  • Hello! I am an AI chatbot trained to assist you with the Mettler Toledo CalcPac-M For balances PM Operating instructions. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
Operating instructions
METTLER TOLEDO
CalcPac-M
2
x
1
The METTLER TOLEDO CalcPac-M
The METTLER TOLEDO CalcPac-M is for arithmetically processing weights and numbers.
The system consists of a METTLER TOLEDO PM balance, the CalcPac-M applications package
and a printer (optional).
Features of the system:
Printout identification : identification number, data, time, code
Choice of decimal : fixed decimal point 0-6 places, exponential format,
places fixed decimal of balance
Basic operations : +, -, x, ÷,
, , 2x, 1/-, x
2
Statistics : +, -, n, X, s
complete totals printout
Constants : 4 contstants incl. operation sign storable
3 contstants can be protected against accidental alteration.
Dynamic calculation with multiplier constant F1 and current
display reading after disconnection of terminal.
To insert the program cassette
Caution: Power cable must be unplugged from balance.
Detach cover (3) from left-hand side of balance.
Now revealed is the standard cassette (2) fitted in the balance before delivery.
This must be replaced by the CalcPac-M program cassette.
Pull out grip of cassette about 2 cm (until it stops), then use grip to withdraw cassette
completely.
Insert CalcPac-M program cassette in the cassette guide, and push it right in.
Put on cover (3) and press home. If the cover will not snap in place, the program cassette
has not been pushed in fully.
Note: Please keep the standard cassette; it may be uweful when using the balance later
without CalcPac-M.
3
To connect the terminal and other units
The two receptacles at the rear of the balance are intended for the following equipment:
Upper receptacle : equipment with RS232C or CL interface, e.g. GA44 Printer (for
GA44 in configuration register, sector I-Face; select standard
setting: S.Stb, b 2400 and pause 1 s)
Lower receptacle (5) : terminal and GM units with adapter plug
The coding pins on the receptacles prevent incorrect connection.
To select date and printout
The configuration register of the program offers new possibilities allowing the CalcPac-M to be
ideally matched to features peculiar to the particular application. A difference from the standard
balance software is that in the -Unit- sector of the configuration file one can now select:
Date : EU-date = day, month, year = standard setting
US-date = month, day, year
Printout : Sel.Prot = automatic printout only for total, other printouts only with [PRINT]
key; = standard setting
Au-Prot = printout automatic on entry and with [PRINT] key.
Software identification
When the balance is switched on, the display shows, for example, this identification.
When a printer is connected, this is also printed out.
4
Functions of general system keys
Switch to secondary Execute a secondary function or
key function begin an entry: after the function is selected the corresponding lamp stays on until the entry is
terminated.
Result/Terminate key To terminate operations or entries.
Clear (erase) Erases all operations and displayed result. Balance returns to weighing mode.
However, constants F1 to F4, register, ID no., date, time and number of decimal places are
retained.
Clear entry Erases the number just entered, but not the operation chosen. Display shows the last entry or
the last result.
Weight transfer Transfers current weight to a calculation (e.g. [5] [x] [ ] [=] ).
or display This key also resets to weighing mode.
Print out The displayed value is printed; time, date and identification only if entered. The [PRINT] key
cannot be used while entering.
stable weight value (only steady values transferred, sector I-Face: S.Stb)
unstable weight value (all values transferred, sector I-Face: S.All)
value calculated from stable weight
value calculated from unstable weight
91.42 g
D 100.21 g
* 137.15
D* 154.67
5
To select display format
The weight is always displayed with the decimal places of balance. For calculated values the format can be changed.
Function Key sequence Remarks
Select decimal No. of decimal places:
places [SET] [.] n n = 0 to 6 : fixed dec. pt
7 : exponential format
: fixed dec. pt of balance
For printing display format with
[SET] [PRINT] “constants”, page 9.
To enter printout text
Function Key sequence Remarks
Identification [SET] [ID NR] [1] [2] [.] [3] [4] [=] Any combination of 0, 1, …9.
number max. 14 digits or points.
Clear printout text by overwriting
with “0”
Date [SET] [DATE] [5] [.] [9] [.] [8] [6] [=] Select EU/US date “Preparation”,
page 2.
Erase as above.
Time [SET] [TIME] [1] [4] [.] [3] [5] [.] [3] [0] [=] Enter: hours.minutes.seconds
Erase as above.
Printout identi- [SET] [CODE] [4] [5] [6] [.] [7] [8] [9] [=] Same as with identification number.
fication code The figure is printed once, but not
stored.
6
Calculating with the CalcPac-M
The CalcPac-M can be used for processing weights and/or numbers.
Entries are made as with normal pocket calculators.
The displayed reading is processed. Entry is terminated with the operation keys [+] [-] [x] [÷] and the result key [=].
With entered and processed values and also “frozen” weights the balance display shows a * at top left. The result is displayed until a key is pressed
again. Return to weighing mode with [ ] or [C].
Function Key sequence (examples) Display Printout with Au Prot
Addition [5] [0] 50
[+] * 50.00
[7] [0] [=] 70
* 120.00
Subtraction Current value 100.00 g
[+] * 100.00 g
[6] [0] [0] 600
[-] * 700.00 g
Multiplication [3] [0] [0] 300
[x] * 400.00 g
[5] 5
[÷] * 2000.00
Division [4] 4
[=] * 500.00
* 50.00
+ * 70.00
= * 120.00
100.00 g
+ * 600.00
= * 700.00 g
- * 300.00
= * 400.00 g
x * 5.00
= * 2000.00
/ * 4.00
= 3 500.00
7
x
1
Function Key sequence (examples) Display Printout with Au Prot
Square root [] 100
*10
Reciprocal 100
[ ] * 0.01
Square 100
[x] [=] * 10000.0
Change sign 100
[-] [=] * -100
Double value 100
[+] [=] * 200
* 100.00
SQR * 10.00
* 100.00
1/X * 0.01
* 100.00
X2 * 10000.0
* 100.00
+/- * -100.00
* 100.00
2x * 200.00
8
Total and statistics
Function Key sequence Display Remarks
Add current Display shows current value, but total
value to is printed.
total 98.53 g
[+] 98.53 g
[+] 95.93 g
[+] 140.20 g
Subtract current [SET] [(-)] 140.20 g
value from total
(e.g. to correct [+] 101.64 g
wrong entries)
Print statistics [SET] [CLEAR ]
and clear
1 + 98.53 g
2 + 194.46 g
3 + 334.66 g
2 - 194.46 g
3 + 296.10 g
-------------------
SUM 296.10 g
n3
MEAN 98.700 g
Std 140.229 g
----CLEAR SUM-----
+ 1st weight
+ 2nd weight
+3rd weight
- 3rd weight
Total weight
Number of fig ures
added
Average
Standard
deviation
Statistics erased
9
To display figure counter, total, average and standard deviation
During a totalling operation it is possible to display the following values or process them like numerical entries:
Function Keys Remarks
Figure counter [SET] [(n)] Number of summands.
Up to 99 values can be
added per total.
Total [SET] [()] Current total
Average [SET] [(x)]
Standard [SET] [(s)]
deviation
If required, the keys can be marked with the felt-tip pen provided.
x
=
1
n
x
i
i=1
n
S
2
=
1
n-1
(x
i
- x)
2
i=1
n
X
X
10
To enter, print and secure the constants F1 to F4
Using the function keys [F1] to [F4] it is possible to store constants, with or without their arithmetical sign (+, -, x, ÷), and process them together
with the current reading.
Function Key sequence (examples) Display Remarks
To enter constants [SET] [F1] Set F1 Standard setting: +0.0
[x] [1] [0] [=] * 10.00
Storage 161.83 g To store final and interim results
function [SET] [F4] [=] Set F4 (memory function).
* 161.83 g For this function it is best to use the [F4] key,
as its memory cannot be blocked with the
jumper (“secure” position). Values can thus
be stored at any time.
To print constants [SET] [PRINT] Current
and display reading
format
To secure con- [SET] [SAVE] Current The constants F1 to F4 can now also be
stants and dis- reading secured against power failure. On
play format completing the entry, constants F1 to F3
can be protected against accidental
change by setting the jumper. For
procedure, see “Configuring” in balance
operating instructions.
F1:x 10.00
F2: 0.10
F3:+ 25.30
F4: 163.83
Format : Balance
Constants
Display
format
11
Calculating with constants F1 to F4
To calcutate with the contstants, simply press the appropriate function key. The current reading is then automatically processed accordingly,
and the result is displayed.
