Vaisala Multichannel User Interface User manual

  • Hello! I am an AI chatbot trained to assist you with the Vaisala Multichannel User Interface User manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
IM-EN-MI-C
Instruction Manual
Multichannel User Interface
MI
Document/Revision No. IM-EN-MI/C Eecve: June 5, 2020
PUBLISHED BY
Vaisala Oyj
Vanha Nurmijärvene 21, FI-01670 Vantaa, Finland
P.O. Box 26, FI-00421 Helsinki, Finland
+358 9 8949 1
Visit our Internet pages at www.vaisala.com.
General safety consideraons
Always power o the Mulchannel user interface MI before accessing the modules
behind the display. Vaisala recommends the installaon of an external power switch.
Warning! If this device is used against manufacturer’s instrucons, user safety may be
compromised.
Warranty
For standard warranty terms and condions, see www.vaisala.com/warranty.
Please observe that any such warranty may not be valid in case of damage due to nor-
mal wear and tear, exceponal operang condions, negligent handling or installaon,
or unauthorized modicaons. Please see the applicable supply contract or Condions
of Sale for details of the warranty for each product.
Technical support
Contact Vaisala technical support at helpdesk@vaisala.com. Provide at least the fol-
lowing supporng informaon:
• Product name, model, and serial number
• Name and locaon of the installaon site
• Name and contact informaon of a technical person who can provide further infor-
maon of the problem
For more informaon, see www.vaisala.com/support.
© Copyright Vaisala 2020. All rights reserved.
Disposal
When wishing to dispose of an obsolete instrument or any parts of an in-
strument, please observe local and naonal regulaons and requirements
for the disposal of electrical and electronic equipment.
Symbols and terms used in this manual:
This indicates a warning. It provides safety precauon informaon needed to
avoid injury while operang the refractometer system.
This indicates that something is important for the operaon of the refractome-
ter system.
Note. Notes contain addional informaon and hints.
This product manual is delivered to the end user with a Vaisala K-PATENTS® product.
Informaon in this manual is subject to change without noce. When the manual is
changed, a revised copy is published at hp://www.kpatents.com/
Document/Revision No. IM-EN-MI/C Eecve: June 5, 2020
Table of contents
1 Descripon . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
2 Mounng . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
3 Connecons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3.1 Connecng the power cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3.1.1 Power terminals for AC power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3.1.2 Power terminals for DC power . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3.1.3 External mains switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3.2 Refractometer cable connecons . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.3 Ethernet cable connecon . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.4 Module cards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.5 Connecons for external control of Recipe app . . . . . . . . . . . . . . . . 13
4 Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4.1 Main display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4.2 Acvang and removing apps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
4.3 Managing users and passwords . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
4.4 System log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
4.5 Ethernet connecons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
4.5.1 MI ethernet connecon . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
4.5.2 Refractometer connecon . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
4.5.3 Modbus/TCP connecon . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
4.5.4 Ethernet/IP connecon . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
5 Apps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
5.1 Shared funconality . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
5.1.1 Applicaon trend . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
5.1.2 Applicaon log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
5.2 Refractometer app . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
5.2.1 Diagnoscs display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
5.2.2 Viewing and adjusng parameters . . . . . . . . . . . . . . . . . . . . . . . . . 37
5.2.3 Refractometer vericaon . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
5.2.4 Collecng eld sample points . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
5.3 Recipe app . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
5.3.1 Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
5.3.2 Acvang Recipe app . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
5.3.3 Creang a new recipe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
5.3.4 Modifying a recipe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
5.3.5 Selecng recipe switching mode . . . . . . . . . . . . . . . . . . . . . . . . . . 45
5.3.5.1 Switching recipe manually . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
5.3.5.2 Recipe idencaon numbers for remote control . . . . . . . . . . . 46
5.3.5.3 Switching recipe via General I/O module inputs . . . . . . . . . . . . 47
5.3.5.4 Switching recipe via Modbus/TCP . . . . . . . . . . . . . . . . . . . . . . . 48
5.3.5.5 Switching recipe via Ethernet/IP . . . . . . . . . . . . . . . . . . . . . . . . 48
5.3.6 Using Recipe app to make unit conversions . . . . . . . . . . . . . . . . . 48
5.3.6.1 Creang a recipe for unit conversion . . . . . . . . . . . . . . . . . . . . . 48
5.3.6.2 Displaying several unit conversions simultaneously . . . . . . . . . 49
5.3.7 Conguring mA output for Recipe app . . . . . . . . . . . . . . . . . . . . . 49
6 Module cards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
6.1 Installaon . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
6.2 Conguraon . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
6.3 Relay module card PR-50023 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
6.3.1 Conguring relays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
6.3.2 Nocaon of limit exceeding . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
6.4 mA output module PR-50021 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
6.4.1 Conguring mA outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
6.5 General I/O module PR-50024 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
6.5.1 Connecng inputs/outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
6.5.2 Conguring the general I/O module card . . . . . . . . . . . . . . . . . . . 66
6.5.2.1 Using the general I/O module to trigger prism wash . . . . . . . . 67
7 Prism wash . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
7.1 Conguring prism wash . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
7.1.1 Wash cycle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
7.1.2 Seng prism wash parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
8 Specicaons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
8.1 Model code . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
8.2 Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
A EU declaraon of conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
© Copyright Vaisala 2020. All rights reserved.
B Soware licence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
C Command selecon tree for Refractometer applicaon . . . . . . . . . . . . . . . . 83
D Command selecon tree for sengs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
MI instrucon manual
Document/Revision No. IM-EN-MI/C Eecve: June 5, 2020
1 Descripon 1
© Copyright Vaisala 2020. All rights reserved.
1 Descripon
The Mulchannel user interface MI is an industrial computer designed to add funcon-
ality to the PR-43 refractometers. The MI sends and receives data and it can also be
used to control the process.
The Mulchannel user interface MI connects with up to 4 Vaisala K-PATENTS® PR-43
refractometers. With module cards, it can also be used to add external inputs and
outputs (e.g. relays and mA outputs) to your refractometer system.
The MI is operated by the touch-screen display. The stainless steel enclosure gives a
good environmental protecon and allows also outdoor installaons, especially with
the included rain cover. The outer door is easy to open by one hand and the handle
can be modied for padlock to prevent unauthorized access. The module cards can be
reached by opening the inner door. The MI has IP66 rang when the outer door is open
if the inner door (display panel) is xed. With outer door closed the rang is IP67.
Figure 1.1 Mulchannel user interface MI
2 MI instrucon manual
Document/Revision No. IM-EN-MI/C Eecve: June 5, 2020
2 Mounng
The Mulchannel user interface MI can be mounted on a wall with wall mounng or
on a table with a table stand. The mounng locaon should be easily accessible. If
mounted outdoors, the MI has to be protected from rain and direct sunlight.
Display on eye-level
External
mains
switch
100-240VAC4-20 mA
Ethernet
1-4 refractometers
Figure 2.1 Mounng guide
2 Mounng 3
© Copyright Vaisala 2020. All rights reserved.
Figure 2.2 Enclosure dimensions
4 MI instrucon manual
Document/Revision No. IM-EN-MI/C Eecve: June 5, 2020
for refractometers
Figure 2.3 Enclosure dimensions
2 Mounng 5
© Copyright Vaisala 2020. All rights reserved.
1
1. Bracket screwed to wall by 4 screws.
Screw hole dimensions 325x254mm
325 mm (12.8 inches)
254mm
(10 inches)
1
1
1
Figure 2.4 Mounng the wall bracket
1. Four mounting screws
loosely attached
2. Top screws inserted into top bracket verticalmounting slots
3. Lower screws slide into bottom bracket horizontal mounting slots
4. Tighten mounting screws x4
width a 19 mm ring spanner
1
2
3
4
Figure 2.5 Mounng the MI onto the wall bracket
Note: Be careful when mounng, the MI weighs 13 kilos (28.5 lbs).
6 MI instrucon manual
Document/Revision No. IM-EN-MI/C Eecve: June 5, 2020
Important: Do not drill mounng holes in the enclosure as that will aect the protec-
on class of the enclosure and damage the electronics.
The MI is shipped with a rain cover. Vaisala recommends that it is installed even indoors
to protect the top of the door from dust and dirt.
1. Screw out top attaching screws to allow
clearance for rain cover to slot in place
1
2. Retighten screws
2
Figure 2.6 Mounng the rain cover
The table stand is very simple to mount.
Figure 2.7 Table stand
2 Mounng 7
© Copyright Vaisala 2020. All rights reserved.
The MI door handle can be modied to make space for a padlock to prevent unautho-
rized access.
Remove handle center cover
and remove segment indicated
by break-line on the back of cover
Replace center cover once
indicated segment has been
removed, making sure the opening
aligns with the channel for the padlock shackle
Figure 2.8 Locking the MI
8 MI instrucon manual
Document/Revision No. IM-EN-MI/C Eecve: June 5, 2020
3 Connecons
Figure 3.1 Mulchannel user interface connecons
3 Connecons 9
© Copyright Vaisala 2020. All rights reserved.
3.1 Connecng the power cable
Warning! Make sure there is no power in your cable. Then open the enclosure and
the inner door. You need a TX25 screwdriver to open the inner door retaining screws.
1
2
3
1.Push door gently
2.Lift handle 45 deg up
3.Open the door
Figure 3.2 Opening the enclosure
Remove inner door retaining screws
and open inner door
Figure 3.3 Opening the inner door
Unscrew one of the cable glands and push the power cable through. Connect the power
cable to the power terminals at lower right of the motherboard (see below). Screw on
the cable gland. Close the inner door and turn power on.
3.1.1 Power terminals for AC power
The primary AC power is connected to a separate terminal strip 1/2/3 in the lower right-
hand corner of the motherboard. The three terminals are marked 1.L, 2.N and 3
(protecve earth). The power terminal 3/ is directly connected to the exposed metal
parts of the MI. Wiring to the terminals shall be 1.5mm
2
minimum. The protecve fuse
in the building system shall comply with the local requirements.
10 MI instrucon manual
Document/Revision No. IM-EN-MI/C Eecve: June 5, 2020
1.L 2.N
3
Brown
Blue
Yellow-green
Figure 3.4 AC power terminals on the motherboard
3.1.2 Power terminals for DC power
The DC power is connected to a terminal stripe in the lower right-hand corner of the
motherboard. The three terminals are marked 1+, 2- and 3 (protecve earth). The
power terminal is directly connected to the exposed metal parts of the MI. The 24V
DC power to this terminal stripe shall be supplied from a secondary circuit which is
double or reinforced insulated from the mains supply within the limits for a limited-
energy circuit (maximum 200 VA/U) according to the IEC 61010-1.
3.1.3 External mains switch
External
mains switch
Figure 3.5 Installing external main switch
3 Connecons 11
© Copyright Vaisala 2020. All rights reserved.
Important: The MI does not have a built-in power switch. The system is always pow-
ered when connected to a power source. Vaisala recommends mounng an external
mains switch to control the MI’s power supply (Figure 3.5). The external power switch
should be easy to access. It is also recommended to use a 10A/230V circuit breaker for
power supply.
MULTI CHANNEL USER INTERFACE MI
INSTRUMENTS
M12 CONNECTORS
L
N
MAIN POWER
SWITCH PR-10900
L
N
+
-
mA Output 1
4-20mA
POWER SUPPLY
100-240VAC/50-60Hz
4158
-
+
-
+
OPTIONAL 24V POWER
SUPPLY FOR MI
24VDC
ETHERNET
M12-4pin, D-code, M
ETHERNET
M12-4pin, D-code, M
PR-8440 Ethernet cable for interfaces
MI has 8 module slots. Any of module cards
can be installed into any free slot. There could
be multiple module cards in slots depending on
external I/O needs.
PR-50021 mA output module
PR-50023 Relay output module
PR-50021 mA output module
PR-50023 Relay output module
PR-8430
Platform
4 cable
SLOT 2 SLOT 4 SLOT 6 SLOT 8
SLOT 1 SLOT 3 SLOT 5 SLOT 7
1...4 x
L
N
SOLENOID VALVE FOR
PNEUMATICALLY OPERATED
SHUT-OFF VALVE ASSEMBLY:
PR-3340-24/110/240
1...4 x
9
N
L
SAFETY SWITCH
PR-7060
8
7654321
N
L LL
POWER SUPPLY
100-240VAC/50-60Hz
or 24VDC
L
N
1...4 x
1...4 x
Grounding screw
PR-43-...-UN/AX/IA refractometer
(For Intrinsic refractometer see further wiring drawing 4118)
1...4 x
Figure 3.6 Mulchannel user interface wiring
12 MI instrucon manual
Document/Revision No. IM-EN-MI/C Eecve: June 5, 2020
3.2 Refractometer cable connecons
The refractometers are connected to the MI with a M12 male-female 8 pins A-coded
cable PR-8430, maximum length 90 m. The refractometer connectors are located un-
derneath the MI enclosure as a group of four. Refractometer connecons can be made
while the MI is powered. A refractometer can be connected to any of the four connec-
tors.
3.3 Ethernet cable connecon
The lemost connector underneath the MI is for ethernet output. It is for a M12 male
4-pins D-coded cable. The cable can be connected while the MI is powered.
If you wish to connect the MI with e.g. a laptop, you can purchase Interconnecng
cable PR-8330 with a M12-4 pin connector in one end and a RJ-45 connector in the
other end.
3.4 Module cards
For the installaon of the module cards, see Chapter 6 “Module cards”.
/