 3500 Series Genetic Analyzers
Normalization: What is it?
This feature normailzes
signal variations associated
with instrument, capillary
array, sample salt load, and
injection variability among
capillaries and instruments
Normalization can be
applied during primary
analysis of the data
Before normalization After normalization
13
HID Validation:
The 3500 Series Genetic Analyzers
Normalization Target (NT)
The expected average RFU for eleven peaks in the
GeneScan 600 LIZ®v2 size standard used for normalization
Normalization Factor
Normalization Target divided by the Average peak height of
the subset of peaks in the GeneScan 600 LIZ®v2 size
standard used for normalization
Normalization Factor Threshold
The passing range for Normalization Factor (default range is
0.3 to 3.0)
Normalization: Definitions
14
HID Validation:
The 3500 Series Genetic Analyzers
Sources of Variation by Normalization Method
No Norm Factory only Int Std only Factor+IntStd
Injection to Injection
Cap to Cap
Within
Instrument to Instrument
Total
39%
33%
25%
14%
31%
24%
24%
13%
20%
19%
12%
11%
0%
5%
10%
15%
20%
25%
30%
35%
40%
%CV
Levels of Normalization
Sources of Variation
Instrument to
Instrument largest
source of variation
Internal standard
Normalization
greatest reduction
in variation
Sensitivity and
GeneScan™ 600 LIZ®Size Standard v2.0
16
HID Validation:
The 3500 Series Genetic Analyzers
Sensitivity Study:
Average Peak Height across a Range of DNA Input Amounts AmpFℓSTR®
MiniFilerTM PCR Amplification Kit
Normalization Method
Factory Only
Factory + Internal Std
0 750 50 250 125
H1126
H1152
H1161 OO7
10000
5000
15000
00.125ng 0.25ng 0.50ng 0.75ng
10000
5000
15000
0
0.125ng 0.25ng 0.50ng 0.75ng
Genomic DNA #2
Control DNA Standard
007
Genomic DNA #1
Genomic DNA #3
Peak Height
17
HID Validation:
The 3500 Series Genetic Analyzers
Sensitivity Study:
Average Peak Height by Dye for 1ng DNA Input AmpFℓSTR
®Identifiler®
PCR Amplification Kit
2000
4000
8000
10000
6000
12000
14000
Internal Standard Normalization
2000
2000
Disabled Enabled
Peak Height
18
HID Validation:
The 3500 Series Genetic Analyzers
Size Standard Peak Height Consistency Study:
Five AmpFℓSTR®PCR Amplification Kits
GeneScan™ 600 LIZ® Size Standard v2.0
6000
5000
4000
3000
2000
1000
0
Lot 1 Lot 2 Lot 3
GS600 Peak Height
Identifiler®
Identifiler® Direct
Identifiler® Plus
MinifilerTM
NGMTM
7000
GS600 Lot Number
Peak Height
19
HID Validation:
The 3500 Series Genetic Analyzers
Size Standard Peak Height Consistency Study:
%CV for 5 Different AmpFSTR®Kits
GeneScan™ 600 LIZ® Size Standard v2.0
% CV
Size Standard Peak Height Consistency · 6 instruments · 3 lots of size standard
20
HID Validation:
The 3500 Series Genetic Analyzers
Internal Standard Normalization:
Injection to Injection Peak Height Consistency
GeneScan™ 600 LIZ®Size Standard v2.0
% CV
Normalization Method
Factory Only
Factory + Internal Std
40
30
20
10
0Lot 1 Lot 2 Lot 3
Size Standard Peak Height Consistency · 6 instruments · 3 lots of size standard
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30