OPERATING
AND
SERVICE
MANUAL
HP
11793A
M
I
C
R
0
WAVE
CONVERTER
September
1990
1 1793-9001 1
111111111111111111111111111111111111111111111111111111111111ll111
HEWLETT
PACKARD
CERTl FlCATlON
Hewlett-Packard Company certifies
that
this
product met its published specifications
ut
the time of shipment
from the factory. Hewlett-Packard further certifies that its calibration measurements are traceable to the United
States National Bureau of Standards, to the extent allowed by the Bureau’s calibration facility, and to the
calibration facilities of other International Standards Organization members.
WARRANTY
This Hewlett-Packard instrument product
is
warranted against defects in material and workmanship for a period
of one year from date
of
shipment. During the warranty period, Hewlett-Packard Company will at its option,
either repair or replace products which prove to be defective.
For warranty service or repair, this product must be returned
to
a service facility designated by HP. Buyer shall
prepay shipping charges to HP and
HP
shall pay shipping charges
to
return the product
to
Buyer. However, Buyer
shall pay all shipping charges, duties, and taxes for products returned
to
HP from another country.
HP warrants that its software
and
firmware designated by HP for use with an instrument will execute
its
programming instructions when properly installed on that instrument. HP does not warrant that the operation
of the instrument, or software, or firmware will be uninterrupted or error free.
LIMITATION OF WARRANTY
The foregoing warranty shall not apply
to
defects resulting from improper or inadequate maintenance by Buyer,
Buyer-supplied software or interfacing, unauthorized modification or misuse, operation outside
of
the environmental
specifications for the product, or improper site preparation or maintenance.
NO OTHER WARRANTY
IS
EXPRESSED OR IMPLIED. HP SPECIFICALLY DISCLAIMS THE IMPLIED
WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
EXCLUSIVE
REMEDIES
THE REMEDIES PROVIDED HEREIN ARE BUYERS SOLE AND EXCLUSIVE REMEDIES. HP SHALL
NOT BE LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL
DAMAGES, WHETHER BASED ON CONTRACT, TORT, OR ANY OTHER LEGAL THEORY.
ASSISTANCE
Product maintenance agreements and other customer assistance agreements are available
for
Hewlett-Packard
products.
For
any assistance, contact your nearest Hewlett-Packard Sales and Service Office. Addresses are provided at
the back of
this
manual.
Agilent Technologies.
Inc
24001
E
Mission
Liberty take, WA
99019
www
agilent
corn
e.
,
.'::is:.
.
Ag
i
I
en
t
Techno
Io
g
i
es
:
-
:
.
Innovating the
HP
Way
.*a*..
June
8,2000
Dear Customer,
As of November
1,
1999,
four
of
Hewlett-Packard's businesses, test and measurement,
semiconductor products, health care solutions, and chemical analysis became a new company,
Agilent Technologies. Now, many of your Hewlett-Packard products and services are in the care of
Agilent Technologies.
At Agilent Technologies, we are working diligently to make this transition as smooth as possible for
you. However, as a result of this transition, the products and related documentation contained in this
shipment may be labeled with either the Hewlett-Packard name and logo, the Agilent Technologies
name and logo, or a combination of both. Information in this package may refer to Hewlett-Packard
(HP), but applies to your Agilent Technologies product. Hewlett-Packard and Agilent branded
products with the same model number are interchangeable.
Whatever logo you see, the information, products, and services come from the same reliable source.
If
you have questions about Agilent Technologies products and services, please visit our website at
httR://www.aqilent.com.
Sincerely,
Rebranding
Team
HP
11793A
MICROWAVE
CONVERTER
SERIAL
NUMBERS
Operating and Service Manual Part
No.
11793-90011
Microfiche Operating and Service Manual Part
No.
11793-90012
Printed.
MAY
1985
This manual applies directly to instruments with serial
numbers prefixed 2520A.
With changes described
in
Paragraph
31,
this manal
also applies
to
instruments with serial numbers pre-
fixed
2407A.
For
additional important information about serial
numbers, see INSTRUMENTS
COVERED
BY
THIS
MANUAL in Paragraph
4.
First Edition
OHEWLElT-PACKARD COMPANY 1985
EAST 24001
MISSION
AVENUE, TAF C-34, SPOKANE, WASHINGTON, U.S.A., 99220
1
Regulatory
Information
(Updated
March
1999)
1
Regulatory Information (Updated March
1999)
Safety Considerations for this Instrument
WARNING
This product
is
a
Safety Class
I
instrument (provided with a
protective earthing ground incorporated in the power cord). The
mains plug shall only be inserted
in
a socket outlet provided with
a
protective earth contact. Any interruption of the protective
conductor inside or outside
of
the product is likely to make the
product dangerous. Intentional interruption is prohibited.
Whenever it
is
likely that the protection has been impaired, the
instrument must be made inoperative and be secured against any
unintended operation.
If
this instrument is to be energized via an auto transformer (for
voltage reduction), make sure the common terminal
is
connected to
the earth terminal of the power source.
If
this product
is
not used as specified, the protection provided by
the equipment could be impaired. This product must be used in
a
normal condition (in which all means
for
protection are intact) only.
No
operator serviceable parts in this product. Refer servicing to
qualified personnel. To prevent electrical shock, do not remove
covers.
Servicing instructions are
for
use by qualified personnel only.
To
avoid electrical shock, do not perform any servicing unless you are
qualified to do
so.
The opening of covers
or
removal of parts is likely to expose
dangerous voltages. Disconnect the product
from
all
voltage sources
while it
is
being opened.
The power cord
is
connected to internal capacitors that my remain
live for
5
seconds after disconnecting the plug from
its
power supply.
For Continued protection against fire hazard, replace the line fuse(s)
only with
250
V
fuse(s)
or
the same current rating and type
(for
example, normal blow or time delay).
Do
not use repaired fuses
or
short circuited fuseholders.
Always
use the three-prong ac power cord supplied with this
product. Failure to ensure adequate earth grounding by not using
this cord may cause product damage.
This product
is
designed
for
use in Installation Category
I1
and
Pollution Degree
2
per
IEC
1010
and
IEC
664
respectively.
FOR
INDOOR
USE
ONLY.
This product has autoranging line voltage input, be sure the supply
voltage
is
within the specified range.
3
Model
Safety Considerations
SAFETY
CONSIDERATIONS
GENERAL
This product and related documentation must be
reviewed before operation for familiarization with
safety markings
and
instructions.
This product
is
a Safety Class
I
instrument (pro-
vided with
a
protective earth terminal).
BEFORE APPLYING POWER
Verify that the product
is
set
to
match the available
line voltage and the correct fuse is installed.
SAFETY EARTH GROUND
An uninterruptible safety earth ground must be
provided from the main power source
to
the prod-
uct input wiring terminals, power cord, or supplied
power cord set.
SAFETY SYMBOLS
Instruction manual symbol: the product
A
will be marked with this symbol when
it
is
necessary for the user
to
refer
to
the instruction
manual (refer
to
Table of Contents).
Indicates hazardous voltages.
1,
Indicates earth (ground)
terminal.
f
-
The WARNING sign denotes
a
hazard. It calls attention to a
procedure, practice, or the like, which,
if
not cor-
rectly performed
or
adhered
to,
could result in per-
sonal injury.
Do
not proceed beyond
a
WARNING
sign until the indicated conditions are
fully
under-
stood and met.
The CAUTION sign denotes
a
hazard. It calls attention
to
an
operating procedure, practice, or the like, which,
if
not correctly performed or adhered
to,
could result
in
damage
to
or destruction of
part
or
all
of the
product.
Do
not proceed beyond a CAUTION sign
until the indicated conditions are fully understood
and met.
