Innovatis Cedex HiRes User manual

Type
User manual
User Manual
Cedex HiRes 1.10 2005
User Manual Cedex HiRes Version 1.10 2005
2
Contents
Contents
CONTENTS ............................................................................................................3
1 RIGHTS..........................................................................................................7
Copyright ...........................................................................................................................................7
Disclaimer..........................................................................................................................................7
FCC Notification................................................................................................................................7
Safety Instructions............................................................................................................................7
2 INTRODUCTION............................................................................................9
3 DESCRIPTION OF CEDEX HIRES..............................................................11
3.1 The Analyzer ..........................................................................................................................11
3.2 The Measurement Procedure...............................................................................................12
The Trypan Blue Exclusion Method ........................................................................................12
3.3 Data Analysis.........................................................................................................................12
4 INSTALLATION AND OPERATION OF THE CEDEX HIRES.....................15
4.1 Installing Cedex HiRes..........................................................................................................15
Contents of Delivery................................................................................................................15
4.2 Choosing a Suitable Location..............................................................................................16
4.3 Preparing the Analyzer.........................................................................................................16
The Reagent Kit.......................................................................................................................16
Installing or Changing the Reagent Kit ..............................................................................16
Refilling the Containers by the User ..................................................................................18
4.4 Preparation and Configuration of the Analysis Software by the Cedex
Administrator
.........................................................................................................................19
5 WORKING WITH THE CEDEX HIRES ........................................................21
5.1 Cedex HiRes - General Instructions....................................................................................21
Switching on............................................................................................................................21
Logging in at the Computer Operating System Level .............................................................21
Starting the Cedex HiRes Software.........................................................................................21
Checking the Analyzer.............................................................................................................22
The Logout ..............................................................................................................................23
Automatic Logout.....................................................................................................................23
Shutting down the Workbench ................................................................................................23
5.2 Description of the Cedex HiRes Workbench......................................................................23
Menu Bar.................................................................................................................................24
File Menu............................................................................................................................24
Special Menu......................................................................................................................24
Setup Menu........................................................................................................................26
The Menu: Users................................................................................................................27
The Menu: Help..................................................................................................................27
User Identification area............................................................................................................27
Data Source Identification Area...............................................................................................27
Measure Button..................................................................................................................27
View Button........................................................................................................................27
Reprocess Button...............................................................................................................27
Sample Properties Area ..........................................................................................................28
Process Parameters Area .......................................................................................................28
Status Area..............................................................................................................................28
User Manual Cedex HiRes Version 1.10 2005 3
Reagent Area.......................................................................................................................... 28
Cell Image Area...................................................................................................................... 28
Result Data Area .................................................................................................................... 28
Liquid Management Button..................................................................................................... 28
Emergency Stop Button.......................................................................................................... 29
Assign and Start Buttons........................................................................................................ 29
5.3 Description of Access Rights.............................................................................................. 29
5.4 Carrying out a Measurement............................................................................................... 29
Performing a Prime................................................................................................................. 30
Choosing the Operating Mode and Specifying Sample Data................................................. 30
Entering a Comment............................................................................................................... 31
Defining the Measurement and Analysis (Process) Parameters............................................ 31
Preparing the Sample and Attaching it to the Sample Port.................................................... 32
Starting Measurement ............................................................................................................ 33
End of Measurement .............................................................................................................. 34
5.5 Interrupting a Measurement ................................................................................................ 35
5.6 Switching off the Cedex HiRes............................................................................................ 35
Carrying out an LM ShutDown ............................................................................................... 35
6 WORKING WITH THE MULTI SAMPLER................................................... 37
6.1 Introduction........................................................................................................................... 37
6.2 Working with the Multi Sampler.......................................................................................... 38
The Multi Sampler SpecialCup............................................................................................... 38
Changing the default position of the SpecialCup.................................................................... 39
6.3 Preparing the Measurement................................................................................................ 39
Using the Multi Sample Tray................................................................................................... 39
Ejecting the Multi Sample Tray............................................................................................... 40
Calibrating the Measurement System with a Focus Adjustment............................................ 40
Performing a Prime................................................................................................................. 41
6.4 Carrying out a Measurement with the Multi Sampler........................................................ 41
The Multi Sampler Data Entry Window................................................................................... 41
Starting a Measurement ......................................................................................................... 42
Disruption of Measurements................................................................................................... 43
Modifying Data Entry, Changing Samples.............................................................................. 43
Non-Linear Measurements in a Sample Series...................................................................... 43
Interrupting or Stopping a Measurement................................................................................ 44
6.5 Results................................................................................................................................... 44
6.6 The Multi Sampler Result Table (Print Results)................................................................. 44
Printout of the Measurement Data.......................................................................................... 45
7 ANALYSIS RESULTS ................................................................................. 47
7.1 Description of the Analysis Results................................................................................... 47
Viable Cell Density Value ....................................................................................................... 47
Total Cell Density Value ......................................................................................................... 47
Viability Parameter Value ....................................................................................................... 47
Total Cell Count Value............................................................................................................ 47
Average Compactness Value................................................................................................. 47
Average Diameter Value......................................................................................................... 48
Aggregate Rate Value ............................................................................................................ 48
Standard Deviation Value....................................................................................................... 48
7.2 Viewing Individual Analysis Results after Completion of the Measuring Process........ 48
7.3 Graphical Display of the Analysis Results......................................................................... 48
Enlarging the Diagrams.......................................................................................................... 48
Aggregate Histogram.............................................................................................................. 49
Measurement Statistics Histogram......................................................................................... 51
4 User Manual Cedex HiRes Version 1.10 2005
Contents
Compactness Histogram.........................................................................................................52
Diameter Histogram.................................................................................................................52
Cultivation Time Chart.............................................................................................................54
Areas in the Cultivation Time Chart Window .....................................................................55
The Cultivation Time Chart Diagram ...............................................................................55
The Data Selection Area..................................................................................................56
The View Area .................................................................................................................57
Saving a View..................................................................................................................57
Loading a View ................................................................................................................57
Deleting a View................................................................................................................58
General Area......................................................................................................................58
Printing User defined Results..........................................................................................58
Exporting the Results.......................................................................................................59
Entering a Comment........................................................................................................59
Hiding the Cultivation Time Chart Window......................................................................59
The Extras Area.................................................................................................................59
Data Series Area................................................................................................................61
List of Measurement Series in the Data Series Area.......................................................61
New Button ......................................................................................................................61
The Details Button ...........................................................................................................61
The Display of the Measurement Series in the Diagram.................................................61
Measurement Series Offset.............................................................................................62
Grouping Measurement Series ...............................................................................................62
Selection of Data Series Window.......................................................................................62
Select Source Area..........................................................................................................63
The Buttons in the Selection of Data Series Window......................................................63
The Selected Measurements Area..................................................................................64
7.4 Exporting the Data.................................................................................................................65
The Export Window............................................................................................................65
Output File Name Area....................................................................................................65
Output Format Area.........................................................................................................66
Data Series Mixing Area..................................................................................................66
