wtw PurCon IS sample preparation Operating instructions

Type
Operating instructions
ba43133e06 05/2009
PurCon IS
Sample preparation
Operating manual
PurCon IS
50
ba43133e06 05/2009
Accuracy when going to
press
The use of advanced technology and the high quality of our instruments
are the result of continuous development. This may result in differences
between this operating manual and your instrument. Also, we cannot
guarantee that there are absolutely no errors in this manual. Therefore,
we are sure you will understand that we cannot accept any legal claims
resulting from the data, figures and descriptions.
Copyright
© Weilheim 2009, WTW GmbH
Reprinting - even as excerpts - is only allowed with the explicit written
authorization of WTW GmbH, Weilheim.
Printed in Germany.
PurCon IS Contents
51
ba43133e06 05/2009
PurCon IS - Contents
1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
1.1 General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
1.2 Structure and function . . . . . . . . . . . . . . . . . . . . . . . . . . 54
1.2.1 Filter unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
1.2.2 Control unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
1.2.3 Permeate suction line . . . . . . . . . . . . . . . . . . . . 57
1.2.4 Permeate line . . . . . . . . . . . . . . . . . . . . . . . . . . 57
2 Safety instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
2.1 Authorized use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
2.2 Name plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
2.3 General safety instructions . . . . . . . . . . . . . . . . . . . . . . . 60
3 Commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
3.1 Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
3.2 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
3.2.1 Control unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
3.2.2 Filter unit and permeate suction line . . . . . . . . . 63
3.2.3 Terminal box and permeate line . . . . . . . . . . . . 68
4 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
4.1 Operating elements . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
4.2 Starting sample preparation . . . . . . . . . . . . . . . . . . . . . . 74
4.3 Taking out of operation . . . . . . . . . . . . . . . . . . . . . . . . . 75
5 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
5.1 Cleaning the filter unit . . . . . . . . . . . . . . . . . . . . . . . . . . 78
5.1.1 Removing and installing the filter . . . . . . . . . . . 78
5.1.2 Cleaning the filter unit . . . . . . . . . . . . . . . . . . . . 80
5.2 Displacing the pump tubings . . . . . . . . . . . . . . . . . . . . . 82
5.3 Replacing pump tubings . . . . . . . . . . . . . . . . . . . . . . . . . 84
5.4 Cleaning the air filter of the compressor . . . . . . . . . . . . 86
5.5 Cleaning the control unit . . . . . . . . . . . . . . . . . . . . . . . . 87
6 What to do if ... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
6.1 Replacing the filter plates . . . . . . . . . . . . . . . . . . . . . . . . 90
6.2 Replacing the permeate tubes (suction side) . . . . . . . . . 90
Contents PurCon IS
52
ba43133e06 05/2009
7 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
7.1 General data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
7.2 Dimensions and weights . . . . . . . . . . . . . . . . . . . . . . . . . 93
7.3 Data of permeate supply . . . . . . . . . . . . . . . . . . . . . . . . . 94
8 Accessories and consumable materials . . . . . . . . . . . 95
PurCon IS Introduction
53
ba43133e06 05/2009
1 Introduction
1.1 General information
PurCon IS sample preparation prepares the sample for the analysis of
component substances with analyzers. PurCon IS uses the most mod-
ern technologies that enable largely maintenance-free and long-term
operation.
In comparison to conventional sample preparations, the maintenance
costs of the PurCon IS are low. Due to the low dwell time of the sample
in the PurCon IS, the preparation is carried out without any effect on
existing component substances. The sample preparation is carried out
continuously. The permeate is free from germs and total suspended
solids so that analyzers can be operated with very high reliability.
The maintenance costs of the analyzers are reduced considerably if
PurCon IS is used for the sample preparation.
Introduction PurCon IS
54
ba43133e06 05/2009
1.2 Structure and function
PurCon IS consists of the following main components:
z Filter unit
z Control unit
1.2.1 Filter unit
Fig. 1-1 Filter unit
1 Framework
2 Filter
3 Air shower
4 Permeate suction pipeline
5 Suction tube
4
1
5
2
3
PurCon IS Introduction
55
ba43133e06 05/2009
The filter unit is immersed in the sample. The filter unit has 2 or 4 filters
depending on the version. The front and back of the filter are used as
a filter membrane. The sample is sucked into the inside of the filter by
the permeate pump of the control unit. The filtered sample ("permeate")
is forwarded to the analyzer by the control unit.
BubbleClean procedure
for the self cleaning
A powerful jet of air is blown onto the outside of the filter from below.
This forces a continuous flow of sample that inhibits the pollution of the
filter membrane. This considerably reduces the overheads for cleaning
the filter.
The filter unit is connected with the control unit via the permeate suction
pipeline. Amongst other things, the control unit contains the air com-
pressor for the self cleaning and the permeate pump (see section
1.2.2 C
ONTROL UNIT).
