  • Hello! I am an AI chatbot trained to assist you with the Roche Multiplate 5 Analyzer User manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
Multiplate
®
analyzer
Operator’s Manual version 4.0
Software version 2.06
Roche Diagnostics
Multiplate® analyzer · Software version 2.06 · Operator’s Manual · version 4.0
2
Document information
Revision history
Edition notice
This publication is intended for users of the Multiplate
®
analyzer.
Every effort has been made to ensure that all the
information contained in this publication is correct at the
time of publishing. However, the manufacturer of this
product may need to update the publication information
as output of product surveillance activities, leading to a
new version of this publication
Where to find information The documentation is divided into the following parts:
• Part A — System Description: The system
description provides information on how the analyzer
is installed, on its safe operation, and on its hardware
and software.
• Part B — Operation: The operation part describes
how to perform the various tasks that are required
when analyzing samples, and how to set up the
analyzer.
• Part C — Maintenance: The maintenance part
provides instructions on how to ensure problem-free
operation of the system.
• Part D — Troubleshooting: The troubleshooting part
provides assistance on how to react to messages that
are displayed on screen and on how to deal with
exceptional situations.
q Keep this Operator’s Manual in a safe place to
ensure that it is not damaged and remains available for
use.
o This Operator’s Manual should be easily accessible at
all times.
Manual version Revision version Software version Revision date Change description
1.0 2.03.11 April 2012 First publication.
1.0 1.0 2.03.13 October 2012 Available languages changed
2.0 2.04 May 2013 Changes in the user interface
3.0 2.05 June 2019 LIS connectivity
4.0 2.06 April 2020 Windows 10 operating system
New layout
Kiosk mode
Supported printer
Menu items
y Revision history
Roche Diagnostics
Multiplate® analyzer · Software version 2.06 · Operator’s Manual · version 4.0
3
Training Do not carry out operation tasks or maintenance actions
unless you have received training from Roche
Diagnostics. Leave tasks that are not described in the
user documentation to trained Roche Service
representatives.
Images The screenshots and hardware images in this publication
have been added exclusively for illustration purposes.
Configurable and variable data in screenshots, such as
tests, results, or path names visible therein must not be
used for laboratory purposes.
Warranty Any customer modification to the system renders the
warranty or service agreement null and void.
For conditions of warranty, contact your local sales
representative or refer to your warranty contract partner.
Always leave software updates to a Roche Service
representative, or perform such updates with their
assistance.
Copyright © 2012–2020, F. Hoffmann-La Roche Ltd. All rights
reserved.
License information
Multiplate
®
analyzer software is protected by contract
law, copyright law, and international treaties. Multiplate
®
analyzer contains a user license between F. Hoffmann-La
Roche Ltd. and a license holder, and only authorized
users may access the software and use it. Unauthorized
use and distribution may result in civil and criminal
penalties.
Open-source and commercial software
Multiplate
®
analyzer may include components or
modules of commercial or open-source software.
Copyright and other notices and licensing information
regarding such Software programs can be found on the
Multiplate
®
analyzer.
This open-source and commercial software and
Multiplate
®
analyzer as a whole can constitute a device
regulated in accordance with applicable law. For more
detailed information, refer to the corresponding user
documentation and labeling.
Note that the respective authorization is no longer valid
according to the corresponding legislation should any
unauthorized changes be made to Multiplate
®
analyzer.
Trademarks The following trademarks are acknowledged:
Roche Diagnostics
Multiplate® analyzer · Software version 2.06 · Operator’s Manual · version 4.0
4
Multiplate is a trademark of Roche. Transferpette is a
trademark of BRAND.
All other trademarks are the property of their respective
owners.
Approvals
The Multiplate
®
analyzer meets the requirements laid
down in:
Directive 98/79/EC of the European Parliament and of the
Council of 27 October 1998 on in vitro diagnostic medical
devices.
Directive 2011/65/EU of the European Parliament and of
the Council of 8 June 2011 on the restriction of the use of
certain hazardous substances in electrical and electronic
equipment.
The Multiplate
®
analyzer is manufactured and tested
according to the following international standards:
• IEC 61326-2-6
• 61010-1
• 61010-2-101
Compliance with the applicable directive(s) is provided by
means of the Declaration of Conformity.
The following mark demonstrates compliance:
Contact address
For in vitro diagnostic use.
Complies with the provisions of the applicable EU directives.
