  • Hello! I've reviewed the user manual for the CO2Meter TR250Z Oxygen Sensor, which includes models CM-0134, CM-0134-WT, CM-0150, CM-0160, CM-0160-WT, and CM-0161. This manual covers important aspects like calibration methods, power requirements, and communication protocols such as MODBUS. I'm ready to answer your questions about setting up and using these oxygen sensors, or any other details covered in the provided document.
  • What is the recommended calibration frequency for the oxygen sensor?
    How can I calibrate the sensor?
    What type of power supply is needed for the unit?
    What is the life expectancy of the sensor?
    Which communication protocols are supported?
© CO2Meter, Inc.
TR250Z Oxygen Sensor
User Manual
Models
CM013425%O2Transmitter
CM0134‐WT25%O2TransmitterwithTubeSampling
CM015025%O2withMountShield
CM016095%O2Transmitter
CM0160WT95%O2TransmitterwithTubeSampling
CM016195%O2withMountShield
TR250Z Oxygen Transmitter – Manual
2
Table of Contents
TABLEOFCONTENTS................................................................................................................2
WELCOME...............................................................................................................................3
IMPORTANTSAFEGUARDS.......................................................................................................3
PACKAGECONTENTS...............................................................................................................4
SOFTWARE..............................................................................................................................4
MINIMUMSYSTEMREQUIREMENTS........................................................................................4
DESCRIPTIONS.........................................................................................................................5
THEORYOFOPERATION..............................................................................................5
STANDARDOPERATION...........................................................................................................6
TERMINALCONNECTORS.............................................................................................6
POWERINGTHEUNIT..................................................................................................7
010VOUTPUT(OPTIONAL).......................................................................................7
LCDDISPLAY............................................................................................................7
CALIBRATION...........................................................................................................................7
CALIBRATINGUSINGMODBUS....................................................................................7
CALIBRATINGUSINGGASLAB........................................................................................7
MANUALCALIBRATION...............................................................................................8
MODBUS/RS485.................................................................................................................10
INPUTREGISTERS....................................................................................................10
HOLDINGREGISTERS...............................................................................................10
DEVICESPECIFICATIONS.........................................................................................................11
SUPPORT...............................................................................................................................12
WARRANTY...........................................................................................................................12
RETURNS...............................................................................................................................12
LIABILITY...............................................................................................................................13
CONTACTUS..........................................................................................................................13
TR250Z Oxygen Transmitter – Manual
3
Welcome
ThankyouforpurchasingtheTR250ZOxygenSensor.TheTR250Zisalonglife25%or95%
oxygenleveltransmitterthatprovidesmultipleanaloglinearoutputsofO2concentrations.It’s
designedtobearugged,maintenancefreeaddonforOEMsorcanbeusedasadesktopdevice.
Inaddition,itcanbecalibratedinnormalair,orcanuseacalibrationgasofanyknownoxygen
concentration.
Ourgoalistomakehighquality,lowcostgasdetectionandindoorairqualityproducts
availabletoyou.
Pleasetakesometimetoreadthroughthismanualinordertogetthemostoutofyourmeter.
Makesuretovisitourwebsiteregularlytofindmoreinformationaboutthisproductandto
downloadthesoftwarefree.Onceyoudownloadandinstallthesoftware,youcanusethe
“Checkforupdates”functionundertheHelpmenutosearchfornewupdates.Pleasealsopay
specialattentiontotheimportantsafeguardsshownonthepagesahead.
Important Safeguards
Toreducetheriskoffire,electricalshockand/orinjurytopersons,basicsafetyprecautionsshould
alwaysbefollowedwhenusingelectricalappliances,includingthefollowing:
1. READALLINSTRUCTIONSBEFOREUSINGTHISDEVICE!
2. Useonlythesuppliedpowersupplytooperatetheunit.
3. Ensurethatwhensamplinginaclosedenvironmentthatthetubesaresecurelyfashionedtothe
device.