Introduction PurCon IS
56
ba43133e06 05/2009
1.2.2 Control unit
Main components of the
control unit
Fig. 1-2 Control unit (example with heating line for the permeate line)
1 Operating panel
2 Control electronics
3 Permeate valve (3/2 port valve)
4 Air compressor
5 Permeate suction tube (back)
6 Permeate line (front)
7 Permeate pump
12
3
4
6
7
5
PurCon IS Introduction
57
ba43133e06 05/2009
The operation and control of PurCon IS is carried out by the control
unit. It contains all the control and monitoring facilities that are required
for the permeate extraction and forwarding. The air compressor for the
BubbleClean self cleaning system is also contained in the control unit.
Permeate pump The permeate is fed by a peristaltic pump with four pump channels. The
pump output can be adjusted in four stages.
1.2.3 Permeate suction line
The permeate suction line connects the filter unit with the control unit.
It consists of three components:
z 2 permeate hoses (inside diameter 4 mm). These are used to feed
the permeate from the membrane plates.
z Encasing tube. It is used as a protective covering for the two perme-
ate tubes. At the same time, the air shower of the filter unit is sup-
plied with compressed air from the air compressor via the encasing
tube.
Integrated protection
against freezing
The compressed air flowing through the encasing tube is warmed by
the air compressor. This ensures that, when outside temperatures are
low, the permeate does not freeze on its way through the permeate
hoses.
1.2.4 Permeate line
The permeate line forwards the permeate from the control unit to the
analyzers. It consists of the following parts:
z Encasing tube.
z Permeate hose (inside diameter 2 mm)
z Terminal box
z Heating line (option).
Terminal box On the analyzer side, the permeate tube is connected to a box (terminal
box). The permeate is delivered to the analyzer or overflow vessel via
a hose nipple.
Protection against
freezing
(option)
The permeate can be protected against freezing with the aid of a heat-
ing line. The temperature monitoring is carried out with the aid of a tem-
perature sensor that is integrated in a freely placeable cable box
("temperature measuring box"). The temperature measuring box is
connected with the control unit. If there is any danger of the permeate
freezing, the control electronics switch the heating line on.
Introduction PurCon IS
58
ba43133e06 05/2009
PurCon IS Safety instructions
59
ba43133e06 05/2009
2 Safety instructions
This operating manual contains essential instructions that must be fol-
lowed during the commissioning, operation and maintenance of the
instrument.
Consequently, all responsible personnel must read this operating man-
ual carefully before working with the instrument. The operating manual
must always be kept available within the vicinity of the instrument.
General safety
instructions
Safety instructions in the operating manual are indicated by the
warning symbol (triangle) in the left column. The signal word
(e. g. "CAUTION") indicates the danger level:
WARNING
indicates instructions that must be followed precisely in order to
prevent serious danger to personnel.
CAUTION
indicates instructions that must be followed precisely in order to
avoid slight injuries to personnel or damage to the instrument or
the environment.
Other labels
Note
indicates notes that draw your attention to special features.
Note
indicates cross-references to other documents, e.g. component oper-
ating manuals.
2.1 Authorized use
Authorized use consists exclusively of the preparation of samples for
the analysis of component substances using analyzers. The technical
specifications given in chapter 7 T
ECHNICAL DATA must be followed. Any
other use is considered to be unauthorized.
2.2 Name plate
The name plate of the PurCon IS is located on the inside of the door of
the control unit housing.
Safety instructions PurCon IS
60
ba43133e06 05/2009
2.3 General safety instructions
This instrument is constructed and inspected in accordance with the
relevant guidelines and norms for electronic measuring instruments
(see chapter 7 T
ECHNICAL DATA).
It left the factory in a safe and secure technical condition.
Function and
operational safety
The failure-free function and operational safety of the instrument is only
guaranteed if the generally applicable safety measures and the special
safety instructions in this operating manual are followed during its use.
The failure-free function and operational safety of the PurCon IS is only
guaranteed under the environmental conditions that are specified in
chapter 7 T
ECHNICAL DATA.
If the instrument was transported from a cold environment to a warm
environment, the formation of condensate can impair the functioning of
the instrument. In this case, wait for the instrument to adjust to room
temperature before recommissioning.
Safe operation If safe operation is no longer possible, the instrument must be taken out
of operation and secured against inadvertent operation.
Safe operation is no longer possible if the instrument:
z has been damaged in transport
z has been stored under adverse conditions for a lengthy period of
time
z is visibly damaged
z no longer operates as described.
If you are in any doubt, contact the supplier of your instrument.
PurCon IS Commissioning
61
ba43133e06 05/2009
3 Commissioning
3.1 Scope of delivery
PurCon IS/2-230
PurCon IS/4-230
z Control unit
z Temperature measuring box
z Filter unit with 2 or 4 filters
z Permeate suction line
z Operating manual
PCIS-PL... Permeate line with terminal box (different lengths, with or without hea-
ting line). Permeate line with terminal box must be ordered seperately.
3.2 Installation
CAUTION
The mounting of the PurCon IS may only be carried out by special-
ists authorized by WTW. WTW reserves the right to reject warranty
claims if this condition is violated.
