Cirrus CR:514 User manual

Type
User manual

This manual is also suitable for

Cirrus CR:514 is a handheld acoustic calibrator designed to calibrate sound level meters. It provides a stable and accurate 94 dB sound pressure level at 1 kHz, making it suitable for calibrating a wide range of sound level meters.

The CR:514 is easy to use, with a simple on/off switch and a clear indicator light. It is powered by a single 9v alkaline battery, which provides approximately 15 hours of continuous use.

The CR:514 is a valuable tool for anyone who needs to calibrate sound level meters, including:

  • Acoustical consultants
  • Industrial hygienists
  • Environmental engineers
  • Audiologists

Cirrus CR:514 is a handheld acoustic calibrator designed to calibrate sound level meters. It provides a stable and accurate 94 dB sound pressure level at 1 kHz, making it suitable for calibrating a wide range of sound level meters.

The CR:514 is easy to use, with a simple on/off switch and a clear indicator light. It is powered by a single 9v alkaline battery, which provides approximately 15 hours of continuous use.

The CR:514 is a valuable tool for anyone who needs to calibrate sound level meters, including:

  • Acoustical consultants
  • Industrial hygienists
  • Environmental engineers
  • Audiologists
CR:514 & CR:515 Acoustic Calibrator
Page 1
CR:514 & CR:515
Acoustic Calibrator
User Manual
This manual, the software to which it relates, the
program code and drawings are all:
© Copyright Cirrus Research plc 1989-2013
CR:514 & CR:515 Acoustic Calibrator
Page 2
The content of this manual, any illustrations, technical information and descriptions within this document
were correct at the time of going to print. Cirrus Research plc reserves the right to make any changes
necessary, without notice, in line with the policy of continuing product development and improvement.
No part of this publication may be duplicated, reprinted, stored in a data processing system or transmitted
by electronic, mechanical, photographic or other means, or recorded, translated, edited, abridged or
expanded without the prior written consent of Cirrus Research plc.
No liability is accepted for any inaccuracies or omissions in this manual, although due care has been taken
to ensure that is it complete and accurate as possible.
Accessories supplied by Cirrus Research plc have been designed for use with the instrumentation
manufactured by Cirrus Research plc. No responsibility is accepted for damage caused by the use of any
other parts or accessories.
In order to take account of a policy of continual development, Cirrus Research plc reserves the right to
change any of the information contained in this publication without prior notice.
Produced by Cirrus Research plc, Acoustic House, Bridlington Road, Hunmanby, North Yorkshire, YO14
0PH, United Kingdom.
© Copyright Cirrus Research plc 2013
Reference Number 01/13/CR514&515/03
Document Printing Date Tuesday, 08 January 2013
CR:514 & CR:515 Acoustic Calibrator
Page 3
Operation. .................................................................................................................... 4
Switching on the Calibrator ............................................................................................. 4
Permanent-on Mode ................................................................................................................. 4
Calibrating a Sound Level Meter. ..................................................................................... 5
Background Noise .......................................................................................................... 5
Stabilisation .................................................................................................................. 5
Changing the Battery .................................................................................................... 6
Battery type. ................................................................................................................ 6
Specification. ................................................................................................................ 7
Appendix 1 Technical Information ............................................................................. 8
Appendix 2 Free Field Correction ............................................................................... 9
Microphone Correction Values .................................................................................................... 9
Example ................................................................................................................................. 9
Appendix 3 - CE Declaration of Conformity ................................................................. 10
Appendix 4 Type Approval Certification ...................................................................... 11
Warranty Information. ............................................................................................... 13
Cirrus Research Offices ......................................................... Error! Bookmark not defined.
CR:514 & CR:515 Acoustic Calibrator
Page 4
Operation.
Switching on the Calibrator
Press the Power Button on the end of the Calibrator to switch the unit on. The Indicator
will illuminate to show that the unit is operating.
The calibrator will automatically switch off after 5 minutes to preserve battery power.
To switch off the calibrator manually, press the power button again and the indicator
