  • Hello! I am an AI chatbot trained to assist you with the Endres+Hauser SS2100i-2 TDLAS Gas Analyzer Operating instructions. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
Products Solutions Services
Operating Instructions
SS2100i-2 TDLAS Gas Analyzer
ATEX/IECEx: Zone 1
BA02197C/66/EN/0
2.23
70210290
Operating Instructions SS2100i-2 TDLAS Gas Analyzer
2 Endress+Hauser
SS2100i-2 TDLAS Gas Analyzer Operating Instructions
Endress+Hauser 3
Table of Contents
1 About this document ............................ 4
1.1 Warnings .................................................................. 4
1.2 Symbols on the device............................................. 4
1.3 U.S. export compliance ........................................... 4
1.4 Standard documentation ........................................ 4
2 Introduction .......................................... 6
2.1 How to use this manual .......................................... 6
2.2 General warnings and cautions ............................. 6
2.3 Manufacturer address ............................................ 8
2.4 About the gas analyzers ......................................... 9
2.5 Getting familiar with the analyzer ...................... 12
3 Safety ................................................... 24
3.1 Potential risks affecting personnel ..................... 24
3.2 Explosion hazard ................................................... 24
4 Installation .......................................... 25
4.1 What should be included in the shipping box .... 25
4.2 Lifting/carrying the analyzer ............................... 25
4.3 Inspecting the analyzer ........................................ 25
4.4 Hardware and tools for installation .................... 26
4.5 Mounting the analyzer ......................................... 26
4.6 Protective chassis ground connections ............... 26
4.7 Opening the analyzer enclosure covers .............. 27
4.8 Cable installation .................................................. 27
4.9 Cable glands ........................................................... 27
4.10 Connecting the solenoid valves ........................... 28
4.11 Application of gland lubricant ............................. 31
4.12 Connecting the signals and alarms ..................... 32
4.13 Configuring the RS-232/RS-485 converter........ 34
4.14 Connecting electrical power to the analyzer ...... 34
4.15 Connecting the gas lines ...................................... 36
4.16 Changing the 4-20 mA current loop mode ........ 37
5 Specifications ...................................... 39
5.1 Exd accessory conditions of use .......................... 40
6 Maintenance and troubleshooting ... 42
6.1 Gas leaks ................................................................ 42
6.2 Excessive sampling gas temperatures and
pressures ................................................................ 42
6.3 Electrical noise ...................................................... 42
6.4 Contamination ...................................................... 42
6.5 Cleaning the mirrors ............................................. 43
6.6 Replacing the pressure sensor ............................. 47
6.7 Flame arrestor replacement and safety ............. 56
6.8 Peak tracking reset procedure ............................. 59
6.9 Instrument issues .................................................. 59
6.10 Service .................................................................... 61
6.11 Service repair order ............................................... 61
6.12 Packing, shipping, and storage ........................... 61
6.13 Storage ................................................................... 62
6.14 Disclaimers ............................................................ 62
6.15 Warranty ................................................................ 62
7 Analyzer parts .................................... 63
8 Wiring diagrams ................................. 72
Operating Instructions SS2100i-2 TDLAS Gas Analyzer
4 Endress+Hauser
1 About this document
1.1 Warnings
Structure of Information Meaning
WARNING
Causes (/consequences)
If necessary, consequences of non-compliance
(if applicable)
‣ Corrective action
This symbol alerts you to a dangerous situation. Failure to avoid the dangerous
situation can result in a fatal or serious injury.
CAUTION
Causes (/consequences)
If necessary, consequences of non-compliance
(if applicable)
‣ Corrective action
This symbol alerts you to a dangerous situation. Failure to avoid this situation
can result in minor or more serious injuries.
NOTICE
Cause/situation
If necessary, consequences of non-compliance
(if applicable)
‣ Action/note
This symbol alerts you to situations which may result in damage to property.
Table 1. Warnings
1.2 Symbols on the device
Symbol
Description
The Laser Radiation symbol is used to alert the user to the danger of exposure to hazardous visible laser radiation
when using the Raman Rxn2 system.
The High Voltage symbol that alerts people to the presence of electric potential large enough to cause injury or
damage. In certain industries, high voltage refers to voltage above a certain threshold. Equipment and conductors that
carry high voltage warrant special safety requirements and procedures.
