Specifications
Specifications are subject to change without notice.
Specification Details
Dimensions (W x D x H) 60 x 21 x 98.1 cm (26.6 x 8.3 x 38.6 in.)
Enclosure Rating: NEMA 4X/IP66
Material: Cold-rolled epoxy powder-coated steel, optional
stainless steel (304)
Weight 54 kg (120 lb)
Power requirements Single phase, 115 or 230 VAC ±10%, 50/60 Hz (not rated for
BI or multi phase)
Power cable 18–12 AWG
Maximum power consumption 300 W
Fuses IEC127 Sheet III Type 2: 1 A anti-surge, 250 V; 2 A anti-
surge, 250 V; 4 A anti-surge, 250 V; 4 A quick blow, 250 V
Pollution degree/overvoltage category 2/II
Operating temperature 5 to 40 °C (41 to 104 °F)
Operating altitude 2000 m (6570 ft) maximum
Storage temperature 5 to 40 °C (41 to 104 °F)
Maximum relative humidity 80% to 31 °C decreasing linearly to 50% at 40 °C
Relays 5 output relays and 1 input relay, 3 A at 250 VAC; 0.5 A at
30 VDC
Analog outputs Two analog outputs, user configurable, optically isolated, self
powered. Maximum resistive load 600 Ω.
Analysis method UV persulfate oxidation with acid sparging for TIC removal
followed by CO
2
NDIR detector measurement
Measurement range (each model has a specific
range)
Standard: 0–5 to 0–20,000 mg/L TOC
Turbo: 0–50,000 µg/L (0-50 mg/L) TOC
Response time Standard: T90 ≤ 8 minutes; T20 ≤ 3 minutes (range-
dependent)
Turbo: T90 ≤ 5 minutes (0–5 mg/L); T20 ≤ 3 minutes
Accuracy/Repeatability/Linearity Standard: ±2%, full-scale undiluted; ±4%, full-scale diluted
Turbo: ≤ 4% or 8 µg/L (whichever is greater)
Method detection limit Standard: ≤ 0.015 mg/L at 0–5 mg/L 25 °C (77 °F)
Turbo: ≤ 5 μg/L at 0–5000 μg/L
Signal drift (60 days) < 2% with auto clean and auto calibration
Sample pressure At atmospheric pressure
Flow rate 25–200 mL/minute
25–60 mL/minute with an external cooler
Sample temperature 2 to 70 °C (36 to 158 °F)
Carrier gas Clean, CO
2
-free air at 2.8 bar (40 psi) minimum to 3.8 bar
(55 psi) maximum. Recommended 3.1 bar (45 psi).
English 3