will extinguish to show that the unit is switched off.
Permanent-on Mode
For some applications there may be a need to have the calibrator switched on
continuously. To allow for this, the calibrator can be turned on by pressing and holding
the power button for three seconds.
Release the button and the indicator will flash to show that the unit is in permanent-on
mode. Press the power button to switch off the calibrator.
CR:514 & CR:515 Acoustic Calibrator
Page 5
Calibrating a Sound Level Meter.
Push the microphone of the Sound Level Meter into the cavity at the end of the
calibrator. Ensure the microphone is fully inserted into the cavity and is past the ‘O’ ring
seals. The microphone should be parallel to the body of the calibrator. Also ensure that
the small bleed-hole next to the microphone cavity is not blocked as this could cause
damage to the microphone.
Most modern Sound Level Meters have electronic calibration with the level adjusted
automatically. Adjust the Sound Level Meter to the correct level where applicable.
When correcting the value generated by the calibrator a correction for the type of
microphone capsule may need to be applied (see Appendix 2)
Background Noise
In order for the calibrator to operate as intended, the ambient acoustic noise level
should be no greater than 80dBA.
Stabilisation
In order for the sound pressure level and frequency to stabilise after switching the
calibrator on when coupled to a microphone, a period of at least 3 seconds should be
allowed before performing a calibration.
CR:514 & CR:515 Acoustic Calibrator
Page 6
Changing the Battery
The CR:514 & CR:515 acoustic calibrators use a single 9v alkaline battery. This type of
battery is known as 6F22 or NEDA 1604. It is also commonly known as PP3.
1. Unscrew the screw holding the battery cover on, using a coin.
2. The battery, type 6F22 (PP3) can now be eased out of its holder and replaced.
The battery should be eased out terminal side first by pushing against the spring
at the other end. Ensure that the battery is inserted with the correct polarity
with the negative terminal at the contact with the larger cutout.
Battery type.
The battery should be an alkaline battery, not an ordinary dry cell. The battery is 9
volts when new and will operate the calibrator down to 6.4 volts. When the battery
voltage is below 6.6 volts but above 6.4 volts, the power LED will flash to indicate that
the battery voltage is low. When the battery voltage is below 6.4 volts the calibrator
will not turn on. A discharged battery may allow switch-on but will soon drop in voltage
and indicate low battery or switch off.
CR:514 & CR:515 Acoustic Calibrator
Page 7
Specification.
Frequency 1kHz ± 1%
Sound Level 94dB re 20μPa
Standardisation CR:514 - IEC 60942:2003 Class 2
CR:515 - IEC 60942:2003 Class 1
Distortion Less than 2%
Operating Humidity 25 to 90% Relative Humidity
Operating Static Pressure 65 kPa to 108kPa
Operating Temperature -10
o
C to +50
o
C
Storing Temperature -20
o
C to +60
o
C
Effective Volume 6.19 cm
3
± 0.2 cm
3
Cavity Diameter 0.525 inch
Battery 1 x 9v 6F22 (Neda 1604)
Battery Life Approx 15 Hours Continuous Use
Battery Voltage 9v Nominal (10v Maximum, 6.4v Minimum)
Weight with Battery 185g
Dimensions 135mm x Ø48mm
CR:514 & CR:515 Acoustic Calibrator
Page 8
Appendix 1 Technical Information
The normal mode of operation of the calibrator is with the unit switched on.
When the LED indicates the unit is switched on this produces the greatest radio
frequency emissions.
The calibrator continues to function after exposure to contact discharges up to 4kV and
air discharges up to 8kV, for both positive and negative voltages relative to earth
ground.
The calibrator conforms to IEC 60942:2003 for a modulated root-mean-square
electromagnetic field strength of 10 V/m.
The maximum susceptibility to power and radio frequency fields is with the cavity facing
away from the emitter with the battery compartment facing the table, the antenna
polarisation horizontal and the calibrator switched on.
CR:514 & CR:515 Acoustic Calibrator
Page 9
Appendix 2 Free Field Correction
When calibrating a microphone which is to be used for free field measurements, a small
correction may be necessary to compensate for the difference between the
microphone's free field response at 'zero degrees' or 'head-on' incidence and the
pressure level generated by the calibrator.
The correction is typically -0.3dB for ½ inch microphones (making the effective
calibration level 93.7dB).
The table below shows the correction values for the standard microphones of Cirrus
Research plc.
Calibration corrections are listed below for the Cirrus Research plc ½" Capsules and
three microphone capsules commonly used in Calibration Laboratories:
Microphone Correction Values
Microphone Type Calibration Correction Effective Calibration Level
MK:202 -0.3dB 93.7 dB
MK:215 -0.3dB 93.7 dB
MK:216 -0.3dB 93.7 dB
MK:226 -0.3dB 93.7 dB
MK:224 -0.3dB 93.7 dB
B&K 4134 0dB 94.0 dB
B&K 4180 0dB 94.0 dB
B&K 4192 0dB 94.0 dB
Example
An example of the procedure used to calculate the value for an MK:224 microphone is
shown below :
Level = 94.0dB + Microphone Correction
Level = 94.0dB + ( -0.3dB)
Level = 93.7dB
Different microphones will have different correction values. Please check the operation
manual for the Sound Level Meter or microphone concerned for details.
CR:514 & CR:515 Acoustic Calibrator
Page 10
Appendix 3 - CE Declaration of Conformity
Cirrus Research plc Hunmanby UK
CE Certificate of Conformity
Manufacturer: Cirrus Research plc
Acoustic House, Bridlington Road
Hunmanby, North Yorkshire, YO14 0PH
United Kingdom
Telephone +44 1723 891655
Equipment Description
The following equipment manufactured after 1
st
January 2007:
CR:514 Acoustic Calibrator
CR:515 Acoustic Calibrator
Along with their standard accessories
According to EMC Directives 89/336/EEC and 93/98/EEC
meet the following standards
EN 61000-6-3 (2001)
EMC : Generic emission standard for residential, commercial and light industrial
environments.
EN 61000-6-1 (2001)
EMC : Generic immunity standard for residential, commercial and light industrial
environments.
Signed Dated 1
st
January 2007
S. ORourke
Director
CR:514 & CR:515 Acoustic Calibrator
Page 11
Appendix 4 Type Approval Certification
CR:514 & CR:515 Acoustic Calibrator
Page 12
CR:514 & CR:515 Acoustic Calibrator
Page 13
Warranty Information.
1. This document is a summary of the full warranty document and explains the Cirrus
Research plc warranty in ordinary English; not in legal or complex terms.
2. The warranty covers any acoustic instrument such as a sound level meter, acoustic
calibrator, real time acoustic analyser or personal sound exposure meter (dosemeter)
manufactured by Cirrus Research plc after March 1st 2007.
3. The warranty covers all faults on the instrument except the microphone and the
display for the period defined in para (4) below, including minor accidental damage
except to the microphone or display.
4. The period of the warranty is 2 (two) years or 104 weeks from the date of purchase as
a new instrument from Cirrus Research plc or their formally approved distributors OR 130
weeks from the date the instrument passed its final manufacturing inspection at Cirrus
Research plc - whichever is the shorter.
5. Any rechargeable battery only has the battery manufacturer’s one year warranty.
6. No warranty is offered for used equipment unless a special arrangement is made and a
written confirmation of the warranty is given by Cirrus Research plc.
7. On completion of the “Routine Verification” by Cirrus Research plc, the instrument will
automatically be given an additional free one year warranty.
8. There will be a charge for this routine verification and the price is published in the
Service Price List.
9. It follows that should the instrument be routinely verified by Cirrus Research plc every
year, the warranty is effectively continuous to a maximum of 15 (fifteen) years from the
date of purchase.
10. Cirrus Research endeavour to ensure stocks of instrument components for the full
fifteen year period but do not guarantee to do so as certain components do become
obsolete or discontinued.
11. If a sub-component becomes obsolete and stocks are depleted then Cirrus Research
will endeavour to facilitate a repair but will not offer the same length guarantee.
12. In the event of any dispute on the terms of the warranty Cirrus Research plc will
accept pendulum arbitration by the United Kingdom Institute of Acoustics Ltd.
13. The warranty does not in any way reduce any legal right of the buyer or user of the
sound level meter; it is in addition to all legal rights determined by the European Union.
CR:514 & CR:515 Acoustic Calibrator
Page 14
Cirrus Research Offices
The addresses given below are the Cirrus Research plc offices. Cirrus Research plc also have
approved distributors and agents is many countries worldwide. For details of your local
representative, please contact Cirrus Research plc at the address below. Contact details for
Cirrus Research authorised distributors and agents are also available from the Internet Web site
at the address shown below.
Main Office
Cirrus Research plc
Acoustic House
Bridlington Road
Hunmanby
North Yorkshire
United Kingdom
YO14 0PH
Telephone: +44 (0)1723 891655
Fax: +44 (0)1723 891742
E-mail: sales@cirrusresearch.co.uk
Web Site: www.cirrusresearch.co.uk
Germany
Cirrus Research plc Deutschland
Arabella Center
Lyoner Strasse 44 48
D-60528 Frankfurt
Germany
Tel: +49 (0)69 95932047
Fax +49 (0)69 95932049
Email: vertrieb@cirrusresearch.de
Web: www.cirrusresearch.de
France
Cirrus France Ltd
40 Bis Avenue Gabriel Fauré
09500 Mirepoix
France
Tel: +33 5 61 67 40 01
Fax: +33 5 61 67 40 56
Email: sales@cirrusresearch.fr
Web: www.cirrusresearch.fr
Spain
CIRRUS RESEARCH S.L.
Travessera de Gracia, 62 4° 7ª
08006 Barcelona
SPAIN
Tel: (34) 933 622 891
Email: info@cirrusresearch.es
Web: www.cirrusresearch.es
Cirrus Environmental
Unit 2 Bridlington Road Industrial Estate
Hunmanby
North Yorkshire
YO14 0PH
United Kingdom
Tel: +44 (0) 1723 891722
Email: sales@cirrus-environmental.com
Web: www.cirrus-environmental.com
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14

Cirrus CR:514 User manual

Type
User manual
This manual is also suitable for

Cirrus CR:514 is a handheld acoustic calibrator designed to calibrate sound level meters. It provides a stable and accurate 94 dB sound pressure level at 1 kHz, making it suitable for calibrating a wide range of sound level meters.

The CR:514 is easy to use, with a simple on/off switch and a clear indicator light. It is powered by a single 9v alkaline battery, which provides approximately 15 hours of continuous use.

The CR:514 is a valuable tool for anyone who needs to calibrate sound level meters, including:

  • Acoustical consultants
  • Industrial hygienists
  • Environmental engineers
  • Audiologists

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI