Thermo Fisher Scientific ID™ Self Pack ® Packing Owner's manual

Type
Owner's manual
Operating Instructions
1
In This Document
This document describes how to pack your own ImmunoDetection™
(ID™) Sensor Cartridges for use in Perfusion Immunoassays,
automated assays that selectively capture biomolecules in a rapid,
reproducible, flow-through format.
You can immobilize the ligand before or after packing the sensor
cartridge. If you are immobilizing ligand before packing, refer to the
Immobilizing Ligand for Perfusion Immunoassays document provided
before continuing.
Related documents
You receive two documents with the POROS® Self Pack® media:
Immobilizing Ligand for Perfusion ImmunoassaysRefer
to this document before packing the cartridge, if you want to
immobilize ligand on a batch of media. Refer to this document
after packing if you want to immobilize on the packed cartridge.
Performing Unlabeled Perfusion ImmunoassaysRefer to
this document after immobilizing ligand and packing the
cartridge.
1 Product Description
The ID Self Pack Packing Device for ImmunoDetection Sensor
Cartridges lets you pack and repack empty PEEK 2.1 mmD/30 mmL
cartridges with POROS affinity Self Pack media. By packing your own
ID cartridges, you get increased reproducibility between cartridges,
flexibility in immobilization chemistry, and unsurpassed affordability.
You purchase empty cartridges, ID Self Pack media, and reagents
separately from the packing device.
In These Instructions
These instructions describe how to use the packing device and related
hardware. They explain the following:
How to assemble the packing device
How to adapt the packing device to different LC instruments
How to pack your own cartridges
How to unpack a cartridge
Packing Device Parts
The Self Pack packing device is made of corrosion-resistant
316L stainless steel that can be used with salt buffers. The packing
device consists of these parts:
Cap (center cover and lock nut)
O-ring
Chamber
A backpressure regulator is included with the packing device. Figure 1
shows the parts of the packing device and the backpressure regulator.
Figure 1 Packing Device Parts
Flat surfaces on the packing device chamber allow you to grasp the
chamber with a wrench if necessary, but hand tightening is usually
sufficient.
PB1000362
Chamber
O-ring Cap
Backpressure regulator
Outlet End Inlet End
Lock
nut Center
cover
Section Page
1 Product Description...............................................................1
2 Additional Items Needed to Pack Your Cartridge..................2
3 Preparing Your Equipment for Packing.............................2
4 Packing a Cartridge.............................................................4
5 Running the Packed Cartridge...........................................6
6 Unpacking a Cartridge ........................................................6
7 Storing Packing Device and Cartridge..............................6
8 Accessories, Spare Parts, and Ordering Information ............6
9 Technical Support..................................................................7
ID Self Pack®
Packing Device
for
ImmunoDetection
Cartridges
2
The inlet end of the chamber connects to your LC instrument. It
includes the following:
A screw-on 2-piece cap with a 10-32 female fitting. You can
adapt this fitting to M-6 metric systems (such as the Pharmacia
FPLC® system) or 1/4-28 systems by using the fitting adaptor
kit supplied with the packing device.
An O-ring seated in a depression around the rim of the inlet
opening. The O-ring lets you hand-tighten the cap, even when
you use the packing device with high-pressure instruments.
The outlet end is the narrow end of the chamber, where:
You can connect an empty 2.1 mm diameter PEEK cartridge.
A white Teflon® washer recessed in the outlet end of the
packing chamber creates a proper seal when a cartridge is
screwed in place. (This washer is not visible in Figure 1.)
Fitting Adaptor Kit
The fitting adaptor kit adapts the Self Pack packing device to M-6
metric or 1/4-28 LC systems. (For ordering information, see Table 2 on
page 6.) The kit provides these fittings:
2 red Fingertight fittings
2 beige ferrules (barrel shaped)
2 black metric nuts
2 green nuts
2 blue ferrules (cone shaped)
2 pieces of orange 0.020-inch I.D. PEEK tubing
To use the fitting adaptor kit, do the following:
1. Connect one end of the orange tubing to an M-6 or 1/4-28 system.
For M-6 systems: Slip a black nut over the tubing followed by
a blue ferrule, with the conical end pointing toward the nut.
Connect to the LC system and finger tighten.
For 1/4-28 systems: Slip a green nut over the end of the
tubing, followed by a blue ferrule, with the conical end pointing
away from the nut. Connect to the LC system and finger
tighten.