I
WARNING
1
Interrupting
the
protective (grounding) conductor
(inside
or
outside
the
instrument)
or
disconnecting
the protective earth terminal will cause a potential
shock
hazard that could result
in
personal injury.
(Grounding one conductor of a two conductor outlet
is
not sufficient protection).
Whenever
it
is
likely that
the
protection has been
impaired,
the
instrument
must
be made inopera-
tive and be secured against any unintended
operation.
If
this
instrument
is
to be energized via an auto-
transformer (for voltage reduction), make sure the
common terminal
is
connected to the earth termi-
nal of
the
power source.
Servicing instructions are
for
use by
service-trained personnel only.
To
avoid dangerous
electric shock, do not perform any servicing unless
qualified to do
so.
Adjustments described
in the
manual are per-
formed with power supplied to the instrument
while protective covers are removed. Energy avail-
able at many points may,
if
contacted, result
in
per-
sonal
injury.
Capacitors inside the instrument may still be
charged even
if
the instrument
has
been discon-
nected from its source of supply.
For
continued protection against fire hazard, re-
place
the
line fuse(s) only with
250V
fuse(s) of the
same current rating and type (for example, normal
blow, time delay, etc.).
Do
not use repaired fuses
or
short-circuited fuseholders.
..
11
Microwave
Converter
Model
11793A
MODEL 11793A
IF
INPUT CABLE
(HP Part Number 11 793-60006)
FREQUENCY CONTROL CABLE
(HP Part Number 8120-1838)
--
LO INPUT CABLE
(HP Part Number 11793-60009)
LINE POWER CABLE
Figure
1.
HP
Model
II793A
Microwave
Converter
with
Accessory
Cables
0
Model
1
1793A
Microwave Converter
LO
Amplitude Range
+8
dBm to
+13
dBm
$7
dBm to
+13
dBm
0 dBm to
+5
dBm
1.
GENERAL INFORMATION
Frequency
(GHz)
Conditions
2
to
18
18
to
26.5
18
to
26.5
Without Option
001, 011
or
021
Without Option
001, 011
or
021
With Option
001, 011
or
021
This Operating and Service Manual contains information about initial inspection, operation, performance
tests, adjustments, troubleshooting and repair of the
HP
Model
1
1793A
Microwave Converter.
2,
Specifications
The
HP
1
1793A
Microwave Converter does not have "hard specifications" against which the instrument
can be tested for performance standards or limits. Instead, the Local Oscillator
(LO)
Requirements are
listed in Table
1.
These requirements are the performance standards or limits against which the
instrument can be tested. The supplemental characteristics listed in Table
2
are not warranted
requirements but are typical characteristics included as additional information for the user.
Table
2.
Supplemental Characteristics
RF
Input Connector:
Precision
3.5
mm male
LO
Input Connector:
Precision
3.5
mm male
IF
Output Connector:
Type-N female
Rear Panel Control Connector:
BNC
female
Dimensions:
Height:
88
mm
(3.5
in.)
Width:
425
mm
(16.7
in.)
Depth:
528
mm
(20.8
in.)
HP System
II
Module Size:
Height:
88
mm
(3.5
in.)
Width:
1
Modular Width
Depth:
508
mm
(20
in.)
1
Model 11793A Microwave Converter
1
I
Not
installed Standard Option
010
Option
020
installed Option
001
Option
011
Option 021
6.
Options and Accessories
Electrical Options
The options are summarized in Table
3.
The following list describes each option:
Option
001.
Option
010.
Adds an 18 to 26.5
GHz
LO
amplifier to the standard instrument.
Moves the LO INPUT to the right side of the front panel.
Option
011.
Moves the
LO
INPUT to the right side of the front panel and adds the 18 to 26.5
GHz
LO
amplifier.
Option
020.
Moves all front-panel connectors to the rear panel.
Option
021.
Moves all front-panel connectors to the rear panel and adds the 18 to 26.5 GHz
LO
amplif ier.
Table
3.
HP
11
793A Option
Guide
Front-Panel Placement Rear Panel
of
LO
Input Placement
for
I
Len Side Right Side
All
Connectors
LO
Amplifier
Mechanical Options
Instruments with Serial Prefix 2407A have System
11
cabinet parts and hardware that are English thread.
Instruments with Serial Prefix 2520A and higher have System
I1
cabinet parts and hardware that are
METRIC thread.
When ordering Option 907, 908, or 909 for instruments with serial number prefix 2407A (English
thread), use the HP part number in parenthesis.
Option
907
(Front-Handle Kit). Ease of handling is increased with front-panel handles.
To retrofit a
standard instrument, order
HP
part number 5061-9688 (5061-0088).
Option
908
(Rack-Flange Kit). All necessary hardware and
installation instructions are included to
enable mounting
of
the Microwave Converter in a rack with 482 mm (standard 19 in.) spacing.
To
retrofit
a standard instrument, order HP part number 5061-9674 (5061-0074) which has standard 4.45 cm (1.75
in.) hole spacing.
Option
909
(Rack-Flange and Front-Handle Combination Kit). This kit
is
not simply a front-handle kit
and rack-flange kit packaged together. The combination is made up of unique parts which include both
functions.
To
retrofit a standard instrument, order HP part number
SO6
1-9675
(SO6
1-0075) which has
4.45 cm (1.75 in.) hole spacing and is a standard flange.
Microwave Converter Model 11793A
Option 907
Front Handle Kit
Option 908 Option 909
Rack Flange
Kit
Rack Flange and Front
Handle Combination
Kit
Figure
2.
Rack
Mounting
Options
Accessories Supplied
The accessories supplied with the Microwave Converter are shown in Figure 1.
a.
b.
C.
d.
e.
The line power cable may be supplied in several plug configurations, depending on the destination of
the original shipment. Refer to the Paragraph 12
"Power Cables"
in this manual.
Fuses with a 0.7SA rating for both
11
5
and 230 Vac (HP part number 21 10-0063) are supplied.
The control cable
(
HP
part number 8 120- 18 38) is a 30 cm
(1
2 in.) flexible cable with male BNC 50
ohm connectors on each end. It connects to the frequency offset mode output of the measuring
instrument.
The
LO
cable (HP part number 11793-60009) is a 36 cm (14.5 in.) flexible cable with
3.5
mm
female end connectors. The
LO
cable includes a 3.5 mm male to Type-N male adapter. The cable
connects to tlie output
of
the
LO
source. Options
001,
01
1, and 021 delete the adapter.
The
IF
cable (HP part number 11793-60006)
is
a 25 cm (10 in.) flexible cable with Type-N male
50
ohm end connectors. The cable connects to the
RF
input of the measuring instrument.
4
Model
11793A
Microwave Converter
Instrument Type
Measuring Instrument
Sensor Module
7.
Recommended
Test
Equipment
Minimum Specification Suggested Model
Contains specific circuitry to control the HP 11793A.
Must have a frequency range of 26.5 GHz and be compatible
with the measuring instrument.
HP 8901 B or HP 8902A'
HP 11792A'
Table
4
lists the test equipment recommended to check, adjust, and troubleshoot the Microwave
Converter. The Minimum Specifications column describes the essential requirements for each piece of test
equipment. If substitute equipment is used, it must meet or exceed the minimum specifications.
Microwave LO Source
Microwave Signal
Source
Table
4.
Recommended Test Equipment
See
LO
Amplitude Range in Table 1,
LO
Requirements. Note
that the required output power depends on the
LO amplifier
option in the HP
11
793A.
