Roche cobas u 701 User manual

Type
User manual
cobas u 701 microscopy
analyzer
Operator’s Manual Version 2.4
Software Version 2.3
Roche Diagnostics
cobas u 701 microscopy analyzer · Software Version 2.3 · Operator’s Manual · Version 2.4
2
Document information
Edition notice This information is intended for operators and
administrators of the cobas u 701 microscopy analyzer.
Every effort has been made to ensure that all the
information is correct at the time of publishing. However,
Roche reserves the right to make any changes necessary
without notice as part of ongoing product development.
Customer modifications to the instrument may impair
instrument safety and may lead to malfunction, incorrect
measurements, and incorrect results. Any customer
modification to the instrument will render the warranty or
service agreement null and void.
Intended use The cobas u 701 microscopy analyzer is a fully
automated urine microscopy system intended for the in
vitro quantitative determination of erythrocytes and
leukocytes, the semi-quantitative determination of
squamous and non-squamous epithelial cells, bacteria,
and hyaline casts and the qualitative determination of
pathological casts, crystals, yeasts, mucus, and sperm in
urine.
These measurements are useful in the evaluation of renal,
urinary, hepatic and metabolic disorders. This system is
intended to be used by trained operators in clinical
laboratories.
Manual version Software version Revision date Change description
1.0.0 2.0 May 2014 First publication
1.0.0 2.1 January 2015 Software upgrade. No content changes in operator’s manual.
2.0.0 2.2 July 2015 o Illustrations were adapted to latest hardware and software.
o New: Optional input connection unit.
o New: Optional connection to external water supply.
o Improved result presentation.
o New: Emergency stop feature
o New: Definition of notification periods.
o Changes to range tables and cross-check rule
configuration.
o New: Definition of particle subclasses.
o Various improvements in configuration features.
o New: Procedure for adjusting probe action to different
tubes and racks.
2.1.0 2.2.3 July 2016 o Improved working with Sample sequence number
mode.
o Improved manual image analysis.
o Definition of STAT racks.
2.2 2.3 March 2018 o New version (2) of control unit.
o Windows 10 operating system.
2.4 2.3 March 2020 o New layout.
o Linguistic and minor content adjustments.
y Revision history
Roche Diagnostics
cobas u 701 microscopy analyzer · Software Version 2.3 · Operator’s Manual · Version 2.4
3
Copyright © 2014–2020, Roche Diagnostics GmbH. All rights
reserved.
Trademarks The following trademarks are acknowledged:
COBAS, COBAS U, and LIFE NEEDS ANSWERS are
trademarks of Roche.
All other product names and trademarks are the property
of their respective owners.
Instrument approvals The cobas u 701 microscopy analyzer meets the
requirements laid down in:
Directive 98/79/EC of the European Parliament and of
the Council of 27 October 1998 on in vitro diagnostic
medical devices
Directive 2011/65/EU of the European Parliament and
of the Council of 8 June 2011 on the restriction of the
use of certain hazardous substances in electrical and
electronic equipment
cobas u 701 microscopy analyzer is manufactured and
tested according to the international safety standards
below:
UL 61010-1, 2
nd
Edition
IEC 61010-1, 2
nd
Edition
IEC 61010-2-081, 1
st
Edition
IEC 61010-2-101, 1
st
Edition
CAN/CSA C22.2 No. 61010 2nd Edition
EN IEC 61326-1 2nd Edition
EN IEC 61326-2-6 2nd Edition
The Operator’s Manual meets the European Standard DIN
EN ISO 18113-3.
Compliance with the applicable directives is provided by
means of the Declaration of Conformity. The following
marks demonstrate compliance:
For in vitro diagnostic use.
Complies with the provisions of the applicable EU directives.
Issued by Underwriters Laboratories, Inc. (UL) for Canada and
the US.