The ETL Listed Mark provides proof of product compliance with North American safety standards. Authorities Having
Jurisdiction (AHJ) and code officials across the US and Canada accept the ETL Listed Mark as proof of product
compliance to published industry standards.
The WEEE symbol indicates that the product should not be discarded as unsorted waste but must be sent to separate
collection facilities for recovery and recycling.
The CE Marking indicates conformity with health, safety, and environmental protection standards for products sold
within the European economic area (EEA).
Table 2. Symbols
1.3 U.S. export compliance
The policy of Endress+Hauser is strict compliance with U.S. export control laws as detailed on the website of the
Bureau of Industry and Security at the U.S. Department of Commerce.
1.4 Standard documentation
All documentation is available:
On the USB provided with the analyzer
Endress+Hauser’s website: www.endress.com
Each analyzer shipped from the factory is packaged with documents specific to the model that was purchased. This
document is an integral part of the complete document package, which also includes:
SS2100i-2 TDLAS Gas Analyzer Operating Instructions
Endress+Hauser 5
Part Number
Document Type
Description
TI01670C
SS2100i-2 Technical
Information
Planning aid for your device. This document contains information for the analyzer
including system design with sample conditioning components and inlet/outlet
points, certificates and approvals, and product technical data.
XA02694C SS2100i-2 Safety
Instructions
Requirements for installing or operating the SS2100i-2 TDLAS Gas Analyzer
related to personnel or equipment safety.
GP01180C
Description of Device
Parameters (NS 5.14)
Provides the user with an overview of the NS 5.14 firmware functionality.
For additional instruction manuals, please refer to the following:
For custom orders, go to the Endress+Hauser website for the list of local sales channels who can provide the
requested order-specific documentation:
https://endress.com/contact
or
https://addresses.endress.com/
For standard orders, go to the Endress+Hauser website to download the published documentation:
www.endress.com
Operating Instructions SS2100i-2 TDLAS Gas Analyzer
6 Endress+Hauser
2 Introduction
Endress+Hauser SS2100i-2 products use SpectraSensors TDLAS technology and are high-speed, diode-laser based
extractive analyzers designed for extremely reliable monitoring of very low (trace) to standard concentrations of
specific components in various background gases. To ensure that the analyzer performs as specified, it is important
to closely review the sections of this manual. This manual contains a comprehensive overview of the SS2100i-2
analyzer and step-by-step instructions for:
Inspecting the analyzer
Mounting and installing the analyzer
Troubleshooting the system
2.1 How to use this manual
Take a moment to familiarize yourself with this Operating Instruction by reading the "Table of Contents".
Some options and accessories are available for the SS2100i-2 analyzers. This manual addresses the most common
options and accessories.
Images, tables, and charts are included to provide a visual understanding of the analyzer and its functions. Special
symbols are used to provide the user with key information regarding the system configuration and/or operation. Pay
close attention to this information.
2.1.1 Who should read this manual
This manual should be read and referenced by anyone installing, operating, or having direct contact with the
analyzer.
2.1.2 Conventions used in this manual
In addition to the symbols and instructional information, this manual is created with “hot links” to enable the user to
quickly navigate between different sections within the manual. These links include table, figure and section
references and are identified by a pointing finger cursor when rolling over the text. Simply click on the link to
navigate to the associated reference.
2.2 General warnings and cautions
Instructional icons are provided in this manual and on the SS2100i-2 unit to alert the user of potential hazards,
important information and valuable tips. Following are the symbols and associated warning and caution types to
observe when servicing the analyzer. Some of these symbols are provided for instructional purposes only and are not
labeled on the system.
2.2.1 Safety warning label
The warning label shown below will be affixed to the front side of all analyzer enclosures that contain sample gas.
SS2100i-2 TDLAS Gas Analyzer Operating Instructions
Endress+Hauser 7
Hazards may vary by stream composition. One or more of the following conditions may apply.
Description
Flammable. Gases used in the processing of this analyzer may be extremely flammable. Any work in a hazardous
area must be carefully controlled to avoid creating any possible ignition sources (e.g., heat, arcing, sparking, etc.).
Toxins
. Endress+Hauser analyzers measure a variety of gases, including high-level H2S. Follow all safety protocols
governing toxic gases and potential leaks.
Inhalation
. Inhaling toxic gases or fumes may cause physical damage or death.
CAUTION
Technicians are expected to follow all safety protocols established by the customer that are necessary for
servicing or operating the analyzer. This may include, but is not limited to, lockout/tagout procedures, toxic gas
monitoring protocols, personal protective equipment (PPE) requirements, hot work permits and other
precautions that address safety concerns related to performing service or operation on process equipment
located in hazardous areas.
2.2.2 Equipment labels
Description
Warning statement for
hazardous voltage
. Contact may cause electric shock or burn. Turn off and lock out system
before servicing.
Failure to follow all directions may result in damage or malfunction of the analyzer.
Maximum voltage and current specifications for the fuse closest to label.
PROTECTIVE EARTH GROUND
Symbol indicates the connection point of the ground wire from the main power
source.
FUNCTIONAL EARTH GROUND
Symbol indicates grounding points intended primarily for troubleshooting.
2.2.3 Instructional symbols
Description
General notes and important information concerning the installation and operation of the analyzer.
Failure to follow all directions may result in fire.
INVISIBLE LASER RADIATIONAvoid exposure to beam. Class 3b Radiation Product. Refer servicing to the
manufacturer-qualified personnel.
Failure to follow all directions may result in damage or malfunction of the analyzer.
Maximum voltage and current specifications for fuses.
Operating Instructions SS2100i-2 TDLAS Gas Analyzer
8 Endress+Hauser
2.2.4 Special safety symbols used on the equipment
Special safety symbols and labeling are used on the equipment to alert the user to potential hazards and important
information associated with the analyzer. Every symbol and label has significant meaning that should be heeded.
Symbol
Description
ENERGIZED FUSE WARNING
Do not remove or replace fuse when energized.
INVISIBLE LASER RADIATIONAvoid exposure to beam. Class 3b Radiation Product. Refer
servicing to the manufacturer or qualified personnel.
DO NOT REMOVERemoval of the seal and/or disassembly of pieces traversed by label voids
the warranty.
CLASS 1 LASER PRODUCTInvisible laser radiation when open. Avoid direct exposure to the
beam.
CLASS 3B LASER PRODUCTInvisible laser radiation. Avoid direct exposure to beam. Class 3b
laser product.
CLASS 3B LASER WARNING
Class 3B invisible laser radiation when open. Avoid direct
exposure to the beam.
2.3 Manufacturer address
Endress+Hauser
11027 Arrow Route
Rancho Cucamonga, CA 91730
United States
www.endress.com
SS2100i-2 TDLAS Gas Analyzer Operating Instructions
Endress+Hauser 9
2.4 About the gas analyzers
Endress+Hauser SS2100i-2 analyzers employ SpectraSensors tunable diode laser absorption spectroscopy (TDLAS) to
measure the concentration of single compounds in gas mixtures. The tunable diode laser (TDL) absorption
spectrometers in the SS2100i-2 operate in the near- to short-wavelength infrared. Each compact sensor consists of a
TDL laser light source, sample cell, and detector configured for high sensitivity measurement of a particular
component within the gas-phase constituents in the stream. The sensor is controlled by microprocessor-based
electronics with embedded software that incorporates advanced operational and data processing algorithms.
In its simplest form, a diode laser absorption spectrometer typically consists of a sample cell with a mirror at one end,
and a mirror or window at the opposite end, through which the laser beam can pass, as shown in the figure below.
The laser beam enters the cell and reflects off the mirror(s) making one or more trips through the sample gas and
eventually exiting the cell where the remaining beam intensity is measured by a detector. With the SS2100i-2
analyzers, sample gas flows continuously through the sample cell ensuring that the sample is always representative
of the flow in the main pipe.
0.8 m cell 8 or 28 m cell
Figure 1. Schematic of a typical tunable diode laser absorption spectrometer
1. Laser 5. Detector
2. Pressure sensor 6. Window/near mirror
3. Outlet 7. Inlet
4. Optical head 8. Far mirror
Due to their inherent structure, the molecules in the sample gas each have characteristic natural frequencies (or
resonances). When the output of the laser is tuned to one of those natural frequencies, the molecules with that
resonance will absorb energy from the incident beam. That is, as the beam of incident intensity, I0(λ), passes through
the sample, attenuation occurs via absorption by the trace gas with absorption cross section σ(λ). According to the
Beer-Lambert absorption law, the intensity remaining, I(λ), as measured by the detector at the end of the beam path
of length / (cell length x number of passes), is given by
()=()exp [()]
where N represents the species concentration. Thus, the ratio of the absorption measured when the laser is tuned on-
resonance versus off-resonance is directly proportional to the number of molecules of that species in the beam path,
or
4
1 5
6
7
2
8 3
4
1 5
6
7
2
3 8
Operating Instructions SS2100i-2 TDLAS Gas Analyzer
10 Endress+Hauser
N = 
()()
()
Figure 3 below shows the typical raw data from a laser absorption spectrometer scan including the incident laser
intensity, I0(λ), and the transmitted intensity, I(l), for a clean system and one with contaminated mirrors (shown to
illustrate the system’s relative intensity to mirror contamination).
Fig 2. Typical raw signal from a laser diode absorption spectrometer with and without mirror contamination
A. Resonance absorption Blue line: raw signal, I (l)
B. Red line: incident energy I0(l) Green line: raw signal, I (l) (contaminated mirrors)
The positive slope of raw data results from ramping the current to tune the laser, which not only increases the
wavelength with current, but also causes the corresponding output power to increase. By normalizing the signal by
the incident intensity, any laser output fluctuations are canceled, and a typical, yet more pronounced, absorption
profile results, as shown in figure 4 below.
Fig 3. Typical normalized signal from a laser diode absorption spectrometer
Contamination of the mirrors results solely in lower overall signal. However, by tuning the laser off-resonance as
well as on-resonance and normalizing the data, the technique self-calibrates every scan resulting in measurements
that are unaffected by mirror contamination.
Wavelength [a.u.]
Wavelength [a.u.]
1.0
0.99
0.98
0.97
0.96
0.95
Wavelength [a.u.]
Signal [a.u.]
3.0
2.5
2.0
1.5
1.0
0.5
0.0
A
SS2100i-2 TDLAS Gas Analyzer Operating Instructions
Endress+Hauser 11
2.4.1 Differential TDLAS
SpectraSensors differential TDLAS technology involves subtracting two spectrums from one another. A “dry”
spectrum, a response from the sample when the analyte of interest has been completely removed, is subtracted from
the “wet” spectrum, a response from the sample when the analyte is present. The remainder is a spectrum of the
pure analyte. This technology is used for very low or trace measurements and is also useful when the background
matrix changes over time.
2.4.2 Wavelength modulation spectroscopy (WMS) signal detection
Endress+Hauser takes the fundamental absorption spectroscopy concept a step further by using a sophisticated
signal detection technique called wavelength modulation spectroscopy (WMS). When employing WMS, the laser
drive current is modulated with a kHz sine wave as the laser is rapidly tuned. A lock-in amplifier is then used to
detect the harmonic component of the signal that is at twice the modulation frequency (2f). Refer to the figure
below, which is shown in arbitrary units (a.u.). This phase-sensitive detection enables the filtering of low-frequency
noise caused by turbulence in the sample gas, temperature and/or pressure fluctuations, low-frequency noise in the
laser beam or thermal noise in the detector.
Fig 4. Typical normalized 2f signal showing species concentration proportional to height.
A. 2f peak height
B. Normalized 2f signal
With the resulting low-noise signal and use of fast post-processing algorithms, reliable parts per million (ppm) or
parts per billion (ppb) detection levels are possible (depending on target and background species) at real-time
response rates.
All Endress+Hauser TDLAS gas analyzers employ the same design and hardware platform. Measuring different trace
gases in various mixed hydrocarbon background streams is accomplished by selecting a different optimum diode
laser wavelength between 700-3000 nm, which provides the least amount of sensitivity to background stream
variations.
2.4.3 Sample conditioning system
The analyzer may be integrated with a sample conditioning system (SCS) designed to meet the sample condition
requirements for the analyzer while preserving sample integrity and minimizing lag time. Refer to the SCS Operating
Instruction for more information.
