Thermo Fisher Scientific Large-Capacity Reach-In CO2 Incubator User manual

Type
User manual
Visit us online to register your warranty
www.thermoscientific.com/labwarranty
User Manual
Forma 29 cu ft Reach-In Incubator
Model 3950 and 3951
Operating and Maintenance Manual 7003950 Rev. 27
Thermo Scientific
Preface
Reach-In Incubator i
Manual Number 7003950
27 41847 11/01/17 Added Intended and NonIntended use statement bpg
26 40087/IN-4743 7/27/16 Changed the label in the remote alarm artwork to a note - pg 6-1 ccs
25 40639 5/3/16 Updated for UL (G Smith), added risk assessment info ccs
24 31580/IN-4675 10/5/15 Motor chg - part number, expl dwg, schematic ccs
23 31289 7/18/14 Removed CE reference on pg 7-3 ccs
22 30374/IN-4419 4/22/14 Updated 1900008-06 drawing - pg 8-11 (left hand door swing) ccs
21 29318//IN-4272 12/14/12 Clarified usage on exploded drawing 3950-08-4 list ccs
20 28773/IN-4231 10/8/12 Updated Figure 1-6 to current ccs
* T/C is a thermal conductivity sensor.
** All units are 50/60 Hz.
Covered
Models CO2Sensor* Voltage**
3950 T/C 100-120
3951 T/C 200-230
Thermo Scientific
ii Reach-In Incubator
Preface
Contains Parts and Assemblies
Susceptible to Damage by
Electrostatic Discharge (ESD)
CAUTION
Important Read this instruction manual. Failure to read, understand and follow the instructions in this manual
may result in damage to the unit, injury to operating personnel, and poor equipment performance. s
Caution All internal adjustments and maintenance must be performed by qualified service personnel. s
Warning If the incubator is not used in the manner specified in this operating manual, the protection provided
by the equipment design may be impaired. s
Caution When operating this unit under high temperatures and/or relative humidity (RH), caution must be
observed when the unit is shut down, unexpectedly or by choice. Condensate can form on the interior of the
chamber as it cools. This includes the heating element surfaces. Heavy condensate can bridge the ceramic
insulators and cause a temporary electrical short to ground. This electrical circuit is protected by a circuit
breaker. However, to avoid this contingency if these conditions occur, open the door of the unit and allow it to
stand open until the water vapor disperses. s
Intended Use: The Thermo Scientific Large Capacity Reach in CO2 Incubators are intended to maintain an
optimal environment for the incubation of cells and tissues for scientific research and clinical applications.
These models maintain temperature and carbon dioxiode (CO2) levels as set by the operator with increased
relative humidity.
Non-Intended Use: The 310 Series Direct Heat Incubators are not intended for use where electrical or physical
contact with the patient is established. Not intended for use by the general public. Not intended for use within
the patient environment. Not to be used outdoors. Not intended to be operated in potentially explosive
environments or for use with flammable materials.
Material in this manual is for information purposes only. The contents and the product it describes are subject
to change without notice. Thermo Fisher Scientific makes no representations or warranties with respect to this
manual. In no event shall Thermo be held liable for any damages, direct or incidental, arising out of or related to
the use of this manual.
©2000 Thermo Fisher Scientific. All rights reserved.
Contains Parts and Assemblies
Susceptible to Damage by
Electrostatic Discharge (ESD)
CAUTION
Thermo Scientific Reach-In Incubator iii
Preface
Important operating and/or maintenance instructions. Read the accompanying text carefully.
Potential electrical hazards. Only qualified persons should perform procedures associated with this
symbol.
Equipment being maintained or serviced must be turned off and locked off to prevent possible injury.
Lifting Hazard Warning. The incubator weighs more than 200lbs (91kgs). Take adequate safety
measures when moving this device.
WEEE Compliance: Thermo Fisher Scientific has contracted with companies for recycling/disposal in
each EU Member State. For further information, send an email to [email protected].
4Always use the proper protective equipment (clothing, gloves, goggles, etc.)
4Always dissipate extreme cold or heat and wear protective clothing.
4Always follow good hygiene practices.
4Each individual is responsible for his or her own safety.
Thermo Scientificiv Reach-In Incubator
Preface
Do You Need Information or Assistance on
Thermo Scientific Products?
If you do, please contact us 8:00 a.m. to 6:00 p.m. (Eastern Time) at:
1-740-373-4763 Direct
1-800-438-4851 Toll Free, U.S. and Canada
1-877-213-8051 FAX
http://www.thermofisher.com Internet Worldwide Web Home Page
service.led.marietta@thermofisher.com Tech Support Email Address
Certified Service Web Pagewww.unitylabservices.com
Our staff can provide information on pricing and give you quotations. We canSales Support
take your order and provide delivery information on major equipment items or make
arrangements to have your local sales representative contact you. Our products are listed on the
Internet and we can be contacted through our Internet home page.
Our staff can supply technical information about proper setup, operation orService Support
troubleshooting of your equipment. We can fill your needs for spare or replacement parts or
provide you with on-site service. We can also provide you with a quotation on our Extended
Warranty for your Thermo Scientific products.
Whatever Thermo Scientific products you need or use, we will be happy to discuss your
applications. If you are experiencing technical problems, working together, we will help you
locate the problem and, chances are, correct it yourself...over the telephone without a service
call.
When more extensive service is necessary, we will assist you with direct factory trained
technicians or a qualified service organization for on-the-spot repair. If your service need is
covered by the warranty, we will arrange for the unit to be repaired at our expense and to your
satisfaction.
Regardless of your needs, our professional telephone technicians are available to assist you
Monday through Friday from 8:00 a.m. to 6:00 p.m. Eastern Time. Please contact us by
telephone or fax. If you wish to write, our mailing address is:
Thermo Fisher Scientific (Asheville) LLC
401 Millcreek Road, Box 649
Marietta, OH 45750
International customers, please contact your local Thermo Scientific distributor.
Thermo Scientific Reach-In Incubator v
Preface
Warranty Notes
Information You Should Know Before Requesting Warranty Service
Locate the model and serial numbers. A serial tag is located on the unit itself.
For equipment service or maintenance, or with technical or special application inquiries, contact Technical
Services at 1-800-438-4851 or 1-740-373-4763 (USA and Canada). Outside the USA, contact your local
distributor.
Repairs NOT Covered Under Warranty
Calibration of control parameters. Nominal calibrations are performed at the factory; typically ±1°C for
temperature, ±1% for gases, and ±5% for humidity. Our service personnel can provide precise calibrations as
a billable service at your location. Calibration after a warranty repair is covered under the warranty.
Damage resulting from use of improper quality water, chemicals or cleaning agents detrimental to
equipment materials.
Service calls for improper installation or operating instructions. Corrections to any of the following
are billable services:
1) electrical service connection
2) tubing connections
3) gas regulators
4) gas tanks
5) unit leveling
6) room ventilation
7) adverse ambient temperature fluctuations
8) any repair external to the unit
Damage resulting from accident, alteration, misuse, abuse, fire, flood, acts of God, or improper
installation.
Repairs to parts or systems resulting from unauthorized unit modifications.
Any labor costs other than that specified during the parts and labor warranty period, which may
include additional warranty on CO2sensors, blower motors, water jackets, etc.
Reach-In Incubator viThermo Scientific
Table of Contents
Installation and Start-Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Control Panel Keys, Displays, & Indicators . . . . . . . . . . . . . . . . . . . . 1-2
Keypad Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Unit Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
Preliminary Cleaning and Disinfecting . . . . . . . . . . . . . . . . . . . . . . 1-4
Installing the Shelves . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-5
Leveling the Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
Attaching the Drain Lines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-5
The Water Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-6
Filling the Humidity Reservoir . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Connecting the Unit to Electrical Power . . . . . . . . . . . . . . . . . . . . 1-7
Connecting the CO2Gas Supply . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
Incubator Start-Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-9
Setting the Operating Temperature . . . . . . . . . . . . . . . . . . . . . . . . 1-9
Setting the Overtemp Setpoint . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-10
Setting the CO2Setpoint . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-10
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Calibrating the Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Calibrating the CO2System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Disable the ADD WATER Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Turn All Audible Alarms ON/OFF . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Set an Access Code . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-2
Set a Low Temp Alarm Limit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Enable Temp Alarms to Trip Relay Contacts . . . . . . . . . . . . . . . . . . .3-3
Set Low CO2Alarm Limit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Set High CO2Alarm Limit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Enable CO2Alarms to Trip Relay Contacts . . . . . . . . . . . . . . . . . . . .3-4
Set New Zero No. for New CO2Sensors . . . . . . . . . . . . . . . . . . . . . .3-5
Set New Span No. for CO2Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
Enable the High Humidity Feature . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
Select a Primary Tank w/ Gas Guard Option . . . . . . . . . . . . . . . . . . 3-6
Disable the Gas Guard System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-7
Set RS485 Address . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-7
Section 1
Section 2
Section 3
vii Reach-In Incubator Thermo Scientific
Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Sensor Fault Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Preventive Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-1
Cleaning the Glass Doors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-1
Cleaning the Humidity Reservoir . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-2
Replacing the Power Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Discarding/Removing Incubator from Service . . . . . . . . . . . . . . . . . . 5-3
Factory Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
Connect the Remote Alarm Contacts . . . . . . . . . . . . . . . . . . . . . . . . .6-1
Connect the RS485 Interface (P/N 190523) . . . . . . . . . . . . . . . . . . . 6-2
Connect Analog Output Boards (P/N 190512, 190543, & 190544) .6-3
CO2Gas Guard (P/N 1900000) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4
Connect CO2Gas Supplies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-5
De-activate Gas Guard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
Operation of CO2Gas Guard . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-5
Inner Doors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-6
Shaker Support Shelves . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-6
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
Spare Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1
Assembly Drawings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-2
Exploded Parts Drawings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-3
Electrical Schematics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-1
Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-1
Table of Contents
Section 4
Section 5
Section 6
Section 8
Section 7
Section 9
Section 10
Reach-In Incubator 1-1Thermo Scientific
Section 1 Installation and Start-Up
Figure 1-1. Front View Figure 1-2. Side View
Figure 1-3. Back View
Figure 1-4. Top View
1-2 Reach-In Incubator Thermo Scientific
Section 1
Installation and Start-Up
Figure 1-4. Control Panel Components
Tank Low Indicator - Used with Gas Guard, lights when gas tank is low,
if applicable.
Silence - Mutes the audible alarm. See Section 4 for ringback times.
Alarm Indicator - Light pulses on/off during an alarm condition in the
cabinet.
Mode Select Switch - Used to select Run, Setpoints, Calibration and
System Configuration Modes.
Message Center - Displays system status.
Mode Select Indicators -
Run: Run Menu
Set: Set Points Menu
Cal: Calibrate Menu
Config: Configuration Menu
Up and Down Arrows - Increases or decreases the number values, toggles
between parameter values of the selected mode.
Enter - Stores the changed value into computer memory.
Heat Indicator - Lights when power is applied to the heaters.
Temp Display Displays temperature continuously
Scroll for Parameters Arrows - Moves the operator through the parame-
ters of the selected mode. Right arrow goes to next, left arrow goes to
previous parameter.
CO2Inject Indicator - Lights when CO2is being injected into the incu-
bator.
% CO2 Display - Displays CO2percentage continuously.
Control Panel Keys,
Displays, & Indicators
Tank Low
Reach-In Incubator 1-3Thermo Scientific
Section 1
Installation and Start-Up
The Model 3950 Series Reach-In Incubator has four basic modes that
allow incubator setup: Run, Setpoints, Calibration and System
Configuration.
Run is the default mode during normal operation.
Set is used to enter system setpoints.
Calibration is used to calibrate various system parameters.
Configuration allows for custom setup of various options.
The chart below shows the selections under each of the modes.
Keypad Operation
RUN SETPOINT CALIBRATION CONFIGURATION
Default Mode Temperature Temp Offset Audible
Overtemp CO2 Cal Access Code
CO2 Temp Lo Alarm
Temp Relay
CO2 Lo Alarm
CO2 Hi Alarm
CO2 Relay
CO2 Z & S #'s
RH High On/Off
H20 Alarm On/Off
Tank Select
Gas Guard
RS485 Address
**Base Unit Displays
**Option Displays
Caution Units must be installed against a wall or similar structure in an
area of minimum ambient temperature fluctuation. A minimum of six (6)
inches clearance is required at the top and back of the incubator plus a
minimum three (3) inch clearance on each side for electrical, water and gas
hook-ups. At least eight (8) inches clearance above the cabinet is required
for service access. s
Locate the unit on a firm level surface capable of supporting the unit
weight of approximately 500 lbs.
Position the incubator away from doors and windows and heating and air
conditioning ducts.
Warning This incubator weighs approximately 500 lbs. Have sufficient
personnel available when moving. s
Disinfect all interior surfaces with a general-use laboratory disinfectant,
such as quaternary ammonium, to remove any residue from the production
of the incubator. Rinse thoroughly with sterile distilled water, then 70%
alcohol. Dry with a sterile cloth as needed. Disinfect the shelf channels and
shelves, then rinse with distilled water before installing.
Caution Before using any cleaning or decontamination method except
those recommended by the manufacturer, users should check with the
manufacturer that the proposed method will not damage the equipment.
Accidental spills of hazardous materials on or inside this unit are the
responsibility of the user. s
The shelves may be installed at any level in the incubator. Install a shelf
channel on each side. With the tabs
pointing up, attach the channel by
locating the rivet into a slotted hole, far
end first. Pull the channel toward the
front and slide the front rivet on the
channel into the slotted hole and press
down. Make sure that the channels are
opposite each other so that the installed
shelf will be level.
Place a bubble-type level on a shelf inside the incubator. Adjust the feet as
needed; counterclockwise to lengthen or clockwise to shorten. Level the
unit front-to-back and left-to-right.
1-4 Reach-In Incubator Thermo Scientific
Section 1
Installation and Start-Up
Preliminary Cleaning and
Disinfecting
Unit Installation
Leveling the Unit
Installing the Shelves
Figure 1-5. Shelf Channel
Attaching the Drain Lines Note Drain connections on the back of the unit are shipped capped. If
humidity is not required, the caps should be left on the drains. s
If manual or gravity feed method is used for humidity in the cabinet,
connecting the drains is optional.
If an in-house pure water supply is used to fill the reservoir, it is
recommended that the drains be connected to a floor drain.
For best operation of the incubator, sterilized distilled, demineralized or
de-ionized water should be used in the humidity reservoir. Water purity
should be in the resistance range of 50K to 1M Ohm/cm, or a
conductivity range of 20.0 to 1.0 uS/cm. Refer to ASTM Standard
D5391-93 or D4195-88 for measuring water purity.
Distillation systems, as well as some types of reverse osmosis water purity
systems, can produce water in the quality range specified. Tap water is not
recommended as it may contain chlorine, which can deteriorate the
stainless steel. Tap water may also have a high mineral content, which
would produce a build-up of scale in the reservoir. High purity or ultra
pure water is not recommended as it is an extremely aggressive solvent and
will deteriorate the stainless steel. High purity water has a resistance of
above 1M to 18M Ohm. Even high purity water can contain bacteria and
organic contaminants. Water should always be sterilized or treated with a
decontaminant, safe for use with stainless steel as well as safe for the
product, prior to being introduced into the humidity reservoir.
Reach-In Incubator 1-5Thermo Scientific
Section 1
Installation and Start-Up
Back of Unit
1/4” Water Inlet fitting
to 1/4” ID tubing
to in-house water supply
shut-off valve (30 PSIG max.)
Chamber Drain (Overflow) -
3/8” elbow with 3/8” hose barb fitting.
Install with fitting sloping slightly downward.
Condensate Drain -
3/8” fitting
to 3/8” ID hose
Back View
Floor
Figure 1-6. Top View of Fitting Connections
The Water Supply
In-house Pure Water
Supply
Filling the Humidity
Reservoir
Caution Distilled or de-ionized water used in the humidity reservoir must
be within a water quality resistance range of 50K to 1M Ohm/cm to
protect and prolong the life of the stainless steel. Use of water outside the
specified range will decrease the operating life of the unit and may void
warranty. s
The incubator is equipped with a liquid level float switch that detects low
water levels in the humidity reservoir and produces an alarm condition. As
all applications do not require humidity, this alarm can be disabled. To
disable the ADD WATER alarm, see Section 3. If humidity is required in
the current application, press the Silence key on the control panel to mute
the alarm while the humidity reservoir is being filled.
There are three methods for filling the humidity reservoir; an in-house
pure water supply, the manual and the gravity feed methods.
All incubators provide a connection for using an in-house water supply.
Connect a low pressure (not more than 30 PSI) water line, equipped with
a manual shut-off valve, to the inlet fitting on the float valve reservoir. The
1/4” hose connection is located on the rear of the incubator cabinet. With
the water turned on, a float valve inside the incubator regulates the water
level in the humidity reservoir. When the reservoir is full, a 10-second
audible tone is heard and the ADD WATER message no longer displays.
This fill system consists of a funnel attached to the side duct sheet and
tubing located inside the incubator. Disinfect the funnel and rinse with
sterile water before using. This funnel may be moved to a preferred
location by unscrewing the well nuts and disconnecting it from the
sidewall. Tighten the nuts after relocation. The holes allotted for funnel
use are slightly larger and in a centralized area. See Figure 1-6.
To fill the reservoir, pour sterile distilled water (approximately 2.5 gallons)
into the funnel. When the float switch trips, a 10-second audible tone
sounds and the ADD WATER message no longer displays. Add 1 gallon
after the 10-second tone is heard.
1-6 Reach-In Incubator Thermo Scientific
Section 1
Installation and Start-Up
The Water Supply
(continued)
Manual Method
An optional carboy water fill system is also available. The carboy option
kit includes an instruction sheet that details the installation of the system.
See the serial tag on the side of the unit or the Specifications section for
electrical specifications. Refer to the electrical schematics at the end of this
manual.
Warning Connect the incubator to a grounded, dedicated circuit. The
power cord connector is the mains disconnect device for the incubator.
Position the incubator so the unit can be easily disconnected. s
Plug the provided power cord into the power inlet connector on the back
of the cabinet, then into a grounded, dedicated electrical circuit.
The Model 3950 Series also has an internal outlet located on the right side
of the interior back wall. The outlet is to provide power (230W
maximum) to accessory equipment. This outlet is not to be used when the
temperature is above 40°C and the RH more than 50%.
Reach-In Incubator 1-7Thermo Scientific
Section 1
Installation and Start-Up
Gravity Feed Method
Pour in sterile,
distilled water
Center Front
Unit Sidewall
Figure 1-6. Manual Feed Method with Funnel
Connecting the Unit
to Electrical Power
Manual Method
(continued)
Warning High concentrations of CO2gas can cause asphyxiation.
OSHA Standards specify that employee exposure to carbon dioxide in any
8-hour shift of a 40-hour work week shall not exceed the 8-hour time
weighted average of 5000 PPM (0.5% CO2). The short-term exposure
limit for 15 minutes or less is 30,000 ppm (3% CO2). Carbon dioxide
monitors are recommended for confined areas where concentrations of
carbon dioxide gas can accumulate.
This incubator is designed for CO2gas only. Connecting a flammable or
toxic gas could result in a hazardous condition.
Gases other than CO2should not be connected to this equipment. CO2
gas cylinders have a UN1013 label on the cylinder and are equipped with a
CGA 320 outlet valve. Check the gas cylinder for the proper identification
labels. The CO2gas supply being connected to the incubator should be
industrial grade, 99.5% pure. Do not use CO2gas cylinders equipped with
siphon tubes. A siphon tube is used to extract liquid CO2from the
cylinder, which can damage the pressure regulator. Consult with your gas
supplier to ensure that the CO2cylinder does not contain a siphon tube.
Gas cylinders should also be secured to a wall or other stationary object to
prevent them from tipping.
A two-stage CO2pressure regulator is required to be installed on the outlet
valve of the gas cylinder. Input pressure to the incubator must be
maintained at 15 psig (103.4 kPa) for proper performance of the CO2
control system. A single stage CO2pressure regulator will not maintain 15
psig (103.4 kPa) to the incubator as the pressure in the CO2cylinder
decreases; therefore, a two-stage regulator is recommended.
If higher purity CO2is desired inside the incubator (greater than 99.5%
pure), the pressure regulator should be constructed with a stainless steel
diaphragm, along with specifying the purity of the CO2from the gas
supplier. Follow the manufacturer’s instructions to ensure proper and safe
installation of the pressure regulator on the gas cylinder.
Consult your facility safety officer to ensure that the equipment is installed
in accordance with the codes and regulations applicable in your area. s
The CO2gas supply being connected should be industrial grade 99.5%
pure and should not contain siphon tubes. Install a two-stage pressure
regulator at the cylinder outlet. The high pressure gauge at the tank should
have 0-2000 psig range. The low pressure gauge, at the incubator inlet,
should have a 0-30 psig range. Input pressure to the incubator must be
maintained at 15 psig (103.4 kPa), ±5 psig.
Caution Each tank should be connected to one incubator only. s
1-8 Reach-In Incubator Thermo Scientific
Section 1
Installation and Start-Up
Connecting the CO2
Gas Supply
Setting the Operating
Temperature
Incubator Start-Up
The incubator has serrated fittings on the back of the cabinet to connect
the gas supply (Figure 1-2). The fitting is labeled CO2 Inlet #1 Tank.
Secure connections with clamps. Check all fittings for leaks.
For units having the CO2Gas Guard option, refer to Section 6.
With the incubator properly installed, connected to power, the humidity
reservoir filled and the unit connected to a gas supply, system setpoints can
be entered. The following setpoints can be entered in Set Mode:
Temperature, Overtemperature and CO2. To enter Set Mode, press the
Mode key until the Set indicator lights. Press the right and/or left arrow
keys until the proper parameter appears in the message display center. See
Chart 1-1 for more detail.
This incubator has an operating temperature setpoint range of 5.0°C above
ambient to 60.0°C. It is shipped from the factory with a temperature
setpoint of 10.0° C. At this setting, all heaters are turned off. To change
the operating temperature setpoint:
1. Press the Mode key until the Set indicator lights.
2. Press the right arrow until “Temp XX.X” is displayed in the message
center.
3. Press up/down until the desired temperature setpoint is displayed.
4. Press Enter to save the setpoint.
5. Press the Mode key until the Run indicator lights for Run mode or
press the right/left arrow keys to go to next/previous parameter.
Caution Any equipment placed inside chamber must be rated for unit
operating temperature and humidity. s
Reach-In Incubator 1-9Thermo Scientific
Section 1
Installation and Start-Up
Connecting CO2Gas
Supply (continued)
Warning The independent overtemp system is designed as a safety to
protect the incubator only. It is not intended to protect or limit the
maximum temperature of the cell cultures or customer’s equipment inside
the incubator if an overtemp condition occurs. s
All Model 3950 Series incubators are equipped with a secondary
temperature monitoring system to monitor the air temperature inside the
cabinet. This system is designed as a safety device to turn off all heaters in
the event of a temperature control failure. The temperature control point
in the incubator will be approximately ±1°C of the overtemp setpoint.
The overtemp setpoint is set by the factory (default) at 40°C. However, the
overtemp can be reset over a range from 0.5°C above the operating
temperature setpoint to 65°C.
If the incubator’s operating temperature setpoint is set above the overtemp
setpoint, the overtemp setpoint will automatically update to 1°C above the
temperature setpoint. It is recommended that the overtemp setpoint be
maintained at 1°C over the operating temperature setpoint.
To set the Overtemp setpoint:
1. Press the Mode key until the Set indicator lights.
2. Press the right arrow until Otemp XX.X is displayed in the message
center.
3. Press the up or down arrow until the desired Otemp setpoint is
displayed.
4. Press Enter to save the setting.
5. Press the Mode key until the Run indicator lights, or press the right or
left arrow to go to the next or previous parameter.
All CO2cells are calibrated at the factory at 37°C, medium humidity and
10% CO2. Therefore, if a temperature setpoint of 37°C has been entered,
the humidity reservoir has been filled and the CO2control is set to run
between 0-10%, the CO2setpoint may be entered immediately. Otherwise,
it is important to allow the unit 12 hours to stabilize at the temperature
setpoint before entering the CO2setpoint.
1-10 Reach-In Incubator Thermo Scientific
Section 1
Installation and Start-Up
Setting the Overtemp
Setpoint
Setting the CO2Setpoint
All models of the incubator have a CO2setpoint range of 0.0% to 20.0%.
The incubator is shipped from the factory with a CO2setpoint of 0.0%.
At this setting, all CO2control and alarms are turned off.
To change the CO2setpoint:
1. Press the Mode key until the Set indicator lights.
2. Press the right arrow until “CO2 XX.X” is displayed in the message
center.
3. Press the up/down arrow until the desired CO2setpoint is displayed.
4. Press Enter to save the setpoint.
5. Press the Mode key until the Run indicator lights to go to Run mode,
or press the right/left arrow keys to go to next/previous parameter.
Reach-In Incubator 1-11Thermo Scientific
Section 1
Installation and Start-Up
Setting the CO2Setpoint
(continued)
1-12 Reach-In Incubator Thermo Scientific
Section 1
Installation and Start-Up
Press MODE to light SET
Silence Mode
Mode
Mode
SET POINTS
TEMP XX.X C
CO2 XX.X%
36.9
4.9
Enter
Enter
Enter
Numbers increase
Numbers increase
Numbers increase
Press MODE to move
to CALIBRATE
mode
Numbers decrease
Numbers decrease
Numbers decrease
Press Enter
to save setting
Press Enter
to save setting
Press Enter
to save setting
Run Set Cal Config Heat Temp
Inject % CO2
Scroll for Parameters
Scroll for Parameters
Scroll for Parameters
Mode OTEMP XX.X C EEnntteerr
Scroll for Parameters
Press to return
to previous parameter
Press to return
to previous parameter
Press to return
to previous parameter
Percent CO2
Over
Temperature
Operating
Temperature
To Se t:
Set Mode
Chart 1-1
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60