Insertion of program cassette
– Caution: The power cable (1) of the balance must be disconnected.
– Loosen cover (3) on left side of balance and remove.
Behind the cover the factory-installed standard cassette (2) is located. This must be
replaced by the GoldPac-M program cassette.
– Pull out handle of the inserted cassette by approx. 2 cm (to stop). Subsequently, completely
withdraw cassette by the handle.
– Unplug and remove standard cassette.
– Insert GoldPac-M program cassette into the cassette slide and push firmly until impacting
stop.
– Replace cover (3), press to snap into position. If the program cassette prevents proper
fitting of cover, it is not yet completely inserted.
Note: Please save standard cassette, it might be needed for later use of the balance
without GoldPac-M.
Connecting terminal and other instruments
Two connectors located at the rear of the balance are intended for connecting the following
instruments:
Upper connector (4) :
Instruments with RS232C or CL interface, e.g., printer GA44 (for
GA44 select in configuration file: Sector I-Face, accessing
Standard Setting - S.Stb, b 2400 and pause 1 s)
Lower connector (5) : Terminal and GM instruments with adapter plug
Coding pins at the connectors prevent improper insertion.