IVD
Roche Diagnostics
cobas u 701 microscopy analyzer · Software Version 2.3 · Operator’s Manual · Version 2.4
4
Contact address
Roche Diagnostics GmbH
Sandhofer Strasse 116
68305 Mannheim
Germany
Made in Hungary
Table of contents 5
Roche Diagnostics
cobas u 701 microscopy analyzer · Software Version 2.3 · Operator’s Manual · Version 2.4
Table of contents
Document information 2
Contact address 4
Table of contents 5
Preface 7
How to use this manual 7
Symbols and abbreviations 9
What is new in publication version 2.2 11
What is new in publication version 2.1.0 11
What is new in publication version 2.0.0 12
System description
1 Safety
Safety classification 21
Safety precautions 22
Safety summary 26
Safety labels on the equipment 33
Disposal of the equipment 39
2 Introduction
General description 43
Daily operation 52
3 Hardware
Safety 61
Main components 63
Covers 64
Connectors 66
Power switches 72
Input and output buffers 74
Tubes, racks, and rack trays 78
Liquid containers 83
Solid waste container 87
Rack transport unit 88
Fluid system 89
Sample handling 95
Cuvette handling 96
Barcode reader 101
Barcodes 102
Radio frequency identification 103
Technical specifications 105
4 Software
Introduction 115
Key screen elements 115
Key work areas 121
Tabs 123
Displaying information 125
Working with lists (tables) 128
Entering information 129
Working with QC charts 131
External keyboard and mouse 132
Wizards 133
Color coding 136
Online help 137
Operation
5 Operation
Safety 145
Short guide to routine testing 147
Routine operating tasks 150
Result handling 168
Non-routine situations 179
Urine sediment image evaluation 196
Managing patients 211
Routine maintenance actions 213
At the end of the shift 225
Checking the microscope focusing
mechanism 240
QC tasks 242
Additional operating tasks 253
6 Configuration
User management 259
System settings: Defining the test
environment 263
System configuration: Defining the operating
environment 285
Managing racks 299
Adjusting the probe action 301
Maintenance
7 Maintenance
Safety 307
Routine maintenance 308
Miscellaneous maintenance actions 311
6 Table of contents
Roche Diagnostics
cobas u 701 microscopy analyzer · Software Version 2.3 · Operator’s Manual · Version 2.4
Troubleshooting
8 Troubleshooting
Exceptional situations 325
Screenshots 327
Log files 328
No microscope check results can be
generated 330
Detached barcode labels 331
Recovering from an irregular stop 332
Emergency stop 333
When you have accidentally pulled the waste
drawer during operation 335
Recovering from a power outage 336
Safety interlock 337
Clogged inlet water filter 337
Blocked floats 340
Glossary
9 Glossary
Index
Index 349
Roche Diagnostics
cobas u 701 microscopy analyzer · Software Version 2.3 · Operator’s Manual · Version 2.4
7
Preface
The cobas u 701 microscopy analyzer (called microscopy
analyzer in this documentation) is a fully automated urine
microscopy system intended for the in vitro quantitative
determination of erythrocytes and leukocytes, the semi-
quantitative determination of squamous and non-
squamous epithelial cells, bacteria, hyaline casts, and the
qualitative determination of pathological casts, crystals,
yeasts, mucus, and sperm in urine.
These measurements are useful in the evaluation of renal,
urinary, hepatic and metabolic disorders. This system is
intended to be used by trained operators in clinical
laboratories.
The analyzer can process up to 116 samples per hour.
How to use this manual
q Keep this Operator’s Manual in a safe place to
ensure that it is not damaged and remains available for
use.
This Operator’s Manual should be easily accessible at all
times.
Throughout this documentation, images of screens are
included for illustration purposes. They are not
necessarily identical with what you see on your analyzer.
To help you find information quickly, there is a table of
contents at the beginning of the manual and of each
chapter. In addition, a comprehensive index can be found
at the end.
Roche Diagnostics
cobas u 701 microscopy analyzer · Software Version 2.3 · Operator’s Manual · Version 2.4
8
Content overview The documentation is divided into the following parts:
Part A — System Description: The system
description provides information on the safe operation
of the system, on its hardware and software, and an
overview of how it is used.
Part B — Operation: The operation part describes
how to perform the various tasks that are required
when analyzing samples.
Part C — Maintenance: The maintenance part
provides information on how to ensure continuous
problem-free operation of the system.
Part D — Troubleshooting: The troubleshooting part
provides assistance on how to deal with exceptional
situations.
Part E — Glossary: The glossary explains key terms
used in the user interface and in this documentation.
Part F — Index: The index provides a quick way to
information, it consists of an alphabetical list of key
words that lead to the relevant information within this
documentation.
Roche Diagnostics
cobas u 701 microscopy analyzer · Software Version 2.3 · Operator’s Manual · Version 2.4
9
Symbols and abbreviations
Visual cues are used to help locate and interpret
information quickly. This section explains the conventions
used for this purpose.
Symbols used in this publication The following symbols are used:
Symbols used on the instrument The following symbols are used:
Symbol Explanation
o List item.
q
Tip. Extra information on correct use or useful
hints.
Start of a procedure.
I Extra information within a task.
f Result of an action within a task.
d Materials that are required for a task.
j Prerequisites of a task.
u
Topic. Used in cross-references to topics and
for related topics containing further
information.
w
Figure. Used in figure titles and cross-
references to figures.
y
Table. Used in table titles and cross-references
to tables.
y Symbols used in the publication
Symbol Comment
Safety alert.
Electrical and electronic equipment marked
with this symbol are covered by the European
directive WEEE.
Label on the water container for external
water supply indicating monthly cleaning.
y Symbols used for easy recognition of information
Roche Diagnostics
cobas u 701 microscopy analyzer · Software Version 2.3 · Operator’s Manual · Version 2.4
10
Abbreviations The following abbreviations are used:
Symbol Comment
Global trade item number
Date of manufacture
Serial number
y Symbols used for easy recognition of information
Abbreviation Explanation
A ampere
BAC bacteria
CRY crystals
CSA Canadian Standards Association
CSV character separated values
e.g. exempli gratia – for example
EC European Community
EMC electromagnetic compatibility
EN European standard
HPF high-power field
HYA hyaline casts
Hz hertz
i.e. id est – that is to say
IEC International Electrical Commission
IVD in vitro diagnostic directive
LAS laboratory automation system
LIS laboratory information system
max. maximum
min. minimum
mm millimeter
MUC mucus
n/a not applicable
NEC non-squamous epithelial cells
nm nanometer
PAT pathological casts
QC quality control
RBC red blood cells
RD Roche Diagnostics
y Abbreviations
Roche Diagnostics
cobas u 701 microscopy analyzer · Software Version 2.3 · Operator’s Manual · Version 2.4
11
What is new in publication version 2.2
New control unit version The control unit was replaced with a new version (version
2).
Windows 10 operating system The internal PC uses the Windows 10 operating system.
Upgraded test strip tray and transporter The test strip tray and test strip transporter were
enhanced.
Replacing of reference plate The reference plate can only be replaced by a Roche
Service representative.
What is new in publication version 2.1.0
Illustrations and screenshots Illustrations and screenshots were adapted to reflect the
latest hardware and software.
Optional components Colored labels for Roche 5-position racks are now
available.
u Optional components (110)
Working with Sample sequence number mode The procedure for working with Sample sequence
number mode has been improved.
u Defining the sample sequence number ranges (153)
u Rerunning tests when working with Sample sequence
number mode (182)
RFID radio frequency identification
SEC squamous epithelial cells
SPRM sperm
STAT short turn around time
UL Underwriters Laboratories Inc.
UPS uninterrupted power supply
V volt
VA volt-ampere
VAC volt alternating current
Wwatt
WBC white blood cells
YEA yeasts
Abbreviation Explanation
y Abbreviations
Roche Diagnostics
cobas u 701 microscopy analyzer · Software Version 2.3 · Operator’s Manual · Version 2.4
12
STAT and routine results are marked when
working with Sample sequence number mode
If you work with Sample sequence number mode,
sample IDs for routine test results are marked with an “N”
preceding the sample sequence number, and “E” for
STAT test results.
u Validating results (169)
Manual particle classification In a given image, you can now classify multiple particles
in one go. You can also remove several classifications in
one go.
u To assign or reclassify particles (189)
u To remove the classification from particles (190)
Working with Sample sequence number mode The information on defining the validation method has
been adapted.
u Defining the validation method (270)
STAT racks You can now define dedicated STAT racks.
u To define a STAT rack (299)
What is new in publication version 2.0.0
Illustrations and screenshots Illustrations and screenshots were adapted to reflect the
latest hardware and software.
Automatic rack feed, input connection unit You can optionally connect the analyzer to a laboratory
automation system by replacing the input buffer with an
input connection unit. This allows automatic rack feed to
the analyzer.
u Input connection unit (76)
u Operation with input connection unit (161)
u To load a priority rack when working with an
LAS (164)
u To perform a QC measurement when working with an
LAS (244)
u To clean the inlet water filter (external water
supply) (339)
u To loosen the floats in the water container for external
water supply (341)
Roche Diagnostics
cobas u 701 microscopy analyzer · Software Version 2.3 · Operator’s Manual · Version 2.4
13
External water supply You can optionally connect the analyzer to an external
laboratory water supply system. When using this feature,
the liquid waste is drained off directly to the laboratory
liquid waste system.
u Water container for external water supply (84)
u Liquid waste with external water supply (86)
u To clean the inlet water filter (external water
supply) (339)
u To loosen the floats in the water container for external
water supply (341)
Image zoom function You can now zoom in on individual microscope images.
u Using the zoom function (126)
Technical specifications Some values were adjusted.
External conditions
Power requirements
Water quality
Wash solution
Cleaning solution
u Technical specifications (105)
Safety information Input connection unit
Information on the proper use of the following items was
added:
u Input connection unit (76)
u Water container for external water supply (84)
u Racks (79)
u To adjust the probe action (301)
u Safety labels on the equipment (33)
Overview work area The categorizing of the task list has been improved.
An emergency stop button (E. Stop) was added. Use
this function if, for some reason, all activities on the
analyzer must be stopped immediately or if the
analyzer is stuck in either the Operating or Init
status.
When working with a laboratory automation system,
the Priority rack button is available on the Overview
work area.
u Emergency stop (333)
u To load a priority rack when working with an
LAS (164)
Roche Diagnostics
cobas u 701 microscopy analyzer · Software Version 2.3 · Operator’s Manual · Version 2.4
14
Result presentation To assist people with color vision deficits, in addition
to the colors hatching is displayed.
New symbols and data alarms were introduced to
provide more information about the status of the
result.
u Color coding (136)
u Checking the status of processing (165)
u Validating results (169)
u To review QC results (251)
QC charts The QC charts function has been improved.
u Working with QC charts (131)
Manually analyzing microscope images The descriptions of manually analyzing images was
adapted and extended.
u Manually analyzing images (186)
Printing and exporting information The feature was adapted and applied consistently.
Some changes:
In the result report, the results of subclasses follow
immediately the results of their main class.
Screenshots are no longer part of the problem report,
instead, you can save them separately.
Particle labels can now be saved with the images.
u Printing and exporting information, generating
reports (255)
Emergency stop An emergency stop feature has been introduced for
situations when all activities on the analyzer must be
stopped immediately or if the analyzer is stuck in either
the Operating or Init status.
u Emergency stop (333)
Cleaning the pipetting stage area The procedure for cleaning the pipetting stage area was
adapted.
u Cleaning the pipetting stage area (235)
Cleaning the centrifuge chamber The procedure for cleaning centrifuge chamber was
adapted.
u Cleaning the centrifuge chamber (236)
Warning intervals You can now define how long before the event a warning
is issued for expiry of materials and maintenance actions.
u Defining when notifications should be
generated (291)
Roche Diagnostics
cobas u 701 microscopy analyzer · Software Version 2.3 · Operator’s Manual · Version 2.4
15
QC materials The conditions for making changes to QC materials have
changed.
u To change QC material data (247)
u To make test parameter related changes (248)
u To include or exclude tests from the QC
measurements (249)
Defining QC materials using the RFID reader The procedure for defining QC materials using the RFID
reader was adapted.
u To define a new QC material by reading the RFID
tag (246)
Report definition The items have been regrouped and complemented.
u Defining the look, content, and handling of
reports (293)
Range tables and cross-check rules The procedures for defining range tables and cross-
check rules were simplified.
u Defining range tables (274)
u Defining cross-check rules (271)
Particle subclasses You can now examine particles that cannot be classified
as a subclass of an existing main class. Such particles are
defined as subclasses of the new generic main class
Others.
u Defining particle subclasses (282)
Semi-quantitative reporting of RBC and WBC A description was added on how to proceed if you want
to display RBC and WBC results on a semi-quantitative
level.
u Displaying RBC and WBC results on a semi-
quantitative level (278)
Importing system settings You can now import system settings that were generated
using software versions other than the current one.
u Importing and exporting system settings (297)
Adjusting the probe action To enable the use of different racks and tubes, a function
is now provided to adjust the probe action to the changed
dimensions.
u To adjust the probe action (301)
Roche Diagnostics
cobas u 701 microscopy analyzer · Software Version 2.3 · Operator’s Manual · Version 2.4
16
Roche Diagnostics
cobas u 701 microscopy analyzer · Software Version 2.3 · Operator’s Manual · Version 2.4
System description
1 Safety .......................................................................................................................... 19
2 Introduction .............................................................................................................. 41
3 Hardware................................................................................................................... 59
4 Software...................................................................................................................113
1 Safety
Roche Diagnostics
cobas u 701 microscopy analyzer · Software Version 2.3 · Operator’s Manual · Version 2.4
19
Table of contents
Safety 1
In this chapter you find information about the safe
operation of the equipment.
In this chapter
1
Safety classification . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Safety precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Operator qualification . . . . . . . . . . . . . . . . . . . . . . . 22
Safe and proper use of the analyzer . . . . . . . . . . . 22
Personal safety . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Personal protective equipment . . . . . . . . . . . . . 23
Accuracy and precision of measured results . 23
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Environmental conditions . . . . . . . . . . . . . . . . . 24
Approved parts . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Third-party software. . . . . . . . . . . . . . . . . . . . . . 24
Miscellaneous safety precautions . . . . . . . . . . . . . 24
Safety summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Warning messages . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Electrical safety . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Biohazardous materials . . . . . . . . . . . . . . . . . . . 27
Waste . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Barcode readers . . . . . . . . . . . . . . . . . . . . . . . . . 28
Foam, bubbles or films on sample . . . . . . . . . . 28
Data security . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Caution messages . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Mechanical safety. . . . . . . . . . . . . . . . . . . . . . . . 30
Working solutions . . . . . . . . . . . . . . . . . . . . . . . . 30
Insoluble contaminants in samples . . . . . . . . . 31
Influence of vibrations . . . . . . . . . . . . . . . . . . . . 31
Malfunction due to interfering
electromagnetic fields . . . . . . . . . . . . . . . . . . . . 31
Fatigue due to long hours of operation . . . . . . 31
Notices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Moving parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
1 Safety
Roche Diagnostics
cobas u 701 microscopy analyzer · Software Version 2.3 · Operator’s Manual · Version 2.4
20
Table of contents
Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Spillage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Excessive ambient humidity . . . . . . . . . . . . . . . 33
Influence of vibrations . . . . . . . . . . . . . . . . . . . . 33
Safety labels on the equipment . . . . . . . . . . . . . . . . . . 33
Analyzer views . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Front view . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Back view. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Solid waste compartment . . . . . . . . . . . . . . . . . 36
Input connection unit. . . . . . . . . . . . . . . . . . . . . 38
Disposal of the equipment . . . . . . . . . . . . . . . . . . . . . . 39
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133
  • Page 134 134
  • Page 135 135
  • Page 136 136
  • Page 137 137
  • Page 138 138
  • Page 139 139
  • Page 140 140
  • Page 141 141
  • Page 142 142
  • Page 143 143
  • Page 144 144
  • Page 145 145
  • Page 146 146
  • Page 147 147
  • Page 148 148
  • Page 149 149
  • Page 150 150
  • Page 151 151
  • Page 152 152
  • Page 153 153
  • Page 154 154
  • Page 155 155
  • Page 156 156
  • Page 157 157
  • Page 158 158
  • Page 159 159
  • Page 160 160
  • Page 161 161
  • Page 162 162
  • Page 163 163
  • Page 164 164
  • Page 165 165
  • Page 166 166
  • Page 167 167
  • Page 168 168
  • Page 169 169
  • Page 170 170
  • Page 171 171
  • Page 172 172
  • Page 173 173
  • Page 174 174
  • Page 175 175
  • Page 176 176
  • Page 177 177
  • Page 178 178
  • Page 179 179
  • Page 180 180
  • Page 181 181
  • Page 182 182
  • Page 183 183
  • Page 184 184
  • Page 185 185
  • Page 186 186
  • Page 187 187
  • Page 188 188
  • Page 189 189
  • Page 190 190
  • Page 191 191
  • Page 192 192
  • Page 193 193
  • Page 194 194
  • Page 195 195
  • Page 196 196
  • Page 197 197
  • Page 198 198
  • Page 199 199
  • Page 200 200
  • Page 201 201
  • Page 202 202
  • Page 203 203
  • Page 204 204
  • Page 205 205
  • Page 206 206
  • Page 207 207
  • Page 208 208
  • Page 209 209
  • Page 210 210
  • Page 211 211
  • Page 212 212
  • Page 213 213
  • Page 214 214
  • Page 215 215
  • Page 216 216
  • Page 217 217
  • Page 218 218
  • Page 219 219
  • Page 220 220
  • Page 221 221
  • Page 222 222
  • Page 223 223
  • Page 224 224
  • Page 225 225
  • Page 226 226
  • Page 227 227
  • Page 228 228
  • Page 229 229
  • Page 230 230
  • Page 231 231
  • Page 232 232
  • Page 233 233
  • Page 234 234
  • Page 235 235
  • Page 236 236
  • Page 237 237
  • Page 238 238
  • Page 239 239
  • Page 240 240
  • Page 241 241
  • Page 242 242
  • Page 243 243
  • Page 244 244
  • Page 245 245
  • Page 246 246
  • Page 247 247
  • Page 248 248
  • Page 249 249
  • Page 250 250
  • Page 251 251
  • Page 252 252
  • Page 253 253
  • Page 254 254
  • Page 255 255
  • Page 256 256
  • Page 257 257
  • Page 258 258
  • Page 259 259
  • Page 260 260
  • Page 261 261
  • Page 262 262
  • Page 263 263
  • Page 264 264
  • Page 265 265
  • Page 266 266
  • Page 267 267
  • Page 268 268
  • Page 269 269
  • Page 270 270
  • Page 271 271
  • Page 272 272
  • Page 273 273
  • Page 274 274
  • Page 275 275
  • Page 276 276
  • Page 277 277
  • Page 278 278
  • Page 279 279
  • Page 280 280
  • Page 281 281
  • Page 282 282
  • Page 283 283
  • Page 284 284
  • Page 285 285
  • Page 286 286
  • Page 287 287
  • Page 288 288
  • Page 289 289
  • Page 290 290
  • Page 291 291
  • Page 292 292
  • Page 293 293
  • Page 294 294
  • Page 295 295
  • Page 296 296
  • Page 297 297
  • Page 298 298
  • Page 299 299
  • Page 300 300
  • Page 301 301
  • Page 302 302
  • Page 303 303
  • Page 304 304
  • Page 305 305
  • Page 306 306
  • Page 307 307
  • Page 308 308
  • Page 309 309
  • Page 310 310
  • Page 311 311
  • Page 312 312
  • Page 313 313
  • Page 314 314
  • Page 315 315
  • Page 316 316
  • Page 317 317
  • Page 318 318
  • Page 319 319
  • Page 320 320
  • Page 321 321
  • Page 322 322
  • Page 323 323
  • Page 324 324
  • Page 325 325
  • Page 326 326
  • Page 327 327
  • Page 328 328
  • Page 329 329
  • Page 330 330
  • Page 331 331
  • Page 332 332
  • Page 333 333
  • Page 334 334
  • Page 335 335
  • Page 336 336
  • Page 337 337
  • Page 338 338
  • Page 339 339
  • Page 340 340
  • Page 341 341
  • Page 342 342
  • Page 343 343
  • Page 344 344
  • Page 345 345
  • Page 346 346
  • Page 347 347
  • Page 348 348
  • Page 349 349
  • Page 350 350
  • Page 351 351
  • Page 352 352
  • Page 353 353
  • Page 354 354
  • Page 355 355
  • Page 356 356