Keys Display Printout with Au Prot
100.00
[F1]
* 10000.0
e.g. F1: x 100 [F2]
F2: ÷ 10 * 1000.0
F3: - 300
F4: + 500 [F3]
* 700.0
[F4]
* 1200.00
Dynamic processing of multiplier contstant F1 without terminal
Having secured the multiplier constant F1 with [SET] [SAVE] and removed the terminal, multiplier constant F1 can be processed dynamically together
with the current value, and the result displayed.
Current weight
Select calculation with
multiplier constant F1 (e.g. F1 = x30)
Current weight processed with multi-
plier constant F1
Return to weight unit selected
under –Unit 1– (e.g. grams)
Current weight
* 100.00
F1: x * 100.00
= * 10000.0
F2: / * 10.00
= * 1000.00
F3: - * 300.00
= * 700.00
F4: + * 500.00
= * 1200.00
Press control bar until
display changes.
Display changes
automatically
until
until
12
Dynamic processing of constant F1 with terminal
Having stored the constant F1 together with corresponding arithmetical operation signs, (+, -, x, :) the constant F1 with the according operation
sign can be processed dynamically together with the current weight.
Example
Storing the
constant F1
Activates dynamic
processing
With a new entry, dynamic processing is deactivated.
X
13
Appendix
Err 0 Division by 0 Function not defined
Err 1 Time exceeded on entry Entry has not been completed in last 30 s or so, or a weight needed for processing has not
or value transfer been transferred.
Err 2 Root of a negative number Function not defined
Err 3 n > 99 for or - with n = 0 Number of summands > 99, or subtract a value from the total if total counter is at 0.
Err 4 Date or time not complete Date: terminate entry of day and month with [.].
EU date: DD.MM.YY
US date: MM.DD.YY
Time: terminate entry of hours and minutes with [.].
Err 5 Alteration of constants F1 to F3 Constants F1 to F3 cannot be redefined unless the jumper is in the “unsecured” position.
with jumper in “secured” ( balance operating instructions, “Configuring”)
position.
All these error signals are shown for about 5 seconds. The balance responds by showing the weight. The cause of the error signal is not
indicated.
Other error signals do not originate from the CalcPac-M, and are explained in the operating Instructions for the balance concerned.
14
But what if …
... display shows Err (or) ? Error signals, see above
... is shown? Entry is not yet possible because the system is busy printing or waiting for a steady weight
value
... is shown or Overflow, i.e. reading exceeds 999 999 with fixed format (to remedy, switch to exponential
printed) format with [SET] [.] [7]) or a number of 999.9 E 9 with exponential format.
... is shown? The balance is in input mode and is waiting for a value to be entered.
... the asterisk appears at top left A special sign on the balance. The value displayed is not the weight on the pan (perhaps the
on the display? result of a calculation). The display is blocked until a key is pressed. Return to weighing mode with
[C] or [ ].
... the system is in an undefined Switch balance off and on again. Entered values and system definitions will be preserved, but
condition) the total register readings will be lost.
... the GA44 Printer produces only Wrong baud rate (transmission speed), see balance operating Instructions.
?
... the GA44 Printer does not print The print interval is wrongly set (correct is 1 second, see balance operating Instructions) or the
every line? printing speed is too low. The time to advance the paper 20 cm should be 23...27 s (see instructions
for GA44 )
... the printer won‘t stop printing? Set configuration of the interface to S.Stb (Send Stable values), see balance operating
instructions.
... dynamic processing of constant No multiplier constant (xF1) entered at F1, or F1 has not been secured with [SET] [SAVE].
F1 with the current reading is not
possible?
15
To give your METTLER TOLEDO product an assured future:
METTLER TOLEDO Service preserves the quality, measurement
accuracy and value of METTLER TOLEDO products for years to come.
Please send for full details of our attractive service conditions.
Thanks in advance
Subject to technical changes and to the availability
of the accessories supplied with the instruments.
© Mettler-Toledo GmbH 2000 702559 Printed in Switzerland 0001/2.12
Mettler-Toledo GmbH, Laboratory & Weighing Technologies, CH-8606 Greifensee, Switzerland
Phone +41-1-944 22 11, Fax +41-1-944 30 60, Internet: http://www.mt.com
P702559
/