Additional Information Area .............................................................................................66
CSV Options Area............................................................................................................67
Export Template Area......................................................................................................67
Buttons in the Export Template Area...............................................................................67
Available Export Templates for Excel..............................................................................68
Export Structure Area......................................................................................................68
Start Export Button...........................................................................................................69
Hide Button......................................................................................................................69
7.5 Viewing and Selecting Cell Images .....................................................................................69
Viewing Individual Cell Images................................................................................................69
Viewing a Cell Image Using the IP ResultViewer....................................................................70
Excluding Images from the Evaluation Process/Including Excluded Images..........................72
7.6 Printing the Analysis Results...............................................................................................72
7.7 Viewing and Evaluating the Data Audit Trail File...............................................................73
7.8 Editing Existing Analysis Results/Data Sets......................................................................73
Viewing Existing Analysis Results/Modifying Measurement Parameters ...............................74
Reprocessing Existing Data Sets............................................................................................75
8 GENERAL FUNCTIONS OF THE SYSTEM ................................................77
8.1 Specifying Target Directories for the Analysis Data .........................................................77
8.2 Changing your own Password.............................................................................................77
8.3 Viewing your own Access Rights........................................................................................77
8.4 Data Entry Fields for the Data Source Indentification Area..............................................78
8.5 Specifying your own Default Measurement Parameters...................................................78
User Manual Cedex HiRes Version 1.10 2005 5
9 ADMINISTRATOR FUNCTIONS................................................................. 79
9.1 Administration of User Profiles........................................................................................... 79
9.2 Adjustment of the Image Processing Operator’s Counting Routine using the Live-
Operator................................................................................................................................. 79
Description of the Individual Operator Parameters ................................................................ 80
9.3 Securing Data........................................................................................................................ 82
10 MAINTENANCE OF THE ANALYZER........................................................ 83
10.1 Cleaning and Maintenance Check List............................................................................... 83
10.2 Daily and, if necessary, manual Maintenance and Cleaning Routines........................... 84
Checking the Reagent Containers when manually refilling.................................................... 84
Cleaning Routines Carried out by the Liquid Management.................................................... 84
The Daily Events Area....................................................................................................... 85
The Cleaning Area............................................................................................................. 85
Automatic Request for Cleaning........................................................................................ 87
The Syringe Handling Area............................................................................................... 87
Carrying out a Cleaning Routine....................................................................................... 87
Interrupting/Terminating a Cleaning Routine .................................................................... 88
Verifying the cleaning with the help of the Chamber Viewer............................................. 88
10.3 Syringe Maintenance............................................................................................................ 89
Removing the Syringe ............................................................................................................ 89
Assembling and Installing the Syringe.................................................................................... 90
10.4 Maintenance of the Syringe Guide and Spindle................................................................ 91
10.5 Maintenance of the Flow chamber, Microscope Optic, 8-Way Valve and the
remaining Hardware Components
...................................................................................... 92
APPENDIX........................................................................................................... 92
Contact Details ............................................................................................................................... 92
Technical Specifications................................................................................................................ 93
Suppliers ......................................................................................................................................... 94
Material Safety Data Sheets .......................................................................................................... 94
INDEX .................................................................................................................. 95
6 User Manual Cedex HiRes Version 1.10 2005
Ri
g
hts
1
1 Rights
Copyright
Cedex
®
Operating Instructions. Copyright
©
2005 innovatis AG. All rights reserved. Operating
instructions and manuals are protected by copyright. No part of this document may be
reproduced, edited or translated without prior written consent of innovatis AG except as
permitted under the terms of the copyright.
Disclaimer
The information in this document is subject to change without prior notice and should not be
construed as a commitment by innovatis AG.
Innovatis AG accepts no responsibility or liability for any errors or inaccuracies in this document
or for any damage occurring during delivery, performance or use of this product.
Windows XP and Windows 2000
®
are registered trademarks of the Microsoft Corporation. All
other trademarks are trademarks or registered trademarks of the respective companies.
Cedex ® is a registered trademark of innovatis AG.
FCC Notification
WARNING
Changes or modifications to this unit not expressly approved by the party responsible for
compliance may void the user‘s authority to operate the equipment.
Note: This equipment has been tested and found to comply with the limits for a Class A digital
device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable
protection against harmful interference when the equipment is operated in a commercial
environment. This equipment generates, uses and can radiate frequency energy and, if not
installed and used in accordance with the instruction manual, may cause harmful interference to
radio communications. Operating this equipment in a residential area is likely to cause harmful
interference, in which case the user will be required to correct the interference at his own
expense.
Safety Instructions
WARNING
It is important that the following safety instructions and guidelines be observed in order to
ensure safe and reliable operation of Cedex HiRes:
Place the device on a dry and sturdy surface, free of vibrations.