4. Donotoperatewithablockedoffsamplepath.
5. Donotoperatethedeviceifitismalfunctioning.
6. Donotinstallthedeviceoutsideinexposedconditions.Duetothenatureofthedesignthe
enclosurecannotbemadewatertightandtheunitmustnotbeexposedtowater.
7. Donotoperatethedevicewiththecoverdetached.
SAVE THESE INSTRUCTIONS!
TR250Z Oxygen Transmitter – Manual
4
Package Contents
Pleaseverifythatyourpackagecontainsthefollowingitemsbeforeusingtheunit:
Contents:
(1)Meter
(1)CertificateofCalibration
(1)UserManual
(1)Powersupply
(1)2MeterUSBtoRS485Cable
(1)ScreenBreather
Software
OurfreeGaslab®softwareletsyouautomatethecalibrationprocessaswellaslogdata.Whileitisnot
requiredtousethesensor,youmayfindituseful.ToinstallGaslab,gototheDownloads”page(see
linkbelow)atCO2Meter.com.Forgeneralinformationaboutthesoftware,refertotheGasLab®User
Manual.http://www.co2meter.com/pages/downloads
IMPORTANT:YOUMUSTINSTALLTHESOFTWAREBEFORECONNECTINGYOURDEVICETOYOURCOMPUTER
Byinstallingthesoftwarefirst,youwillensurethattheproperdriversareautomaticallyinstalledon
yourcomputerwhenyouplugintheUSBdeviceforthefirsttime.
Minimum System Requirements
Toutilizeourfreesoftware,yourcomputermustmeetthefollowingrequirements:
WindowsXPSP3orhigher
Microsoft.NetFramework3.5SP1
Pentium4(ornewer)operatingat2.4Ghzorfaster
1GBofRandomAccessMemory(RAM)
Harddiskspacewithatleast20Megabytes(MB)free(200+MBrecommendedforlogsand
applicationfiles)
Softwareiscompatiblewith64bitoperatingsystemsandisfullytestedwithWindows.
ToinstallGaslab,clickonthe“InstallNow”linkandthenselect“Run”whenpromptedbyyourbrowser
(asshownonError!Referencesourcenotfound.).Followthestepsandinstructionspromptedbyyour
computer’soperatingsystem.Makesureyouhaveadministratorprivilegesinordertoinstallthis
program.
TR250Z Oxygen Transmitter – Manual
5
Descriptions
DisplayMaininterfacetoviewcurrentsensorreadingsandmenustate.Willblink“STRT”upon
applyingpowerwhilethesensorwarmsupandstabilizes.
LEDsareinforlowerrightcorneroftheunitonthemainPCBandarelabeledS,EandP.
SLEDStatusLED.Blinkseverytimeacompletemeasurementcycleisperformedandthevalueis
updated.Serve’sasaheartbeatfortransmitteroperation.
ELEDErrorLED.Willlightifanyabnormalconditionsaredetectedwithinthedevice,primarilythe
sensorfallingoutofitsrecommendedoperatingspecificationswilltriggerthislamp.
PLEDPowerLED.Willbeoncontinuouslywhenpowerisappliedtotheunit.
Buttons‐TheunithasthreesurfacemountbuttonsabovetheLEDsonthePCB.Theycanbegentlyand
carefullypressedwithyourfinger.Thesebuttonsallowyoutoaccessallaspectsofsensorsoperation
andtestorcalibratedifferentfunctionalmodules.
Theory of Operation
Thezirconiaoxygensensorutilizessolidstateelectrochemicalreactionstoproduceavoltage
proportionaltothecurrentoxygen.Thisvoltageisthendigitalizedandfilteredtoausablevalue.The
oxygensensorhasalsobeenfactorycalibratedtomeetspecifications.
TR250Z Oxygen Transmitter – Manual
6
Standard Operation
TheO2transmitterscomesfactorycalibratedforeasyinstallation.Duringitslifecyclecalibrationmay
becomeperiodicallynecessary.WerecommendperformingaspanO2calibrationeverymonthfor
guaranteedoperationandensuredaccuracy.Noothercalibrationisnecessaryduringthelifecycleofthe
sensor.
Terminal Connectors
1. VDC=G‐G+
2. O2420mA=SIGGND
3. RS485INTERFACE=ABGND
4. RESET=RST
TR250Z Oxygen Transmitter – Manual
7
Powering the Unit
Theunitcanbepoweredbyhardwiring24VDCpowerthroughtheMDCM4FP2MCable(included)but
samplingunitscanalsobepoweredbytheincluded24VDCInternationalwallpowersupply.
0-10V Output (Optional)
Togeta10voltspan(210voltinpractice)froma420mAline,adda500ohmresistorfromsignalto
ground.500ohmsx.004amps=2voltsand500ohmsx.020amps=10volts.
Installtheresistoratthereceivingend,asthiseliminatesallvoltagedropsfromresistanceinthewire.
420mAand210Vareusedsoitcanbedeterminedifthetransmitterisactuallypresent.Lessthan4mA
or2V:notransmitter,notconnectedornosignal.
LCD Display
TheLiquidCrystalDisplay(LCD)screenshowsthefollowingfeatures:
CO2inpercentageformat(##.##%)for100%
Calibration
Thecalibrationprocessvariesdependingonthetypeofunitandwhetherithasoptionaldatalogging
functionalityornot.Allunitsarefactorycalibratedwithmultiplereferencepointsofgas,andhavebeen
verifiedtobeaccuratewithintheirspecificfunctionalitybeforeshipment.However,iftheunitis
severelyjoltedorotherwisemechanicallydisturbed,thesensorcandriftrequiringrecalibration.All
calibrationproceduresfollowasinglepointcalibrationroutinethateffectivelyshiftsthezeropointof
theO2sensor.
CalibrationmaybeperformedviaMODBUS,manually,orbyaUSBconnectiontoaPCrunningGasLab®
software.
Calibrating using MODBUS
CalibrationviaMODBUSrequiresanRS485cableandassumesknowledgeofpropercoding.Register
valuesarelistedattheendofthismanual.
Calibrating using Gaslab
InstalltheGaslabsoftwarefirst,thenconnectthesensortoyourPCusingastandardminiUSBcable.
Calibrationcanbeperformedusingeither0%CO2calibrationgas(typicallynitrogen)orusingafresh
sourceofair,assumedtobeapproximately20.9%O2.
Followtheonscreenpromptsspecifictoyoursensortoperformthecalibration.
TR250Z Oxygen Transmitter – Manual
8
Manual Calibration
Manualcalibrationisperformedusingthe3onboardbuttonsandtheservicemenuonthedisplay.To
entertheservicemenuthebuttonsmustbepressedinafactorysetorder.Thiscodecanbecustomized
beforeshipmentbutbydefaultwillbe123.Thebuttonsmustbepressedwithinahalfsecondofeach
other,ortheunitwillreturntostandardoperatingmode.Whenintheservicemenutheunitwillnot
updatethecurrentO2readings.Onceyouenterthemenuthedisplaywillread“1234”orfor1second,
thentheProductIdentificationCodewillbedisplayed.Thiscodeisimportant,asitwillidentifywhich
manualcalibrationprocedurebelowyoushoulduse.
Product Identification Code
Afterpressingbuttons123,theunitwilldisplay“1234”,followedbya4digitProductIdentification
Code(PIC).Youmustknowyoursensor’sPICtoproperlycalibrate.
1. IfthePICis“1195”or“1225”,followthe1195/1225CalibrationProcedure
2. IfthePICis“1295”or“1395”,followthe1295/1395CalibrationProcedure
1195/1225/1425 Calibration
Procedure
Fresh Air Calibration
1. Allowsensortostabilizeforatleast1hour
2. Ensurethesensorisexposedtofreshair
3. Press123toentertheservicemenu.Thedisplaywillflash1234andthePIC
4. Press2toentertheOxygensubmenu
5. Press2againtostartthecalibration
6. Press3toSave.
Note:Steps36mustbedonequicklyorthesensorwillexitthecalibrationmode.Youcanminimizethis
byrememberingthesequence123223toperformmanualfreshaircalibration.
Zero Point Calibration using 100% Nitrogen
1. Allowsensortostabilizeforatleasthour
2. Apply100%NitrogenCalibrationGastoSensor
3. Wait5minutesforstablereadings
4. Press123toenterservicemenu.Thedisplaywillflash1234andthePIC
5. Press2toentertheOxygensubmenu
6. Press1tocalibrateto100%Nitrogen
7. Press3toSave.Nonumberswillbedisplayeduntilthesensorisexposedtooxygenagain
Note:Steps4‐7mustbedonequicklyorthesensorwillexitthecalibrationmode.Youcanminimize
thisbyrememberingthesequence123213toperformmanualfreshaircalibration.
TR250Z Oxygen Transmitter – Manual
9
1295/1395/1495 Calibration
Procedure
Fresh Air Calibration
1. Allowsensortostabilizeforatleast1hour
2. Ensurethesensorisexposedtofreshair
3. Press123toentertheservicemenu.Thedisplaywillflash1234andthePIC
4. Press2toentertheOxygensubmenu
5. Press1tostartthecalibration
6. Press3toSave.
Note:Steps36mustbedonequicklyorthesensorwillexitthecalibrationmode.Youcanminimizethis
byrememberingthesequence123213toperformmanualfreshaircalibration.
Span Calibration using 70% Oxygen (optional)
Toperformaspancalibration,only70%oxygengascanbeused.Ifyoudonothave70%O2testgas,do
notattemptaspancalibration.
1. Allowsensortostabilizeforatleasthour
2. Apply705OxygenCalibrationGastoSensor
3. Wait5minutesforstablereadings
4. Press123toenterservicemenu.Thedisplaywillflash1234andthePIC
5. Press2toentertheOxygensubmenu
6. Press2tocalibrateto70%Oxygen
7. Press3toSave.Nonumberswillbedisplayeduntilthesensorisexposedtooxygenagain
Note:Steps4‐7mustbedonequicklyorthesensorwillexitthecalibrationmode.Youcanminimize
thisbyrememberingthesequence123223toperformmanualfreshaircalibration.
TR250Z Oxygen Transmitter – Manual
10
Modbus / RS-485
ThesensorsupportMODBUSprotocolsoverRS485,unidirectionaltransmissionat9600kbps.Address0
orFE.
ThefollowingModbusregistersareavailable:
Input Registers
1. O2Value,x10
2. RawADCValue
Holding Registers
1. O2SpanValue(4bytefloatingnumber,part1)
2. O2SpanValue(4bytefloatingnumber,part2)
3. O2PotValue
4. 420mAoutputZeroValue
5. 420mAoutputSpanValue(4bytefloatingnumber,part1)
6. 420mAoutputSpanValue(4bytefloatingnumber,part2)
7. ControlRegister
8. StatusRegister
Thefollowingcommandsareavailable:
0x03ReadHoldingRegister
0x04ReadInputRegister
0x06WriteHoldingRegister
TR250Z Oxygen Transmitter – Manual
11
Device Specifications
MeasuringRange:
25%O2 025%vol.
95%O2 095%vol.
Repeatability:
025%CO2 ±0.05%,±1%of measuredvalue
095%O2 ±0.02%,±1%of measuredvalue
Accuracy
25%/95%±2%fullscale
PowerSupply:
MaximumVoltage 24VDC
MinimumVoltage 20VDC
PowerConsumption 600mA@24VDC
SensorRatings:
LifeExpectancy >15years
MaintenanceInterval Nomaintenancerequired
WarmupTime <15min(instantmeasurements)
Outputs:
O2OutputValue420mA,linearlyscaled
210VDCOptional
RS485 ModbusInterface
TR250Z Oxygen Transmitter – Manual
12
Support
Thequickestwaytoobtaintechnicalsupportisviaemail.Pleasesendallsupportenquiresto
[email protected].Inyouremail,pleaseincludeaclear,concisedefinitionoftheproblemandany
relevanttroubleshootinginformationorstepstakensofar,sowecanduplicatetheproblemandquickly
respondtoyourinquiry.
Warranty
Thisunitcomeswitha90-day limitedmanufacturer’swarranty,startingfromthedatetheunitwas
shippedtothebuyer.
Duringthisperiodoftime,CO2Meter.comwarrantsourproductstobefreefromdefectsinmaterials
andworkmanshipwhenusedfortheirintendedpurposeandagreestofixorreplace(atourdiscretion)
anypartorproductthatfailsundernormaluse.Totakeadvantageofthiswarranty,theproductmustbe
returnedtoCO2Meter.comatyourexpense.If,afterexamination,wedeterminetheproductis
defective,wewillrepairorreplaceitatnoadditionalcosttoyou.
Thiswarrantydoesnotcoveranyproductsthathavebeensubjectedtomisuse,neglect,accident,
modificationsorrepairsbyyouorbyathirdparty.NoemployeeorresellerofCO2Meter.com’sproducts
mayalterthiswarrantyverballyorinwriting.
Returns
Iftheproductfailsundernormaluseduringthewarrantyperiod,anRMA(ReturnMaterial
Authorization)numbermustbeobtainedfromCO2Meter.com.Aftertheitemisreceived,
CO2Meter.comwillrepairorreplacetheitematourdiscretion.
ToobtainanRMAnumber,pleasecallCO2Meter.comat(877)6784259.WhenrequestinganRMA
number,pleaseprovidethereasonforreturnandoriginalordernumber.
Ifwedeterminethattheproductfailedduetoimproperuse(waterdamage,dropping,tampering,
electricaldamageetc.)orabuse,orifitisbeyondthewarrantyperiod,wewillinformyouofthecostto
fixorreplaceyourdevice.
Ifyouarereturningyourdeviceduetoawarrantyclaim(withanRMAnumber)andyoustillhavethe
unitoriginalpackage,pleaseuseittoshipyourunittous.Pleasemakesuretoincludetheprovided
RMAnumberontheoutsideofthebox,preferablyontheshippinglabel.Makesureyousecuretheunit
insidethepackageproperlytopreventanydamageduringtransitthatcouldvoidyourdevice’s
warranty.Finally,pleaseshipyourdevicetotheaddressshownunderthe“ContactUs”sectionbelow.
CO2Meter.comwillnot,underanycircumstances,beresponsibleforyourshipmentexpensesandno
refundwillbeissuedforshippingchargesnecessaryforyoutoshiptheunittous.
TR250Z Oxygen Transmitter – Manual
13
Liability
Allliabilitiesunderthisagreementshallbelimitedtotheactualcostoftheproductpaidto
CO2Meter.com.InnoeventshallCO2Meter.combeliableforanyincidentalorconsequentialdamages,
lostprofits,lossoftime,lostsalesorlossordamagetodata,injurytopersonorpersonalpropertyor
anyotherindirectdamagesastheresultofuseofourproducts.
Contact Us
Weareheretohelp!
Ifthetroubleshootingguideabovedoesn’thelpyousolvingyourproblemorformoreinformation,
pleasecontactususingtheinformationbelow.
(877) 6784259Tollfree(MF9:00am–6:00pmEST)
(866) 4222356
www.co2meter.com
Address:
CO2Meter
131BusinessCenter
BuildingA,Unit3
OrmondBeach,FL32174
/