Roche Diagnostics GmbH
Sandhofer Strasse 116
68305 Mannheim
Germany
Made in Germany
Table of contents 5
Roche Diagnostics
Multiplate® analyzer · Software version 2.06 · Operator’s Manual · version 4.0
Table of contents
Document information 2
Contact address 4
Table of contents 5
Preface 7
Intended use 7
Symbols and abbreviations 7
What is new in publication version 4.0 9
What is new in publication version 3.0 10
What is new in publication version 2.0 12
System description
1Safety
Safety classification 17
Safety precautions 18
Safety summary 21
Data security 30
Safety labels on the equipment 31
Disposal of the equipment 32
2 Installation
Setting up the analyzer 35
Unpacking the analyzer 35
Installing the analyzer 36
Setting up the software 38
Installing the electronic pipette 41
Installing the printer 43
Verifying the installation 47
3Overview
Multiplate technology 51
Components 51
Electronic pipette 54
Technical specification 58
Consumables and accessories 60
4 Software
Introduction 63
Main screen 65
Menu options 73
Operation
5 Operation
Safety 103
Starting the analyzer 105
Using the electronic pipette 106
Performing a measurement 108
Result handling 129
Deleting orders from the worklist 134
Reagent storage conditions and
preparation 135
Multiplate system control 138
Shutting down the analyzer 146
Taking the analyzer out of operation 146
Logging on as administrator (LabAdmin) 147
6 Configuration
Passwords 151
Setting up a printer 152
Adjusting the internal control parameters for
data alarms 153
Configuring the analyzer 154
Changing the language 159
Editing test parameters 160
LIS configuration 164
Electronic QC 165
Maintenance
7 Maintenance
Overview 171
Cleaning and disinfecting the analyzer 173
Maintenance and cleaning of the electronic
pipette 174
Checking the proper functioning of the
printer 185
Checking the temperature 186
Troubleshooting
8 Troubleshooting
Dealing with exceptional situations 191
Troubleshooting the electronic pipette 198
6 Table of contents
Roche Diagnostics
Multiplate® analyzer · Software version 2.06 · Operator’s Manual · version 4.0
Roche Diagnostics
Multiplate® analyzer · Software version 2.06 · Operator’s Manual · version 4.0
7
Preface
Intended use
The Multiplate (multiple platelet function) analyzer is an
instrument designed for in-vitro diagnostic use in
hospitals and other medical and scientific institutions.
Symbols and abbreviations
Product names Except where the context clearly indicated otherwise, the
following product names and descriptors are used.
Symbols used in the publication The following symbols are used:
Product name Descriptor
Multiplate
®
analyzer
analyzer
Multiplate Transferpette electronic pipette or autopipette
y Descriptors
Symbol Comment
o List item
P Start of procedure
S End of procedure
U Cross-reference
f Navigation path indication
Q Tip
%1, %2 Placeholder for software provided
values.
Safety alert
Electrical and electronic equipment
marked with this symbol are covered by
the European directive WEEE.
y Symbols used for easy recognition of information
Roche Diagnostics
Multiplate® analyzer · Software version 2.06 · Operator’s Manual · version 4.0
8
Symbols used on product
Abbreviations The following abbreviations are used:
Symbol Explanation
Date of manufacture
Global Trade Item Number.
y Symbols used on product
Abbreviation Explanation
EC European community
EN European standard
HL7 Health level 7
IEC International electrical commission
IVD In vitro diagnostics
LIS Laboratory information system
LQC Liquid quality control
MEA Multiple electrode aggregometry
n/a Not applicable
PDF Portable document format
PNG Portable network graphics
QC Quality control
XML Extensible markup language
y Abbreviations used
Roche Diagnostics
Multiplate® analyzer · Software version 2.06 · Operator’s Manual · version 4.0
9
What is new in publication version 4.0
Operating system Operating system changed to Windows 10
The operating system has been changed to Windows 10
for new Multiplate installations. The installation
instruction has been changed accordingly.
Layout Content migrated into new page layout.
Kiosk mode
The Multiplate
®
analyzer starts automatically in operator
mode without log on request.
Supported printer New Kyocera printer added.
u Installation of the printer (45)
New menu items The following menu items are directly available and
supported in the Multiplate software:
• Start IrfanView
u See:
Using keyboard shortcuts (63)
Start IrfanView (91)
To generate printouts using IrfanView p (120)
• Start PDF Reader
u See:
Using keyboard shortcuts (63)
PDF Reader (91)
• Logoff
u See:
Using keyboard shortcuts (63)
Roche Diagnostics
Multiplate® analyzer · Software version 2.06 · Operator’s Manual · version 4.0
10
What is new in publication version 3.0
Safety In the following safety sections new safety statements
have been added or existing statements have been
adapted:
u Operator qualification (18)
u Barcode read errors (24)
u Patient misidentification (24)
u Sharps, rough edges, and/or moving parts (21)
u Malicious software and hacker attacks (27)
u Unauthorized access (28)
u Data security (30)
u Data alarms (142)
Roche-provided firewall Information about Roche-provided firewall added.