Commissioning PurCon IS
62
ba43133e06 05/2009
3.2.1 Control unit
Types of mounting The control unit can be mounted in the following ways in the vicinity of
the sampling location:
z Rail mounting with the PCIS-MSG accessory set
z Wall mounting (see below)
Note
Instructions for mounting are given in the mounting instructions of the
accessories. Ordering information for the accessories are given in sec-
tion 8 A
CCESSORIES AND CONSUMABLE MATERIALS.
Wall mounting The housing of the control unit has four attaching eyes on the back for
mounting on a wall. Make sure that the wall has the stability required
for the weight of the control unit. Only use mounting material, which is
suitable for the wall material (screws, dowels, etc.). The drilled hole
spacings are shown in the following diagram:
Fig. 3-1 Drilled hole spacings for the control unit (dimensions in mm)
Note
Make sure that there is sufficient space under the control unit for con-
necting the permeate suction tube and the permeate line.
Rail mounting Alternatively, the control unit can be mounted on a railing.
533
355
PurCon IS Commissioning
63
ba43133e06 05/2009
3.2.2 Filter unit and permeate suction line
Types of mounting The filter unit can be mounted on site in the following ways:
z Chain mounting with the PCIS-MSK accessory set
z Pipe mounting with the PICS-MSS accessory set, e.g. on the railing
Note
Observe the minimum immersion depth of the filter unit (see dimensio-
nal drawing in chapter 7 T
ECHNICAL DATA. The frame tube must be enti-
rely immersed.
Note
Instructions for mounting are given in the mounting instructions of the
accessories. Ordering information for the accessories are given in sec-
tion 8 A
CCESSORIES AND CONSUMABLE MATERIALS.
Connecting the
permeate suction line
with the filter unit
Fig. 3-2 Connecting the permeate suction line with the filter unit
1 Remove the coupling nut (pos. 1 in Fig. 3-2) and the threaded
plugs (pos. 2).
2 Push a hose clip (pos. 3) over the encasing tube (pos. 4).
3 Thread the two suction hoses (pos. 5) through the compressed
air connection (pos. 5).
4 Slip the encasing tube on the compressed air connection and
secure with the hose clip.
4
3
5
6
2
1
Commissioning PurCon IS
64
ba43133e06 05/2009
Fig. 3-3 Connecting the permeate suction hoses
5 Push the threaded plugs (pos. 2) over the two suction tubes up
to the cable gland (pos. 6) and secure with the coupling ring
(pos. 1). The flat side of the threaded plug must face down-
wards.
6 Connect the two suction hoses with the filter as shown in the
following Fig. 3-3. In the case of four filters, connect filters 1
and 3, as well as 2 and 4 to an intake tube via Y pieces.
PurCon IS Commissioning
65
ba43133e06 05/2009
Connecting the
permeate suction line
with the control unit
WARNING
Touching the electrical lines inside the control unit can result in
danger to life from electric shock. Before opening the operating
panel of the instrument, switch it off at the mains switch and pull
out the mains plug.
Fig. 3-4 Preparing the control unit
1 Open the door of the control unit.
2 Remove the screw (pos. 1 in Fig. 3-4) and open the operating
panel (pos. 2).
3 Undo the back cable gland (pos. 3 in Fig. 3-4).
Netz
I
0
0
1
4
3
2
1
3
2
Commissioning PurCon IS
66
ba43133e06 05/2009
Fig. 3-5 Mounting the permeate suction hose
4 Remove the coupling nut (pos. 4 in Fig. 3-5) from the cable
glands and remove the threaded plug (pos. 5).
5 Thread the two suction hoses (pos. 6a and 6b) from below
through the terminal block (pos. 7).
6 Push the encasing tube (pos. 8) through the cable gland into
the terminal block up to the stop. Twist the encasing tube
somewhat while doing so. The encasing tube is pushed in far
enough when the marking on the hose lines up with the cable
gland. Then tighten up the cable gland.
7 Push the threaded plugs (pos. 5) over the two suction tubes up
to the cable gland and secure with a coupling ring (pos. 5). The
flat side of the threaded plug must face upwards.
6a, b
5
4
7
8
PurCon IS Commissioning
67
ba43133e06 05/2009
Fig. 3-6 Connecting the suction lines to the permeate valve
8 Connect the two suction tubes with the free connections of the
permeate valve (pos. 9a and 9b in Fig. 3-6).
9a
9b
Commissioning PurCon IS
68
ba43133e06 05/2009
3.2.3 Terminal box and permeate line
One end of the permeate line is already connected with the terminal
box in the factory. The following main steps are necessary for mounting
the unit:
z Mounting the terminal box in the vicinity of the analyzer
z Connecting the terminal box with the analyzer
z Connecting the free end of the permeate line with the control unit.
Note
The terminal box should be mounted in the immediate vicinity of the
analyzer if possible in order to avoid long distances to the permeate
and the dead times associated with it.
Mounting the
terminal box
Fig. 3-7 Mounting the terminal box
1 Open the terminal box.
2 Screw the terminal box onto a flat wall surface via the four
drilled holes at the corners (Fig. 3-7). Use suitable screws
(max. diameter 5 mm).
3 Close the terminal box.
Permeate output
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50