2. Put a red Fingertight fitting on the other end of the tubing, followed
by a beige ferrule.
The fitting is now ready to be connected to the 10-32 fitting on the
cap of the packing device. Because the cap is heavy, do not attach
it now. Wait until you are ready to attach the cap to the chamber
(see Section 4.2, Filling the Packing Device).
2 Additional Items Needed to Pack Your
Cartridge
In addition to the packing device, to pack a cartridge you need an
empty PEEK cartridge, POROS Self Pack media (purchased
separately), and your own ligand for immobilization.
You can order this equipment from PerSeptive Biosystems. For
ordering information, see Section 8, Accessories, Spare Parts, and
Ordering Information.
PEEK Cartridges
Empty PEEK ID cartridges are sold in packages of 5, 10, or 20. Each
empty PEEK cartridge comes with these parts:
One hollow tube
Two end fittings, which screw onto threads at either end of the
cartridge tube
Two frits that are placed in the end fittings to retain the media
Two end plugs
Two EZ Grip fittings and ferrules that are used to connect the
packed cartridge to a system
End fittings are assembled on the tube, and the end plugs and frits are
installed.
Figure 2 shows a PEEK cartridge without end plugs.
Figure 2 PEEK Cartridge Parts
3 Preparing Your Equipment for Packing
To pack a cartridge, do the following:
Clean the cartridge and packing device
Prepare the slurry solvent and packing solvent
Prepare the packing device
Prepare your LC system
Assemble the packing device and empty cartridge
Position the packing device
Lab Equipment Needed
In addition to the packing device and empty cartridge, you need this
equipment to pack a cartridge:
LC system
Spatula
Glass beaker of 50 ml or greater
Squirt bottle
Laboratory tissue
25 ml graduated cylinder or 20 ml pipette
Deionized water
3.1 Cleaning the Cartridge and Packing Device
NOTE: Observe good laboratory practice. Wear safety glasses during
the following procedures.
The cartridge and packing device are reusable. You clean them
differently depending on whether you have used them before.
Never Used
If the cartridge and packing device have never been used, do the
following:
Cartridge: Unscrew the end fittings and flush the tube and end
fittings with deionized water. Do not reassemble the cartridge.
Set all the pieces aside.
Packing device: Wash or sonicate the packing chamber and
cap with 50% isopropanol, ethanol, or other alcohol to ensure
cleanliness. Set all the pieces aside.
PB100365
End fitting
Frit
Tube
End plug
3
Previously Used
If the cartridge or packing device has been used before, clean it before
each reuse to avoid contamination with packing material left over from
a previous use.
Cartridge: Sonicate the tube, end fittings, and frits in 50%
isopropanol, ethanol, or other alcohol for 10 minutes. Rinse all
pieces thoroughly with deionized water. Set all the pieces
aside.
Packing device: Flush the packing device cap with deionized
water and thoroughly wash the threads at the outlet end of the
chamber. Residual media gets trapped in the threads during
normal use. Wash the O-ring. Flush with organic solvent if you
see any media residue on surfaces after the water wash. Rinse
again with water to remove the organic solvent. Do not
reassemble the packing device. Set all the pieces aside.
3.2 Preparing the Slurry Solvent and
Packing Solvent
To prepare the slurry solvent and packing solvent, do the following:
1. Prepare your slurry solvent and packing solvent (Table 1).
2. Fill the squirt bottle with the slurry solvent.
3.3 Preparing the Packing Device
Make sure the white Teflon washer is installed in the outlet end of the
chamber. The washer is required for a proper seal to pack the
cartridge.
3.4 Preparing Your LC System
To prepare your LC system, do the following:
Prepare the tubing
Set the pressure limit
Prepare the pumps
Preparing the Tubing
Make sure that the fitting on the end of your LC system pump outlet
tubing is compatible with the fitting on the packing device cap. If it is
not, use the fitting adaptor kit. See Fitting Adaptor Kit on page 2.
Do not connect the free end of the pump outlet tubing to the cap of the
packing device yet.
For example, on an FPLC system, do the following:
1. Connect one end of the orange tubing of the fitting adaptor kit to
the outlet of the FPLC mixing chamber with the M-6 black nut. See
Fitting Adaptor Kit on page 2.
