  • Hello! I've reviewed the firmware update instructions for the MRU AMPROplus and OPTIMAX analyzers. This document guides you through the process of updating your device's firmware using an SD card. I am familiar with the steps and requirements for the update, such as the battery charge level and the use of an SD card. If you have any questions about the process or the device, feel free to ask!
  • What should the battery charge be before performing the update?
    What tools are needed for the firmware update?
    What to do if I receive a ZIP file?
AMPRO
plus
/ OPTIMAX
Firmware update
AMPRO
plus
/ OPTIMAX
Firmware update
PLEASE NOTE:
The battery should have more than 50% charge when performing this update.
MRU is not responsible for any malfunction or defects that occur during this procedure!
Tools needed:
SD Card
PC with SD card reader
You have received either a ZIP file or a file ending in FWB via. email
If you have received a ZIP file, then you need to extract (UNZIP) that file first
Copy the unzipped file onto the SD Card
AMPRO
plus
/ OPTIMAX
Firmware update
(1) Rubber cover
(2) IR Window for printer
(3) SD Card slot
(4) USB port
Power up your analyzer
Wait until the zeroing is completed
Insert the SD card with the new firmware into the SD card slot (3)
The analyzer will display that a new
firmware version is available
confirm with OK if you want to make
this update
The analyzer will shut down,
perform the update (condensate
LED is blinking), and then restart
again, and then display that the new
firmware has been installed
Confirm the message with OK, then restart the analyzer
PLEASE NOTE:
Please contact your MRU service department if you have any questions
AMPRO
plus
/ OPTIMAX
Firmware update