Frequency Range:
2
to 26.5GHz
Output Level:
-1
to +1 dBm
HP 86738"
HP 86738
I
To Troubleshoot the Microwave Converter
I
[
Instrument Type
I
Minimum Specification
I
Multimeter DC Volts Range: 1OOV
DC Volts Accuracy: 2%
AC Volts Range: lOOV
AC Volts Accuracy: 10% at 50 Hz
RF Test Equipment
Troubleshooting of the microwave components of the Microwave Converter requires
measurements to 26.5GHz of gain or insertion
loss,
SWR,
power, conversion
loss,
and
frequency. Instruments which are required to perform these tasks include: signal genera-
tors and sweepers, network analyzers, spectrum analyzers, and power meters with ac-
companying couplers, pads, line stretchers, etc. Select the appropriate equipment and
use standard test procedures to make these measurements.
8.
INSTALLATION
9.
Initial
inspection
Inspect the shipping container for damage. If the shipping container or cushioning material
is
damaged,
the shipping materials should be kept until the contents
of
the shipment have been checked for
completeness and the instrument has been checked mechanically and electrically. If there
is
mechanical
damage or if the instrument does not pass the performance tests, notify the nearest Hewlett -Packard
office. Keep the damaged shipping materials
(if
any) for inspection by the carrier and a Hewlett-Packard
representative.
5
Model
1
179
3A
I
8120-1351
8120-1703
Microwave Converter
0
4
Table
2-2.
AC Power Cables Available
8120-1378
8120-1398
8120-1754
8120-1378
8120-1521
8120-1676
8120-2104
8120-0698
8120-1957
8120-2956
8120-4211
8120-1860
z!q
Number
5
5
7
1
6
2
3
6
2
3
7
6
Cable
Color
For
Use
In Country
Cable
Length
(inches)
90
90
Plug
Description
I
250V
Straight'BS1363A
90"
Mint Gray
Mint Gray
United Kingdom,
Cyprus, Nigeria,
Rhodesia,
Singapore
I:".J
00
I
250V
Gray
Gray
Austrailia,
New Zealand
79
87
79
79
Mint Gray
Mint Gray
East and West
Europe, Saudi
Arabia, Egypt,
(unpolarized in
many nations)
Straight'CEE7-Y11
goo
81 20-1 692
I
125V
United States,
Canada, Mexico,
Japan (1OOV or
200V), Philippines,
Taiwan
Straight'NEMA5-15P
90"
Straight'NEMA5-15P
Straight'NEMA5-l5P
90"
Straight'NEMA5-15P
Straight'SEVl 01
1
Type 12
1959-24507
80
80
36
80
80
36
79
Black
Black
Black
Jade Gray
Jade Gray
Jade Gray
Gray
I
250 V
Switzerland
~~
United
States,
Canada
Straight'NEMAG-15P
79
79
Gray
Gray
Denmark
Straight'DHCK107
90"
I
250V
Straight'lEC83-B1 79 Black
South Africa, India
Straig ht'CEE22-VI
(Systems Cabinet Use)
'Part number shown for plug
is
industry identifier for plug only. Number shown for cable
is
HP Part Number for complete
cable including plug.
E
=
Earth Ground; L
=
Line;
N
=
Neutral
7
Microwave Converter Model 11793A
14.
Operating Environment
The operating environment should be within the following limitations:
Temperature
..........................
0°C to +55"C
Humidity
..............................
~95% relative at 40°C
Altitude
...............................
44570 meters
(1
5
000
feet)
15.
Bench Operation
The instrument cabinet
has
plastic feet which are shaped to insure self-alignment when stacked in a
system configuration.
16.
Rack Mounting
Rack mounting information is provided with the rack mounting kits. If the kits were not ordered with the
instrument as options, they may be ordered through the nearest Hewlett-Packard office. Refer to
Paragraph
6,
under "Mechanical Options" for kit information. See Figure 2,
Rack
Mounting Options.
17.
Storage and Shipment
The instrument should be stored in a clean, dry environment. The following environmental limitations
apply to both storage and shipment.
Temperature (Standard)
.......................
-
5
5°C
to
+7
S"C
Temperature (Options 001,
01
1, and 021)
....