2. Reserve the free end of the orange tubing for the red Fingertight
fitting that will connect into the cap of the packing device.
Setting the Pressure Limit
For Perfusion Chromatography or conventional HPLC systems:
Set the pressure limit of your system at 3,000 psi.
For medium-pressure systems: Set the pressure limit at the
maximum allowable on the system. For example, on an FPLC system,
set the pressure limit at 4.0 MPa.
Preparing the Pumps
Flush the pumps and the outlet tubing with the packing solvent listed in
Table 1.
NOTE: On an FPLC system, you must flush both pumps with packing
solvent because both pumps are used during packing.
3.5 Assembling the Packing Device and
Empty Cartridge
To assemble the packing device and cartridge for packing, do the
following:
1. Locate the equipment needed for packing. See Lab Equipment
Needed on page 2.
2. Write the name of your POROS Self Pack media on one of the
blank labels provided with the media.
These labels identify the cartridges and also identify the top of the
cartridge (the end connected to the packing chamber). Use the top
as the inlet end. Attach the label to an end fitting. Reserve this
fitting to be the top end fitting.
3. Set the cartridge end fittings on the countertop so that you can look
down into them.
4. Drop a frit into each end fitting.
5. Pick up one end fitting containing a frit.
6. Screw the cartridge tube into the end fitting until finger-tight.
7. Tighten the end fitting a little more: grip the cartridge tube with one
hand and the end fitting in the other hand and tighten securely.
8. Screw the free end of the cartridge tube into the outlet end of the
packing chamber. Tighten securely with two hands.
9. Set the unused end fitting aside.
10. Screw the backpressure regulator into the cartridge (see Figure 3
on page 4).
3.6 Positioning the Packing Device
1. Clamp the open packing device/cartridge assembly on the front of
your LC instrument. If clamps are not available on the instrument,
use a ring stand. Make sure that the following conditions are met:
The packing device is positioned vertically.
The pump outlet tubing can reach the top of the packing
device.
You can fit a 50 ml beaker under the end of the cartridge.
2. Position your empty beaker below the end of the cartridge.
3. Seat the packing chamber O-ring in the depression around the rim
of the chamber opening (shown in Figure 1 on page 1).
Table 1 Slurry Solvents and Packing Solvents for
Self Pack Media
POROS Self Pack Media Slurry Solvent Packing Solvent
AD 100 % ethanol 5 % ethanol
BA, EP, AL, XL, PG, PA Water Water
4
4 Packing a Cartridge
The packing operation consists of five steps:
Slurrying the media
Filling the packing device
Packing the cartridge
Removing and capping the packed cartridge
Cleaning up
4.1 Slurrying the Media
1. Determine the weight or volume of media to add to the packing
device (100 mg of dry media or 0.2 ml of wet settled media per
cartridge).
NOTE: With wet media of unknown volume, allow it to settle in a
graduated test tube. Note the total volume (Vt) and the volume of
settled media (Vm). Calculate the volume in ml of resuspended
media needed using the following equation:
V = 0.2 x (Vt/Vm)
2. Add 5 ml of the appropriate slurry solvent (see Table 1 on page 3)
to the media and mix.
4.2 Filling the Packing Device
Caution: Do not proceed without setting your system pressure as
described in Section 3.4, Preparing Your LC System.
To fill the packing device, do the following:
1. Add the media slurry into the open packing device.
2. Rinse the pipette or cylinder with a small amount of slurry solvent
to ensure that all of the slurry is added to the packing device.
3. Add slurry solvent to the packing device until it is completely full.
NOTE: Do not stir up the slurry at the bottom of the packing device.
Direct the slurry solvent so that it runs down the inner wall of the
packing device, instead of pouring directly on the slurry.
4. Quickly screw the 2-piece cap onto the packing device: place the
center cover on the chamber, thread the lock nut down over the
cap and hand tighten until snug. Use a lab tissue to wipe any liquid
that comes out through the top as you tighten the cap.
NOTE: Do not overtighten the cap. Overtightening can damage the
threads of the chamber and the cap.
Figure 3 shows a packing device setup on a BioCAD®/SPRINT
System.
Figure 3 Typical Packing Device Setup
5. Attach the pump outlet tubing from the LC system to the cap of the
packing device. Finger tighten the fitting.
4.3 Packing the Cartridge
This section discusses these topics:
Principles of Packing
Packing a Cartridge Using the INTEGRAL® 100Q System,
VISION or BioCAD Workstation, BioCAD/SPRINT System,
BioCAD/RPM® System, or FPLC system
Packing a Cartridge Using Other Systems
Principles of Packing
As a cartridge packs, the backpressure on the cartridge rises.
The optimal packing flow rate is the highest flow rate possible that
allows the following:
Packing to proceed within the pressure limit you set earlier
A sufficient volume of packing solvent to pass through the
cartridge to ensure an efficient and stable bed
The actual packing flow rate is dependent on the system used, the
particle size of the media, and the length of the cartridge.
PB1000367
Packing chamber
Cap
PEEK cartridge
Backpressure
regulator
5
Packing a Cartridge Using the INTEGRAL 100Q System,
VISION or BioCAD Workstation, BioCAD/SPRINT System,
BioCAD RPM Workstation, or FPLC System
NOTE: Monitor the highest sustained pressure that occurs during
packing. Do not exceed this pressure when you run the packed
cartridge.
To pack a cartridge, do the following:
1. Use a total volume of at least 25 ml packing solvent to pack the
cartridge.
2. Set the flow rate for packing the cartridge at 9 ml/min.
3. Turn on the flow and monitor the volume of packing solvent in the
waste beaker below the cartridge. If the pressure is below the
system limit increase the flow rate until a pressure of 2,000 psi is
attained, or the maximum system flow rate is reached, whichever
comes first.
4. Once at least 25 ml of packing solvent has passed through the
cartridge and a constant pressure has been reached, stop the flow.
5. Let the cartridge sit for 3 minutes to allow the pressure to dissipate
evenly across the packed bed.
NOTE: The packing flow rate has been determined to allow you to
pack your cartridge within the pressure limit of your system.
However, variations in frit permeability or system backpressure
may require you to adjust the flow rate slightly.
6. If your system is well below the pressure limit during packing,
increase the flow rate in 0.5 ml/min increments until you approach
the pressure limit. Pack cartridges at the highest flow rate possible
without exceeding the system pressure limit.
7. If you reach the pressure limit before you finish packing the
cartridge, stop the flow, restart the flow at 0.5 ml/min lower than the
original setting, and continue packing. Repeat if necessary until
you are able to pack the cartridge within the pressure limit.
NOTE: Occasionally a defective frit generates excessive
backpressure. If so, replace both frits and start again.
8. Record the highest sustained pressure that occurred during
packing.
Do not exceed this pressure when you run the packed cartridge.
Packing a Cartridge Using Other Systems
NOTE: Monitor the highest sustained pressure that occurs as you pack
the cartridge. Do not exceed this pressure when you run the packed
cartridge.
If you are using systems other than the INTEGRAL Micro Analytical
Workstation, VISION or BioCAD Workstation, BioCAD/SPRINT
System, BioCAD RPM System, or FPLC system, you need to
experiment to find the correct flow rate. Start by setting the flow at
either 9 ml/min or the maximum allowable on the system, whichever is
lower.
When using other systems to pack the cartridge, note the following:
1. Use a total volume of at least 25 ml packing solvent to pack the
cartridge.
2. Turn on the flow and monitor the volume of packing solvent coming
through the cartridge by observing the level in the waste beaker
below the cartridge.
3. If you reach the pressure limit before the required volume of
packing solvent passes through the cartridge, stop the flow, restart
the flow rate at 0.5 ml/min lower than the original setting, and
continue packing.
4. If the pressure limit is again exceeded, lower the flow rate another
0.5 ml/min.
5. Continue lowering the flow rate in 0.5 ml/min increments until at
least 25 ml of packing solvent passes through the cartridge and
you reach a constant pressure; then, stop the flow.
NOTE: Occasionally a defective frit generates excessive
backpressure. If so, replace the frits and start again.
6. Let the cartridge sit for 3 minutes to allow the pressure to dissipate
evenly across the packed bed.
7. Record the final flow rate. Use this flow rate the next time you pack
a cartridge with your system.
8. Record the highest sustained pressure that occurred during
packing.
Do not exceed this pressure when you run the packed cartridge.
4.4 Removing and Capping the Packed Cartridge
To remove the cartridge, do the following:
1. Remove the backpressure regulator from the cartridge.
2. Loosen the end fitting on the cap of the packing device to release
the pressure in the cartridge.
3. Unscrew the cartridge from the packing device. Grasp the bottom
end fitting of the cartridge. (Turn slowly at first to make sure the
end fitting does not unscrew from the cartridge body.) Turn the
cartridge to the left to unscrew it.
A small bubble of excess packing material may cover the top end
of the cartridge tube. If the material is wet, lightly touch the bubble
with a lab tissue or paper towel to draw off the moisture.
4. Draw the flat side of the spatula across the top of the cartridge to
remove the excess material and to level off the top of the packed
bed.
5. With a damp tissue, carefully wipe any remaining packing material
from around the rim at the top of the cartridge tube. Do not touch
the top of the packed bed itself.
6. Holding the unused end fitting upright to ensure that the frit remains
seated, turn the packed cartridge upside down and screw it into the
end fitting. Keep track of the cartridge inlet end.
7. Tighten the end fitting with both hands.
8. Insert the end plugs into the cartridge end fittings.
9. Set the cartridge aside while you clean up from the packing
procedure.
6
4.5 Cleaning Up
To clean up, do the following:
1. Make sure an empty waste beaker is below the cartridge packing
device.
2. Pump several milliliters of packing solvent through the packing
device to push out excess packing material accumulated at the
outlet end of the chamber.
3. Remove the connection at the top of the packing device.
The pressure inside the packing device is released and the entire
contents flush into the beaker.
4. Remove the packing device from the clamp, take off the cap, and
clean all pieces thoroughly with water.
5. Rinse the threads at the outlet end to remove residual media.
NOTE: The O-ring may stick to the cap of the packing device by
surface tension. Use the spatula to pull it out.
5 Running the Packed Cartridge
For information about running the packed cartridge, see the Performing
Unlabeled Perfusion Immunoassays document provided.
6 Unpacking a Cartridge
To unpack a cartridge, do the following:
1. Unscrew one end fitting and remove the frit.
2. Connect the other end of the cartridge to the pump outlet tubing of
your LC system.
3. Position a beaker under the cartridge.
4. Start the flow rate at 5 ml/min.
5. As soon as the media is pushed out of the cartridge into the
beaker, stop the flow.
6. Remove the remaining end fitting.
7. Sonicate the tube, end fittings, and frits in 50% isopropanol,
ethanol, or other alcohol for 10 minutes. Rinse all pieces
thoroughly with deionized water.
7 Storing Packing Device and Cartridge
Cartridge: To store a packed cartridge, see the cartridge
Operating Instructions sheet included with your ID Self Pack
media.
To store an unpacked cartridge, keep the tube, end fittings,
frits, and end plugs together.
Packing Device: Store the packing device with the cap on to
keep the threads from knocking against other equipment. If the
threads are damaged, the device may not seal properly during
packing. Also, damaged threads may cause the cap to lock in
place irreversibly. Ideally, store the capped device in its original
box.
8 Accessories, Spare Parts, and Ordering
Information
Table 2 Spare Parts and Accessories
Description Quantity Part Number
ID Self Pack Packing Device 1 ea 1-9544-00
Chemistry Trial Kit
AD, AL, BA, EP, and XL Self
Pack Media, with reagents and
5 empty cartridges
1 kit 2-3011-00
ID Sensor Cartridge Kits (cartridges with reagents)
AD ID Sensor Kit 1 kit 2-3006-00
AL ID Sensor Kit 1 kit 2-3010-00
BA ID Sensor Kit 1 kit 2-3005-00
EP ID Sensor Kit 1 kit 2-3003-00
PA ID Sensor Kit 1 kit 2-3001-00
PG ID Sensor Kit 1 kit 2-3002-00
XL ID Sensor Kit 1 kit 2-3004-00
ID Sensor Cartridges (cartridges only)
AD ID Sensor Cartridge 1 cartridge 2-1007-00
AL ID Sensor Cartridge 1 cartridge 2-1010-00
BA ID Sensor Cartridge 1 cartridge 2-1005-00
EP ID Sensor Cartridge 1 cartridge 2-1003-00
PA ID Sensor Cartridge 1 cartridge 2-1001-00
PG ID Sensor Cartridge 1 cartridge 2-1002-00
XL ID Sensor Cartridge 1 cartridge 2-1004-00