2.4.4 Determining firmware version
When the analyzer is powered on for the first time, the firmware version will display on the system LCD display for
approximately seven seconds. Refer to Powering Up the Analyzer in the Description of Device Parameters for this
Wavelength [a.u.]
Signal [a.u.]
B
A
Operating Instructions SS2100i-2 TDLAS Gas Analyzer
12 Endress+Hauser
analyzer for operational instructions. The firmware version for each analyzer is also listed on the analyzer calibration
certificate.
2.5 Getting familiar with the analyzer
The SS2100i-2 analyzer consists of two connected enclosures in a standard cable-coupled configuration, as shown in
the figure below.
Fig 5. Cable-coupled analyzer configuration
The left enclosure (electronics enclosure) contains the control electronics and input/output electrical connections, as
shown below. The right enclosure (sample cell enclosure) contains the sample cell and heater, as shown in Figure 8
→ (8/28 m sample cell) , Figure 9 (0.8 m sample cell), or Figure 10 (0.1 m sample cell). The keypad and
LCD display on the cover of the electronics enclosure are the user interface to the analyzer. Power and signal
connections are made via access ports on the bottom of the electronics enclosure. Tube fittings on the right side of
the sample cell enclosure are for sample supply and return connections. Both enclosures are mounted on a Unistrut®
frame.
Cable & Gland Kit
1100002143
SS2100i-2 TDLAS Gas Analyzer Operating Instructions
Endress+Hauser 13
Fig 6
. Dimensions and Mounting for Electronics and sample cell enclosures (cable
-coupled configuration)
Front view
Back view Bottom view
1.
Electronics enclosure
1. 8x mounting feet 1. M25 4-20 mA output board
2.
LCD display
A. 1036.83 mm (40.82 in) 2. M25 digital input/output port
3.
Keypad window
B. 439.93 mm (17.32 in) 3. M25 solenoid valve port
4.
Chassis ground
C. 500.13 mm (19.69 in) 4. M20 power input port
5.
24x cover screws
5. Breather
6.
Measurement cell enclosure
6. M25 serial com port
7.
Chassis ground
A. 247.90 mm (9.76 in)
8. Gas in
9. Gas out
A 1066.8 mm
(42.0 in)
B.
469.9 mm
(18.5 in)
C.
670 mm
(26.38 in)
A
B B
C
A
A
B B
C
Front View Back View
Bottom View
1
2
3
4
5
6
1
6
7
2
8
3
9
4
5
1
Operating Instructions SS2100i-2 TDLAS Gas Analyzer
14 Endress+Hauser
Fig 7. Components in SS2100i-2 electronics enclosure
1. 4-20mA AI board 11. Relay control board
2. Fuse (F3) 12. Temperature control board
3. High temperature limit thermostat 13. ARM9 control electronics
4. Analyzer power supply 14. RS-232 to RS-422/485 converter
5. Relays 15. Solid-state relay
6. AC line filter 16. Temperature controller
7. Fuses (F4-F10) 17. Auxiliary power supply
8. Main circuit breaker 18. AC terminal block for sample cell heater
9. Ground bus bar 19. Ground terminal blocks
10. 4-20mA AO board 20. Terminal blocks
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
SS2100i-2 TDLAS Gas Analyzer Operating Instructions
Endress+Hauser 15
Fig 8. Components on an 8 or 28 m sample cell panel enclosure
1. Heater 6. Transition plate thermistor
2. High temp. limit thermostat set at 70 °C 7. Sample temperature thermistor
3. AC terminal block for heater 8. Measurement cell
4. Terminal blocks 9. Pressure sensor
5. Fan
1
2
3
4
5
6
7
8
9
Operating Instructions SS2100i-2 TDLAS Gas Analyzer
16 Endress+Hauser
Fig 9. Components on a 0.8 m sample cell panel assembly
1. Heater 6. Transition plate thermistor
2. High temp. limit thermostat set at 70 °C 7. Pressure sensor
3. AC terminal block for heater 8. Measurement cell
4. Terminal blocks 9. Sample temperature thermistor
5. Fan
1
2
3
4
5
6
7
8
9
SS2100i-2 TDLAS Gas Analyzer Operating Instructions
Endress+Hauser 17
Fig 10. Components on a 0.1 sample cell panel enclosure
1. Heater 6. Transition plate thermistor
2. High temp. limit thermostat set at 70 °C 7. Measurement cell
3. AC terminal block for heater 8. Pressure sensor
4. Terminal blocks 9. Sample temperature thermistor
5. Fan
In the electronics enclosure, the analyzer power supply provides power to the analyzer control electronics and relays
controlling valves. The analyzer control electronics drive the laser, collect the signal and analyze the spectra.
Powered relays control valves while unpowered relays serve as alarm contacts. An AC line filter is used to condition
the input power.
The auxiliary switch-mode power supply provides power to the heater temperature controller and the RS-232/RS-
485 converter. The power supply is rated for 1.3 A at 24VDC output at ambient temperatures Ta 60 °C. The
operational state is indicated by LEDs on the front face, where green means the output voltage is on and within
specification, and red means the output voltage is on but below specification.
The thermostat prevents the temperature inside the enclosure from getting too hot. The thermostat is preset at the
factory to open the power circuit if the temperature inside the analyzer enclosure exceeds 70 ± 4 °C. The power
1
2
3
4
5
6
7
8
9
Operating Instructions SS2100i-2 TDLAS Gas Analyzer
18 Endress+Hauser
circuit will remain open until the manual reset button (located between the two wire terminals) on the thermostat is
pressed or the temperature drops approximately 30% below the set-point.
A DIN rail near the bottom holds fuse terminal blocks, the main breaker, and terminal blocks for all external
connections. The temperature controller controls the heater in the sample cell enclosure via the solid-state relay.
NOTICE
See Figure 7→ for locating fuses. If you need to replace a fuse, use only the same type and rating of fuse as the
original as listed in the tables below.
In the sample cell enclosure, the sample cell is the actual TDLAS spectrometer through which the gas sample flows.
The sample cell is equipped with a pressure sensor and thermistor to monitor the thermodynamic conditions of the
sample. The heater maintains the inside of the analyzer enclosure at a constant temperature with air circulated by
the fan.
Drawing Reference
Description
Rating
F3
Miniature Fuse, 5 x 20 mm, Time Delay
250 VAC 1.6 A
F41 Miniature Fuse, 5 x 20 mm, Time Delay 250 VAC 0.5 A
F5141, F61, F71, F81
Miniature Fuse, 5 x 20 mm, Time Delay
250 VAC 0.1 A
F91 Miniature Fuse, 5 x 20 mm, Time Delay 250 VAC 1 .0A
F101 Miniature Fuse, 5 x 20 mm, Time Delay 250 VAC 1.2 A
Fuse specifications for 240 VAC systems
Drawing Reference Description Rating
F3
Miniature Fuse, 5 x 20 mm, Time Delay
250 VAC 1.6 A
F41 Miniature Fuse, 5 x 20 mm, Time Delay 250 VAC 0.5 A
F5
141
, F6
1
, F7
1
, F8
1
Miniature Fuse, 5 x 20 mm, Time Delay
250 VAC 0.1 A
F91 Miniature Fuse, 5 x 20 mm, Time Delay 250 VAC 1.0 A
F101
Miniature Fuse, 5 x 20 mm, Time Delay
250 VAC 2.0 A
Fuse specifications for 120 VAC systems
1 Housed in fuse terminal blocks. Illuminated LED indicates blown fuse.
SS2100i-2 TDLAS Gas Analyzer Operating Instructions
Endress+Hauser 19
Fig 11. Label placement on exterior of enclosures
1. Chassis ground label 4. Chassis ground label
2. Manufacturer label 5. Port label
3. Class 1 laser product label
1
2
3
4
5
5
Operating Instructions SS2100i-2 TDLAS Gas Analyzer
20 Endress+Hauser
Fig 12. Label placement on electronics panel assembly
1. Electric shock warning label 6. Functional ground earth label
2. Energized fuse warning label 7-8. General warning label
3. General warning label 9. Electric shock warning label
4. Electric shock warning label 10. Functional earth ground label
5. Fuse rating label 11. Protective earth ground label
1
2
3
4
5
6
7
8
9
10
11
/