Barcode reader Information about hand-held barcode reader added.
Supported printers Installation information about the Lexmark MS310d
printer added.
u Installing the printer (43)
Supported barcode types A list of supported barcode types was added.
u Barcodes (59)
Service menu The Service menu was removed from the menu bar. All
menu commands that are not service related have been
moved to the Configuration menu.
u Configuration menu (94)
LIS support and Worklist window LIS support was added. Use the new Worklist window to
do the following:
• Download orders from the LIS.
• Assign orders to channels.
• Start autopipette measurements.
• Upload results to the LIS.
• Manage orders.
u Worklist (79)
u Performing measurements using the worklist (108)
u Sending results to LIS from measurements created on
the analyzer (127)
u Deleting orders from the worklist (134)
Roche Diagnostics
Multiplate® analyzer · Software version 2.06 · Operator’s Manual · version 4.0
11
The LIS communication is configured in the LIS
configuration window.
u LIS configuration (164)
Electronic QC Information about the Electronic QC window was added.
u Electronic QC (97)
Reruns Use the worklist to rerun an order. Rerun orders are
indicated in the Test field in the Worklist window with
Rerun and the rerun number (for example Rerun 1).
u Rerunning an order (128)
Result deletion Information on how to delete result data was added.
u Deleting result data (133)
Multiplate passwords The purpose of the level 1 and level 2 Multiplate
passwords has been changed: the level 1 password is an
operator password and the level 2 password is an
administrator password. Both passwords have different
access levels for the configuration options.
u Passwords (151)
Test ID A unique test ID used on the LIS was introduced.
u Test ID (161)
Roche Diagnostics
Multiplate® analyzer · Software version 2.06 · Operator’s Manual · version 4.0
12
What is new in publication version 2.0
Main screen redesigned The main screen of the Multiplate software has been
optimized and redesigned.
Target range added A target range has been added for each test. The target
range can be displayed on the main screen.
Graphic zoom factors expanded The zoom factors for the channel graphics have been
extended to 6 levels covering up to 600 AU.
Menus and windows restructured The Multiplate menus and the corresponding windows
have been restructured. The former Service menu has
been split into two menus: Configuration for user
configuration tasks and Service for Roche Service tasks.
Multiplate tests Manufacturer defined tests cannot be edited anymore.
Only user defined tests are fully editable.
User interface terms reworked The terminology of the user interface terms (all the text
that is visible on screen) has been reworked and adapted
to Roche standards.
Operating system changed to Windows 7 The operating system has been changed to Windows 7
for new Multiplate installations. The installation
instruction has been changed accordingly.
Keyboard shortcuts updated Keyboard shortcuts have been updated for all menus.
Printer installation procedure Procedures for installing new printers have been added to
the manual.
Roche Diagnostics
Multiplate® analyzer · Software version 2.06 · Operator’s Manual · version 4.0
System description
1 Safety .......................................................................................................................... 15
2 Installation................................................................................................................. 33
3 Overview .................................................................................................................... 49
4 Software..................................................................................................................... 61
1 Safety
Roche Diagnostics
Multiplate® analyzer · Software version 2.06 · Operator’s Manual · version 4.0
15
Table of contents
Safety 1
In this chapter you find information about the safe
operation of the equipment.
In this chapter
1
Safety classification . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Safety precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Operator qualification . . . . . . . . . . . . . . . . . . . . . . . 18
Safe and proper use of the analyzer . . . . . . . . . . . 18
Miscellaneous safety precautions . . . . . . . . . . . . . 19
Safety summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Warning messages . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Sharps, rough edges, and/or moving parts. . . 21
Electrical safety . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Biohazardous materials . . . . . . . . . . . . . . . . . . . 22
Waste . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Fire . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Barcode read errors . . . . . . . . . . . . . . . . . . . . . . 24
Patient misidentification. . . . . . . . . . . . . . . . . . . 24
Caution messages . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Reagents and other working solutions . . . . . . 25
Cross contamination of samples . . . . . . . . . . . 25
Fatigue due to long hours of operation . . . . . . 26
Malfunction due to interfering
electromagnetic fields . . . . . . . . . . . . . . . . . . . . 26
Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Malicious software and hacker attacks. . . . . . 27
Unauthorized access . . . . . . . . . . . . . . . . . . . . . 28
Notices. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Circuit breakers and fuses . . . . . . . . . . . . . . . . 29
Spillage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Data security . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Safety labels on the equipment . . . . . . . . . . . . . . . . . . 31
1 Safety
Roche Diagnostics
Multiplate® analyzer · Software version 2.06 · Operator’s Manual · version 4.0
16
Table of contents
Disposal of the equipment . . . . . . . . . . . . . . . . . . . . . . 32
1 Safety
Roche Diagnostics
Multiplate® analyzer · Software version 2.06 · Operator’s Manual · version 4.0
Safety classification 17
Safety classification
This section explains how precautionary information is
presented in this manual.
The safety precautions and important user notes are
classified according to the ANSI Z535.6 Standard.
Familiarize yourself with the following meanings and
icons:
!
Generic hazard statements
r The safety alert symbol by itself without a signal word
is used to promote awareness to hazards which are
generic or to direct the reader to safety information
provided elsewhere in the document.
These symbols and signal words are used for specific
hazards:
WARNING
!
Warning
r Indicates a hazardous situation which, if not avoided,
could result in death or serious injury.
CAUTION
!
Caution
r Indicates a hazardous situation which, if not avoided,
could result in minor or moderate injury.
NOTICE
Notice
r Indicates a hazardous situation which, if not avoided,
could result in damage to equipment.
Important information that is not safety relevant is
indicated by the following symbol:
q Tip
Indicates additional information on correct use or useful
tips.
1 Safety
Roche Diagnostics
Multiplate® analyzer · Software version 2.06 · Operator’s Manual · version 4.0
18 Safety precautions
Safety precautions
!
Generic hazard statements
r Particular attention must be paid to the following
safety precautions. If these safety precautions are
ignored, the operator may suffer serious or fatal injury.
Each precaution is important.
Operator qualification
The intended users of the analyzer are appropriately
trained healthcare professionals operating the analyzer in
a laboratory setting.
Operators are required to have a sound knowledge of
relevant guidelines and standards as well as the
information and procedures contained in the Operator’s
Manual.
• Do not carry out operation and maintenance unless
you have been trained by Roche Diagnostics.
• Carefully follow the procedures specified in the
Operator’s Manual for the operation and maintenance
of the analyzer.
• Leave maintenance, installation or service that is not
described in the Operator’s Manual to trained Roche
Service representatives.
• Follow standard laboratory practices, especially when
working with biohazardous material.
• Perform regular security and privacy awareness
trainings for staff handling personal data in a
compliant way and according to the privacy principles
that are mandatory by customer regulations.
Safe and proper use of the analyzer
Personal protective equipment • Be sure to wear appropriate protective equipment,
including, but not limited to, safety glasses with side
shields, fluid resistant lab coat, and approved
disposable gloves.
• Wear a face shield if there is a chance of splash or
splatter.
1 Safety
Roche Diagnostics
Multiplate® analyzer · Software version 2.06 · Operator’s Manual · version 4.0
Safety precautions 19
Accuracy/precision of measured results An incorrect measuring result may lead to an error in
diagnosis, therefore posing danger to the patient.
• Do not use consumables that have exceeded their
expiration date, otherwise inaccurate data may be
obtained.
• For diagnostic purposes always assess the results in
conjunction with the patient’s medical history, clinical
examination, and other findings.
Installation • Leave installation tasks that are not described in the
Operator’s Manual to trained Roche Service
representatives.
• Follow the installation instructions carefully.
Operating conditions • Operation outside of the specified ranges may lead to
incorrect results or malfunction of the analyzer.
u See Environmental conditions (58).
• Use the analyzer indoors only and avoid heat and
humidity.
u See Environmental conditions (58).
• Perform maintenance according to the specified
intervals to maintain the required operating conditions
for the analyzer.
• Keep the Operator’s Manual in a safe place to ensure
that it is not damaged and remains available for use. It
must be easily accessible at all times.
Approved parts Use of non-approved parts or devices may result in
malfunction and may render the warranty null and void.
Only use parts and devices approved by Roche.
Approved accessories Use of non-approved accessories may result in
malfunction. Only use accessories approved by Roche.
u See Consumables and accessories (60).
Third-party software Installation of any third-party software that is not
approved by Roche Diagnostics may result in incorrect
behavior of the analyzer. Do not install any non-approved
software. Non-compliance will void the warranty.
u See Setting up the software (38).
Miscellaneous safety precautions
Power interruption A power failure or momentary drop in voltage may
damage the analyzer or lead to data loss. Perform regular
backups of measurement results. Do not switch off power
while the computer equipment accesses the hard disk.
1 Safety
Roche Diagnostics
Multiplate® analyzer · Software version 2.06 · Operator’s Manual · version 4.0
20 Safety precautions
Analyzer unused for an extended period of
time
If the analyzer is not used for an extended period of time,
it must be shut down and switched off. Handle the
reagents as described in their respective package inserts.
For further information, call your Roche Service
representative.
/