Self Pack Media
AD (Adsorbed) 500 mg 2-3114-00
AL (Aldehyde) 500 mg 2-3115-00
BA (Streptavidin) 1 ml 2-3111-00
EP (Epoxide) 500 mg 2-3116-00
PA (Protein A) 1 ml 2-3117-00
PG (Protein G) 1 ml 2-3118-00
XL (Cross-linked Protein G) 1 ml 2-3119-00
Reagents
Loading Buffer 1 pack 2-2101-00
Elution Buffer 1 pack 2-2102-00
AD ID Reagent Pack 1 set 2-3120-00
AL ID Reagent Pack 1 set 2-3121-00
EP ID Reagent Pack 1 set 2-3122-00
XL ID Reagent Pack 1 set 2-3124-00
Empty Cartridges
2.1 mmD/30 mmL PEEK Pkg of 5 2-3008-05
2.1 mmD/30 mmL PEEK Pkg of 10 2-3008-10
2.1 mmD/30 mmL PEEK Pkg of 20 2-3008-20
Accessories, Packing Device
Fitting Adaptor Kit 1 ea 1-9532-00
O-Ring 1 ea 1-9108-00
7
9 Technical Support
PerSeptive Biosystems is dedicated to helping you use Perfusion
Chromatography and ID media to the fullest extent possible. Our
biochromatographers, bioprocess engineers, and applications
development laboratories are available for support ranging from
telephone consultation to full method development.
PerSeptive Biosystems offers a full line of ID media for Perfusion
Chromatography in the reversed-phase, ion exchange, affinity, and
other chromatographic modes. Please contact your PerSeptive
Biosystems representative for technical and ordering information.
PerSeptive Biosystems publishes a continuing series of Application
and Technical Notes, highlighting specific purification solutions and
technical aspects of Perfusion Chromatography. Please contact
PerSeptive Biosystems directly for a publication list.
For further details or for answers to questions on ID Self Pack,
Perfusion Chromatography, or other products, please contact
PerSeptive Biosystems. Refer to the back cover of this document for
the number of the appropriate office.
Accessories, Cartridge
Frits, 2.1 mm PEEK Pkg of 5 1-9124-05
EZ Grip Fittings (SS) Pkg of 5 P5-1011-05
Table 2 Spare Parts and Accessories (Continued)
Description Quantity Part Number
500 Old Connecticut Path
Framingham, MA 01701
Tel: (508) 383-7700
Tel: (800) 899-5858
Center of Excellence
www.pbio.com
Headquarters
850 Lincoln Centre Drive
Foster City, CA 94404
Tel: (650) 638-5800
Tel: (800) 345-5224
www.pebiosystems.com
For international office locations, please call the headquarters or refer to our web site at www.pebiosystems.com.
PerSeptive Biosystems
PE Biosystems
PerSeptive Biosystems, POROS, Perfusion Chromatography, BioCAD, INTEGRAL, RPM, Self Pack, and the fractal icon are registered trademarks of PerSeptive
Biosystems, Inc. ID, ImmunoDetection, SPRINT, and VISION are trademarks of PerSeptive Biosystems, Inc.
Luer-Lok is a registered trademark of Becton-Dickinson and Company. Tris is a registered trademark of Rohm and Haas Company. FPLC is a registered trademark of
Pharmacia Biotech AB. Teflon is a registered trademark of E. I. Du Pont de Nemours. EZ is a trademark of Scientific Systems, Inc.
POROS® products and perfusive chromatography are covered by US patents 5,019,270; 5,030,352; 5,228,989; 5,384,042; 5,552,041; 5,605,623; and 5,833,861; issued to
PerSeptive Biosystems, Inc.; foreign patents pending. Chromatography materials of certain pore geometries, including wide-pore supports, can permit perfusive
chromatography, that is, separations where the rate of intraparticle convection exceeds the rate of diffusion. Use of any such chromatography media that permit
performance at high linear velocity, without license, constitutes infringement of US patent 5,019,270 and US patent 5,228,989. A limited license to use the patented
perfusive chromatography process is granted with the purchase of POROS® products from PerSeptive Biosystems. The license may not be transferred except with the
transfer of POROS® columns or media, and terminates upon expiration of the useful life of the product.
Part Number 8-0122-40-0795, Rev. 2, 3/00 ©2000 Printed In USA Printed on recycled paper.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8

Thermo Fisher Scientific ID™ Self Pack ® Packing Owner's manual

Type
Owner's manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI