Knauer AZURA Recycling kit for pump P6.1L Owner's manual

  • Hello! I have reviewed the user manual regarding the Recycling kit A58208 for the AZURA Pump P 6.1L. This document describes how to convert the pump for peak recycling, including installation instructions for both isocratic and HPG configurations, and for pumps with or without degassers. I'm ready to answer your questions about the device or installation procedures.
  • What is the order number for the recycling kit?
    Which pump does this recycling kit work with?
    Does this kit work with pumps that have a degasser?
    What is the purpose of the recycling kit?
AZURA Recycling upgrade kit V6753, version 1.1 Page 1
Recycling kit for AZURA® Pump P 6.1L
The recycling kit (order no. A58208) can be used to convert the AZURA® Pump P 6.1L isocratic or
HPG for peak recycling. Connect the parts as shown in the figure below:
To install the recycling kit, proceed as follows:
Disconnection of the inlet tube to the pump:
1. Remove the fitting of the pump head inlet (see AZURA P 6.1 L manual) together with
the attached inlet tube.
If you have the HPG version of the AZURA P 6.1L, remove the pump head inlet of pump A.
Note: In case you have an AZURA P 6.1L with degasser, also remove the fitting on the
outlet side of the degasser for pump A.
Installation of the recycling kit:
1. Connect the outlet bushing P3830G with the pump head inlet. Therefore, follow the instruc-
tions to undo and tighten of the pump head inlet fitting as described in the pump manual.
Note: When installing the kit at an HPG pump, the kit has to be connect to pump A.
2. Connect the stainless-steel capillary with the last but one port of the fraction valve according
to the figure below:
P1233 +
P0954
A7200
A2511
1/10 M2528+
M2529
A0116
A0134
P3830G
To the pump
To the multiposition valve
To the degasser/eluent
AZURA Recycling upgrade kit V6753, version 1.1 Page 2
3. Insert the tube in the eluent bottle.
Note: If you have a pump with degasser, connect the flat bottom fitting (M2528 and M2529)
with the degasser outlet for pump A.
Peak recycling
Waste Collecting
Solvent recycling
HPLC System
/