Avoid transporting the device as much as possible once it has been assembled.
Prevent the device from getting wet.
Protect the device from dust and smoke.
Do not expose the device to direct sunlight.
Ensure there is sufficient air circulation around the Cedex system during operation
(particularly around the computer and monitor).
Operate within a constant temperature range of 50°F – 95°F (10°C – 35°C).
User Manual Cedex HiRes Version 1.10 2005 7
For constant performance it is recommended to keep the analyzer permanently on and
only shut down the software application on a regular basis.
Cedex HiRes should only be used with the supplied or specified devices (Cedex HiRes
Analyzer, Cedex HiRes software, monitor, computer, Multi Sampler, Reagent Tray).
Do not install any additional software on the corresponding computer.
Only trained personnel in a controlled laboratory environment should operate the Cedex
system.
All services and repairs should be performed only by authorized service technicians.
Always ensure that there are sufficient reagents in the reagent containers.
Clean the containers before reusing them.
Always observe the safety regulations when handling Trypan Blue, detergent, disinfectant
or acids and bases.
Waste contaminated with tissue or cells, e.g. sample cups, pipette tips, etc., must be
disposed of according to the recommended rules for handling biohazard materials.
Any leaks in the liquid management system must be repaired immediately. Contact your
distributor or authorized service technician.
Always ensure that there is a sample cup attached to the single sample port.
Only use Cedex Detergent or a detergent with identical specifications for the cleaning
process in the liquid management system (see “Suppliers”, page 94).
Only use sample cups from suppliers specified on page 94 for the conveyance of cell
samples to the sample port.
Other sample cups with other catalog numbers can cause irreparable damage to
Cedex HiRes!
Always wear protective clothing, particularly gloves, when handling samples.
To prevent damage to the flow chamber, no particles larger than 80 µm should enter the
measurement system.
Particles (especially microcarriers) larger than 90 µm should not gain access into the
Analyzer because they cause irreparable damage and render the device unfunctional.
Dirt or dust particles should not gain access into the device.
Ensure that all equipment required for sample preparation, e.g. sample cups, pipette tips,
etc., are stored in a dust-free environment since coarse dust particles can impair the flow
chamber’s functionality.
Check that the sample cup is free of contamination before use. Only clean sample cups
can guarantee optimal functioning of the Cedex.
Ensure that the Trypan Blue solution is free of contamination, otherwise filter prior to use.
Innovatis recommends the use of the Cedex HiRes Reagent Kit (see “Suppliers”, page
94).
Secure your measurement data, as well as the Cedex.ini and CedexSt.ini files (see
Cedex HiRes Administrator Manual).
NOTE
The software and hardware of your Cedex HiRes can differ from the information described here,
but this should not make a difference to the operational functions in general.
WARNING
8 User Manual Cedex HiRes Version 1.10 2005
Ri
g
hts
1
The manufacturer is not responsible or liable for any damages resulting from improper use or
failure to comply with the regulations in the Operating Instructions.
User Manual Cedex HiRes Version 1.10 2005 9
2 Introduction
The Cedex HiRes determines the cell density and viability of cell suspensions automatically.
The analysis is based on the Trypan Blue exclusion method whereby cell suspensions are
stained and analyzed. The obtained data allows for the evaluation of the overall state of the cell
culture, including the specific growth rate and doubling time of a cell population, as well as the
documentation of the cultivation process.
The evaluation of the samples is based on automatically obtained digital images that are
analyzed using analysis software on a separate computer (Cedex HiRes control unit). All the
necessary steps – right from the preparation of the samples to the recording and evaluation of
the image data - are automatically performed by the Analyzer. Recognition and differentiation of
cells as either viable or dead are done under standardized conditions and, therefore, provide
very precise and objective results.
Many parameters, including Viable Cell Density, Total Cell Density, Viability, Total Cell Count,
Average Compactness and Average Diameter are analyzed from a single image result data
(Image), as well as from all of the selected images of a sample (Overall). Standard Deviation is
also calculated, but only from all of the selected images of a sample (Overall). In addition, the
Aggregate Rate, which shows the percentage of the total cells that appear in aggregates
(clusters), is determined for both single images (Image) and all of the images (Overall).
Graphical representations of the measurement results can be viewed in the Measurement
Statistics Histogram, Diameter Histogram, Compactness Histogram, Aggregate Histogram and
Cultivation Time Chart.
In the Measurement Statistics histogram, individual images can be compared or else a number
of images can be grouped together to give a better overview.
The Diameter Histogram shows the diameter distribution of the living cells (Cell Diameter
Histogram). In the current version, this function was extended with an additional diameter
histogram (Diameter Histogram for all Objects), which shows
all objects detected such as cells,
debris, air bubbles and viable and dead cell aggregates.
The deviation of the shape of living cells from that of an ideal sphere is represented in the
Compactness Histogram.
The Aggregate Histogram gives a quick overview of the amount of aggregation found in the cell
culture.
The Cultivation Time Chart displays the growth curves and the cultivation parameters in a time-
dependent manner. The results of the Cultivation Time Chart can be separately printed out. For
further processing of the results in other programs, data export in formats such as DIF (Data
Interchange Format, Excel- compatible) and CSV (Comma Separated Values, generally
applicable) is possible.The Cultivation Time Chart now includes calculations for the doubling
time Tau [T] and specific growth-rate [µ].
The Cedex 1.10. version (Cedex HiRes) can be operated on the Windows XP Professional and
Windows 2000 Professional operating systems. Secure data handling, as well as a carefully
designed user management system, guarantee the compatibility with CFR 21, Part 11.
Access security for Cedex HiRes has been stepped up. Access is allowed only for authorized
persons. The workstation can be automatically blocked after a certain number of consecutive
entries of a false password via the cedex.ini (see Administrator Handbook for more details).
Image recognition has been further improved. The use of a Live-Operator now makes it possible
to directly observe the effects on the analysis of a sample when the operator is adjusted.
10 User Manual Cedex HiRes Version 1.10 2005
Introduction
2
User Manual Cedex HiRes Version 1.10 2005 11
3 Description of Cedex HiRes
The Cedex HiRes consists of two functional units, the Analyzer (see Figure 1, page 11 ) and the
Control Unit (computer or laptop equipped with HiRes software).
The preparation and processing of the samples takes place in the Analyzer.
The analysis software is installed on a computer and controls the Analyzer; it also controls the
analysis and evaluation of the images obtained.
Single sample port with sample cup
Syringe module with 8-way valve
Tray with the Cedex Reagent Kit
(containers with Trypan Blue,
detergent, disinfectant, distilled
or de-ionized water, waste)
ON/OFF Switch
Figure 1
Cedex Analyzer
3.1 The Analyzer
The Analyzer consists of two main components, the hardware for image capturing (scanner)
and the liquid management system.
The flow chamber is specially developed to analyze a defined volume of cell suspension. The
Liquid Management system is responsible for the staining of the sample as well as for the filling
and cleaning of the flow chamber and the liquid feed system.
The volumes of sample and dye, as well as specific cleaning solutions required for each
cleaning routine, are precisely controlled via the syringe module with the 8-way distribution
valve.
The individual components of the liquid management system are easily accessible to allow for
trouble-free maintenance and control.
12 User Manual Cedex HiRes Version 1.10 2005
Description of the Cedex HiRes
3
3.2 The Measurement Procedure
The determination of the number of viable cells (cells/mL) and cell viability of cell suspensions
(the percentage of viable cells of the total cell sample) is a fully automated and highly
standardized procedure based on the manual Trypan Blue Exclusion Method. A defined volume
of Trypan Blue is added by the syringe to the sample cup containing the cell suspension. The
cell suspension and dye are optimally mixed by controlled pump actions between the syringe
and sample cup and then incubated for the exact same time for each measurement. This
minimizes effects on the viability measurements that might be caused by the toxicity of the
Trypan Blue.
After the incubation period, the stained cell suspension in the syringe is then transported in one
portion via the capillary tubes and the 8-way valve into the precision flow chamber, whose
image is recorded with the scanner. The scanned image of the flow chamber is then broken up
into smaller images for evaluation. By default, 150 images are analyzed. 20 images of previous
Cedex versions correspond to 155 Cedex HiRes images.
Once the measurement procedure is completed, a cleaning routine is automatically performed
by the Liquid Management. It is only possible to begin a new measurement after the cleaning
routine has finished.
The Trypan Blue Exclusion Method
The Trypan Blue exclusion method is based on the principle that viable cells will not take up the
dye, while dead or dying cells have a compromised membrane that allows for the uptake of the
dye. Thus, dead cells will appear darkly stained in a microscope.
NOTE
Trypan Blue has a greater affinity to serum proteins than to cellular proteins. If the microscopic
image has a dark background due to high quantities of stained serum proteins, then centrifuge
the cell sample and re-suspended it in a protein free medium or physiological saline solution
prior to counting.
3.3 Data Analysis
The analysis software and its graphical user interface (GUI), known as the Workbench,
enables the user to work with Cedex HiRes.
The Cedex HiRes software is installed on a computer provided by the customer. If a computer is
part of delivery, the software is by default pre-installed. The software analyzes the digital image
information extracted by the scanner.
The analysis software recognizes structures and identifies particles, thus differentiating cells
from cell debris and dirt particles. Viable and dead cells are differentiated according to their
degree of Trypan Blue staining.
The image processing is integrated in the Cedex HiRes Workbench in such a way that it allows
the user or Administrator to check on its function in the system constantly. The analysis process
cannot be altered during data recording in order to allow for consistent and objective
measurement results.
User Manual Cedex HiRes Version 1.10 2005 13
The Cedex HiRes can register several thousand cells per minute depending on cell density. The
rapid processing of image data allows for a significant increase in random checks, which greatly
improves the reliability and statistical accuracy compared to manual counting.
This analysis method provides information on, e.g., cell viability and morphology (diameter,
compactness).
14 User Manual Cedex HiRes Version 1.10 2005
Description of the Cedex HiRes
3
User Manual Cedex HiRes Version 1.10 2005 15
4 Installation and Operation of the Cedex HiRes
4.1 Installing Cedex HiRes
The installation of Cedex HiRes is done in the following steps:
Cedex HiRes is delivered and assembled by a service technician or a sales partner
authorized by innovatis AG.
The computer provided by the user must be equipped with the operating system Windows
XP Professional or Windows 2000 Professional. For further information, please refer to
the Cedex HiRes Administrator Manual.
The Cedex software and Cedex Security Pack should be installed by your system
administrator before the service technician or sales partner can install the Analyzer. The
relevant instructions will be delivered once the contract is confirmed.
Your system administrator should, if desired, have set up net access and/or printer
installation before the service technician or sales partner assembles Cedex HiRes.
The system technician or sales partner installs the Analyzer and checks the security
settings in the software.
The system technician or sales partner will check the Cedex entries, perform the focus
adjustment and ensure that Cedex HiRes is ready to use. Finally, he/she will create a
current status report of the device.
The system administrator is responsible for setting up the user account for the operating
system, for example CedexSA.
If required, the system administrator activates the Security Pack in order to control access
to the applications and desktop.
The Administrator for the Cedex then sets up Cedex user accounts and assigns them the
required rights (see Cedex HiRes Administrator Manual).
Contents of Delivery
1 x Cedex HiRes Analyzer
1 x USB hardware key (Dongle)
Power Supply adaptor 100-240V AC, 100W
Power supply cable (country specific)
Cords:
2 x Firewire cables, one with a 4 point and one with a 6 point connection, for connecting
the Analyzer to the Control Unit (PC)
1 x USB cable
Cedex HiRes Install CD
1 x User Manual incl. License agreement, 1 x Administrator Manual
Reagent Kit (Cat.No. 7 61 50 00 00 see Suppliers, page 94)
Set of empty Reagent Kit bottles
Sample cups (4 mL) (Cat.No. 6 00 05 00 00, see Suppliers, page 94)
Cedex Calibration Beads (Cat.No. 7 60 00 00 16 see Suppliers, page 94)
2 x ferrite cores for use with Fire Wire cable
16 User Manual Cedex HiRes Version 1.10 2005
Installation and Operation of Cedex HiRes
4
If the PC is
not provided by the customer but delivered by innovatis AG instead:
1 x PC or 1 x laptop (customer choice) including manual and DVD-burner with Windows
XP
®
or Windows 2000
®
and Cedex Analysis software
1 x keyboard
1 x mouse (only with PC)
1 x monitor incl. the manual
1 x recovery image cd
NOTE
The software and hardware of your Cedex HiRes may differ from the information described
here, but this should not make a difference to the operational functions in general.
NOTE
A printer is not included with the Cedex HiRes shipment. Follow the instructions for installing a
printer in the Cedex HiRes Administrator Manual.
4.2 Choosing a Suitable Location
The Cedex HiRes is highly sensitive and consists of top-quality electronic and optical
components. To guarantee optimal and reliable performance of the Cedex HiRes, select the
appropriate location according to the safety regulations described in the first section.
Avoid further transportation of Cedex HiRes after the professional installation.
WARNING
The transportation locks for the scanner must be attached to the inside of the casing before
Cedex HiRes is transported once again. Please contact an innovatis authorized service
technician in such a case. Vibrations or shaking must also be avoided when transporting the
Analyzer over short distances. Focus adjustment must be done after transportation (see Cedex
HiRes Administrator Handbook).
4.3 Preparing the Analyzer
The Reagent Kit
To ensure optimal performance of the Analyzer, the reagents must be free of impurities such as
dust, crystals, biological material or fluff. Innovatis AG’s Cedex Reagent Kit is manufactured
under GMP (Good Manufacturing Practice) conditions and guarantees optimal handling with the
best possible safety. One Reagent Kit pack is sufficient for 100 measurements, including the
standard cleaning routines.
Installing or Changing the Reagent Kit
Proceed as follows to install or change the Reagent Kit:
1. Perform an LM Shutdown if Liquid Management system has been primed (see “Carrying
out an LM ShutDown”, page 35).
User Manual Cedex HiRes Version 1.10 2005 17
2. Open the Reagent Kit chamber in Cedex HiRes by pushing the front upwards until it
clicks into place (see , page 17).
3. Remove the old Reagent Kit with the tray and dispose of the old reagent bottles if
necessary.
4. Open the new Reagent Kit and remove the covers from the individual containers.
5. Place the new Reagent Kit containers in the appropriate positions in the Reagent Tray
(see , page 17).
6. Close the front cover by pulling it all the way down. When the front cover is closed, the
capillaries are automatically correctly positioned in the individual containers.
7. Start the Cedex software to confirm that the Reagent Kit has been changed (see “”, page
22).
8. Confirm the changing of the Reagent Kit in the Setup menu of the Cedex HiRes
Workbench (see , page 17). The Reagent Control display on the Cedex HiRes
Workbench will turn from red to green.
9. Perform a Prime (see “Performing a Prime”, page 30).
10. Do not forget to attach a clean sample cup to the sample port.
Container for Detergent
Container for Trypan Blue
Container for disinfectant
Waste container
Container for distilled or de-ionized water
Figure 3
Position of the Reagent Tray with the individual containers
Figure 5
Confirming change of the Reagent Kit
18 User Manual Cedex HiRes Version 1.10 2005
Installation and Operation of Cedex HiRes
4
NOTE
For reliable functioning of the reagent control (correct number of possible measurements,
reporting an impending overflow of the waste container), it is critical that all of the reagent
containers are correctly filled to the line with the appropriate liquid (or, in the case of the waste
container, emptied) all at the same time (see “Refilling the Containers by the User”, page 18).
The changing of the Reagent Kit must then be confirmed in the software as soon as the
Reagent Tray has been put back into the chamber. Also, it is important that the sample cup is
filled with only the prescribed sample volume (+/-10 %) and that there is an empty cup in the
sample port when cleaning functions are being carried out. Failure to do all of the above will
result in a mis-match between the calculated reagents available as shown in the Reagent
Control and the actual reagent amounts in the containers or space available in the waste
container.
Refilling the Containers by the User
In order to ensure the optimal functioning of the Analyzer, the Reagent Kit containers must be
thoroughly cleaned before refilling (liquid change by the user is recommended once a week).
Impurities such as dust and fluff or contamination with micro-organisms may thus be avoided.
WARNING
If you are using your own reagents, it may be necessary to filter them before use.
Innovatis AG recommends using the Cedex HiRes Reagent Kit (see “Suppliers”, pg. 94).
For refilling the containers with reagents, follow the instructions below:
1. Rinse all reagent containers well before refilling! Make sure that there are no impurities
such as dust or fluff in the containers.
2. Fill the detergent container to the required level with Cedex Detergent (see label).
WARNING
Only Cedex Detergent or a detergent with identical specifications should be used for the cleaning
procedures in order to prevent damage to the sensitive surface of the Liquid Management
components (see “Suppliers”, page 94).
1. Fill disinfectant or 70% Isopropyl in the disinfectant container to the required level (see
label).
2. Fill distilled or de-ionized water in the water container to the required level (see label).
3. Fill 0.2% Trypan Blue solution in the Trypan Blue container to the required level (see label).
Innovatis AG recommends using 0.2%Trypan Blue solution. If necessary, dilute a more
concentrated Trypan Blue solution with buffer (Phosphate Buffered Saline (PBS), pH 7.4).
4. Put all of the reagent containers, as well as the emptied waste container, in the appropriate
positions in the Reagent Tray (see , page 17).
5. There should always be a sample cup attached to the single sample port (see , page 33).
The sample cup (4 mL) is correctly attached when it is in an upright position beneath the
sample port and the capillary tube almost reaches the bottom of the sample cup.
NOTE
If the Trypan Blue is completely used up during measurements and then refilled, a Prime must
be performed again before running the next sample (see “Performing a Prime”, page 30). If a
Prime is not performed, the first measurement after refilling of Trypan Blue will be labeled
User Manual Cedex HiRes Version 1.10 2005 19
Invalid (“Invalid Measurement” in red will appear in the Result Data window, see , page 25 ).
Subsequent measurements will be valid.
4.4 Preparation and Configuration of the Analysis Software by
the Cedex Administrator
The operating system of the PC will automatically start after the individual components of the
Cdex HiRes have been switched on in the correct order (first the Analyzer and then the
computer, see “Switching on”, page 21). The Cedex HiRes administrator has to execute the
following operations before the Cedex HiRes software can be used routinely:
secure access rights and define the Administrator password.
create user profiles and delegate access rights.
define target directories for the analysis data.
perform the focus adjustment.
Detailed information can be found in the Cedex HiRes Administrator Manual.
NOTE
For security reasons, the Cedex HiRes Administrator is only accorded administrative rights.
He/she cannot carry out, reprocess or view measurements.
After the successful configuration of a user profile, the access rights of the user can be defined.
Access to the following areas can then be determined:
Available work areas
Default work area
Cell Types / Operators (software routines for image recognition)
User Management
System
Detailed information can be found in the Cedex HiRes Administrator Manual.
20 User Manual Cedex HiRes Version 1.10 2005
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98

Innovatis Cedex HiRes User manual

Type
User manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI