SICK GM901, Measuring probe version Operating instructions

Category
Measuring & layout tools
Type
Operating instructions
O P E R A T I N G I N S T R U C T I O N S
GM901-02
CO Measuring Device with Measuring Probe
Installation, Operation, Maintenance
Described product
GM901-02
with measuring probe
Manufacturer
SICK AG
Erwin-Sick-Str. 1
79183 Waldkirch
Germany
Legal information
This work is protected by copyright. Any rights derived from the copyright shall be
reserved for SICK AG. Reproduction of this document or parts of this document is only
permissible within the limits of the legal determination of Copyright Law. Any modifica‐
tion, abridgment or translation of this document is prohibited without the express writ‐
ten permission of SICK AG.
The trademarks stated in this document are the property of their respective owner.
© SICK AG. All rights reserved.
Original document
This document is an original document of SICK AG.
2
O P E R A T I N G I N S T R U C T I O N S | GM901-02 8008932/Y153/3-0/2016-03 | SICK
Subject to change without notice
Contents
1 About this document........................................................................ 6
1.1 Function of this document....................................................................... 6
1.2 Scope of application................................................................................. 6
1.3 Target group.............................................................................................. 6
1.4 Symbols and document conventions...................................................... 6
1.4.1 Warning symbols...................................................................... 6
1.4.2 Warning levels / Signal words................................................. 7
1.4.3 Information symbols................................................................ 7
2 Safety information............................................................................ 8
2.1 Basic safety information........................................................................... 8
2.2 Intended use............................................................................................. 9
2.2.1 Purpose of the device.............................................................. 9
2.2.2 Responsibility of user.............................................................. 10
3 Product description........................................................................... 11
3.1 Product identification............................................................................... 11
3.2 Product features....................................................................................... 11
3.3 GM901-02 layout...................................................................................... 12
3.4 Measuring probes..................................................................................... 13
3.5 Evaluation unit.......................................................................................... 16
3.6 Accessories .............................................................................................. 16
4 Important information on GPP measuring probes....................... 19
5 Transport and storage....................................................................... 20
5.1 Transport safeguards............................................................................... 20
6 Mounting............................................................................................. 21
6.1 Information on installing the sender/receiver unit and measuring
probe......................................................................................................... 21
6.2 Overview of the installation steps (duct-side preparation).................... 22
6.3 Scope of delivery...................................................................................... 22
6.3.1 Check scope of delivery.......................................................... 22
6.3.2 Checking the delivery state..................................................... 22
6.4 Installing the flange with tube................................................................. 23
6.5 Installing the evaluation unit................................................................... 25
7 Electrical installation........................................................................ 26
7.1 Safety information for electrical installation........................................... 26
7.2 Connection overview................................................................................ 27
7.2.1 Connection diagram for standard version............................. 27
7.2.2 Electrical wiring on the evaluation unit.................................. 28
7.2.3 Connection diagram with terminal box (detached evalua‐
tion unit)................................................................................... 30
CONTENTS
8008932/Y153/3-0/2016-03 | SICK O P E R A T I N G I N S T R U C T I O N S | GM901-02
3
Subject to change without notice
7.3 Uninstalling the electrical system............................................................ 32
8 Commissioning.................................................................................. 33
8.1 Safety information for commissioning.................................................... 33
8.1.1 Important information on GPP measuring probes................ 35
8.2 Overview of commissioning work............................................................ 35
8.3 Material required ..................................................................................... 35
8.4 Overview of commissioning steps........................................................... 36
8.5 Criteria for switching on............................................................................ 36
8.6 Zero adjust................................................................................................ 37
8.6.1 Manual SPAN test (optional)................................................... 37
8.6.2 Online SPAN test according to EPA Guidelines...................... 39
8.6.3 Automatic SPAN test................................................................ 41
8.7 Basic parameters set............................................................................... 43
8.8 Installing the sender and receiver........................................................... 44
8.9 Connecting the sender and receiver with a CAN line............................. 48
8.10 Installing the weatherproof cover............................................................ 49
9 Operation............................................................................................ 51
9.1 Safety......................................................................................................... 51
9.2 Operating and display elements.............................................................. 51
9.2.1 LEDs.......................................................................................... 51
9.2.2 Operator panel of the evaluation unit GM901...................... 52
9.2.3 Setting the display contrast.................................................... 53
9.3 Entering the password.............................................................................. 54
9.4 Menus........................................................................................................ 55
9.4.1 Measuring mode...................................................................... 55
9.4.2 Diagnosis.................................................................................. 55
9.4.3 Parameters (settings).............................................................. 56
9.4.4 Calibration (Cal)....................................................................... 70
9.4.5 Service menu "Maint".............................................................. 73
10 Maintenance...................................................................................... 76
10.1 Safety......................................................................................................... 76
10.2 Preparatory work....................................................................................... 77
10.3 Visual inspection....................................................................................... 77
10.4 Maintenance plan..................................................................................... 77
10.4.1 Maintenance Table for sender/receiver unit......................... 78
10.4.2 Maintenance Table for GPP measuring probe....................... 78
10.4.3 Maintenance Table for purge air unit..................................... 78
10.4.4 Maintenance Table for GMP measuring probe...................... 78
10.4.5 Final maintenance work.......................................................... 78
11 Troubleshooting................................................................................. 79
11.1 Warning messages................................................................................... 79
11.2 Malfunction messages............................................................................. 80
CONTENTS
4
O P E R A T I N G I N S T R U C T I O N S | GM901-02 8008932/Y153/3-0/2016-03 | SICK
Subject to change without notice
11.3 Troubleshooting on the sender................................................................ 81
11.4 Troubleshooting on receiver..................................................................... 82
11.5 Troubleshooting on the evaluation unit................................................... 83
11.6 Communication fault between evaluation unit and receiver................. 84
11.6.1 Sensor values for trouble-free operation............................... 84
11.6.2 Remote Diagnosis, GM901 Sensor Values............................ 85
11.7 Measures after purge air failure (when using the GMP measuring
probe)........................................................................................................ 86
12 Decommissioning............................................................................. 87
12.1 Safety information on shutting down...................................................... 87
12.2 Materials and tools required.................................................................... 88
12.3 Disassembling the sender and receiver.................................................. 88
12.4 Storage...................................................................................................... 89
12.5 Disposal..................................................................................................... 89
13 Technical data.................................................................................... 90
13.1 Technical data........................................................................................... 90
13.1.1 Operating data......................................................................... 90
13.1.2 Technical Data - Measuring Probes........................................ 91
13.2 Dimension drawings................................................................................. 92
13.2.1 Dimension drawing, sender/receiver unit with GMP meas‐
uring probe............................................................................... 92
13.2.2 Dimension drawing, sender/receiver unit with GPP meas‐
uring probe............................................................................... 92
13.2.3 Dimension drawing, evaluation unit....................................... 93
13.2.4 Dimension drawings, accessories.......................................... 94
14 Ordering information........................................................................ 96
14.1 Spare parts............................................................................................... 96
14.2 Accessories............................................................................................... 96
15 List of tables....................................................................................... 97
CONTENTS
8008932/Y153/3-0/2016-03 | SICK O P E R A T I N G I N S T R U C T I O N S | GM901-02
5
Subject to change without notice
1 About this document
1.1 Function of this document
These Operating Instructions describe:
System components
Start-up
Operation
Maintenance work required for reliable operation
Troubleshooting
1.2 Scope of application
These Operating Instructions are valid only for the gas in-situ measuring device with the
designation GM901 with measuring probe.
They are not valid for the other gas in-situ measuring devices from SICK.
1.3 Target group
This Manual is intended for persons installing, operating and maintaining the device.
Operation
The device should only be operated by skilled technicians who, based on their technical
training and knowledge as well as knowledge of the relevant regulations, can assess
the tasks given and recognize the hazards involved.
Installation and maintenance
Skilled technicians may be required at certain times during installation and mainte‐
nance.
Please observe the information at the beginning of the respective Sections.
1.4 Symbols and document conventions
1.4.1 Warning symbols
Symbol Significance
Immediate hazard of severe personal injury or death
Hazard (general)
Hazard through electrical voltage
Hauard through toxic substances
Hazard for the environment/nature/organic life
Table 1: Warning symbols
1 ABOUT THIS DOCUMENT
6
O P E R A T I N G I N S T R U C T I O N S | GM901-02 8008932/Y153/3-0/2016-03 | SICK
Subject to change without notice
1.4.2 Warning levels / Signal words
DANGER
Risk or hazardous situation which will result in severe personal injury or death.
WARNING
Risk or hazardous situation which could result in severe personal injury or death.
CAUTION
Hazard or unsafe practice which could result in less severe or minor injuries.
Notice
Hazard which could result in property damage.
Note
Hints
1.4.3 Information symbols
Symbol Significance
Important technical information for this product
Important information on electrical or electronic functions
Table 2: Information symbols
ABOUT THIS DOCUMENT 1
8008932/Y153/3-0/2016-03 | SICK O P E R A T I N G I N S T R U C T I O N S | GM901-02
7
Subject to change without notice
2 Safety information
2.1 Basic safety information
ATTENTION
Responsibility of the operator in case of device malfunctions or failure
If the GM901-02 is used as a sensor in combination with control technology, the opera‐
tor must ensure that a failure or malfunction on the GM901-02 cannot lead to unal‐
lowed damage or hazardous operating states.
ATTENTION
Responsibility for system safety
The person setting the system up is responsible for the safety of the system in which
the device is integrated.
ATTENTION
Responsibility of the operator for correct usage of the device
Basis of this Manual is the delivery of the device according to the preceding project
planning (e.g., based on the SICK application questionnaire) and the relevant delivery
state of the device (see delivered system documentation).
b
Contact SICK Customer Service if you are not sure whether the device corresponds
to the state defined during project planning or to the delivered system documenta‐
tion.
DANGER
Health risk when safety information is not available
Improper installation or maintenance work can lead to accidents with serious conse‐
quences for the health of persons carrying out these tasks.
b
Keep these Operating Instructions and the associated instructions available for
reference.
b
Read and observe these Operating Instructions.
b
Observe all safety instructions.
b
If anything is not clear: Please contact SICK Customer Service.
b
Use the device only as described in "Intended use". The manufacturer bears no
responsibility for any other use.
b
Carry out the specified maintenance work.
b
Do not attempt any work on or repairs to the device unless described in this Man‐
ual. Do not remove, add or change any components in or on the device unless
such changes are officially allowed and specified by the manufacturer. Failure to
observe these precautions could result in:
Voiding the manufacturer's warranty.
The device becoming dangerous.
b
Observe all relevant regulations concerning protective clothing.
2 SAFETY INFORMATION
8
O P E R A T I N G I N S T R U C T I O N S | GM901-02 8008932/Y153/3-0/2016-03 | SICK
Subject to change without notice
Electrical safety
WARNING
Endangerment of electrical safety during installation and maintenance work when the
power supply is not switched off
b
Before starting the work on the device, ensure the power supply can be switched
off using a power disconnector switch/circuit breaker in accordance with DIN EN
61010.
b
Make sure the power disconnector switch is easily accessible.
b
An additional disconnecting device is mandatory when the power disconnector
switch cannot be accessed or only with difficulty after installation of the device
connection.
b
The power supply may only be activated again after the end of the work or for test
purposes by personnel carrying out the work under consideration of valid safety
regulations.
WARNING
Endangerment of electrical safety through power cable with incorrect rating
Electrical accidents can occur when the specifications for replacement of a removable
power cable have not been adequately observed.
b
Always observe the exact specifications in the Operating Instructions (Technical
Data Section) when replacing a removable power cable.
DANGER
Danger of electrical accidents
Incorrect performance of electrical work could result in serious electrical accidents.
b
Only let the work described in the following be carried out by electricians familiar
with potential hazards.
Grounding the device
CAUTION
Device damage through incorrect or missing grounding
During installation and maintenance work, it must be ensured that the protective
grounding to the devices and/or lines involved is effective in accordance with EN
61010-1.
2.2 Intended use
2.2.1 Purpose of the device
The GM901-02 only serves to monitor CO concentrations of gases in industrial plants.
The GM901-02 measures continuously directly in the gas duct (in-situ).
SAFETY INFORMATION 2
8008932/Y153/3-0/2016-03 | SICK O P E R A T I N G I N S T R U C T I O N S | GM901-02
9
Subject to change without notice
2.2.2 Responsibility of user
Designated users
see "Target group", page 6.
Correct project planning
Basis of this Manual is the delivery of the device according to the preceding
project planning (e.g., based on the SICK application questionnaire) and the rele‐
vant delivery state of the device (see delivered system documentation).
w
Contact SICK Customer Service if you are not sure whether the device corre‐
sponds to the state defined during project planning or to the delivered system
documentation.
Correct use
b
Use the device only as described in "Intended use".
The manufacturer bears no responsibility for any other use.
b
Carry out the specified maintenance work.
b
Do not attempt any work on or repairs to the device unless described in this Man‐
ual.
Do not remove, add or change any components in or on the device unless such
changes are officially allowed and specified by the manufacturer.
Failure to observe these precautions could result in:
°
Voiding the manufacturer's warranty.
°
The device becoming dangerous.
Special local requirements
In addition to the information in these Operating Instructions, follow all local laws, tech‐
nical rules and company-internal operating directives applicable wherever the device is
installed.
Read the Operating Instructions
b
Read and observe these Operating Instructions.
b
Observe all safety instructions.
b
If anything is not clear: Please contact SICK Customer Service.
Retain documents
These Operating Instructions must be
b
Available for reference.
b
Passed on to new owners.
2 SAFETY INFORMATION
10
O P E R A T I N G I N S T R U C T I O N S | GM901-02 8008932/Y153/3-0/2016-03 | SICK
Subject to change without notice
3 Product description
3.1 Product identification
Product name GM901-02
Device version CO measuring device with probe
Manufacturer SICK AG Erwin-Sick-Str. 1 • D-79183 Waldkirch • Germany
Type plates
Sender and receiver: Under the optics tube
Measuring probe: On the electronics housing
Table 3: Product identification
3.2 Product features
The in-situ gas analyzer GM901-02 serves for continuous measurement of gas
concentrations in industrial plants.
The GM901-02 is an in-situ measuring system which means measuring is done
directly in the gas carrying duct.
Measuring components: CO and reference value temperature.
Measuring principle: Infrared spectroscopic gas filter correlation.
PRODUCT DESCRIPTION 3
8008932/Y153/3-0/2016-03 | SICK O P E R A T I N G I N S T R U C T I O N S | GM901-02
11
Subject to change without notice
3.3 GM901-02 layout
Enter meas
diag par cal maint
Operation
Service
Warning
Malfunction
GM 901
Evaluation Unit
1
2
3
4
5
6
á
8
9
7
ß à
â ã ä
å
Figure 1: Device component overview
System components Line
1
GM901-02 sender
9
CAN connection, sender-receiver
2
GM901-02 receiver
ß
Signal line, GM901 receiver to connection unit
3
T-piece with beam splitter
à
Power supply 115 V/230 V AC
4
Measuring probe
á
Line to temperature sensor PT1000
5
Temperature sensor PT1000
â
Power supply 115 V/230 V AC
6
Connection unit (optional) for distance analyzer - EvU >
20 m
ã
Signal cable, max. 1000 m (CAN bus and temperature)
7
Evaluation unit
ä
Analog output signals: 1 input, 1 output
Status signals: 1 input, 3 outputs
8
Service PC
å
RS232 interface (Service)
Table 4: GM901-02 layout
NOTE
If the evaluation unit is installed detached, the 20 mA converter must be installed near the device.
3 PRODUCT DESCRIPTION
12
O P E R A T I N G I N S T R U C T I O N S | GM901-02 8008932/Y153/3-0/2016-03 | SICK
Subject to change without notice
3.4 Measuring probes
Gas diffusion probe (GPP)
Figure 2: GPP measuring probe with ceramic filter
Characteristics:
Suitable for applications with dry process gas
Measuring probe material:
Standard: Stainless steel (1.4571/316 Ti)
On request: 1.4539
No moving parts
Active elements: Heaters of the optical surfaces and their controls
Temperature sensor PT1000
Power supply via a plug
Filter element for filtering dust particles
°
With ceramic filter: Dust removal for particles > 1 μm
EPA compliant (EPA Guideline CFR 40, Part 60 or Part 75): The measuring cham‐
ber can be filled with known concentrations of test gases
SPAN and zero point test possible: Usage as "zero path" by purging the measuring
chamber with air or N
2
possible
Pressure-resistant up to 200 mbar differential pressure to duct
No special flow conditions in duct necessary
PRODUCT DESCRIPTION
3
8008932/Y153/3-0/2016-03 | SICK O P E R A T I N G I N S T R U C T I O N S | GM901-02
13
Subject to change without notice
GMP measuring probe - probe with open measuring gap
Figure 3: GMP measuring probe with open measuring gap and PT1000 tem‐
perature sensor
Characteristics:
Very short response times
High temperature stability
Continuous purge air necessary
Air outlet in the duct 90° to gas flow
Temperature sensor PT 1000 for continuous measurement of medium tempera‐
tures in the active measuring path of the probe
Gas inlet and outlet with the GMP measuring probe
The GMP probe has a closing device towards the opening for sample gas and is oper‐
ated using the lever on the probe flange. This ensures a correct purge air function with‐
out an analyzer.
1
2
3
4
Figure 4: GMP measuring probe closing device
1
Purge air outlet
2
Open measuring path (measuring gap)
3
Closing device lever
4
Closing device set to position "open"
3 PRODUCT DESCRIPTION
14
O P E R A T I N G I N S T R U C T I O N S | GM901-02 8008932/Y153/3-0/2016-03 | SICK
Subject to change without notice
GM901-02 with GMP probe and purge air unit
1
2
3
4
6
5
7
Figure 5: GMP measuring probe with purge air unit on the duct
1
GM901-02 sender
2
GM901-02 receiver
3
T-piece with beam splitter
4
Lever for closing device of the sample gas opening
5
GMP measuring probe
6
Purge air hose
7
Purge air unit SLV4
Measuring probes in comparison
Measuring probe feature GMP (open probe) GPP (dry)
Type Measuring path open in flow
direction; purge air guidance
with outlet aligned 90° to gas
flow
Gas diffusion probe with
ceramic filter, for dry sample
gas
Max. gas temperature 430 °C 430 °C
Gas check according to EPA
specification possible
- Yes
Purge air supply required Yes No
Heating on optical interfaces
in the probe
- Yes, with integrated control
Flow rate of sample gas 1 ... 40 m/s < 40 m/s
Suitable for wet sample gas Yes No
Maximum duct pressure ±120 hPa, depending on
purge air supply
±200 hPa with safety clamps
on the device flange seal
Measurable components CO
System response time (T90) ≥ 5 s ≥ 120 s
Duct diameter > 360 mm > 300 mm
Table 5: GMP and GPP measuring probes in comparison
PRODUCT DESCRIPTION 3
8008932/Y153/3-0/2016-03 | SICK O P E R A T I N G I N S T R U C T I O N S | GM901-02
15
Subject to change without notice
Measuring probe feature GMP (open probe) GPP (dry)
Dust concentration < 2 g/m3 a.c. < 30 g/m3 a.c.
Probe lengths available [m] 1.1 1.1
Active measuring paths availa‐
ble [mm]
250/500
Table 5: GMP and GPP measuring probes in comparison
3.5 Evaluation unit
Evaluation unit with connection line (4 m)
The evaluation unit (EvU) serves in the measuring system as user interface and pre‐
pares and outputs the measured values and performs control and monitoring func‐
tions. The EvU can be located in the vicinity of the sender/receiver unit. It can also be
located up to about 1000 meters from the sampling point, e.g. installed in the switch
center or monitoring center, and performs the following functions:
Output of measured values, computed data and operating states
Communication with the peripheral equipment
Output of error messages and other status signals
Access during service (diagnosis)
3.6 Accessories
Connection unit
For CAN bus with 24 V power supply
To extend the distances 19 ... 1000 m between analyzer and evaluation unit
Weatherproof cover
The weatherproof cover is used when the measuring system is operated outdoors. It is
available as an accessory.
Figure 6: Weatherproof cover
3 PRODUCT DESCRIPTION
16
O P E R A T I N G I N S T R U C T I O N S | GM901-02 8008932/Y153/3-0/2016-03 | SICK
Subject to change without notice
Purge air unit
1
6
5
4
3
2
Figure 7: Purge air unit components
1
Low-pressure monitor
2
Air filter
3
Quick-release clamps
4
Purge air blower
5
Purge air hose
6
Hose clamps
The purge air unit serves to feed purge air when using the open measuring probe
(GMP). It supplies filtered ambient air to the purge air fixture and protects the optical
surfaces of the sender and the receiver against contamination.
DANGER
Further information on the purge air unit, see Operating Instructions of the purge air
unit.
Optical adjustment device
To align the sender and receiver
PRODUCT DESCRIPTION 3
8008932/Y153/3-0/2016-03 | SICK O P E R A T I N G I N S T R U C T I O N S | GM901-02
17
Subject to change without notice
Figure 8: Optical adjustment device
CO test cells with holder (SPAN test)
For the yearly drift control (zero point and sensitivity)
Figure 9: CO test cell with holder in transport case
Deliverable test cells
Test cells are available depending on the application-specific measuring ranges, meas‐
uring paths and test points (e.g. 70%).
3 PRODUCT DESCRIPTION
18
O P E R A T I N G I N S T R U C T I O N S | GM901-02 8008932/Y153/3-0/2016-03 | SICK
Subject to change without notice
4 Important information on GPP measuring probes
CAUTION
Observe the following information during commissioning of the GPP measuring probe
to avoid damage to the measuring system and to achieve stable measurements.
Check the filter surface visually for damage.
Warm the probe up for about 30 minutes beforehand so that the temperature of
the window and triple reflector is sufficient to prevent condensation on the optical
surfaces.
Gas can escape at the measuring probe during overpressure processes when the
1/4" plug is open or loose. Air is suctioned into the measuring probe during low-
pressure processes. The measured value in the chamber then drops depending on
the low-pressure in the process, possibly down to zero.
Do not damage the filter when inserting the probe in the flange.
IMPORTANT INFORMATION ON GPP MEASURING PROBES 4
8008932/Y153/3-0/2016-03 | SICK O P E R A T I N G I N S T R U C T I O N S | GM901-02
19
Subject to change without notice
5 Transport and storage
5.1 Transport safeguards
Check the transport safeguards
b
Check all transport safeguards for damage.
Remove the transport safeguards from the measuring probe (GMP)
Figure 10: Transport safeguards on the GMP probe
1
Protective stickers
2
Protective caps
1 Remove the protective stickers.
2 Remove the caps.
5
TRANSPORT AND STORAGE
20
O P E R A T I N G I N S T R U C T I O N S | GM901-02 8008932/Y153/3-0/2016-03 | SICK
Subject to change without notice
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98