13
5.2 Closed-loop control
If the FlowPac is used to control mass flows, the system can be set up as follows:
• Place empty container or vessel on the weighing platform.
• Enter the following data using the keypad:
- Time xxxxxx <Enter> Enter time
- Date xxxxxx <Enter> Enter date
- ID xxxxxx <Enter> Enter identification number
- Code xxxxxx <Enter> Enter code number
- Prt x <Enter> Enter record interval [min]
- Valve % 50 <Enter>
- Mode xxxx <Enter> Select operating mode (see section 4.)
- Tol % xx <Enter> Enter tolerance limits
- Max xxx <Enter> Enter upper limit
- Min xxx <Enter> Enter lower fill limit
- →0← Tare
- M-Tare 0 <Enter> Enter manual tare if need be
- Timer xxx <Enter> Enter duration of operation with timer key [min]
• Store entries in a fail-safe manner with the storage key.
• Fill containers/vessels.
• Set up pump/metering devices and adjust.
• Enter desired set value in grams per hour and store in a fail-safe manner.
• Check that delivery tubing is free from bubbles and then press the Preset key.
• Press the Start key.
Notes:
• To obtain good results, the pump delivery rate during operation should be at least 15% of the
max. pump capacity (including allowance for any loss of accuracy).
• In operation with analog-controlled pumps, the GA37 D/A Converter is attached to the RS232C
interface. The converter is connected between the scale and the printer.
• For correct operation, it is essential to enter the data shown in boldface.
Regulated mass flow
Pump
Refilling
Valve
Supply
Record printer
Alarm
RS232C
or CL
GM bus
GM bus
Entry
GM54