-25°C to +7S"C
Humidity
......................................
49
5%
relative
Altitude
.......................................
15
300
meters
(SO
000
feet)
18.
Packaging
Original Packaging. Containers and materials identical to those used in factory packaging are available
through Hewlett -Packard offices. If the instrument
is
being returned to Hewlett -Packard for servicing;
attach a tag indicating the type of service required, return address, model number, and full serial number.
Also, mark the container
FRAGILE
to ensure careful handling. In any correspondence refer to the
instrument by model number and full serial number.
Other Packaging. The following general instructions should be used for repackaging the instrument with
commercially available materials.
a. Wrap the instrument
in
heavy paper
or
plastic. (If shipping to a Hewlett-Packard office or
service center, attach a tag indicating the type of service required, return address, model number,
and full serial number.)
b.
Use a strong shipping container.
A
double-wall carton made of 2.4 MPa
(350
psi) test material is
adequate.
8
Model
11
793A
Microwave Converter
c.
Use enough shock-absorbing material; a
7.5
to
10.0
cm
(3
to
4
in.) layer around all sides of the
instrument to provide a firm cushion and to prevent movement in the container. Protect the
front panel with cardboard.
d. Seal the shipping container securely.
e. Mark the shipping container
FRAGILE
to ensure careful handling.
19.
Operation
Before the Microwave Converter is plugged in,
all
protective earth
terminals, extension cords, auto -transformers, and devices connected
to
it
should be connected
to
a protective earth grounding socket. Any
interruption
of
the protective earth grounding will cause a potential
shock hazard that could result in personal injury.
Before the Microwave Converter is plugged
in,
it must be set
to
the
same line voltage as the power source
or
damage
to
the instrument
may result.
The
HP
11
793A
Microwave Converter is designed to be operated remotely by the measuring instrument
in its frequency offset mode. Operation of the Microwave Converter is, therefore, under control of the
measuring instrument.
20.
Operating Instructions
To
operate the Microwave Converter, plug
in
the power cord and connect the signal and control cables as
indicated in Figure
3,
Microwave Converter Interconnections
and
Test Set-up.
Refer to Table
1,
LO
Requirements,
for
LO
power requirements.
This instrument does
not
have an
orl/off
switch
or
front-panel
power indicator light. Power is supplied
to
the instrument as soon as
the power cord is plugged in
to
the line voltage source.
9
Microwave Converter Model
11793A
FREPUENCY
OUTPUT
OFFSET TTL
0
I
:::::
D00000
0000
0
0
0000
00
0
0000
0
0
0
000000
00000
Q
HP
6902A
MEASURING RECEIVER
SI
II
FREOUENCY
I
OFFSET MODE RF
INPUT INPUT
I
OR
1
Figure
3.
Microwave Convertor Interconnections and Typical Test Setup
21.
Panel Features
Front-panel features consist of the
LO
input, RF input, and the IF output connectors. For Options
020
and
021,
the connectors are located on the rear panel. See Figure
4,
Front-Panel Features
for the
standard locations of front-panel connectors.
Rear-panel features are the Line Voltage Select Switch, Fuse, Power Cord and the Frequency Offset input
connectors. See Figure
5,
Rear -Panel Features.
22.
Operator Checks
Operator’s checks consist of insuring that a fuse of proper value is installed, that the power cord is plugged
in, that the signal/control lines are connected, and that both the measuring instrument and the external
LO
are connected and turned on.
23.
Operator Maintenance
The only maintenance needed by the Microwave Converter is replacement of the primary power fuse.
To replace the fuse, insert a standard screwdriver into the fuseholder and push in against the spring. Turn
counterclockwise (while pushing with a slight pressure) until the fuseholder is released. Remove and
replace the fuse with
a
new fuse of identical value.
10
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50

HP HP 11793A User manual

Category
PC/workstation barebones
Type
User manual
This manual is also suitable for

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI